510(k) Data Aggregation

    K Number
    K090031

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-04-16

    (101 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Actherm Infrared Ear Thermometers ACT8000 Series are used to measure body temperature from auditory canal. This device is intended for household and hospital use on people of all age, and is used with or without probe cover medACCU2010.

    Device Description

    Infrared Ear Thermometers (Model ACT8000 Series) are electronic Actherm thermometers using a thermopile as the temperature sensor's electric signal is then calculated and displayed by a MicroController. These thermometers display the temperature decimal. Model ACT8000 Series has the same indication for use. However, appearance difference exists between ACT8000 Series. Dual scale, Backlight, Scan LED, W/O cover, following with the model number indicates scale switchable, with Backlight, with Scan completed indicator, without probe cover.

    The digital thermometer comprises: a thermopile for temperature sensing, a buzzer for sounding effect, a Micro Controller and a LCD display for calculating and displaying the target temperature digitally.

    The system uses two 1.5V DC batteries for the power supply and the battery power is automatically checked by the MicroController and displayed in LCD if the battery is exhausted.

    Actherm Infrared Ear Thermometers ACT8000 Series can be combined with accessories, weighting base and wall mount. For the weighting base, the whole set of body temperature measurement devices, including infrared ear thermometer, digital thermometer, and probe cover, can be put on it, For the wall mount, it can be fixed on the wall and the cradle of infrared ear thermometer can be put on this wall mount. Also, the probe cover boxes can be hanged on the hook that under the wall mount or put inside the cradle.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Actherm Infrared Ear Thermometer (K090031):

    Device: Actherm Infrared Ear Thermometer ACT8000 Series and its Probe Cover medACCU2010

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Meet ASTM Standard Specification for Infrared Thermometer for Intermittent Determination of Patient Temperature (ASTM E 1965-98: 2003)Confirmed to meet ASTM E 1965-98: 2003
    Meet EN60601-1 requirementsConfirmed to meet EN60601-1
    Meet EN60601-1-2 requirementsConfirmed to meet EN60601-1-2
    Meet EN12470-5 requirementsConfirmed to meet EN12470-5
    AccuracyConfirmed to meet specified accuracy
    PrecisionConfirmed to meet specified precision
    RepeatabilityConfirmed to meet specified repeatability
    Biological CompatibilityIdentical materials used in other legally marketed devices under the same use conditions (implies biological compatibility is met)

    Note: The document explicitly states that "Bench testing confirmed accuracy, precision and repeatability measurements specified in the labeling." However, the specific quantitative values for these criteria (e.g., ± 0.2°C) are not provided in this summary. It simply states they were met according to the labeling.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated for any of the studies (bench or clinical).
    • Data Provenance: Not explicitly stated. The document mentions "bench testing contained in this submission and clinical testing supplied." It does not specify the country of origin or if these studies were retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Experts: Not explicitly stated.
    • Qualifications: Not explicitly stated.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not explicitly stated.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • MRMC Study: No, an MRMC comparative effectiveness study was not conducted or reported in this summary. This device is a thermometer, which typically does not involve human readers interpreting AI output in the same way imaging devices might.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop):

    • Standalone Performance: Yes, the performance criteria listed (accuracy, precision, repeatability) are inherently standalone performance measures for the device itself.
      • "Bench testing confirmed accuracy, precision and repeatability measurements specified in the labeling."
      • The thermometer operates autonomously to measure temperature.

    7. Type of Ground Truth Used:

    • Ground Truth: For the bench testing (accuracy, precision, repeatability), a "ground truth" would typically be established using a reference thermometer or a calibrated temperature source with known accuracy (e.g., a blackbody radiator for infrared thermometers).
    • For "clinical testing," the ground truth for body temperature would likely be established by a highly accurate reference method, such as a rectal thermometer or an oral thermometer with known accuracy, under controlled clinical conditions. The document does not specify the exact method used for clinical testing ground truth.

    8. Sample Size for the Training Set:

    • Sample Size: Not applicable. This device is an infrared ear thermometer, a hardware medical device that measures temperature. It does not use "training sets" in the context of machine learning or AI algorithms. Its performance is based on its physical design, calibration, and adherence to established engineering and medical device standards.

    9. How the Ground Truth for the Training Set Was Established:

    • Ground Truth Establishment: Not applicable, as there is no "training set" in the context of this device.