K Number
DEN170081

Validate with FDA (Live)

Date Cleared
2018-04-20

(203 days)

Product Code
Regulation Number
866.3378
Type
Direct
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification and differentiation of microorganisms cultured from human specimens.

The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.

Device Description

The MBT-CA System consists of the Microflex LT/SH mass spectrometer, reference library, kit reagents (US IVD HCCA, US IVD Bacterial Test Standard), US IVD 48 Spot Target or MBT Biotarget 96 US IVD plate, and software. The MALDI Biotyper CA System with closed safety covers is a Class 1 Laser product. With the safety cover opened it becomes a Class 4 Laser product.

The MALDI Biotyper CA System reference library was established by analyzing the type strain from each claimed species combined with 4 to ~30 additional strains from the same species provided by clinical laboratories or commercial strain collections. Currently a total of 3029 strains (covering 334 species / groups with 294 bacteria plus 40 yeasts) are contained in the clinically validated MBT-CA library.

Implementation methodology, construction parameters and quality assurance protocols use a standard operating protocol for generation of reference entries and all testing parameters are the same.

MBT-CA microorganism identification is based on isolate MALDI spectra using Bruker reference libraries with a 1:1 comparison of unknown MALDI spectra against each single entry of a given reference library. During a single identification event, an unknown MALDI spectra is compared against each single reference entry producing individual log(score) results. This number of log(scores) is sorted based on their value and the highest one is used to generate the final result. The addition of new reference entries does not influence the already included entries. If no reference entries are removed within a library update the log(score) calculation remains unchanged for the same MALDI spectra.

MALDI Biotyper CA System client software displays a user-interface which guides the user through the MALDI Biotyper CA System workflow. The MALDI Biotyper CA System client also interfaces to the flexControl software for automated acquisition of mass spectra on the microflex LT/SH instrument.

The MALDI Biotyper CA System server communicates with the MALDI Biotyper CA System client and the MBT-DB server. It performs preprocessing on acquired spectra, and matches peak lists against the Main Spectrum (reference pattern, (MSP)) for matching and calculates the score value (log (score)).

The MBT-DB server stores all information for the MALDI Biotyper CA System. The MBT-DB maintains spectra data (creation information and mass/intensity lists), project data (results of defined and executed runs), method data (parameter lists for spectra preprocessing and identification), user management data, reference patterns and other peak lists plus additional maintenance data.

GTPS firmware communicates with the flexControl PC software, controls and monitors the vacuum, moves the sample carrier and performs the docking of the target plate, controls and monitors high voltages in the ion source, generates trigger signals, and monitors instrument status.

The flexControl acquisition software communicates with the MALDI Biotyper CA System client, loads automatic run jobs, communicates with the GTPS firmware, communicates with the laser in the microflex LT/SH instrument, sets the acquisition parameters in the digitizer and reads the acquired data from the digitizer, performs automated data acquisition, evaluates acquired spectra, adjusts the laser power during automatic data acquisition, performs a re-calibration of the time-of-flight to mass transformation, stored acquired spectra on disk and performs source cleaning. The flexControl software does not display a user interface.

The optional Honeywell (Hyperion 1300g) Barcode Reader USB cable is connected to the MALDI Biotyper CA System computer. The barcode reader scans the unique ten-digit target ID which appears in the Target ID box on the target plate. After the target ID has been entered, the a new Run page opens and the ten-digit target ID appears as the Plate ID and is appended to the Run name. Sample identifications are entered into the computer corresponding to the target plate position for that run.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the MALDI Biotyper CA System, extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes performance for the addition of Candida auris to the existing MBT-CA reference library, rather than a full de novo clearance study of the entire device. Therefore, the "acceptance criteria" here refer to the performance required for the inclusion of this new organism into the established system.

Criteria (for C. auris identification)Reported Device Performance (C. auris identification)
High confidence organism ID (log(score) $\ge$ 2.0)22 out of 22 (100%) correctly identified
Low confidence organism ID (log(score) $\ge$ 1.7 - <2.0)0
Incorrect MBT-CA ( $\ge$ 1.7) / No ID0
Overall Performance for C. auris100% successful identification

Additionally, for analytical specificity (cross-identification studies):

Criteria (Analytical Specificity)Reported Device Performance
No cross-identification for existing claimed organisms after C. auris library updateIn silico: 100% identical results for 6822 log(scores) of a subset of stored spectra. Wet testing of cleared organisms: No influence of new C. auris reference entries and no cross-identification observed for 360 spectra/log(scores) from 10 cleared species.
No cross-identification for Research Use Only (RUO) organismsWet testing of RUO organisms: None of the RUO organisms were identified with the MBT-CA libraries after the C. auris update. No influence of new C. auris reference entries and no cross-identification observed for 9 RUO species.

2. Sample Size Used for the Test Set and Data Provenance

  • For C. auris performance evaluation:

    • Number of C. auris isolates: 28 isolates for performance evaluation, 22 of which were used for generating the truth tables (excluding 6 strains used for reference entries).
    • Total C. auris spectra: 22 strains * 3 sample preparations (DT, eDT, Ext) * 8 spots = 528 spectra for truth table counting. (The document states 888 spectra for identification across (37 strains * 3 sample preparations * 8 spots), where 37 strains includes 28 C. auris and 9 related yeasts).
    • Data Provenance: CDC & FDA Antibiotic Resistance Isolate Bank (USA), isolates from (b) (4) (Europe/US likely), and field strains (unspecified location).
  • For Analytical Specificity (wet testing of claimed organisms):

    • Sample Size: 10 "cleared species" (e.g., E. coli, K. pneumoniae, S. aureus, C. albicans, C. glabrata).
    • Total Spectra: 10 species * 3 sample prep techniques * 3 replicates * 2 instruments = 360 spectra/log(scores).
    • Data Provenance: Not explicitly stated but inferred to be well-characterized strains from resource centers (e.g., ATCC).
  • For Analytical Specificity (wet testing of RUO organisms):

    • Sample Size: 9 RUO species.
    • Total Spectra: Not explicitly stated (marked as (0) pectra / log(scores) which could be a typo or redaction for a large number). Each species tested with 3 replicates, 3 sample prep techniques, on 2 instruments.
    • Data Provenance: Not explicitly stated but inferred to be well-characterized strains from resource centers.
  • For Analytical Specificity (in silico evaluation):

    • Sample Size: 6822 log(scores) from a subset of stored spectra for previously claimed organisms.
    • Data Provenance: Not explicitly stated but refers to historical data from previous studies (K130831, K142677, K163536).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts or their qualifications for establishing the ground truth for the Candida auris test set. However, it indicates that the ITS sequence was determined for the 6 strains used for the new reference library entries, and implies that the remaining C. auris isolates were well-characterized from reputable sources (CDC & FDA, other resource centers). The "Reference Algorithm" is used as the comparative method for C. auris performance. Given this is a microbiology identification device, ground truth is typically established by established phenotypic, genotypic (e.g., DNA sequencing), or biochemical methods, often confirmed by expert microbiologists.

