Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K163018

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-23

    (53 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is used in palliation of malignant neoplasms in the biliary tree

    Device Description

    The Zilver 635® Biliary Stent (ZILBS-635) consists of the following two components: a selfexpanding metal stent and stent delivery system. The stent is a laser cut self-expanding openended cylinder of nitinol. It is a slotted tube that is designed to provide support while allowing conformance to the duct upon deployment. Post-deployment, the stent is designed to impart an outward radial force upon the inner lumen of the duct, establishing patency in the stented region, thereby maintaining flow in the duct. The ZILBS-635 has radiopaque markers at both ends to assist in fluoroscopic visualization of the stent position. This stent delivery system like the predicate devices allows for co-axial deployment of the stent delivery system accepts a 0.035 inch wire guide. The stent delivery system includes a handle assembly and an introducer assembly. The handle assembly consists of a handle, a wire guide FLLA, holding bushing, a cannula, a check flo hub FLLA/cap and a safety tab. The introducer assembly consists of an outer sheath, an inner carrier and tip. The stent delivery system has an endoscopic pusher ring located on the inner carrier near the proximal end of the loaded stent. The outer sheath has a radiopaque maker band located within the wall of the outer sheath at the distal end. The stent delivery system is supplied with a syringe for flushing through the wire guide FLLA and the check flo hub FLLA.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Zilver 635® Biliary Stent, seeking substantial equivalence to previously cleared predicate devices. Since this is a submission for a medical device and not an AI/ML product, the information you've requested (acceptance criteria, study details for AI/ML performance, sample sizes, ground truth establishment, expert qualifications, etc.) is not applicable or present in the provided text.

    The document focuses on demonstrating that the Zilver 635® Biliary Stent (ZILBS-635) is substantially equivalent to existing predicate devices (Zilver® 635™ Biliary Stent and Zilver® Biliary Stent) for the palliation of malignant neoplasms in the biliary tree.

    Here's an overview of the information that is available in the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria and reported device performance in the context of AI/ML or a clinical study evaluating effectiveness against specific performance metrics for the device in vivo. Instead, it refers to "Performance Data" which includes:

    • Biocompatibility evaluation: Conducted in accordance with FDA's Use of International Standard ISO 10993-1.
    • Performance testing: "simulated use, deployment testing, dimensional testing, joint strength testing, MRI and shelf life testing were performed as per Cook's design control system."

    The specific quantitative results or acceptance criteria for these tests are not provided in this summary. The general conclusion is that "The non-clinical data supports the safety of the subject device and demonstrates that the Zilver 635® Biliary Stent (ZILBS-635) is safe and effective and should perform as intended in the specified use conditions."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and is not present in the document. The document describes non-clinical performance testing (e.g., biocompatibility, simulated use, mechanical testing) rather than a clinical study with a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and is not present in the document, as it pertains to medical devices and not an AI/ML diagnostic or prognostic tool.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and is not present in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and is not present in the document, as it pertains to medical devices and not an AI/ML diagnostic or prognostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable and is not present in the document.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    This information is not applicable and is not present in the document. For a medical device like a stent, "ground truth" would typically refer to the physical and mechanical properties meeting engineering specifications, and evidence of safety and performance in simulated and potentially clinical environments. The document refers to "biocompatibility evaluation," "simulated use, deployment testing, dimensional testing, joint strength testing, MRI and shelf life testing" as the basis for demonstrating safety and effectiveness.

    8. The sample size for the training set

    This information is not applicable and is not present in the document.

    9. How the ground truth for the training set was established

    This information is not applicable and is not present in the document.


    In summary, the provided text is a regulatory submission for a medical stent, focusing on demonstrating substantial equivalence to predicate devices through engineering and non-clinical performance testing. It does not contain any information related to AI/ML device performance or the specific criteria and studies you've asked about in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051124

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-07-12

    (70 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cook® Zilver® 635™ Biliary Stent is indicated for use in palliation of malignant neoplasms in the biliary tree.

    Device Description

    The Zilver® 635 " Biliary Stent is a self-expanding, nitinol stent designed for excellent radial strength and optimal longitudinal flexibility. Constructed from a series of interconnected Z-shaped segments, the stent conforms to the shape of the biliary system and provides circumferential scaffolding throughout the stent's length. Gold radiopaque markers on each end of the Zilver® 635™ Biliary Stent, along with one radiopaque marker at the distal end of the delivery system, allow precise positioning of the stent. The stent's interconnected Z-shaped segments also keep foreshortening to a minimum.

    The Zilver® 635" Biliary Stent comes preloaded in a 6 French delivery system. The stent is deployed with the use of a simple hand held device. The stent is available in unrestrained outer diameters of 4, 5, 6, 7, 8, 9 and 10 mm and in lengths of 20, 30, 40, 50, 60 and 80 mm. The 9, 10, 12, and 14 mm diameter stent with a cannula outside diameter of 1.6 mm is an addition to this product line.

    AI/ML Overview

    The Cook Zilver® 635™ Biliary Stent is a medical device and thus does not involve AI/ML. Much of the requested information (e.g., sample size for training set, number of experts, adjudication method) is not applicable to this type of medical device as it is not an AI/ML device.

