    Device Name
    BIOSIGN FLU A+B
    Date Cleared
    2010-11-10

    (693 days)

    Product Code
    Regulation Number
    866.3328
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioSign® Flu A+B test is an in vitro rapid qualitative test that detects influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, and nasopharyngcal aspirate/wash specimens obtained from patient with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. A negative test result is presumptive and it is recommended these results be confirmed by viral culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

    Device Description

    BioSign® Flu A+B is an immuno-chromatographic test for the rapid, qualitative detection of influenza A and B. The test device has two test lines, thereby allowing the separate identification of type A and/or type B viral antigens from the same specimen.

    In the test procedure, a specimen is collected and placed into the Extraction Well of the test device containing extraction solution for one minute, during which time antigen is extracted from disrupted virus particles. The test device is then raised, tapped and laid back down onto a level surface to allow the solution in the Extraction Well to migrate through the pads containing lyophilized detector antibodies conjugated to gold dye and then through the test membrane. If influenza antigens are present in the specimen, they will react with anti-influenza antibody coupled to gold dye particles, migrate through the membrane as antigen-antibody-dye complexes, bind to the immobilized anti-influenza antibody on the membrane, and generate a colored line in the Test line position (A and/or B). The rest of the sample and unbound/bound dye complexes continue to migrate to the Control line position, where antibody to the antiinfluenza antibody is immobilized, and anti-influenza antibody-unbound/bound dye complexes form the Control line (internal procedural control).

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary, extracting the requested information:

    The device under review is the BioSign® Flu A+B / Status™ Flu A & B, an immuno-chromatographic test for the rapid, qualitative detection of influenza A and B antigens.


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are implicitly set by the reported performance (sensitivity and specificity) when compared against viral culture as the reference method. The document presents performance for different sample types (nasopharyngeal aspirate, nasopharyngeal swab, nasal swab).

    Table of Acceptance Criteria and Reported Device Performance:

    Sample Type & AntigenAcceptance Criteria (Implicit from Study Results)Reported Device Performance (Sensitivity)Reported Device Performance (Specificity)
    Nasopharyngeal AspirateHigh sensitivity and specificity required for clinical utility.
    Flu A (Sensitivity)-95.3% (95% CI: 92.1-98.5%)-
    Flu A (Specificity)--85.7% (95% CI: 83.3-88.1%)
    Flu B (Sensitivity)-91.6% (95% CI: 83.6-99.6%)-
    Flu B (Specificity)--97.5% (95% CI: 96.5-98.5%)
    Nasopharyngeal SampleHigh sensitivity and specificity required for clinical utility.
    Flu A (Sensitivity)-89.6% (95% CI: 84.0-95.2%)-
    Flu A (Specificity)--77.0% (95% CI: 74.2-79.8%)
    Flu B (Sensitivity)-86.8% (95% CI: 81.4-92.2%)-
    Flu B (Specificity)--92.9% (95% CI: 91.2-94.6%)
    Nasal Swab SampleHigh sensitivity and specificity required for clinical utility.
    Flu A (Sensitivity)-91.7% (95% CI: 78.2-97.1%)-
    Flu A (Specificity)--75.2% (95% CI: 70.2-79.6%)
    Flu B (Sensitivity)-82.4% (95% CI: 59.0-93.8%)-
    Flu B (Specificity)--88.3% (95% CI: 84.4-91.3%)

    Note: The document does not explicitly state numerical "acceptance criteria" thresholds, but rather reports the observed performance and implies these values were sufficient for regulatory clearance. The wide confidence intervals in some cases suggest that the performance might vary.


    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Total Patients: 862
      • Nasopharyngeal Aspirate: 253 cases
      • Nasopharyngeal Sample: 251 cases
      • Nasal Swab Sample: 358 cases
        (Note: The sum of individual sample types is 862, matching the total number of patients, indicating these might be overlapping or distinct subsets of samples from the same patient pool, or simply different sample types evaluated from the overall patient population.)
      • Archived Samples: 80 samples for both Flu A and Flu B verification (Columbia NY Presbyterian Hospital).
    • Data Provenance:
      • Country of Origin: USA (samples collected at 5 sites in the USA).
      • Retrospective or Prospective: Prospective clinical study, with additional testing on "Archived Samples" which would be considered retrospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used to establish the ground truth or their qualifications. The ground truth method was viral culture, and it's common for laboratory professionals to perform and interpret viral cultures, but no details on their expertise are provided.


    4. Adjudication Method for the Test Set

    The primary reference method for initial comparison was viral culture. For samples that produced discrepant results between the BioSign® Flu A+B test and viral culture, a third, more definitive method was used for resolution: proFLU plus by Prodesse (real-time RT-PCR, or PCR). This constitutes an adjudication method, where ambiguous cases are resolved by a higher-tier test. The document refers to the results from the PCR as "further analyzed" or used to identify cases that were "positive by both BioSign and PCR" or "negative by both BioSign and PCR".


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study design is typically used for image-based diagnostic aids where human readers interpret results with and without AI assistance to assess the impact of AI on reader performance. The BioSign® Flu A+B is a rapid diagnostic test, not an AI-powered diagnostic imaging device.


    6. Standalone (Algorithm Only) Performance Study

    Yes, a standalone performance study was done for the device. The reported sensitivity and specificity values directly reflect the performance of the BioSign® Flu A+B device itself, without any human interpretation adjustments beyond standard operation and reading of a rapid diagnostic test. The entire clinical study section, including the performance tables, represents the standalone performance of the device against the ground truth.


    7. Type of Ground Truth Used

    The type of ground truth used was primarily viral culture, which is a laboratory-based method for identifying and growing viruses. For discrepant results, real-time RT-PCR (proFLU plus by Prodesse) was used as a confirmatory or adjudicating ground truth.


    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" or its sample size. This type of device (rapid diagnostic test) is typically developed and optimized during its R&D phase using various sample sets, but the clinical study described is primarily for validation/testing of the finalized device.


    9. How the Ground Truth for the Training Set Was Established

    As no explicit training set is described, the method for establishing ground truth for a training set (if one existed in the R&D phase) is not provided. The ground truth for the clinical test set was established by viral culture (and PCR for discrepants), as described in point 7.