4. Adjudication Method for the Test Set

The document does not describe an explicit adjudication method (like 2+1 or 3+1) involving multiple human readers for setting the ground truth for the test set. Instead, it relies on:

  • Established sources for isolates (CDC & FDA Antibiotic Resistance Isolate Bank, other resource centers).
  • ITS sequence determination for reference strains.
  • "Reference Algorithm" as the gold standard for comparison (likely a combination of genotypic and phenotypic characterization).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is described in this document. The device is for automated microorganism identification, so human reader improvement with AI assistance is not directly applicable in the same way it would be for image interpretation tasks. The comparison is between the device's automated identification and a reference method.

6. Standalone Performance Study (Algorithm Only)

Yes, a standalone performance study was done. The entire evaluation focuses on the performance of the MALDI Biotyper CA System (algorithm plus instrument) in identifying microorganisms without human-in-the-loop during the identification process itself. Users "aid in the diagnosis" using the output, but the identification is automated. The log(score) output directly reflects the algorithm's confidence in its standalone identification.

7. Type of Ground Truth Used

The ground truth used for the Candida auris evaluation was based on:

  • Genetic Sequencing: ITS sequence determination for the 6 C. auris strains used for reference library generation.
  • Established Reference Materials: Isolates obtained from reputable sources like the CDC & FDA Antibiotic Resistance Isolate Bank, which are generally well-characterized by various methods (genotypic and phenotypic). The document refers to it as the "Reference Algorithm" in performance tables.

For the analytical specificity studies, "well-characterized set of cleared species" and "well-characterized set of RUO species" from resource centers implies established identification through various, often molecular, methods.

8. Sample Size for the Training Set

The document describes the content of the reference library (which serves as the training/reference data for the algorithm), but does not specifically refer to it as a "training set" in the context of machine learning model development. Instead, it's a reference database.

  • Total Reference Entries (database): 3029 strains covering 334 species/groups (294 bacteria + 40 yeasts).
  • C. auris specific reference entries: 6 strains.
  • How reference entries are built: Each entry is based on analyzing the type strain from each claimed species combined with 4 to ~30 additional strains from the same species.

9. How the Ground Truth for the Training Set Was Established

The ground truth for the reference library (training set) entries was established by:

  • Analyzing type strains: For each claimed species.
  • Analyzing additional strains: 4 to ~30 additional strains per species from clinical laboratories or commercial strain collections.
  • Standard Operating Protocol: Implementation methodology, construction parameters, and quality assurance protocols use a standard operating protocol for the generation of reference entries.
  • DNA Sequencing: For the newly added C. auris strains, ITS sequencing was performed for all 6 strains used for the new reference library entries. It's also mentioned as a requirement for "all type strains and at least 20% of the non-type strains of a species detected by the device" to be characterized by DNA sequence analysis (as part of special controls for this device type). This implies that a significant portion of the existing library entries also have DNA sequencing as part of their ground truth establishment.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR

DECISION MEMORANDUM

A. DEN Number:

DEN 170081

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the MALDI Biotyper CA System

C. Measurands:

See Indications for Use

D. Type of Test:

A mass spectrometer system for clinical use for the identification and differentiation of microorganisms is a qualitative in vitro diagnostic device intended for the identification and differentiation of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer and a spectral database. The system acquires, processes and analyzes spectra to generate data specific to a microorganism(s). The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.

E. Applicant:

Bruker Daltonik GmbH

F. Proprietary and Established Names:

Trade Name: MALDI Biotyper CA System

Common Names: MALDI Biotyper CA (MBT-CA) System, MBT-CA

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.3378
    1. Classification:

{1}------------------------------------------------

Class II (Special Controls)

    1. Product code(s):
      QBN
    1. Panel:
      83- Microbiology

H. Indications for Use:

    1. Indications for use:
      The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time of flight (MALDI-TOF) for the identification and differentiation of microorganisms cultured from human specimens.

The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.

