Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN170081

    Validate with FDA (Live)

    Date Cleared
    2018-04-20

    (203 days)

    Product Code
    Regulation Number
    866.3378
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification and differentiation of microorganisms cultured from human specimens.

    The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.

    Device Description

    The MBT-CA System consists of the Microflex LT/SH mass spectrometer, reference library, kit reagents (US IVD HCCA, US IVD Bacterial Test Standard), US IVD 48 Spot Target or MBT Biotarget 96 US IVD plate, and software. The MALDI Biotyper CA System with closed safety covers is a Class 1 Laser product. With the safety cover opened it becomes a Class 4 Laser product.

    The MALDI Biotyper CA System reference library was established by analyzing the type strain from each claimed species combined with 4 to ~30 additional strains from the same species provided by clinical laboratories or commercial strain collections. Currently a total of 3029 strains (covering 334 species / groups with 294 bacteria plus 40 yeasts) are contained in the clinically validated MBT-CA library.

    Implementation methodology, construction parameters and quality assurance protocols use a standard operating protocol for generation of reference entries and all testing parameters are the same.

    MBT-CA microorganism identification is based on isolate MALDI spectra using Bruker reference libraries with a 1:1 comparison of unknown MALDI spectra against each single entry of a given reference library. During a single identification event, an unknown MALDI spectra is compared against each single reference entry producing individual log(score) results. This number of log(scores) is sorted based on their value and the highest one is used to generate the final result. The addition of new reference entries does not influence the already included entries. If no reference entries are removed within a library update the log(score) calculation remains unchanged for the same MALDI spectra.

    MALDI Biotyper CA System client software displays a user-interface which guides the user through the MALDI Biotyper CA System workflow. The MALDI Biotyper CA System client also interfaces to the flexControl software for automated acquisition of mass spectra on the microflex LT/SH instrument.

    The MALDI Biotyper CA System server communicates with the MALDI Biotyper CA System client and the MBT-DB server. It performs preprocessing on acquired spectra, and matches peak lists against the Main Spectrum (reference pattern, (MSP)) for matching and calculates the score value (log (score)).

    The MBT-DB server stores all information for the MALDI Biotyper CA System. The MBT-DB maintains spectra data (creation information and mass/intensity lists), project data (results of defined and executed runs), method data (parameter lists for spectra preprocessing and identification), user management data, reference patterns and other peak lists plus additional maintenance data.

    GTPS firmware communicates with the flexControl PC software, controls and monitors the vacuum, moves the sample carrier and performs the docking of the target plate, controls and monitors high voltages in the ion source, generates trigger signals, and monitors instrument status.

    The flexControl acquisition software communicates with the MALDI Biotyper CA System client, loads automatic run jobs, communicates with the GTPS firmware, communicates with the laser in the microflex LT/SH instrument, sets the acquisition parameters in the digitizer and reads the acquired data from the digitizer, performs automated data acquisition, evaluates acquired spectra, adjusts the laser power during automatic data acquisition, performs a re-calibration of the time-of-flight to mass transformation, stored acquired spectra on disk and performs source cleaning. The flexControl software does not display a user interface.

    The optional Honeywell (Hyperion 1300g) Barcode Reader USB cable is connected to the MALDI Biotyper CA System computer. The barcode reader scans the unique ten-digit target ID which appears in the Target ID box on the target plate. After the target ID has been entered, the a new Run page opens and the ten-digit target ID appears as the Plate ID and is appended to the Run name. Sample identifications are entered into the computer corresponding to the target plate position for that run.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the MALDI Biotyper CA System, extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes performance for the addition of Candida auris to the existing MBT-CA reference library, rather than a full de novo clearance study of the entire device. Therefore, the "acceptance criteria" here refer to the performance required for the inclusion of this new organism into the established system.

    Criteria (for C. auris identification)Reported Device Performance (C. auris identification)
    High confidence organism ID (log(score) $\ge$ 2.0)22 out of 22 (100%) correctly identified
    Low confidence organism ID (log(score) $\ge$ 1.7 - <2.0)0
    Incorrect MBT-CA ( $\ge$ 1.7) / No ID0
    Overall Performance for C. auris100% successful identification

    Additionally, for analytical specificity (cross-identification studies):

    Criteria (Analytical Specificity)Reported Device Performance
    No cross-identification for existing claimed organisms after C. auris library updateIn silico: 100% identical results for 6822 log(scores) of a subset of stored spectra. Wet testing of cleared organisms: No influence of new C. auris reference entries and no cross-identification observed for 360 spectra/log(scores) from 10 cleared species.
    No cross-identification for Research Use Only (RUO) organismsWet testing of RUO organisms: None of the RUO organisms were identified with the MBT-CA libraries after the C. auris update. No influence of new C. auris reference entries and no cross-identification observed for 9 RUO species.

