    Device Name
    THERMASSAGE
    Manufacturer
    Date Cleared
    2006-10-12

    (184 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TherMassager is indicated for use in:

    • Relief of minor muscle aches and pains
    • Relief of muscle spasm
    • Temporary improvement of local circulation (i.e., blood circulation)
    • Temporary improvement in the appearance of cellulite
    Device Description

    The TherMassager is a self-contained, 9.0 VIDC electrically powered therapeutic massager. The basic configuration is that of an ergonomically rounded hand-held box-like unit. Weight is approximately 4 pounds. The housing is made from medical grade plastic and contains a suction pump, a massage motor and skin rollers. The outside of the unit is affixed with a strap which may be secured over the operators hand to provide additional grip during treatment application. Three treatment rollers are supplied with each unit. The TherMassager unit has I spiral roller, 1 ribbed roller and 1 lobed (contoured) roller. The rollers are made from medical grade silicone elastomer meeting ISO 10993 biocompatibility standards.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the manner requested. This document is a 510(k) premarket notification for a therapeutic massager, the TherMassager.

    Instead, the document details the device's substantial equivalence to a predicate device (UPC Therapeutic Massager, K990445) based on a comparison of device characteristics and intended use. The core of this submission is to demonstrate that the new device is as safe and effective as a legally marketed device, not to present novel performance data from a specific study against predefined acceptance criteria.

    Therefore, I cannot provide the requested table, sample sizes, expert qualifications, adjudication methods, MRMC study details, standalone performance, or ground truth information because this type of data is not present in the provided 510(k) summary.

    The closest information available related to "acceptance criteria" is the comparison of device characteristics to the predicate device to establish substantial equivalence.

    Here's a breakdown of what can be extracted or inferred based on the document's content, acknowledging the limitations for your specific request:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in a format of "acceptance criteria" versus "reported performance." Instead, the document presents a side-by-side comparison of features between the TherMassager and the predicate device (LPG). Substantial equivalence is the "acceptance criteria" here, meaning the device's characteristics and performance are similar enough to the predicate to be considered safe and effective for the same indications.

    FeatureTherMassager (Reported Device)LPG (Predicate Device)Substantially Equivalent? (SF?)
    Indications for UseRelief of minor muscle aches and painsRelief of muscle spasmTemporary improvement of local circulationTemporary improvement in the appearance of celluliteSAMEYES
    Power RequirementsAC/DC AdaptorInput = 120 VDC - 60 HzOutput = 9 VDC120 VAC -- 60 HzYES
    Weight: MassagerAC/DC Adapter1.8 Kg.5 Kg2 Kg - Handpiece onlyYES
    Leakage Current≤50 µamp<50 µampYES
    Vacuum (Torr)Variable, 760-400 TorrVariable, 50-500 mBar (720 - 385 Torr equivalent)YES
    MaterialsHousing: medical grade plastic - 2401 MT SGRollers silicone - ISO 10993 compliantMedical grade materials, unspecifiedYES
    Massage UnitVariable speed - User controlledVariable according to vacuum freq. and cycle rateYES
    Vacuum frequency & cycle rateConstant - force variableVariable - pulsatesYES
    RollersPoweredPoweredYES

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This is a 510(k) submission based on substantial equivalence to an existing device, not a performance study with a test set of data points. There is no mention of a clinical or analytical study with a test set of samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No ground truth establishment by experts is described as part of this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no test set or adjudication process described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a therapeutic massager, not an AI or imaging diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical therapeutic massager, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The basis for equivalence relies on comparing technical specifications and intended use to an already legally marketed device, not on clinical ground truth or outcomes data for the new device.

    8. The sample size for the training set

    Not applicable. There is no mention of a training set for an algorithm or model.

    9. How the ground truth for the training set was established

    Not applicable. There is no mention of a training set or ground truth establishment.