Bacteria:
Abiotrophia defectivaAchromobacter xylosoxidans
Acinetobacter baumannii / nosocomialis groupAcinetobacter calcoaceticus
Acinetobacter haemolyticusAcinetobacter johnsonii
Acinetobacter juniiAcinetobacter lwoffii
Acinetobacter pittiiAcinetobacter radioresistens
Acinetobacter ursingiiActinomyces europaeus
Actinomyces funkeiActinomyces graevenitzii
Actinomyces hyovaginalisActinomyces meyeri
Actinomyces neuiiActinomyces odontolyticus
Actinomyces orisActinomyces radingae
Actinomyces turicensisActinomyces urogenitalis
Actinotignum schaalii groupAerococcus sanguinicola
Aerococcus urinaeAerococcus viridans
Aeromonas salmonicidaAeromonas hydrophila / caviae group
Aggregatibacter actinomycetemcomitansAggregatibacter aphrophilus
Aggregatibacter segnisAlcaligenes faecalis
Alloiococcus otitisAlloscardovia omnicolens
Bacteria:
Anaerococcus murdochiiAnaerococcus vaginalis
Arthrobacter cumminsiiBacteroides caccae
Bacteroides fragilisBacteroides nordii
Bacteroides ovatus groupBacteroides pyogenes
Bacteroides salyersiaeBacteroides stercoris group
Bacteroides thetaiotaomicron groupBacteroides uniformis
Bacteroides vulgatus groupBifidobacterium breve
Bordetella pertussis bronchiseptica parapertussisBordetella hinzii
Brevibacterium caseiBrevundimonas diminuta group
Burkholderia cepacia complexBurkholderia gladioli
Burkholderia multivoransCampylobacter coli
Campylobacter jejuniCampylobacter ureolyticus
Capnocytophaga ochraceaCapnocytophaga sputigena
Chryseobacterium gleumChryseobacterium indologenes
Citrobacter amalonaticus complexCitrobacter freundii complex
Citrobacter koseriClostridium beijerinckii
Clostridium bifermentansClostridium butyricum
Clostridium clostridioforme groupClostridium difficile
Clostridium innocuumClostridium paraputrificum
Clostridium perfringensClostridium ramosum
Clostridium septicumClostridium sordellii
Clostridium sporogenes /Clostridium botulimim (group I)Clostridium tertium
Corynebacterium accolensCorynebacterium afermentans group
Corynebacterium amycolatumCorynebacterium aurimucosum group
Corynebacterium bovisCorynebacterium coyleae
Corynebacterium diphtheriaeCorynebacterium frenevi
Corynebacterium glucuronolyticumCorynebacterium glutamicum
Corynebacterium jeikeiumCorynebacterium kroppenstedtii
Corynebacterium macginleyiCorynebacterium minutissimum
Corynebacterium mucifaciens /ureicelerivorans groupCorynebacterium propinquum
Corynebacterium pseudodiphtheriticumCorynebacterium pseudotuberculosis
Bacteria:
Corynebacterium resistensCorynebacterium riegelii
Corynebacterium striatum groupCorynebacterium tuberculostearicum
Corynebacterium ulceransCorynebacterium urealyticum
Corynebacterium xerosisCronobacter sakazakii group
Cupriavidus pauculus groupDelftia acidovorans group
Dermabacter hominisDermacoccus nishinomiyaensis
Edwardsiella tardaEikenella corrodens
Elizabethkingia meningoseptica groupEnterobacter aerogenes
Enterobacter amnigenusEnterobacter cloacae complex
Enterococcus aviumEnterococcus casseliflavus
Enterococcus duransEnterococcus faecalis
Enterococcus faeciumEnterococcus gallinarum
Enterococcus hiraeEnterococcus mundtii
Enterococcus raffinosusEscherichia coli
Escherichia hermanniiEscherichia vulneris
Ewingella americanaFacklamia hominis
Finegoldia magnaFluoribacter bozemanae
Fusobacterium canifelinumFusobacterium necrophorum
Fusobacterium nucleatumGardnerella vaginalis
Gemella haemolysansGemella morbillorum
Gemella sanguinisGramulicatella adiacens
Haemophilus haemolyticusHaemophilus influenzae
Haemophilus parahaemolyticus groupHaemophilus parainfluenzae
Hafnia alveiHelcococcus kunzii
Kingella denitrificansKingella kingae
Klebsiella oxytoca RaoultellaornithinolyticaKlebsiella pneumoniae
Klebsiella variicolaKocuria kristinae
Kytococcus sedentariusLactobacillus gasseri
Lactobacillus jenseniiLactobacillus rhamnosus
Lactococcus garvieaeLactococcus lactis
Leclercia adecarboxylataLegionella longbeachae
Bacteria:
Legionella pneumophilaLeuconostoc citreum
Leuconostoc mesenteroidesLeuconostoc pseudomesenteroides
Listeria monocytogenesMacrococcus caseolyticus
Mannheimia haemolytica groupMicrococcus luteus
Micrococcus lylaeMobiluncus curtisii
Moraxella sg Branhamella catarrhalis*Moraxella sg Moraxella osloensis*Moraxella sg Moraxella nonliquefaciens*Morganella morganii
Myroides odoratimimusMyroides odoratus
Neisseria bacilliformisNeisseria cinerea
Neisseria elongataNeisseria flavescens / subflava group
Neisseria gonorrhoeaeNeisseria lactamica
Neisseria meningitidisNeisseria sicca group
Neisseria weaveriNocardia brasiliensis
Nocardia cyriacigeorgicaNocardia farcinica group
Nocardia novaNocardia otitidiscaviarum
Ochrobactrum anthropiOligella ureolytica
Oligella urethralisPantoea agglomerans
Parabacteroides distasonisParabacteroides goldsteinii
Parabacteroides johnsonii / merdae groupParvimonas micra
Pasteurella multocidaPediococcus acidilactici
Pediococcus pentosaceusPeptoniphilus harei group
Peptostreptococcus anaerobiusPlesiomonas shigelloides
Pluralibacter gergoviaePorphyromonas gingivalis
Porphyromonas someraePrevotella bivia
Prevotella buccaePrevotella denticola
Prevotella intermediaPrevotella melaninogenica
Propionibacterium acnesProteus mirabilis
Proteus vulgaris groupProvidencia rettgeri
Providencia stuartiiPseudomonas aeruginosa
Pseudomonas fluorescens groupPseudomonas oryzihabitans
Pseudomonas putida groupPseudomonas stutzeri
Ralstonia pickettiiRhizobium radiobacter
Bacteria:
Rothia aeriaRothia dentocariosa
Rothia mucilaginosaSalmonella sp**
Serratia fonticolaSerratia liquefaciens
Serratia marcescensSerratia odorifera
Serratia plymuthicaSerratia rubidaea
Sphingobacterium multivorumSphingobacterium spiritivorum
Sphingomonas paucimobilis groupStaphylococcus aureus
Staphylococcus auricularisStaphylococcus capitis
Staphylococcus capraeStaphylococcus carnosus
Staphylococcus cohniiStaphylococcus delphini
Staphylococcus epidermidisStaphylococcus equorum
Staphylococcus felisStaphylococcus haemolyticus
Staphylococcus hominisStaphylococcus intermedius
Staphylococcus lentusStaphylococcus lugdunensis
Staphylococcus pasteuriStaphylococcus pettenkoferi
Staphylococcus pseudintermediusStaphylococcus saccharolyticus
Staphylococcus saprophyticusStaphylococcus schleiferi
Staphylococcus sciuriStaphylococcus simulans
Staphylococcus vitulinusStaphylococcus warneri
Staphylococcus xylosusStenotrophomonas maltophilia
Streptococcus agalactiaeStreptococcus anginosus
Streptococcus canisStreptococcus constellatus
Streptococcus dysgalactiaeStreptococcus equi
Streptococcus gallolyticusStreptococcus gordonii
Streptococcus intermediusStreptococcus lutetiensis
Streptococcus mitis / oralis groupStreptococcus mutans
Streptococcus parasanguinisStreptococcus pneumoniae
Streptococcus pyogenesStreptococcus salivarius / vestibularis group
Streptococcus sanguinisStreptococcus sobrinus
Streptococcus thermophilusSutterella wadsworthensis
Trueperella bernardiaeTuricella otitidis
Vagococcus fluvialisVeillonella parvula group

