Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K063544

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-02-14

    (82 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pyrenees Cervical Plate System is indicated for use in anterior screw fixation to the cervical spine (C2 - C7) for the following indications: degenerative disc disease (DDD), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).

    Device Description

    The Pyrenees Cervical Plate System is a spinal fixation system which consists of cervical screws and plates. All of the components are available in a variety of sizes to match more closely the patient's anatomy. Materials: The devices are manufactured from Commercially Pure titanium and titanium alloy per ASTM and ISO standards. Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of cervical segments of the spine.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the "Pyrenees Cervical Plate System". This document focuses on demonstrating substantial equivalence to legally marketed predicate devices through material and functional comparisons, not on efficacy studies involving AI or complex statistical analysis of diagnostic accuracy. Therefore, most of the requested information regarding acceptance criteria, device performance, ground truth, and expert evaluation is not present in this type of submission.

    Here's a breakdown of what can be extracted and what is not available based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    This type of information (specific performance metrics for AI, such as sensitivity, specificity, or AUC) is not applicable to this medical device submission. The submission is for a spinal fixation system, not a diagnostic AI tool.

    Acceptance CriteriaReported Device Performance
    Material CompositionMeets ASTM and ISO standards for Commercially Pure titanium and titanium alloy.
    FunctionalityFunctions as an adjunct to fusion to provide immobilization and stabilization of cervical segments of the spine.
    Substantial EquivalenceDemonstrated to be substantially equivalent to current cleared Pyrenees Cervical Plate (K060442), Synthes CSLP (K971883, K000538), K2M, Inc. Tectonic Anterior Cervical Plate System (K051531), De Puy Acromed PEAK Cervical Plate System (K971730, K926486) and the Interpore Cross Anterior Cervical Plate System (K002592) in design, function, material, and intended use.
    Mechanical TestingTesting in accordance with ASTM F1717 was performed. (Specific results or acceptance thresholds are not detailed in this summary.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    Not applicable. This is a submission for a physical medical device (cervical plate system), not an AI diagnostic tool that uses datasets. The "testing" mentioned refers to mechanical and material compliance testing, not clinical studies with patient data in the context of an AI algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. There is no "ground truth" in the AI diagnostic sense for a cervical plate system. The ground truth for this device would be established through engineering specifications, material science, and possibly in-vitro mechanical testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This relates to expert review of AI diagnostic outcomes, which is not relevant for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is not an AI diagnostic tool and does not involve human readers interpreting images with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For this type of device, the "ground truth" is primarily engineering and material specifications, and performance against established standards (e.g., ASTM F1717 for mechanical properties). Clinical outcomes would be observed post-market but are not part of this 510(k) summary for substantial equivalence.

    8. The sample size for the training set:

    Not applicable. There is no "training set" in the AI sense for this device.

    9. How the ground truth for the training set was established:

    Not applicable. There is no AI training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1