    Intended Use

    AOSept Clear Care Cleaning and Disinfecting Solution is indicated for use in cleaning, daily protein removal, disinfecting, and storing of soft (hydrophilic) or rigid gas permeable (fluoro silicone acrylate and silicone acrylate) contact lenses as recommended by your eye care practitioner.

    Device Description

    AOSept Clear Care Cleaning and Disinfecting Solution is an aqueous solution contains hydrogen peroxide 3% (stabilized with phosphonic acid), sodium chloride, a phosphate buffer system and a non-ionic surfactant.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the AOSEPT Clear Care Cleaning and Disinfecting Solution, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestDevice Performance (AOSept Clear Care)
    CompatibilityOptical Parameters (Paraperm EW RGP Lenses - Power, Base Curve, Diameter)Maintained within specifications after 30 cycles
    Physical Parameters (Paraperm EW RGP Lenses - Visual appearance, % transmittance, microscope)Maintained within specifications after 30 cycles
    Optical Parameters (Fluoroperm 30 RGP Lenses - Power, Base Curve, Diameter)Maintained within specifications after 30 cycles
    Physical Parameters (Fluoroperm 30 RGP Lenses - Visual appearance, % transmittance, microscope)Maintained within specifications after 30 cycles
    Compatibility with Focus Lens Drops (lubricating/rewetting)Successfully demonstrated compatibility
    Compatibility with Aquify Lens Drops (lubricating/rewetting)Successfully demonstrated compatibility
    MicrobiologyISO 14729 (FDA 510(k)) Stand Alone TestProvides effective disinfection based on criteria
    ISO 14729 (FDA 510(k)) Regimen TestProvides effective disinfection based on criteria
    BiocompatibilityFDA's May 1, 1997 Guidance for Industry; Premarket Notification 510(k) Guidance for Contact Lens Care ProductsMeets guidelines; non-toxic and biocompatible (from previous studies)
    Clinical Effectiveness (RGP Lenses)Lenses "clean" after 1-month wear91% reported none or slightly deposited
    Subject reporting lens needs cleaning at follow-up 27% reported lenses needed cleaning
    Usage of lubricating drops prior to insertion every day29% of AOSept Clear Care users
    Comfort grading (0-10 scale) at Follow-up 2Reported high ( > 8.5)
    (Comparative with Aquify vs. Focus Clerz)Aquify users showed higher comfort, fewer lens deposits, better wettability

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical Study for RGP Lenses): The text states, "Two thirds of subjects used the AOSept Clear Care, while and one-third used the control solution." The total number of subjects is not specified, only the ratio.
    • Data Provenance: The RGP lens clinical study was a "one-month prospective randomized trial." The country of origin of the data is not explicitly stated, but the submission is from CIBA Vision Corporation in Duluth, Georgia, USA, suggesting the study was likely conducted in the USA.
    • Other Studies (Compatibility, Microbiology, Biocompatibility): Sample sizes for these preclinical studies are not provided in this summary. The summary mentions these were "previously conducted" and "reviewed by FDA under two 510(k), K013512 and K003345."

    3. Number of Experts and Qualifications for Ground Truth

    This submission does not detail the use of experts to establish ground truth in the typical sense of a diagnostic medical device's performance claims.

    • For the clinical study, a "one-month prospective randomized, trial" was conducted. The "ground truth" for clinical effectiveness (e.g., lens cleanliness, comfort) would be based on subject reporting and potentially investigator observations. The qualifications of the clinical investigators are not specified.
    • For pre-clinical tests (Microbiology, Compatibility, Biocompatibility), the "ground truth" is defined by standardized test methods (e.g., ISO 14729) and regulatory guidelines (e.g., FDA guidance documents, GLP regulations). Performance against these standards serves as the benchmark, rather than expert consensus on individual cases.

    4. Adjudication Method

    • No specific adjudication method (like 2+1 or 3+1) is mentioned or appears applicable given the nature of the device (contact lens solution) and the types of studies performed (compatibility, microbiology, clinical trial for lens care). Clinical endpoints like "lens cleanliness" and "comfort" are typically assessed by the subject or clinical investigator, not through a multi-expert adjudication of a diagnostic finding.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This type of study is relevant for diagnostic imaging AI devices where human readers interpret medical images with and without AI assistance. This device is a contact lens care solution, and its evaluation methodologies do not involve MRMC studies.

    6. Standalone (Algorithm Only) Performance Study

    • Not applicable/done in the context of an algorithm. This product is a chemical solution. The "standalone performance" is demonstrated by the various in vitro and in vivo tests, such as the ISO 14729 Stand Alone Test and Regimen Test for antimicrobial activity, which evaluate the solution's properties independent of human interaction during the test itself, but these tests are not for an AI algorithm.

    7. Type of Ground Truth Used

    • Compatibility: Ground truth is established by objective measurements against pre-defined specifications (e.g., optical power, diameter, visual appearance, % transmittance) and compliance with regulatory guidelines.
    • Microbiology: Ground truth is established by adherence to in vitro antimicrobial efficacy standards as defined by ISO 14729 and FDA 510(k) performance criteria for contact lens disinfecting solutions, measured by reductions in microbial populations.
    • Biocompatibility: Ground truth is established by meeting the guidelines set forth in FDA guidance documents and demonstrating non-toxicity and biocompatibility through standard tests.
    • Clinical: Ground truth for effectiveness with RGP lenses is based on observed clinical outcomes and subject-reported data (e.g., lens cleanliness scores, comfort ratings, need for rewetting drops).

    8. Sample Size for the Training Set

    • Not applicable. This product is a chemical solution, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable. As there is no AI/ML algorithm, there is no training set or associated ground truth establishment process.