{2}------------------------------------------------

{3}------------------------------------------------

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

Bacteria:

Vibrio parahaemolyticus

Weeksella virosa

Yersinia frederiksenii

Yersinia kristensenii

  • = subgenus

sp** = species

  • Yeasts:
Candida albicansCandida auris
Candida boidiniiCandida dubliniensis
Candida duobushaemuloniiCandida famata
Candida glabrataCandida guilliermondii
Candida haemulonisCandida inconspicua
Candida intermediaCandida kefyr
Candida kruseiCandida lambica
Candida lipolyticaCandida lusitaniae
Candida metapsilosisCandida norvegensis
Candida orthopsilosisCandida parapsilosis
Candida pararugosaCandida pelliculosa
Candida tropicalisCandida valida
Candida zeylanoidesCryptococcus gattii
Cryptococcus neoformans var grubii*Cryptococcus neoformans var neoformans*
Cyberlindnera jadiniiGeotrichum candidum
Geotrichum capitatumKloeckera apiculata
Malassezia furfurMalassezia pachydermatis
Pichia ohmeriRhodotorula mucilaginosa
Saccharomyces cerevisiaeTrichosporon asahii
Trichosporon inkinTrichosporon mucoides group
* = variety

Vibrio vulnificus

Yersinia enterocolitica

Yersinia pseudotuberculosis

Yersinia intermedia

    1. Special conditions for use statement(s):
      For in vitro diagnostic use only

{7}------------------------------------------------

The MALDI Biotyper CA System is for prescription use only.

Special instrument requirements:

Microflex LT/SH mass spectrometer

Database: MALDI Biotyper for Clinical Applications (MBT-CA)

Software:

  • MBT-CA System Software Package:
  • MBT-CA System client software displaying the user interface ●
  • MBT-CA System Server ●
  • MBT-CA System DB Server ●
  • flexControl Software Package (GTPS firmware, flexControl acquisition software) ●

Honeywell (Hyperion 1300g) Barcode Reader (optional)

I. Device Description:

  • The MBT-CA System consists of the Microflex LT/SH mass spectrometer, reference library, ● kit reagents (US IVD HCCA, US IVD Bacterial Test Standard), US IVD 48 Spot Target or MBT Biotarget 96 US IVD plate, and software. The MALDI Biotyper CA System with closed safety covers is a Class 1 Laser product. With the safety cover opened it becomes a Class 4 Laser product.
  • The MALDI Biotyper CA System reference library was established by analyzing the type . strain from each claimed species combined with 4 to ~30 additional strains from the same species provided by clinical laboratories or commercial strain collections. Currently a total
  • of 3029 strains (covering 334 species / groups with 294 bacteria plus 40 yeasts) are P contained in the clinically validated MBT-CA library.
  • Implementation methodology, construction parameters and quality assurance protocols use a standard operating protocol for generation of reference entries and all testing parameters are the same.
  • MBT-CA microorganism identification is based on isolate MALDI spectra using Bruker ● reference libraries with a 1:1 comparison of unknown MALDI spectra against each single entry of a given reference library. During a single identification event, an unknown MALDI spectra is compared against each single reference entry producing (b) (4) individual log(score) results. This number of log(scores) is sorted based on their value and the highest one is used to generate the final result. The addition of new reference entries does not influence the already included entries. If no reference entries are removed within a library update the log(score) calculation remains unchanged for the same MALDI spectra.
  • MALDI Biotyper CA System client software displays a user-interface which guides the user through the MALDI Biotyper CA System workflow. The MALDI Biotyper CA System client also interfaces to the flexControl software for automated acquisition of mass spectra on the microflex LT/SH instrument.
  • The MALDI Biotyper CA System server communicates with the MALDI Biotyper CA ● System client and the MBT-DB server. It performs preprocessing on acquired spectra, and

{8}------------------------------------------------

matches peak lists against the Main Spectrum (reference pattern, (MSP)) for matching and calculates the score value (log (score)).

  • The MBT-DB server stores all information for the MALDI Biotyper CA System. The MBT-DB maintains spectra data (creation information and mass/intensity lists), project data (results of defined and executed runs), method data (parameter lists for spectra preprocessing and identification), user management data, reference patterns and other peak lists plus additional maintenance data.
  • GTPS firmware communicates with the flexControl PC software, controls and monitors the vacuum, moves the sample carrier and performs the docking of the target plate, controls and monitors high voltages in the ion source, generates trigger signals, and monitors instrument status.
  • The flexControl acquisition software communicates with the MALDI Biotyper CA System client, loads automatic run jobs, communicates with the GTPS firmware, communicates with the laser in the microflex LT/SH instrument, sets the acquisition parameters in the digitizer and reads the acquired data from the digitizer, performs automated data acquisition, evaluates acquired spectra, adjusts the laser power during automatic data acquisition, performs a re-calibration of the time-of-flight to mass transformation, stored acquired spectra on disk and performs source cleaning. The flexControl software does not display a user interface.
  • The optional Honeywell (Hyperion 1300g) Barcode Reader USB cable is connected to the MALDI Biotyper CA System computer. The barcode reader scans the unique ten-digit target ID which appears in the Target ID box on the target plate. After the target ID has been entered, the a new Run page opens and the ten-digit target ID appears as the Plate ID and is appended to the Run name. Sample identifications are entered into the computer corresponding to the target plate position for that run.

Required Materials Supplied by Bruker

  • US IVD 48 Spot Target [P/N: 8604532] ●
  • US IVD BTS (Bacterial Test Standard) [P/N: 8604530] ●
  • US IVD HCCA portioned [P/N: 8604531] ●

Required Materials that are Not Supplied by Bruker

The following solvents and chemicals are not supplied by Bruker but are required to perform the analysis. For best results, use freshly prepared solutions and chemicals of (b) (4) or MALDI (b) (4) solvents). compatible grade (for example,

  • Standard Solvent (acetonitrile 50%, water 47.5% and trifluoroacetic acid 2.5%) ● [Vendor: (b) (4) or equivalent]
  • Acetonitrile ●
  • (b) (4) water ●
  • Formic acid (FA)
  • (b) (4) Ethanol (EtOH) ●
  • Trifluoroacetic acid (TFA)
  • Sterile Colony Transfer Device ●
  • . Sterile (b) inoculation loops
  • (b) (4) pipette tips ●
  • Suitable pipettes for volumes from ●

(b) (4)

(b) (4)

{9}------------------------------------------------

Image /page/9/Figure/0 description: The image shows a list of laboratory equipment. The list includes plastic tubes, screw-cap micro tubes, a bench-top microcentrifuge, a vortex mixer, and standard laboratory equipment. Some of the text in the image is redacted.

