Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K020979

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-06-27

    (93 days)

    Product Code
    Regulation Number
    864.5400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Sysmex® CA-7000 is as a fully automated, computerized blood plasma coagulation analyzer for in vitro diagnostic use in clinical laboratories. The instrument uses citrated human plasma to perform coagulation tests.

    Device Description

    Not Found

    AI/ML Overview

    The Sysmex® Automated Coagulation Analyzer CA-7000 performed a clinical correlation study and a precision study to demonstrate its performance claims as being similar to predicate devices.

    1. Table of acceptance criteria and the reported device performance:

    The document describes two types of studies: "Summary of Method Comparison Studies Between CA-6000 or BCS" and "Summary of Precision Studies Sysmex® Automated Coagulation Analyzer CA-7000".

    Acceptance Criteria for Method Comparison Studies (Implied via Predicate Equivalence):
    The acceptance criteria for the method comparison studies are implicitly defined by demonstrating substantial equivalence to the predicate devices (Sysmex® Automated Coagulation Analyzer CA-6000 and Behring Coagulation System (BCS™ System)). This is primarily evaluated through the Coefficient of Correlation (r), which should be close to 1 (indicating a strong positive linear relationship), and visual inspection of the Regression Equation.

    Reported Device Performance (Method Comparison Studies):

    TestPredicate DeviceSample Number (n)Coefficient of Correlation (r)Regression Equation
    Prothrombin Time (Innovin®, seconds)CA-60001550.999Y = 0.97X + 0.38
    Prothrombin Time (Innovin®, INR)CA-60001550.999Y = 0.95X + 0.04
    Prothrombin Time (Thromborel® S, seconds)BCS1740.997Y = 1.09X - 1.54
    Prothrombin Time (Thromborel® S, % PT)BCS1680.984Y = 0.98X - 3.40
    Derived FibrinogenCA-60001040.991Y = 1.02X + 0.12
    Activated Partial Thromboplastin TimeCA-60001510.997Y = 1.03X - 0.01
    Fibrinogen (Clauss)CA-60001340.994Y = 0.91X + 0.05
    Factor VIIBCS1240.993Y = 1.14X - 2.66
    Factor VIIIBCS1430.977Y = 1.10X - 4.46
    Protein C, coagulometricBCS1390.994Y = 1.14X - 4.77
    Thrombin TimeCA-60003810.981Y = 0.78X + 2.72
    Batroxobin TimeBCS1690.987Y = 1.02X + 0.75
    Lupus Anticoagulant LA1 Screening ReagentCA-60001360.996Y = 0.92X - 1.25
    Lupus Anticoagulant LA2 Confirmation ReagentCA-60001360.953Y = 0.66X + 10.59
    Lupus Anticoagulant LA1/LA2 RatioCA-60001360.987Y = 1.00X - 0.08
    Antithrombin IIIBCS1660.997Y = 0.96X - 1.07
    Heparin, chromogenicBCS1150.982Y = 1.02X + 0.01
    Plasminogen, chromogenicBCS1420.994Y = 0.96X + 0.55
    a2- Antiplasmin, chromogenicBCS1440.982Y = 0.95X + 3.18
    Protein C, chromogenicBCS1560.996Y = 0.98X - 0.63
    Factor VIII, chromogenicBCS1360.990Y = 1.10X - 1.52

    Acceptance Criteria for Precision Studies:
    The acceptance criteria for precision are not explicitly stated as numerical thresholds (e.g., maximum allowable %CV). However, the intent is to demonstrate acceptable within-run, between-run, and total precision for each assay at different control levels. The reported %CV values should be within generally accepted clinical laboratory standards for these types of assays.

    Reported Device Performance (Precision Studies):

