510(k) Data Aggregation

    K Number
    K081777

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-09-29

    (98 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ventrio Hernia Patch is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias.

    Device Description

    The proposed Ventrio Hernia Patch is a self-expanding, sterile prosthesis, containing two primary layers of mesh stitched with polytetrafluoroethylene (PTFE) monofilament to an expanded polytetrafluoroethylene (ePTFE) sheet. The mesh component is non-absorbable, however, the device contains a fully absorbable recoil ring using AbsorbaFlex Memory Technology, an absorbable polydioxanone (PDO) monofilament. The AbsorbaFlex Memory Technology's PDO monofilament ring provides memory and stability to the device, facilitating ease of initial insertion, proper placement, and fixation of the device. The mesh component of the device is constructed of knitted polypropylene monofilament approximately 0.006" in diameter. The AbsorbaFlex PDO monofilament ring is 0.038" in diameter. It has two donut-shaped layers of knitted polypropylene monofilament mesh 0.006" in diameter on each side of the ring. These layers of mesh are stitched together around the ring using PTFE monofilament. The extra large oval sized patches contain inner and outer PDO monofilament rings. A single layer of expanded ePTFE is attached to the bottom primary layer of polypropylene mesh (posterior mesh). The attachment is accomplished with an interlocking stitch using PTFE monofilament. The peripheral edge of the polypropylene mesh will be heat sealed to the ePTFE layer.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Ventrio Hernia Patch." This summary focuses on the device's characteristics and its substantial equivalence to predicate devices, rather than presenting a traditional clinical study with acceptance criteria and performance metrics in the format requested.

    Therefore, many of the requested fields cannot be directly extracted from the provided text, as this document is a regulatory submission for a hernia patch, not an AI/algorithm-based diagnostic device with performance metrics like sensitivity, specificity, or AUC. The "performance data" mentioned refers to laboratory bench testing and in-vitro/in-vivo resorption studies to demonstrate mechanical strength and resorption characteristics, not diagnostic accuracy.

    However, I will populate the table and answer the questions based on the information that is available in the provided text.


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance
    Material Composition EquivalenceThe Ventrio Hernia Patch contains two primary layers of knitted polypropylene monofilament mesh and an expanded polytetrafluoroethylene (ePTFE) sheet, similar to predicate devices. It uses a fully absorbable recoil ring made of polydioxanone (PDO) monofilament.
    Mechanical StrengthLaboratory bench testing was performed to compare the Ventrio Hernia Patch to the Bard Composix Kugel Hernia Patch.
    Resorption CharacteristicsIn-vitro and in-vivo resorption studies were performed on the PDO monofilament. These studies characterized the mechanical strength and resorption.
    Intended Use EquivalenceThe intended use for the Ventrio Hernia Patch is stated to be the same as the predicate device (Bard Composix Kugel Hernia Patch), "with the exception of the indication for repair of chest wall defects."
    Safety and EffectivenessBased on laboratory testing and in-vitro and in-vivo resorption data, the results show that the proposed device is as safe and effective for its intended use as the currently marketed predicate device.
    Substantial EquivalenceBased on performance data, the proposed device is determined to be substantially equivalent to the predicate device (Bard Composix Kugel Hernia Patch (K061314)).

    2. Sample size used for the test set and the data provenance

    The document does not detail specific sample sizes for "test sets" in a clinical study sense. The performance data relied on "laboratory bench testing" and "in-vitro and in-vivo resorption studies." The provenance of this data (e.g., country of origin, retrospective/prospective) is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document pertains to a medical device (hernia patch), not an AI/algorithm-based diagnostic tool requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for this device would be its physical properties and biological interaction, assessed through laboratory and animal studies, not expert consensus on diagnostic cases.

    4. Adjudication method for the test set

    Not applicable. See reasoning for point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a physical medical device (hernia patch), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document describes a physical medical device (hernia patch), not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance would be established through:

    • Physical/Mechanical Testing: Measuring properties like tensile strength, tear resistance, and overall structural integrity during laboratory bench testing.
    • In-vitro Studies: Assessing material behavior in simulated biological environments.
    • In-vivo Resorption Studies: Evaluating the biological degradation and mechanical stability of the absorbable PDO monofilament in living organisms (likely animal models, though not explicitly stated) over time.

    8. The sample size for the training set

    Not applicable. This document pertains to a medical device (hernia patch), not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. See reasoning for point 8.