J. Standard/Guidance Document Referenced (if applicable):

Not applicable

K. Test Principle:

Organisms to be identified with the MALDI Biotyper CA System are isolated using the appropriate isolation media. Users are instructed to first test the organism using the direct transfer technique (unless specified to perform extraction in the product labeling). If results are less than <2.0 log(score), users are then directed to perform extraction (eDT or EXT) procedure.

  • . Direct Transfer (DT): An individual colony from an overnight subculture plate is transferred to a selected position on an US IVD 48 Spot Target (target). The target is air dried and US IVD HCCA portioned (matrix) is added. The standard solvent (50% acetonitrile / 47.5% H2O / 2.5% trifluoroacetic acid) in the matrix solution extracts proteins (mainly ribosomal proteins, which are present in high concentration) from the microorganisms. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MALDI Biotyper CA System.
  • . extended Direct Transfer (eDT): An isolated colony of bacteria or yeast is transferred as a thin film directly onto a sample position on a cleaned target. The sample spot is overlayed with 1 uL 70% aqueous formic acid directly on the target plate and allowed to dry at room temperature. Afterwards the spots are overlaved with US IVD HCCA portioned (matrix) and allowed to dry. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MALDI Biotyper CA System.
  • Extraction Procedure (Ext): For this purpose, isolated colonies from the overnight subculture . plate are extracted using an ethanol/formic acid procedure. Afterwards they are transferred to the target plate and allowed to dry. Afterwards the spots are overlaid with US IVD HCCA portioned (matrix) and allowed to dry. When dried matrix crystallizes, the inoculated target is ready to be analyzed on the MALDI Biotyper CA System.
  • Samples are analyzed using MALDI (matrix-assisted laser desorption/ionization)-TOF . (time-of-flight) mass spectrometry. The matrix transfers protons onto the extracted proteins and absorbs UV light. After complete drying, the mixture is exposed to laser pulses, resulting in energy transfer from the matrix causing evaporation and release of positively charged intact proteins and peptides ("soft" ionization technique). The ionized molecules are accelerated by electrical potentials through a flight tube to the mass spectrometer, with separation of the particles determined by their mass/charge ratio (m/z). As different proteins/peptides have different masses, ions arrive at the detector at different times (time of flight). The system measures the time (in the nanosecond range) between pulsed acceleration and the corresponding detector signal, and the speed is then converted into an exact molecular mass. The mass-to-charge ratio of an ion is proportional to the square of its drift

{10}------------------------------------------------

time. Highly abundant microbial proteins (mainly ribosomal proteins) result in a mass spectrum with characteristic mass and intensity distribution. It is species-specific for many bacteria and is interpreted as a molecular fingerprint to identify the test organism. Data acquisition is controlled with MALDI Biotyper CA System Software. The spectrum of the unknown organism is first transformed into a peak list. This peak list is compared to the reference peak list of each organism found in the reference library (database) and a log(score) is generated. See Assay Cut-Off.

L. Performance Characteristics:

    1. Analytical performance:
    • a. Addition of C. auris to MBT-CA reference library

A panel of Candida auris isolates and nine (9) other yeast species related to C. auris or commonly misidentified as C. auris were obtained from the CDC & FDA Antibiotic Resistance Isolate Bank (https://wwwn.cdc.gov/arisolatebank/). Additional C. auris isolates were obtained from (b) (4)

(9), (b) (4) (8), (b) (4) (1). Isolates are summarized in Table 1 below.

Twenty-eight (28) C. auris isolates as well as the nine (9) additional yeast species of the CDC panel were used for this study. Each isolate was cultured on sabourauddextrose agar, and spotted 8 times using DT, eDT and Ext sample preparation techniques. and measured on the MALDI instrument. Mass spectra acquisition and MBT-CA System identification were performed using FDA-cleared MBT-CA System software (client version 3.2.12). 888 spectra (37 strains * 3 sample preparations * 8 spots = 888) were used for identification and compared against the:

  • cleared validated MBT-CA library. .
  • . non-clinically validated MBT-CA library,
  • . cleared plus non-clinically validated MBT-CA libraries, and
  • . cleared validated MBT-CA library plus the six (6) new C. auris reference entries.

All 37 isolates used in this study were identified successfully using DT, eDT and Ext sample preparation techniques. Six (6) strains were used for reference library generation and 28 strains of C. auris were analyzed. All 28 strains were used for performance evaluation but only 22 strains were used for generating the truth tables; the six (6) strains used for reference entries were excluded from truth table counting (See Tables 2 and 3).

The ITS sequence was determined for all six strains used for the new reference library entries.

Table 1: All Candida auris and Candida auris-related strains used in this study
Strain IDResource CentreComment
Candida auris(b) (4)reference strain
Candida auris(b) (4)reference strain
Candida auris(b) (4)reference strain
(b) (4)(b) (4)
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisfield strain
Candida aurisreference strain
Candida auris AR0381 CAU 01 CDCCDC, USA“C. auris panel”
Candida auris AR0382 CAU 02 CDCCDC, USA“C. auris panel”
Candida auris AR0383 CAU 03 CDCCDC, USA“C. auris panel”
Candida auris AR0384 CAU 04 CDCCDC, USA"C. auris panel"
Candida auris AR0385 CAU 05 CDCCDC, USA“C. auris panel”
Candida auris AR0386_CAU_06 CDCCDC, USA"C. aurispanel"
Candida auris AR0387_CAU_07 CDCCDC, USA"C. aurispanel"
Candida auris AR0388 CAU 08 CDCCDC, USA"C. aurispanel"
Candida auris AR0389_CAU_09 CDCCDC, USA"C. aurispanel"
Candida auris AR0390_CAU_10 CDCCDC, USA"C. aurispanel"
Candida duobushaemulonii AR0391_CAU_11 CDCCDC, USA"C. aurispanel"
Candida duobushaemulonii AR0392_CAU_12 CDCCDC, USA"C. aurispanel"
Candida duobushaemulonii AR0394_CAU_14 CDCCDC, USA"C. aurispanel"
Candida haemulonii AR0393_CAU_13 CDCCDC, USA"C. aurispanel"
Candida krusei AR0397_CAU_17 CDCCDC, USA"C. aurispanel"
Candida lusitaniae AR0398_CAU_18 CDCCDC, USA"C. aurispanel"
Kodameae ohmeri AR0396_CAU_16 CDCCDC, USA"C. aurispanel"
Saccharomyces cerevisiae AR0399_CAU_19 CDCCDC, USA"C. aurispanel"

