510(k) Data Aggregation

    K Number
    K090257

    Validate with FDA (Live)

    Date Cleared
    2009-05-26

    (113 days)

    Product Code
    Regulation Number
    866.5750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.

    Device Description

    IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document primarily focuses on analytical and clinical performance without explicitly stating numerical acceptance criteria prior to presenting the results. However, we can infer implied acceptance criteria from the reported performance, especially for linearity and specificity (inhibition). For clinical sensitivity and specificity, the provided confidence intervals suggest a critical range.

    Performance MetricAcceptance Criteria (Implied / Inferred)Reported Device Performance
    PrecisionWithin-run and total precision (CV%) generally low for various allergen concentrations. (No explicit numerical target provided.)Within-Run CV: 3.28% - 5.42%Total CV: 4.91% - 6.94%
    Linearity (Slope)Slope of regression (observed vs. expected) close to 1.00. (No explicit numerical range provided, but 95% CI is given.)Slopes: 0.997 - 1.011 (All 95% CIs include 1.00)
    Linearity (Intercept)Intercept of regression (observed vs. expected) close to 0.00. (No explicit numerical range provided, but 95% CI is given.)Intercepts: -0.322 - 0.425 (All 95% CIs include 0.00)
    Specificity (Inhibition)Target % inhibition of 50% met by relevant inhibitor extract in a concentration-dependent fashion.All allergens met the 50% inhibition target, showing concentration-dependent inhibition. Lowest reported inhibition at 0.008 mg/mL was 48.28% (Walnut) and 48.95% (Egg).
    Clinical SensitivityNo explicit numerical target provided, but 95% confidence interval presented.54.4% (95% CI: 48% - 60%)
    Clinical SpecificityNo explicit numerical target provided, but 95% confidence interval presented.96.9% (95% CI: 96% - 98%)
    Overall AgreementNo explicit numerical target provided, but 95% confidence interval presented.87.3% (95% CI: 85% - 89%)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision: For each of the seven allergens, three positive samples and one negative sample were tested. Each sample was assayed in two aliquots per day for 20 different days, totaling 40 measurements per sample.
    • Linearity: For each allergen, two samples were diluted into 5 levels (undiluted + 4 dilutions). The table shows N=12 for each allergen, implying 2 samples * 5 dilutions + 2 undiluted samples.
    • Specificity (Inhibition): A single serum sample or pool of sera was used for each allergen.
    • Clinical Performance:
      • Total Patients: 1,157 individuals.
      • Clinical Data: 261 positive, 896 normal.
      • Data Provenance: The document does not explicitly state the country of origin. It samples were described as "samples from non-atopic individuals and samples from atopic patients with case histories of suspected clinical reactions." This indicates that the data is retrospective clinical data for the "clinical data" ground truth, likely collected from patients presenting with symptoms.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The document does not specify the number of experts or their qualifications.
    • For clinical performance, the ground truth is referred to as "clinical data" which includes "case histories of suspected clinical reactions to the specific allergen or allergy group" and "documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." This suggests that the clinical diagnosis was established by medical professionals, but not explicitly stated as a panel of "experts" for the study itself.

    4. Adjudication Method for the Test Set

    • The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The "clinical data" appears to be the established reference.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study evaluates an in vitro diagnostic (IVD) assay (IMMULITE® 2000 3gAllergy™ Specific IgE Assay), which is a laboratory test, not an AI system that assists human readers. Therefore, the concept of "human readers improve with AI vs. without AI assistance" does not apply.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, the performance presented for the IMMULITE® 2000 3gAllergy™ Specific IgE Assay is standalone (algorithm only). This device is an automated in vitro diagnostic test that quantitatively measures allergen-specific IgE. Its performance is evaluated independently of human interpretation of the assay results, though human clinicians will interpret a patient's overall clinical picture, including the assay results.

    7. The Type of Ground Truth Used

    • Precision and Linearity: Internal quality control materials or spiked samples with known concentrations.
    • Specificity (Inhibition): Known relevant inhibitor extracts and negative control samples.
    • Clinical Performance: "Clinical data" described as "case histories of suspected clinical reactions to the specific allergen or allergy group" and "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." For the egg allergen, comparison studies were done against previously cleared individual egg constituents (egg white and egg yolk allergens). This is best characterized as clinical documentation/diagnosis.

    8. The Sample Size for the Training Set

    • The document does not mention a training set. This is not applicable as the device is a chemiluminescent immunoassay, not a machine learning algorithm that requires a training set. The "assay" itself is the device, and its development involves chemical and biological optimization, not algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this type of device.