510(k) Data Aggregation

    K Number
    K011665

    Validate with FDA (Live)

    Device Name
    BCT SYSTEM
    Manufacturer
    Date Cleared
    2001-08-09

    (71 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BCT™ System is an automated coaqulation analyzer for in vitro diagnostic use in clinical laboratories. The instrument performs the following parameters:

    • Activated Partial Thromboplastin Time (APTT)
    • Antithrombin IIIa
    • Batroxobin
    • D-dimer
    • Deficient Plasmas
    • Derived Fibrinogen
    • Fibrinogen
    • Heparin
    • Lupus Anticoagulants
    • Prothrombin Time (PT)
    • Plasminogen
    • Protein C-clotting
    • Protein C-chromogenic
    • Thrombin Time
    • von Willebrand factor
    Device Description

    The current BCT™ System was originally determined to be substantially equivalent as a coagulation analyzer in 510(k) Premarket Notification K955278. Subsequent to its clearance, the indications for use of the instrument were expanded through two additional Premarket Notifications, K001064 and K001067 for the addition of various analytes. The current BCT™ System was cleared to perform coagulometric, and chromogenic tests, such as the routine tests: prothrombin time, partial thromboplastin time, heparin, and fibrinogen, as well as the special tests: single factor determination, antithrombin IIIa, batroxobin, plasminogen, protein C, and D-dimer. The inclusion of the new testing parameter, lupus anticoagulants (LA), is the subject of this modification. The addition of the new proposed analyte to the instrument was accomplished without modification to the instrument principle, operation, hardware or instruction manual.

    AI/ML Overview

    This document describes the modification of the Dade Behring BCT™ System to include the new testing parameter, lupus anticoagulants (LA). The study demonstrates that the modified BCT™ system is substantially equivalent to the Sysmex® CA-6000 System for measuring Lupus Anticoagulants.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The acceptance criteria for this 510(k) modification are based on demonstrating substantial equivalence to a predicate device (Sysmex® CA-6000 System) through correlation and precision studies. While explicit numerical acceptance criteria (e.g., minimum correlation coefficient) are not stated, the reported performance data is presented to show strong correlation, suggesting the device meets internal benchmarks for equivalence.

    Correlation Study (Method Comparison):

    AssayAcceptance Criteria (Implied: Strong Correlation to Predicate)Reported Performance (Coefficient of Correlation (r))Regression Equation
    LA 1 Screening Reagent (seconds)R-value close to 1.00.968y=0.87x + 4.34
    LA 2 Confirmation Reagent (seconds)R-value close to 1.00.882y=0.76x + 5.46
    LA 1/LA 2 RatioR-value close to 1.00.965y=1.08x + 0.05
    LA 1 Screening Reagent (normalized)R-value close to 1.00.969y=0.89x + 0.11
    LA 2 Confirmation Reagent (normalized)R-value close to 1.00.882y=0.84x + 0.15
    LA 1/LA 2 Ratio (normalized)R-value close to 1.00.964y=0.97x + 0.04

    Precision Study (Implied Acceptance Criteria for %CV based on typical industry standards for coagulation assays):

    AssaySampleAcceptance Criteria (Implied: Low %CV)Reported Performance (Total %CV)
    LA 1 Screening Reagent (seconds)LA Control HighTypically < 5-10%2.0
    LA Control LowTypically < 5-10%2.2
    Control Plasma NTypically < 5-10%1.0
    LA 2 Confirmation Reagent (seconds)LA Control HighTypically < 5-10%1.1
    LA Control LowTypically < 5-10%1.6
    Control Plasma NTypically < 5-10%1.9
    LA 1/LA 2 RatioLA Control HighTypically < 5-10%1.4
    LA Control LowTypically < 5-10%1.3
    Control Plasma NTypically < 5-10%1.2

    Study Conclusion: The consistently high correlation coefficients (most above 0.96, with the lowest being 0.882) and low coefficients of variation (all below 2.2%) demonstrate that the BCT™ System's performance for LA testing is comparable to the predicate device and exhibits good precision. This data supports the claim of substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Sizes:
      • LA 1 Screening Reagent (seconds, normalized): 207 samples
      • LA 2 Confirmation Reagent (seconds, normalized): 208 samples
      • LA 1/LA 2 Ratio (normalized): 206 samples
    • Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It only states that the study "evaluated plasma samples."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This study is a method comparison and precision study for a diagnostic instrument's analytical performance, not a study involving diagnostic image interpretation or complex clinical decision-making by human experts. Therefore, the concept of "experts establishing ground truth" in the traditional sense (e.g., radiologists, pathologists) is not directly applicable.
    • The "ground truth" or reference method for the method comparison study is the Sysmex® CA-6000 System using Dade Behring LA 1 and LA 2 Reagents.
    • For the precision study, the "ground truth" is established by the known values of the control materials used.

    4. Adjudication Method for the Test Set:

    • Not applicable as this is an analytical performance study comparing instrument measurements, not a study requiring human adjudication of diagnostic findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance:

    • Not applicable. This is not an MRMC study and does not involve human readers or AI assistance in diagnostic interpretation. It is a comparison of two automated coagulation analyzers.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    • Yes, this study represents the standalone performance of the BCT™ System for measuring Lupus Anticoagulants. The data presented is the direct output of the instrument's analytical capabilities compared to a predicate device and control materials.

    7. The Type of Ground Truth Used:

    • Method Comparison Study: The ground truth is the measurements obtained from the predicate device, the Sysmex® CA-6000 System, which is a legally marketed device already established for these specific coagulation assays.
    • Precision Study: The ground truth is the assigned values of the control materials (LA Control High, LA Control Low, Control Plasma N).

    8. The Sample Size for the Training Set:

    • The document does not mention a "training set" as this is a traditional analytical validation study for a medical device, not a machine learning or AI algorithm development study. The instrument's method for measuring LA is established, and this study validates its performance.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no "training set" in the context of this type of analytical device validation. The "ground truth" for the performance evaluation (test set) is described in point 7.