{11}------------------------------------------------

{12}------------------------------------------------

Candida auris
Reference Algorithm
MBT-CA ResultHigh resolution speciesLow resolution species/genusNegativeTotal
Positive Organism ID;(High Confidence); log(score) $\ge$ 2.0220022
Positive Organism ID;(Low Confidence); log(score) $\ge$ 1.7 - <2.0000
Incorrect MBT-CA ( $\ge$ 1.7)No ID00n/a
Total2222

Table 2: C. auris Performance (MBT-CA results vs Reference Algorithm)

b. Analytical Specificity:

Interference and specificity studies were previously reported in 510(k) K130831. For the addition of C. auris to the MBT-CA reference library, the following studies were conducted.

In silico evaluation:

To determine whether cross identifications would occur due to the MBT-CA library update (i.e., cleared MBT-CA library + C. auris), a subset of stored spectra for previously claimed organisms was run against the current as well as the updated reference library to verify that performance for these analytes was identical. In this study (0) spectra were compared against the current validated and the updated (b) (4) validated MBT-CA library. The selected spectra covered all claimed species species), each sample preparation technique (b) (4) and a spectru expected results ("no ID", "low confidence ID" and "high confidence ID").

The cross-validation showed 100% identical results. Each single log(score) of the 6822 log(scores) was identical before and after the library update demonstrating that adding a new reference entry does not influence the identification results of prior library versions.

(b)

Wet testing of claimed organisms:

(4) leared species (see Table 3 below) A predefined and well characterized set of s (validated plus non-clinically was tested using the current MBT-CA libr validated) as well as the modified MBT-CA libraries (deletion of three C. auris entries of the non-clinically validated library, inclusion of six new C. auris entries in the validated library). The established extraction and spotting procedures were used to verify that performance for this test set is similar for the modified and the (b)elected, claimed species were previously cleared MBT-CA library. Each of the (4)tion techniques (DT, eDT and each measured in triplicates. All three sample pre-

{13}------------------------------------------------

Ext) were applied independent from the identification success of DT and/or eDT sample preparation. All sample spots were measured on two different instruments. Overall 360 spectra / log(scores) were analyzed. All organisms could be identified using the established spotting algorithm (i.e., DT, eDT, EXT). No influence of the new C. auris reference entries and no cross-identification was observed.

Microorganisms
GroupSpeciesResource CentreStrain ID (b) (4
Gramneg.bacteriaAerobeEscherichia coli
AerobeKlebsiella pneumoniae
AerobePseudomonas aeruginosa
AnaerobeBacteroides fragilis
Grampos.bacteriaAerobeEnterococcus faecalis
AerobeStaphylococcus aureus
AerobeStreptococcuspneumoniae
AnaerobeClostridium perfringens
YeastsCandida albicans
Candida glabrata

Table 3: "Wet Testing" - cleared organisms

Wet testing of RUO organisms:

To verify that no cross-identifications between the Research Use Only (RUO) and the MBT-CA libraries occurs, three replicates of nine (9) species (see Table 4 below) which are in the RUO library (but are not in the claimed library or the non-clinically validated library) were tested against the MBT-CA old and updated libraries. All three sample preparation techniques (DT, eDT and Ext) were applied independent from the identification success of DT and/or eDT sample preparation. All organisms were tested on two instruments. The MBT-CA identification was performed using the current MBT-CA libraries (validated plus non-clinically validated) as well as the modified MBT-CA libraries (deletion of three (3) C. auris entries of the nonclinically validated library, inclusion of six (6) new C. auris entries in the validated library). Overall (0) pectra / log(scores) were analyzed. None of the RUO organisms were (fified with the MBT-CA libraries. No influence of the new C. auris reference entries and no cross-identification was observed.

Microorganisms
GroupSpeciesResource CentreStrain ID
Gramneg.bacteriaAerobeAcinetobacter gerneri(b) (4
Chryseobacteriumchaponense
Flavobacterium flevensePantoea calida
Table 4: "Wet Testing" - RUO organisms

{14}------------------------------------------------

(b) (4
Grampos.bacteriaAerobeMicrococcus flavus
AerobeRothia terrae
bacteriaAnaerobeClostridium magnum
AnaerobePaenibacillus durus
YeastsCandida solani
  • c. Reproducibility: See K130831, K142677 and K163536.
  • d. Linearity/assay Reportable Range: Not applicable, qualitative assay.
  • Traceability, Stability, Expected Values (controls, calibrators, or methods): e.

Calibrator: See K130831, K142677 and K163536.

Controls: See K130831, K142677 and K163536.

Sample Stability after Matrix Overlay: The sample stability study on target plates for Gram-negative bacteria was previously validated and reported in 510(k) K130831. The sample stability on target plates for Gram-positive bacteria and yeasts was validated and reported in 510(k) K142677.

US IVD Bacterial Test Standard (BTS) Stability: BTS Stability was established and described in 510(k) K130831.

HCCA portioned (Matrix) Stability: CCA portioned (Matrix) Stability was established and described in 510(k) K130831.

Target plates stability: Target plate stability was established and described in 510(k) K142677.

Organism Stability: For FDA-cleared media organism stability studies see K130831, K142677 and K163536.

  • f. Detection Limit:
    The Limit of Detection/Dynamic Range study for Gram-negative bacteria was previously performed and reported in K130831. The Limit of Detection/Dynamic Range study for Gram-positive bacteria and yeasts was performed and reported in 510(k) K142677.

g. Influence of Agar Media

The validation of sample preparation of test organism to demonstrate that culture media inoculated onto US IVD 48 spot targets with or without an organism present does not interfere with system performance was previously performed and reported in 510(k) K130831.

