510(k) Data Aggregation

    K Number
    K041307

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-06-14

    (28 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stryker XCEL Anchor System is intended for use in securing soft tissue to bone in such procedures as:

    Shoulder:
    Bankart repair
    SLAP lesion repair
    Rotator cuff repair
    Capsular shift repair
    Biceps tenodesis
    Acromio-clavicular separation

    Knee:
    Medial collateral ligament repair
    Lateral collateral ligament repair
    Joint capsule closure to anterior proximal tibia
    Posterior oblique ligament or joint capsule to tibia repair
    Extra capsular reconstruction/ITB tenodesis
    Patellar ligament and tendon avulsion repairs

    Elbow:
    Biceps tendon reattachment

    Foot & Ankle:
    Achilles tendon repair/reconstruction
    Lateral stabilization
    Medial stabilization

    The Stryker XCEL Anchor System is intended for single-use only.

    Device Description

    The line extension of the Stryker XCEL Anchor System consists of a Poly L-lactic acid (PLLA) anchor pre-threaded with non-absorbable braided polyethylene surgical suture, and pre-assembled on a disposable inserter.

    AI/ML Overview

    The provided document is a 510(k) summary for the Stryker XCEL Anchor System, a medical device intended for securing soft tissue to bone. It's a "line extension" of an existing device, meaning it's substantially equivalent to a previously cleared product.

    Here's the information regarding acceptance criteria and the study, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in terms of performance metrics (e.g., tensile strength, pull-out force with specific thresholds). Instead, the acceptance criterion for this 510(k) "line extension" is primarily substantial equivalence to the predicate device.

    Acceptance Criterion (Implicit)Reported Device Performance
    Substantial Equivalence to Predicate Device (K023013)"The line extension of the Stryker XCEL Anchor System is substantially equivalent in material of construction, overall design, intended use, and safety and efficacy to the predicate device. The subject device was shown to have substantially equivalent performance when compared to the predicate device."
    Sterility (SAL of 10^-6)"The line extension of the Stryker XCEL Anchor System will be provided sterile for single-use (ASTM 4169)... validated to a sterility assurance level (SAL) of 10−⁶."
    Biocompatibility (per ISO-10993-1 and G95-1)"The device is biocompatible per ISO-10993-1 and G95-1."
    Ethylene Oxide Residuals (limits per ANSI/AAMI/ISO 11135)"The device will be sterilized by Ethylene Oxide (ANSI/AAMI/ISO 11135), including limits for Ethylene Oxide residuals..."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide details on specific sample sizes for a test set or the provenance (country of origin, retrospective/prospective) of data used for any specific performance testing. The focus of this 510(k) summary is on demonstrating substantial equivalence rather than presenting detailed raw study data. It implies that performance testing was conducted, but the specifics are not included in this high-level summary.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This document does not describe a clinical study with human patients where expert adjudication of outcomes would be required to establish "ground truth." The evaluation here is for a medical device's substantial equivalence to a predicate, not a diagnostic algorithm. Therefore, information about experts for ground truth is not applicable in this context.

    4. Adjudication Method for the Test Set

    As there's no mention of a clinical study generating a "test set" requiring human interpretation or diagnosis, there's no adjudication method described in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done for this 510(k) application. This type of study is typically for evaluating the impact of an AI-powered diagnostic tool on human reader performance. The Stryker XCEL Anchor System is a physical surgical implant, not a diagnostic AI device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    N/A. This is not an algorithm or AI device, so the concept of standalone performance does not apply.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to established standards and predicate device performance. The type of "ground truth" used for demonstrating substantial equivalence includes:

    • Existing Predicate Device Performance: The primary ground truth is the established safety and efficacy profile of the predicate device (Stryker XCEL Anchor System, K023013). The new device is compared against this.
    • Industry Standards: Compliance with standards like ASTM 4169 (sterility), ANSI/AAMI/ISO 11135 (EO sterilization and residuals), and ISO-10993-1/G95-1 (biocompatibility) serve as ground truth for specific aspects of the device.

    8. The Sample Size for the Training Set

    There is no mention of a "training set" as this is not an artificial intelligence or machine learning device. The term "training set" is not applicable.

    9. How the Ground Truth for the Training Set was Established

    As there is no training set, this question is not applicable.