510(k) Data Aggregation

    K Number
    K063167

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-02-06

    (111 days)

    Product Code
    Regulation Number
    868.1400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EMMA Emergency Capnometer Monitor measures, displays and monitors carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

    The EMMA Emergency Capnometer Analyzer measures and displays carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

    Device Description

    The EMMA Emergency Capnometer is a miniature mainstream infrared gas analysis bench with an integrated user interface. The complete carbon dioxide analyzer is contained within a transducer that is attached to the breathing circuit via the EMMA Airway Adapter.

    AI/ML Overview

    The provided 510(k) summary for the EMMA Emergency Capnometer focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a standalone study with specific performance metrics. Therefore, many of the requested details are not explicitly present in the document.

    Here's an analysis based on the available information:

    Acceptance Criteria and Device Performance

    The documentation does not provide specific quantitative acceptance criteria (e.g., a target accuracy range for CO2 concentration or respiratory rate) with corresponding reported device performance. Instead, it states a general conclusion about equivalence or superiority.

    The key statement is from section 12: "The EMMA Emergency Capnometer demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."

    Acceptance CriteriaReported Device Performance
    Not explicitly stated as quantitative targets. The implicit acceptance criteria are that the device's performance, safety, and effectiveness are at least equivalent to or superior to the predicate devices."Demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified.
      • Data Provenance: The study involved "calibrated gas samples and legally marketed anesthesia and ventilation devices." This suggests a laboratory or bench testing environment. There is no information regarding country of origin or whether the data was retrospective or prospective in a clinical setting.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not mentioned.
      • Qualifications of Experts: Not mentioned. The ground truth was likely established by the "calibrated gas samples" (known concentrations) and possibly measurements from the "legally marketed anesthesia and ventilation devices" (which would have their own validated measurement capabilities).
    3. Adjudication method for the test set:

      • Not applicable/Not mentioned. The study appears to be a direct comparison against a known standard (calibrated gas) and established medical devices, rather than a subjective assessment requiring expert adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Was an MRMC study done? No. This device is an emergency capnometer, a medical measurement device, not an AI-powered diagnostic tool requiring human reader interpretation. The study described is a device-to-device performance comparison.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes, implicitly. The testing described ("Testing in direct comparison to predicates throughout the operating range was conducted using calibrated gas samples and legally marketed anesthesia and ventilation devices") is a standalone performance test of the device itself, without human interpretation of its outputs being part of the primary evaluation.
    6. The type of ground truth used:

      • Ground Truth Type: A combination of "calibrated gas samples" (representing a known, accurate CO2 concentration) and measurements from "legally marketed anesthesia and ventilation devices" (which are themselves considered accurate and reliable benchmarks). This is akin to a reference standard or benchmarking against established, validated measurement systems.
    7. The sample size for the training set:

      • Not applicable/Not mentioned. This device does not appear to be an AI/machine learning product requiring a training set in the conventional sense. Its development and validation are based on established engineering principles for gas analysis.
    8. How the ground truth for the training set was established:

      • Not applicable/Not mentioned, as there is no mention of a training set for an AI/ML model.

    Summary of Device Performance and Equivalence Claim:

    The 510(k) submission for the EMMA Emergency Capnometer establishes substantial equivalence primarily through direct comparative testing against its predicate devices (Tidal Wave Model 610, Novametrix Medical Systems Inc. and VEO Multigas Monitor for Pocket PC, Phasein AB). The "testing vs. predicates" section (11) indicates that the device was compared using "calibrated gas samples and legally marketed anesthesia and ventilation devices" across its operating range. The conclusion (12) states that the EMMA Emergency Capnometer "demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics." This implies that the device's accuracy in measuring CO2 concentration and respiratory rate was found to be at least as good as, if not better than, the predicate devices when tested under controlled conditions.