{15}------------------------------------------------

  • h. Carry-Over/ Cross-Contamination
    The carry-over, cross-contamination and target cleaning study was previously performed and reported in K130831.

  • Assay Cut-off i.
    The assay cut-off remains unchanged as established in K130831. Using statistical analysis, a probability ranking of the organism identification is generated. The probability ranking is represented as a log (score) between 0.00 and 3.00.

Organism identification (direct or extracted) is reported with high confidence if the log(score) is ≥ 2.00. If a direct transfer organism identification log(score) is <2.00; the user is instructed to follow an extraction procedure (eDT or EXT).

After extraction:

  • If the organism identification log(score) is between 1.7 and <2.0, the ● identification is reported as low confidence.
  • If the organism identification log(score) is <1.7, it is reported as 'No ● Identification'.

Some MBT-CA identifications results are non-clinically validated organisms and are displayed in the MBT-CA report in the interest of public health as a means of directing the required additional laboratory testing. Non-clinically validated organism results are created from reference patterns which have not been clinically validated. 'No Identification' is reported based on the validated reference pattern (MSP) library if a result from the non-clinically validated reference pattern (MSP) library was found to yield a higher log(score) value. The non-clinically validated result is added as a comment (grey in square brackets) below the 'No Identification' result. Identification of non-clinically validated organisms must be performed with an alternate laboratory method. Results for non-clinically validated organisms cannot be transmitted from the MBT-CA to the laboratory information system.

The log(score) value ranges defined in the MBT-CA reflect the probability of organism identification. Results should be reviewed by a trained microbiologist and final organism identification should be based on all relevant information available. This information includes, but is not limited to, Gram staining, colony morphology, growth characteristics, sample matrix, or other factors that might impact organism identification.

If an indication for the possibility of cross-matching patterns (i.e., crossidentification) was found during the clinical or analytical performance studies (see K130831, K142677 and K163536), a matching hint was placed for the organism in the package labelling (matching hint table).

2. Comparison Studies:

a. Method comparison with predicate device: Not applicable.

{16}------------------------------------------------

b. Matrix comparison: Not applicable.

    1. Clinical Studies: See K130831, K142677 and K163536.
    1. Clinical cut-off: See Assay Cut-Off.
    1. Expected values/Reference range: See L.1.i.

M. Instrument Name(s)

Microflex LT/SH mass spectrometer

O. Other Supportive Instrument Performance Characteristics Data Not Covered in the "Performance Characteristics" Section above:

See K130831, K142677 and K163536

P. Proposed Labeling:

The labeling is sufficient and satisfies the requirements of 21 CFR parts 801 and 809 as well as the Special Controls for this type of device.

{17}------------------------------------------------

Q. Identified Risks to Health and Mitigation Measures:

Identified Risks to HealthMitigation Measures
Incorrect identification or lack of identificationof a pathogenic microorganismSpecial Controls 1, 2, 3, 4
Failure to correctly interpret test resultsSpecial Control (3)
Failure to correctly operate the instrumentSpecial Control (3)(i), (5)(iv)(H)

Identified Risks to Health and Mitigation Measures

R. Benefit/Risk Analysis:

Summary
Summary ofthe Benefit(s)The primary benefit from this device is more rapid and accurate identification of Candidaauris from cultured material. In the setting of a critically ill patient, more rapididentification of pathogens, such as C. auris, may ensure appropriate antimicrobial use andpatient isolation precautions earlier.
Summary ofthe Risk(s)The primary risk associated with use of this device is incorrect identification of C. auris orother pathogenic microorganisms. If an organism misidentification, or 'no identification'was to occur, patients could potentially experience a delay in effective antimicrobialtherapy, with associated increases in morbidity or mortality. However, since itsintroduction into the clinical microbiology laboratory, MALDI-TOF has been found to behighly accurate for claimed microorganisms, with few reports of misidentificationresulting in patient harm. Overall, risk of patient injury from use of the device is low andshould not be greater than alternatives currently used by clinical microbiology laboratories.
Summary ofOtherFactorsNone.

Conclusions

Do the probable benefits outweigh the probable risks?

The probable benefits of the MBT-CA system outweigh the potential risks in light of the listed special controls and applicable general controls. An identification for C. auris and protocol for future database updates will provide patients access to rapid and accurate laboratory results. The validation data, in association with clinical experience using MALDI-TOF technology, suggest that errors will be uncommon and unlikely to result in patient harm. The risks of patient harm are further mitigated by the proposed special controls, including product labelling, and current laboratory practices, which include alternative identification methods, AST testing, and other diagnostics.

S. Patient Perspectives

{18}------------------------------------------------

This submission did not include specific information on patient perspectives for this device.

T. Conclusion:

The information provided in this De Novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.3378. FDA believes that the stated special controls, and applicable general controls, provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code:QBN
Device Type:Clinical Mass Spectrometry Microorganism Identification and DifferentiationSystem
Class:II (special controls)
Regulation:21 CFR 866.3378
  • (a) Identification. A clinical mass spectrometry microorganism identification and differentiation system is a qualitative in vitro diagnostic device intended for the identification and differentiation of microorganisms from processed human specimens. The system acquires, processes, and analyzes spectra to generate data specific to a microorganism(s). The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infection.
  • (b) Classification. Class II (special controls). Clinical mass spectrometry microorganism identification and differentiation system must comply with the following special controls:
    • (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of what the device detects, the type of results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended, when applicable.
    • (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt with an indication for in vitro diagnostic use.
    • (3) The 21 CFR 809.10(b) labeling must include:
      • A detailed device description, including all device components, control elements (i) incorporated into the test procedure, instruments, ancillary reagents required but not provided, and a detailed explanation of the methodology and all pre-analytical methods for processing of specimens, and algorithm used to generate a final result. This must include a description of validated inactivation procedure(s) that are confirmed through a viability testing protocol, as applicable.
      • (ii) Performance characteristics for all claimed sample types from analytical studies with clinical specimens that include prospective samples and/or, if appropriate, characterized samples.
      • (iii) Performance characteristics of the device for all claimed sample types based on analytical studies, including, but not limited to, limit of detection, inclusivity,

{19}------------------------------------------------

reproducibility, interference, cross reactivity, interfering substances, carryover/cross contamination, sample stability, and additional studies regarding processed specimen type and intended use claims, as applicable.

