510(k) Data Aggregation

    K Number
    K061461

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-12-20

    (208 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAYO Conservative Hip Prosthesis is indicated for cementless use in skeletally mature individuals undergoing primary surgery for total hip replacement. Diagnostic indications include severe hip pain and disabilities due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, slipped capital femoral epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

    Device Description

    Like its predicate, the modified MAYO Conservative Hip Prosthesis is a modular femoral stem intended to replace the hip joint in total hip arthroplasty. It is identical to its predicate in that it features a 12/14 Morse-type taper and uses the same variety of modular femoral heads. It differs only in overall size where the design is intended to address smaller patient anatomies. As with the predicate, the proposed stem is collarless, wedge-shaped, and is designed for use without bone cement. Fixation is achieved by mechanical press fit into the proximal femoral shaft and by biological ingrowth into the fiber metal pads. As with the predicate, the proposed device is available with and without Calcicoat Ceramic Coating (HA/TCP).

    AI/ML Overview

    This device, the MAYO® Conservative Hip Prosthesis, is a traditional medical device, not an AI/ML-driven device. As such, the concept of "acceptance criteria" and performance metrics typical for AI/ML devices (like accuracy, precision, etc.) do not directly apply in the same way.

    However, I can extract the relevant information from the provided text regarding the device and its evaluation as requested, adapting the AI/ML framework where possible to the context of a traditional medical device submission.

    Here's a breakdown based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial Equivalence: Device is equivalent to its predicate device (MAYO Conservative Hip Prosthesis, K030733) in terms of intended use, fundamental scientific technology, and safety/effectiveness for its indications.The modified MAYO Conservative Hip Prosthesis:Is a modular femoral stem intended to replace the hip joint in total hip arthroplasty, similar to its predicate.Features a 12/14 Morse-type taper and uses the same modular femoral heads as the predicate.Differs only in overall size to address smaller patient anatomies.Is collarless, wedge-shaped, and designed for cementless use with mechanical press fit and biological ingrowth into fiber metal pads, like the predicate.Is available with and without Calcicoat® Ceramic Coating (HA/TCP), like the predicate.Has the same intended use for cementless total hip replacement in skeletally mature individuals with diagnostic indications including various forms of arthritis and hip conditions.Changes neither the intended use nor the fundamental scientific technology compared to the predicate.Is packaged, manufactured, and sterilized using the same materials and processes as the predicate.Engineering evaluations demonstrated that the device is equivalent to the predicate.
    Safety and Effectiveness: No new questions of safety or effectiveness are raised.Non-Clinical Performance and Conclusions: "Engineering evaluations demonstrated that the device is equivalent to the predicate." This implies that the device maintained the safety and effectiveness profile of the predicate device. Clinical Performance and Conclusions: "Clinical data and conclusions were not needed for this device." This statement, coupled with the substantial equivalence finding, indicates that the FDA considered the non-clinical data sufficient to demonstrate safety and effectiveness without requiring new human trials.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: Not applicable. This submission relies on "engineering evaluations" and comparison to a predicate device, rather than a "test set" of patient data or images in the AI/ML sense.
    • Data Provenance: Not applicable for a "test set." The evaluation is based on engineering principles and comparison to the predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable. Ground truth for a test set is not relevant for this type of device submission based on engineering evaluations and predicate comparison. The "experts" involved would be the Zimmer Inc. engineers performing the evaluations and the FDA reviewers assessing the submission, but they are not establishing "ground truth" for a dataset.

    4. Adjudication Method for the Test Set

    • Not applicable, as there is no "test set" in the AI/ML sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a traditional hip prosthesis, not an AI-driven medical imaging or diagnostic device. Therefore, MRMC studies and human reader improvement with AI assistance are not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical implantable device, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" in this context is the established safety and effectiveness of the predicate device (MAYO Conservative Hip Prosthesis, K030733) and the engineering principles applied to demonstrate that the modified device is equivalent. It's based on engineering evaluations and regulatory precedent, not pathology, expert consensus on patient data, or outcomes data specifically for this modified device.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not involve a "training set" as it is not an AI/ML algorithm.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable, as there is no "training set" for this device.