    2. Sample Size Used for the Test Set and Data Provenance

    • For C. auris performance evaluation:

      • Number of C. auris isolates: 28 isolates for performance evaluation, 22 of which were used for generating the truth tables (excluding 6 strains used for reference entries).
      • Total C. auris spectra: 22 strains * 3 sample preparations (DT, eDT, Ext) * 8 spots = 528 spectra for truth table counting. (The document states 888 spectra for identification across (37 strains * 3 sample preparations * 8 spots), where 37 strains includes 28 C. auris and 9 related yeasts).
      • Data Provenance: CDC & FDA Antibiotic Resistance Isolate Bank (USA), isolates from (b) (4) (Europe/US likely), and field strains (unspecified location).
    • For Analytical Specificity (wet testing of claimed organisms):

      • Sample Size: 10 "cleared species" (e.g., E. coli, K. pneumoniae, S. aureus, C. albicans, C. glabrata).
      • Total Spectra: 10 species * 3 sample prep techniques * 3 replicates * 2 instruments = 360 spectra/log(scores).
      • Data Provenance: Not explicitly stated but inferred to be well-characterized strains from resource centers (e.g., ATCC).
    • For Analytical Specificity (wet testing of RUO organisms):

      • Sample Size: 9 RUO species.
      • Total Spectra: Not explicitly stated (marked as (0) pectra / log(scores) which could be a typo or redaction for a large number). Each species tested with 3 replicates, 3 sample prep techniques, on 2 instruments.
      • Data Provenance: Not explicitly stated but inferred to be well-characterized strains from resource centers.
    • For Analytical Specificity (in silico evaluation):

      • Sample Size: 6822 log(scores) from a subset of stored spectra for previously claimed organisms.
      • Data Provenance: Not explicitly stated but refers to historical data from previous studies (K130831, K142677, K163536).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts or their qualifications for establishing the ground truth for the Candida auris test set. However, it indicates that the ITS sequence was determined for the 6 strains used for the new reference library entries, and implies that the remaining C. auris isolates were well-characterized from reputable sources (CDC & FDA, other resource centers). The "Reference Algorithm" is used as the comparative method for C. auris performance. Given this is a microbiology identification device, ground truth is typically established by established phenotypic, genotypic (e.g., DNA sequencing), or biochemical methods, often confirmed by expert microbiologists.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (like 2+1 or 3+1) involving multiple human readers for setting the ground truth for the test set. Instead, it relies on:

    • Established sources for isolates (CDC & FDA Antibiotic Resistance Isolate Bank, other resource centers).
    • ITS sequence determination for reference strains.
    • "Reference Algorithm" as the gold standard for comparison (likely a combination of genotypic and phenotypic characterization).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is described in this document. The device is for automated microorganism identification, so human reader improvement with AI assistance is not directly applicable in the same way it would be for image interpretation tasks. The comparison is between the device's automated identification and a reference method.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a standalone performance study was done. The entire evaluation focuses on the performance of the MALDI Biotyper CA System (algorithm plus instrument) in identifying microorganisms without human-in-the-loop during the identification process itself. Users "aid in the diagnosis" using the output, but the identification is automated. The log(score) output directly reflects the algorithm's confidence in its standalone identification.

    7. Type of Ground Truth Used

    The ground truth used for the Candida auris evaluation was based on:

    • Genetic Sequencing: ITS sequence determination for the 6 C. auris strains used for reference library generation.
    • Established Reference Materials: Isolates obtained from reputable sources like the CDC & FDA Antibiotic Resistance Isolate Bank, which are generally well-characterized by various methods (genotypic and phenotypic). The document refers to it as the "Reference Algorithm" in performance tables.

    For the analytical specificity studies, "well-characterized set of cleared species" and "well-characterized set of RUO species" from resource centers implies established identification through various, often molecular, methods.

    8. Sample Size for the Training Set

    The document describes the content of the reference library (which serves as the training/reference data for the algorithm), but does not specifically refer to it as a "training set" in the context of machine learning model development. Instead, it's a reference database.

    • Total Reference Entries (database): 3029 strains covering 334 species/groups (294 bacteria + 40 yeasts).
    • C. auris specific reference entries: 6 strains.
    • How reference entries are built: Each entry is based on analyzing the type strain from each claimed species combined with 4 to ~30 additional strains from the same species.

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the reference library (training set) entries was established by:

    • Analyzing type strains: For each claimed species.
    • Analyzing additional strains: 4 to ~30 additional strains per species from clinical laboratories or commercial strain collections.
    • Standard Operating Protocol: Implementation methodology, construction parameters, and quality assurance protocols use a standard operating protocol for the generation of reference entries.
    • DNA Sequencing: For the newly added C. auris strains, ITS sequencing was performed for all 6 strains used for the new reference library entries. It's also mentioned as a requirement for "all type strains and at least 20% of the non-type strains of a species detected by the device" to be characterized by DNA sequence analysis (as part of special controls for this device type). This implies that a significant portion of the existing library entries also have DNA sequencing as part of their ground truth establishment.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1