  • A detailed explanation of the interpretation of test results for clinical specimens (iv) and acceptance criteria for any quality control testing.
  • (4) The device's labeling must include a prominent hyperlink to the manufacturer's website where the manufacturer shall make available their most recent version of the device's 21 CFR 809.10(b) labeling, which must reflect any changes in the performance characteristics of the device. FDA must have unrestricted access to this website or manufacturers must provide this information to FDA through an alternative method that is considered and determined by FDA to be acceptable and appropriate.
  • (5) Design verification and validation must include:
    • Any clinical studies must be performed with samples representative of the (i) intended use population and compare the device performance to results obtained from an FDA accepted reference method and/or FDA accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol (including predefined statistical analysis plan, if applicable), clinical study report, and results of all statistical analyses.
    • (ii) Performance characteristics for analytical and clinical studies for specific identification processes for the following, as appropriate:
      • (A)Bacteria
      • (B) Yeasts
      • (C) Molds
      • (D)Mycobacteria
      • (E) Nocardia
      • (F) Direct sample testing (e.g., Blood culture)
      • (G) Antibiotic resistance markers
      • (H)Select Agents (e.g., pathogens of high consequence)
    • (iii) Documentation that the manufacturer's risk mitigation strategy ensures that their device does not prevent any device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use (e.g., safety and effectiveness of the functions of the indicated device(s) remain unaffected).
    • (iv) A detailed device description including the following:
      • (A)Overall device design, including all device components and all control elements incorporated into the testing procedure.
      • (B) Algorithm used to generate a final result from raw data (e.g., how raw signals are converted into a reported result).

{20}------------------------------------------------

  • (C) A detailed description of device software, including, but not limited to, validation activities and outcomes.
  • (D) Acquisition parameters (e.g., mass range, laser power, laser profile and number of laser shots per profile, raster scan, signal-to-noise threshold) used to generate data specific to a microorganism.
  • (E) Implementation methodology, construction parameters, and quality assurance protocols, including the standard operating protocol for generation of reference entries for the device.
  • (F) For each claimed microorganism characteristic, each organism must have a minimum of five reference entries (including the type strain for microorganism identification) or, if there are fewer reference entries, a clinical and/or technical justification, determined by FDA to be acceptable and appropriate, for why five reference entries are not needed.
  • (G) All type strains and at least 20 % of the non-type strains of a species detected by the device must be characterized by DNA sequence analysis or, if there are fewer strain sequences, then a clinical and/or technical justification, determined by FDA to be acceptable and appropriate, for the reduced number of strains sequenced must be provided.
  • (H) As part of the risk management activities, an appropriate end user device training program must be offered as an effort to mitigate the risk of failure from user error.

{21}------------------------------------------------

Page Intententionally Left Blank

§ 866.3378 Clinical mass spectrometry microorganism identification and differentiation system.

(a)
Identification. A clinical mass spectrometry microorganism identification and differentiation system is a qualitative in vitro diagnostic device intended for the identification and differentiation of microorganisms from processed human specimens. The system acquires, processes, and analyzes spectra to generate data specific to a microorganism(s). The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infection.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use statement must include a detailed description of what the device detects, the type of results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended, when applicable.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt with an indication for in vitro diagnostic use.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including all device components, control elements incorporated into the test procedure, instrument requirements, ancillary reagents required but not provided, and a detailed explanation of the methodology and all pre-analytical methods for processing of specimens, and algorithm used to generate a final result. This must include a description of validated inactivation procedure(s) that are confirmed through a viability testing protocol, as applicable.
(ii) Performance characteristics for all claimed sample types from clinical studies with clinical specimens that include prospective samples and/or, if appropriate, characterized samples.
(iii) Performance characteristics of the device for all claimed sample types based on analytical studies, including limit of detection, inclusivity, reproducibility, interference, cross-reactivity, interfering substances, carryover/cross-contamination, sample stability, and additional studies regarding processed specimen type and intended use claims, as applicable.
(iv) A detailed explanation of the interpretation of test results for clinical specimens and acceptance criteria for any quality control testing.
(4) The device's labeling must include a prominent hyperlink to the manufacturer's website where the manufacturer must make available their most recent version of the device's labeling required under § 809.10(b) of this chapter, which must reflect any changes in the performance characteristics of the device. FDA must have unrestricted access to this website, or manufacturers must provide this information to FDA through an alternative method that is considered and determined by FDA to be acceptable and appropriate.
(5) Design verification and validation must include:
(i) Any clinical studies must be performed with samples representative of the intended use population and compare the device performance to results obtained from an FDA-accepted reference method and/or FDA-accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol (including predefined statistical analysis plan, if applicable), clinical study report, and results of all statistical analyses.
(ii) Performance characteristics for analytical and clinical studies for specific identification processes for the following, as appropriate:
(A) Bacteria,
(B) Yeasts,
(C) Molds,
(D) Mycobacteria,
(E) Nocardia,
(F) Direct sample testing (
e.g., blood culture),(G) Antibiotic resistance markers, and
(H) Select agents (
e.g., pathogens of high consequence).(iii) Documentation that the manufacturer's risk mitigation strategy ensures that their device does not prevent any device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use (
e.g., safety and effectiveness of the functions of the indicated device(s) remain unaffected).(iv) A detailed device description, including the following:
(A) Overall device design, including all device components and all control elements incorporated into the testing procedure.
(B) Algorithm used to generate a final result from raw data (
e.g., how raw signals are converted into a reported result).(C) A detailed description of device software, including validation activities and outcomes.
(D) Acquisition parameters (
e.g., mass range, laser power, laser profile and number of laser shots per profile, raster scan, signal-to-noise threshold) used to generate data specific to a microorganism.(E) Implementation methodology, construction parameters, and quality assurance protocols, including the standard operating protocol for generation of reference entries for the device.
(F) For each claimed microorganism characteristic, a minimum of five reference entries for each organism (including the type strain for microorganism identification), or, if there are fewer reference entries, a clinical and/or technical justification, determined by FDA to be acceptable and appropriate, for why five reference entries are not needed.
(G) DNA sequence analysis characterizing all type strains and at least 20 percent of the non-type strains of a species detected by the device, or, if there are fewer strain sequences, then a clinical and/or technical justification, determined by FDA to be acceptable and appropriate, must be provided for the reduced number of strains sequenced.
(H) As part of the risk management activities, an appropriate end user device training program, which must be offered as an effort to mitigate the risk of failure from user error.