    Assay / Control LevelnMeanWithin Run %CVBetween Run %CVTotal %CV
    Prothrombin Time (Innovin®, seconds)
    Control Plasma N4011.90.40.20.4
    Ci-Trol® Control Level 34037.20.52.02.1
    Prothrombin Time (Innovin®, INR)
    Control Plasma N401.10.40.20.4
    Ci-Trol® Control Level 3403.40.52.12.1
    Prothrombin Time (Thromborel®S, seconds)
    Control Plasma N4012.00.40.50.6
    Control Plasma P4025.10.91.11.4
    Prothrombin Time (Thromborel®S, % of norm)
    Control Plasma N4091.90.60.60.8
    Control Plasma P4036.11.01.21.5
    Derived Fibrinogen (Innovin®, g/L)
    Control Plasma N401.95.42.25.5
    Path. plasmapool405.73.21.33.3
    Activated Partial Thromboplastin Time (Actin® FSL, seconds)
    Control Plasma N4029.60.70.30.7
    Ci-Trol® Control Level 34070.40.60.40.7
    Fibrinogen (Clauss) (Thrombin, g/L)
    Control Plasma N402.51.40.71.5
    Control Plasma P400.92.51.22.6
    Factor VII (Innovin®)
    Control Plasma N4099.72.43.44.1
    Control Plasma P4031.51.72.42.8
    Factor VIII (Actin® FSL)
    Control Plasma N40104.66.55.28.1
    Control Plasma P4033.85.93.96.8
    Protein C Coagulometric (Protein C Reagent, % norm)
    Control Plasma N40112.53.21.23.2
    Control Plasma P4040.94.32.44.6
    Thrombin Time (Test Thrombin Reagent, seconds)
    Control Plasma N4016.50.61.11.3
    Path. plasmapool4019.73.54.45.5
    Batroxobin Time (Batroxobin Reagent, seconds)
    Control Plasma N4020.11.10.61.2
    Path. plasmapool4058.31.20.71.4
    Lupus Anticoagulant (LA1 Screening, seconds)
    Control Plasma N4036.11.62.73.1
    LA Control High4090.61.51.52.1
    Lupus Anticoagulant (LA2 Confirmation, seconds)
    Control Plasma N4034.61.00.41.0
    LA Control High4040.41.00.91.3
    Lupus Anticoagulant (LA1 / LA2, ratio)
    Control Plasma N401.041.02.52.7
    LA Control High402.241.20.91.5
    Antithrombin III (Berichrom™ Antithrombin III (A))
    Control Plasma N4092.61.60.71.7
    Control Plasma P4031.71.80.71.9
    Heparin (Berichrom™ Heparin Reagent)
    Ci-Trol® Heparin Control Low400.067.36.19.2
    Ci-Trol® Heparin Control High400.221.92.12.8
    Plasminogen (Berichrom™ Plasminogen Reagent, % norm)
    Control Plasma N40101.41.21.41.8
    Control Plasma P4034.61.32.52.8
    α2-Antiplasmin (Berichrom™ α2-Antiplasmin, % norm)
    Control Plasma N4098.81.50.91.7
    Control Plasma P4035.43.21.83.5
    Protein C (Berichrom™ Protein C Reagent, % norm)
    Control Plasma N40101.91.70.81.8
    Control Plasma P4033.32.72.43.5
    Factor VIII Chromogenic (% norm)
    Control Plasma N40107.51.73.84.2
    Control Plasma P4029.61.42.42.8

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size for Test Set (Clinical Correlation Studies):
      • The sample sizes varied by assay, ranging from 104 (Derived Fibrinogen) to 381 (Thrombin Time). All specific sample sizes are listed in the table above under "Sample Number (n)".
    • Data Provenance:
      • The document states: "During those studies, specimens were evaluated from apparently healthy individuals and from patients with different pathological conditions which are expected to affect the results for a particular assay." This suggests a mix of prospective and retrospective samples, likely collected to represent a clinically relevant range.
      • There is no explicit mention of the country of origin for the clinical data. The manufacturing site is Sysmex Corporation, Kobe, Japan, and the distributor is Dade Behring Inc. in the US, but this does not specify the origin of the clinical samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The study compares the new device's readings against predicate devices, implying that the predicate devices' results serve as the reference or "ground truth" for the comparison. There is no mention of independent experts establishing a ground truth for the clinical correlation studies.

    4. Adjudication method for the test set:

    • This information is not applicable as the ground truth was established by comparing to predicate devices, not by expert consensus requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is not applicable. The device is an automated coagulation analyzer, which does not involve "human readers" or "AI assistance" in the context of interpreting medical images or data by human clinicians. It is a laboratory instrument that outputs numerical results.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the entire study focuses on the standalone performance of the Sysmex® Automated Coagulation Analyzer CA-7000. It is an automated system designed to operate "without human-in-the-loop performance" for the actual coagulation test execution and measurement. The results are generated directly by the analyzer.

    7. The type of ground truth used:

    • The "ground truth" for the method comparison studies was the results obtained from the predicate devices (Sysmex® Automated Coagulation Analyzer CA-6000 or Behring Coagulation System (BCS™ System)). This is a form of comparative (or reference method) ground truth where an established, cleared device's performance serves as the benchmark.
    • For precision studies, the concept of "ground truth" is different; it's about the device's internal consistency and reproducibility, which is assessed inherently by statistical analysis of repeated measurements on control samples.

    8. The sample size for the training set:

    • This information is not applicable. This device is a traditional laboratory instrument, not an AI/machine learning model that typically requires a "training set" in the common sense of AI development. It operates based on established chemical and optical principles.

    9. How the ground truth for the training set was established:

    • This information is not applicable for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1