510(k) Data Aggregation

    K Number
    K210109

    Validate with FDA (Live)

    Device Name
    visor2 system
    Date Cleared
    2022-04-08

    (444 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    visor2 is used to aid 3D navigation for positioning of Transcranial Magnetic Stimulation (TMS) coils over preselected brain regions based on data from MRI measurements or scalp landmarks and optional EMG recording in clinical procedures. The visor2 system is compatible with the following commercially available TMS coils of the following TMS companies:

    • · Mag & More: PMD70-aCool Double coil Figure-8 (510575)
    • MagStim Company Ltd.: Double 70mm Remote Coil (3190-00 rev2), Double 70mm Air Film Coil (3910 00);
    • MagVenture A/S: Double 75mm (C-B60), Double Cooled 75mm (Cool-B65), Double 95mm (D-B80), Angulated Double Cooled 97mm (Cool-B70 rev2), Butterfly figure-8 (C-B70)
    • · Neuronetics Inc .: NeuroStar coil
      The visor2 system must be operated by a trained physician and must not be used during surgical operations.
    Device Description

    The visor2 system is a 3D navigation system, supporting external/surface positioning of labeling identified commercially available Transcranial Magnetic Stimulation (TMS) coils based on the anatomical and functional areas for which use of the coil has been FDA cleared and as identified in cleared labeling for the coil. The visor2 system optionally records and processes electrophysiological signals to guide the user in the TMS coil placement. The visor2 system is compatible with TMS stimulators identified in the labeling that meet IEC 60601-1 and IEC 60601-1-2 electrical safety and EMC standards, a compatible optical tracking system, and further compatible accessories. Optionally, a compatible amplifier identified in the labeling may be applied for measuring electrophysiological signals. The visor2 system must be operated by a trained physician, and must not be used during surgical operations.
    The visor2 system consists of a cart, an All-in-One medical computer with touch interface, a mounting plate for an optional amplifier applied for EMG recording and a Tracking camera with mounting arm. There are two fundamental visor2 system variants: single or dual display. the second configuration is available with an additional touch display added next to the All-in-One computer. Both variants are identical with regard to hardware features.

    AI/ML Overview

    The acceptance criteria for the visor2 system and the study results are detailed in the provided text. The device is a 3D navigation system that aids in the positioning of Transcranial Magnetic Stimulation (TMS) coils over preselected brain regions based on MRI measurements or scalp landmarks, with optional EMG recording.

    Here's a breakdown of the requested information:


    1. A table of acceptance criteria and the reported device performance

    The provided text combines the acceptance criteria (defined as "Main PASS criteria:" or implicitly as "Pass" alongside the test method and results) and the reported device performance within the "Bench testing demonstrated that the performance parameters of the visor2 system meets pre-specified requirements and are substantially equivalent to those of the predicate device" section.

    Test / FeatureAcceptance Criteria (Main PASS criteria / Result)Reported Device Performance
    Electrical SafetyPass (IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 +A1:2012)Passed
    Electromagnetic CompatibilityPass (IEC 60601-1-2 Edition 4.0 2014-02)Passed
    Compliance with Special ControlsPass (Review and document compliance with Class II Special Controls guidance document: Repetitive Transcranial Magnetic Stimulation Systems, issued July 26, 2011)Passed
    System performance to specificationsPass (Hardware, software and system evaluations)Passed
    Biocompatibility evaluationPass (The Pointer Tool patient contact materials are identical to the StimGuide TMS coil location/navigation feature in K183376)Passed
    Software verification, validationPass (Compliance with FDA Guidance document: Guidance for the Content of Premarket Submissions for Software Content in Medical Devices, May 11, 2005; including intentional and unintentional cybersecurity risks and mitigations applied)Passed
    Commercial TMS Coil compatibilityPassThe compatible commercially compatible TMS coils are demonstrated to be able to be calibrated and properly applied within the location accuracy defined by the visor2 system.
    Overall accuracy of TMS coil position relative to the head (absolute value)N/A (Worst case upper limit of the position mismatch in positioning the coil relative to the head)Pass, Range <= 10 mm (This appears to be the actual result/performance, implicitly meeting a standard, as the system accuracy below states 4mm as the goal)
    Position accuracy (Tracking camera)Pass, Range <= 10 mmPass, Range <= 10 mm (This is the acceptance criteria and presented as the result)
    Navigation PrincipleThe stimulus position should be within 4mm of the physical setup and the distance between stimulus target and stimulus below 2mm.Main PASS criteria met.
    Coil CompatibilityAccuracy limit in validation workflow 4mmMain PASS criteria met.
    Coil Compatibility - ValidationPassed validationPassed validation
    Tracking System AccuracyDeclaration of Conformity by Northern Digital Inc. and over-all system accuracy better than 4mmAchieved (stated Declaration of Conformity by Northern Digital Inc. and over-all system accuracy better than 4mm)
    System AccuracyThe match between positions and length in over-all system accuracy better than 4mmAchieved (stated match between positions and length in over-all system accuracy better than 4mm)
    Imaging ModalityAccuracy of landmarks to MRI within 2mm.Main criteria met.
    Selection of targets via anatomical and functional landmarks (EMG)10% accuracy in EMG parameters for functional mapping; Over-all position system accuracy better than 4mmFunctional mapping with 10% accuracy in EMG parameters achieved. Overall position system accuracy better than 4mm achieved.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The provided document describes non-clinical performance tests (bench testing). It does not mention a "test set" in terms of patient data or clinical images, but rather tests conducted on the device itself and its components.

    • Sample Size for Test Set: Not applicable in the context of patient data / images. The testing involved various controlled measurements of the device's accuracy and functionality. For some tests, like "Position accuracy (Tracking camera)", it mentions a single marker stepped through "more than 900 positions".
    • Data Provenance: The tests are described as "Bench testing," implying laboratory or engineering testing of the device and its sub-systems.
    • Retrospective or Prospective: Not applicable as no patient data is used. These are performance verification tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The ground truth for these non-clinical performance tests would be established by predefined engineering specifications, calibrated measurement systems, and physical setups, rather than human expert consensus on medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. As no patient data or human interpretation is involved, there is no need for an adjudication method as would be used in a clinical study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done or mentioned. The document explicitly states: "Clinical data are not needed to support substantial equivalence." The device is cleared as an accessory for 3D navigation, focusing on its technical accuracy and compatibility with existing TMS systems rather than direct clinical outcomes or human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The performance metrics provided (e.g., tracking system accuracy, system accuracy, imaging modality accuracy, EMG accuracy) represent the standalone technical performance of the device's navigation and measurement capabilities. While the device is intended for use by a trained physician, the accuracy measurements are objective assessments of the system itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical performance tests was based on:

    • Known nominal positions/rotations/measurements in a physical setup, using calibrated instruments and methods.
    • Pre-defined engineering specifications for parameters like accuracy (e.g., 4mm for system accuracy, 2mm for MRI landmark accuracy, 10% for EMG parameters).
    • Declarations of Conformity from component manufacturers (e.g., Northern Digital Inc. for tracking system accuracy).

    8. The sample size for the training set

    • Not applicable. This device is a navigation system and does not appear to involve machine learning or AI models that require a "training set" in the conventional sense (e.g., for image classification or prediction tasks). Its functionality is based on established principles of optical tracking, 3D spatial registration, and signal processing.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no "training set" for an AI/ML model for this device's function.