510(k) Data Aggregation

    K Number
    K082140

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-09-29

    (61 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cepheid Xpert™ MRSA/SA Blood Culture Assay performed on the GeneXpert® Dx System™ is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from patient positive blood cultures. The assay utilizes automated realtime polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The assay is performed directly on positive blood culture specimens using BD BACTEC™ Plus Aerobic/F blood culture bottles that are determined as Gram Positive Cocci in Clusters (GPCC) or as Gram Positive Cocci in singles (GPC) by Gram stain. The Cepheid Xpert™ MRSA/SA Blood Culture Assay is not intended to monitor treatment for MRSA/SA infections. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing.

    Device Description

    The Cepheid Xpert™ MRSA/SA Blood Culture Assay (Xpert MRSA/SA Blood Culture Assay) is a rapid, automated DNA test for simultaneously detecting MRSA and SA Alber) I a rapids, and culture specimens. The specimens. The specimen consists of an aliquot taken from a positive blood culture bottle for testing with the Xpert MRSA/SA Blood Culture Assay. Using one of the small disposable transfer pipettes provided with the test kit, a single drop of the positive blood culture aliquot (approximately 50 µL) is transferred into the Elution Reagent. The Elution Reagent is vortexed for 10 seconds and the entire contents are transferred to "S" chamber of the disposable fluidic cartridge (the Xpert MRSA/SA Blood Culture Assay cartridge). The two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to different, uniquelylabeled chambers of the Xpert MRSA/SA cartridge. The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off real-time, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fully-automated and completely integrated.

    The GeneXpert Dx System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of MRSA and SA in approximately 50 minutes. Each system has 2 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

    The Xpert MRSA/SA Assay includes reagents for the simultaneous detection of the target organisms, SA and MRSA. The primers and probes in the Xpert MRSA/SA Assay detect nucleic acid sequences of the staphylococcal protein A (spa), the gene for methicillin/oxacillin resistance (mecA), and staphylococcal cassette chromosome (SCCmec) inserted in the SA chromosomal attB site.

    The test includes a Sample Processing Control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are detailed below for the Cepheid Xpert™ MRSA/SA Blood Culture Assay.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document implicitly defines acceptance criteria based on the performance observed in the predicate device and the clinical study results. The clinical study results serve as the primary demonstration of meeting these criteria.

    Metric (MRSA)Acceptance Criteria (Implicit from Predicate/Standard of Care)Reported Device Performance (Xpert MRSA/SA Blood Culture Assay)
    Sensitivity98.2 - 100.0% (BD GeneOhm StaphSR)100.0% (53/53) (95% CI: 93.3% - 100%)
    Specificity98.2 - 100.0% (BD GeneOhm StaphSR negative % agreement)100.0% (196/196) (95% CI: 98.1% - 100%)
    Metric (SA)Acceptance Criteria (Implicit from Predicate/Standard of Care)Reported Device Performance (Xpert MRSA/SA Blood Culture Assay)
    Sensitivity98.8 - 100.0% (BD GeneOhm StaphSR)100.0% (77/77) (95% CI: 95.3% - 100%)
    Specificity96.5 - 100.0% (BD GeneOhm StaphSR negative % agreement)99.4% (171/172) (95% CI: 96.8% - 100%)

    Note: The predicate device's performance is listed as "Positive % Agreement" and "Negative % Agreement," which are equivalent to sensitivity and specificity in this context. The implicit acceptance criteria for the new device are to perform at least comparably to the predicate and the "standard of care" (reference culture results and susceptibility testing).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Clinical Test Set: 249 specimens.
    • Data Provenance: Multi-site prospective investigation study at three US institutions. The data is prospective, collected from subjects whose routine care included blood culture testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    The document does not explicitly state the number of experts or their specific qualifications (e.g., years of experience as a radiologist) used to establish the ground truth for the clinical test set. It mentions "reference culture results and susceptibility testing, the current standard of care" and that susceptibility testing was performed "in accordance with the CLSI documents M2-A9 and M100-S17." This implies that qualified laboratory personnel, following established clinical microbiology guidelines, performed these reference methods.

    4. Adjudication Method for the Test Set:

    The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1). The ground truth was established by "reference culture results and susceptibility testing, the current standard of care." This suggests a direct comparison against the established laboratory gold standard, rather than a consensus among multiple human readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly done in the context of comparing human readers with and without AI assistance. The study compares the device's performance directly against "reference culture results and susceptibility testing," which is the established standard of care for identifying bacterial infections. The device is an automated nucleic acid amplification test, not an AI assistance tool for human interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Yes, a standalone performance study was conducted. The Cepheid Xpert™ MRSA/SA Blood Culture Assay is an automated device that performs sample preparation, amplification, and real-time detection without human intervention once the sample is loaded. The "Overall Results" section (Page 12) directly reports the device's performance in identifying MRSA and SA specimens relative to culture. This refers to the algorithm's performance without a human in the loop for interpretation.

    7. The Type of Ground Truth Used:

    The ground truth used for the clinical study was based on:

    • Reference Culture Results: This involves standard microbiological culture techniques to isolate and identify organisms.
    • Susceptibility Testing: Performed in accordance with CLSI documents (M2-A9 and M100-S17) to determine methicillin/oxacillin resistance, using cefoxitin as a surrogate.

    This represents established microbiological laboratory standards, often considered the "gold standard" for pathogen identification and antimicrobial susceptibility.

    8. The Sample Size for the Training Set:

    The document does not explicitly state a separate "training set" for the clinical validation studies in the way one might for a machine learning algorithm. For analytical studies, various strains were tested:

    • Analytical Inclusivity Study on CDC Staphylococcus aureus Specimens: 25 Staphylococcus aureus strains.
    • Analytical Inclusivity Study on Expanded Panel of Staphylococcus aureus Specimens: 121 additional Staphylococcus aureus strains.
    • "Empty Cassette Variants" Study: 22 Staphylococcus aureus isolates.
    • Analytical Limit of Detection: Replicates of 20 for 6 MRSA isolates and 3 MSSA isolates.
    • Cross-reactivity Study: 105 strains (98 ATCC, 7 NARSA).

    9. How the Ground Truth for the Training Set Was Established:

    For the analytical "training" (or rather, validation) sets mentioned above, the ground truth was established through:

    • Bacterial strain identification, PFGE type, and SCCmec type determined by the CDC (for CDC specimens).
    • Catalase, tube coagulase, and Gram stain for characterizing discordants in the expanded panel study.
    • Methicillin susceptibility assessed by disk diffusion using a cefoxitin disk.
    • Known characteristics and genetic information of the cultured strains (e.g., MRSA, SA, Cfu/test).