Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170113

    Validate with FDA (Live)

    Device Name
    pixium 4343RCE
    Date Cleared
    2017-02-09

    (28 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As a part of a radiographic system, the Pixium 4343RCE is intended to acquire digital radiographic images. The Pixium 4343RCE is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding fluoroscopy, angiography and mammography.

    Device Description

    The pixium 4343RCE is a Stationary x-ray system that converts x-ray patterns into electrical signals. The signals are converted into visible images for use in medical diagnosis. In the device, a cesium iodide scintillator absorbs the input x-ray photons. The scintillator in turn emits visible spectrum photons that illuminate an array of photodetectors that create an electrical charge representation of the x-ray input. A matrix scan of the array converts the integrated charges into a modulated electrical signal.

    The detector is permanently installed and intended to be integrated into an x-ray system, where it constitutes an x-ray receptor for direct x-ray imaging. It is electrically powered by and connected with the x-ray system. The device is connected to the Philips Eleva Workspot to create a complete x-ray imaging chain, and it is intended to be used in Philips x-ray systems such as the DigitalDiagnost.

    AI/ML Overview

    The provided text describes the Philips pixium 4343RCE, a stationary x-ray system. The document is a 510(k) summary for premarket notification to the FDA, asserting substantial equivalence to a predicate device (Philips pixium 4343RC, K131483).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present specific numerical acceptance criteria (e.g., minimum DQE, MTF values that must be met) for the pixium 4343RCE, nor does it provide a direct side-by-side comparison of claimed performance data against such defined criteria. Instead, the acceptance criteria are implicitly stated as demonstrating similar performance to the predicate device and compliance with various standards. The "Discussion" column in the comparison table serves as the "reported device performance" against the implicitly accepted criterion of not negatively impacting safety or effectiveness compared to the predicate.

    Here's a table summarizing relevant performance metrics and the discussion provided, which indicates how the device meets (or is considered to meet) implied acceptance criteria of non-inferiority or slight improvement:

    FeaturePredicate Device pixium 4343RC (K131483)Proposed Device pixium 4343RCE (K170113)Discussion / Performance vs. Implied Acceptance Criteria
    Design Features
    X-Ray AbsorberCsI ScintillatorCsI ScintillatorIdentical; no impact to safety and effectiveness.
    Installation typeStationary, permanently installedStationary, permanently installedIdentical; no impact to safety and effectiveness.
    Readout MechanismThin Film TransistorThin Film TransistorIdentical; no impact to safety and effectiveness.
    Detector Size500 x 490 x 45.5 mm³500 x 490 x 45 mm³Similar, slightly smaller; no impact to safety or effectiveness.
    Detector Weight<14 kg11.7kg ± 0.85kgSimilar, slightly lighter; no impact to safety or effectiveness.
    Image Size (Pixel)2840 x 28742840 x 2874Identical; no impact to safety and effectiveness.
    Image Size (X-ray field)420 x 425 mm²420 x 425 mm²Identical; no impact to safety and effectiveness.
    Distance Image to Rim34.9 mm40mmSimilar, slightly greater; no impact to safety or effectiveness.
    Pixel Size148µm148µmIdentical; no impact to safety and effectiveness.
    Nyquist Frequency3.37 lp/mm3.37 lp/mmIdentical; no impact to safety and effectiveness.
    ADC Digitization16 bit16 bitIdentical; no impact to safety and effectiveness.
    SENR (@ 1 µGy)42.3 dB42.3 dBIdentical; no impact to safety and effectiveness.
    Max X-ray Dose for Linear Response50 µGy50 µGyIdentical; no impact to safety and effectiveness.
    Image Readout Duration1.6 s1.36 sSlightly faster; no impact to safety and effectiveness.
    Number of Modes23Additional default mode included; no impact safety and effectiveness.
    Exposure Window Durations1-8192 ms0-8191 msMinimal decrease; insignificant, no impact to safety or effectiveness.
    Sequence Time Detector6 s6 sIdentical; no impact to safety and effectiveness.
    Image Data16.3 MBytes16.3 MBytesIdentical; no impact to safety and effectiveness.
    Use w and w/o Radiographic Grid?YesYesIdentical; no impact to safety and effectiveness.
    Maximum Lifetime Dose100 Gy100 GyIdentical; no impact to safety and effectiveness.
    Warm-up Duration before Calibration4 h2 hDecreased; no impact to safety or effectiveness.
    Data Interface to Workstation100 Mbit/s Ethernet1 Gbit/s EthernetIncreased; no impact to safety or effectiveness.
    Power Consumption20.4WAvg: 8W. Max: 15WDecreased; no impact to safety or effectiveness.
    Cover Factor (Optical Fill Factor)60.3%63%Essentially the same; no impact to safety or effectiveness.
    Image Quality
    Modulation Transfer Function (MTF)1 lp/mm 64%, 2 lp/mm 32%, 3 lp/mm 17%, Nyquist 13% (3.37 lp/mm)1 lp/mm 62%, 2 lp/mm 35%, 3 lp/mm 19%, Nyquist 15% (3.37 lp/mm)Similar, with some improvements and one minimal decrease; no impact to safety and effectiveness.
    Detective Quantum Efficiency (DQE) at 2 µGy0.05 lp/mm 65%, 1 lp/mm 52%, 2 lp/mm 42%, 3 lp/mm 25%, 3.37 lp/mm 18%0.05 lp/mm 67%, 1 lp/mm 51%, 2 lp/mm 42%, 3 lp/mm 27%, 3.37 lp/mm 18%Similar, with one improvement and one minimal decrease; no impact to safety and effectiveness.

    The study that proves the device meets the acceptance criteria is detailed as "Non-clinical verification and validation tests." These tests cover the intended use, technical claims, requirement specifications, and risk management results. The document explicitly states: "Non-clinical verification and validation test results demonstrate that the pixium 4343RCE... Meets the acceptance criteria and is adequate for its intended use." The basis for this is "compliance with the aforementioned international and FDA-recognized consensus standards and device specific guidance document" and showing "identical or equivalent technological characteristics," "equivalent diagnostic capability," and no "new potential hazards or safety risks" compared to the predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a separate "test set" in the context of data used for a clinical performance evaluation, as it explicitly states that no clinical study was required. The non-clinical verification and validation tests are performed on the device itself (hardware and software) and against established technical specifications and standards (e.g., ISO, IEC, NEMA, FDA guidance). Therefore, there is no mention of a "sample size" of patient data or "data provenance" (country of origin, retrospective/prospective) for a clinical test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Since no clinical study was conducted, there was no "test set" of patient data requiring expert consensus for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable as no clinical test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a digital x-ray detector, not an AI-assisted diagnostic tool for human readers. No MRMC study was performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the device itself (the x-ray detector) as a standalone component without human interpretation being part of the primary evaluation. The "Summary of Non-clinical Performance Data" and the "Comparison" tables effectively serve as the standalone performance assessment. The device's performance metrics (MTF, DQE, SENR, image readout speed, etc.) are evaluated directly for the detector. The conclusion is that "the device is as safe, as effective, and performs as well or better than the predicate device" in terms of its technological characteristics and ability to produce images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance evaluation, the "ground truth" is based on technical specifications, international and FDA-recognized consensus standards, and device-specific guidance documents. For instance, the MTF and DQE values are compared against those of the predicate device, with the implicit ground truth being that comparable or improved values demonstrate equivalent performance.

    8. The sample size for the training set

    Not applicable. The document describes an x-ray detector, not an AI algorithm that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable as there is no training set mentioned for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1