    Here's a breakdown of the acceptance criteria and study information provided for the Cook Zilver® 635™ Biliary Stent, based on the provided text:

    1. Table of acceptance criteria and reported device performance:

    The provided text generally describes the types of tests performed rather than specific numerical acceptance criteria or detailed results for each test. The summary states that the results "provide reasonable assurance that the device has requirements for its use as a biliary stent," indicating that acceptance criteria were met.

    Acceptance Criteria CategoryReported Device Performance (Summary)
    DeploymentDemonstrated reliability
    a. Profile(Implicitly met reliable design)
    b. Deployment Accuracy(Implicitly met reliable design)
    c. Stent length and Change due to Deployment(Implicitly met reliable design)
    d. Stent Diameter(Implicitly met reliable design)
    e. Uniformity of Expansion(Implicitly met reliable design)
    Stent IntegrityDemonstrated reliability
    Radial ForceDemonstrated reliability

    2. Sample size used for the test set and data provenance:

    • Sample Size: Not explicitly stated for each test. The document mentions "tests" in plural, implying multiple units were tested, but no specific count is given.
    • Data Provenance: Not specified, but generally, such tests are conducted internally by the manufacturer (Cook Incorporated) as part of their design verification and validation process in a controlled lab setting. It is not patient data.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • Not Applicable. This is a hardware medical device (stent). The "ground truth" for its performance is established through physical and mechanical testing against engineering specifications, not expert consensus on diagnostic images or clinical outcomes in the same way an AI/ML algorithm's ground truth would be established.

    4. Adjudication method for the test set:

    • Not Applicable. As this is a mechanical device, there is no expert adjudication of results in the context of diagnostic interpretation. Results are typically based on objective measurements and established engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No. This is not an AI/ML diagnostic device, so an MRMC study comparing human readers with and without AI assistance is not applicable.

    6. If a standalone (i.e., algorithm-only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device. The concept of an "algorithm-only" performance does not apply.

    7. The type of ground truth used:

    • The "ground truth" for this device's performance is based on engineering specifications and performance metrics defined by the manufacturer and relevant industry standards for biliary stents (e.g., measurements of deployment accuracy, radial force, crimped profile).

    8. The sample size for the training set:

    • Not Applicable. This is a hardware medical device. There is no "training set" in the context of machine learning. The device design is based on engineering principles and materials science, not data-driven training.

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for a machine learning model, this question does not apply. The design of the stent (e.g., Z-shaped segments, nitinol material) is based on established engineering and biomedical knowledge to achieve desired mechanical properties.
    Ask a Question

    Ask a specific question about this device

    K Number
    K040505

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-16

    (18 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cook Zilver® 635™ Biliary Stent is indicated for use in palliation of malignant neoplasms in the biliary tree.

    Device Description

    The Zilver® 635" Biliary Stent is a self-expanding, nitinol stent designed for excellent radial strength and optimal longitudinal flexibility. Constructed from a series of interconnected Z-shaped seqments, the stent conforms to the shape of the biliary system and provides circumferential scaffolding throughout the stent's length. Gold radiopaque markers on each end of the Zilver" 635 " Biliary Stent, along with one radiopaque marker at the distal end of the delivery system, allow precise positioning of the stent. The stent's interconnected Z-shaped segments also keep foreshortening to a minimum. The Zilver® 635" Biljary Stent comes prefoaded in a 6 French delivery system. The stent is deployed with the use of a simple hand held device. The stent is available in unrestrained outer diameters of 4, 5, 6, 7, 8, 9,and 10 mm and in lengths of 20, 30, 40, 50 and 60 mm.

    AI/ML Overview

    This device is a medical implant, and the provided text describes its 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than a study against a pre-defined acceptance criteria for diagnostic/AI performance. The information you requested, particularly regarding acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert involvement, and comparative effectiveness studies, is not typically part of a 510(k) submission for this type of device (a biliary stent).

    However, I can extract the relevant information pertaining to the device's functional performance testing as described in the submission.

    Here is the information based on the provided text, heavily inferring where direct answers to your questions are not present due to the nature of the submission (device not AI/diagnostic):

    1. Table of Acceptance Criteria (Inferred Functional Performance) and Reported Device Performance

    Acceptance Criteria Category (Inferred)Specific TestDevice Performance (Reported as assurance of reliable design and performance)
    Deploymenta. ProfileAssured reliable design and performance
    b. Deployment AccuracyAssured reliable design and performance
    c. Stent length and Change due to DeploymentAssured reliable design and performance
    d. Stent DiameterAssured reliable design and performance
    e. Uniformity of ExpansionAssured reliable design and performance
    f. Stent IntegrityAssured reliable design and performance
    g. Ease of DeploymentAssured reliable design and performance
    Mechanical PropertiesRadial ForceAssured reliable design and performance
    Finite Element AnalysisAssured reliable design and performance
    Tensile TestingAssured reliable design and performance

    Note: The submission states "The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a biliary stent." This implies that the device met the pre-defined acceptance criteria for each of these functional tests, without detailing the specific numerical thresholds for those criteria in this summary.


    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided text.
    • Data Provenance: Not applicable in the context of device functional testing described. These are laboratory/bench tests, not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth, in the diagnostic sense, is not established for these functional device performance tests. These tests are based on engineering specifications and measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This concept pertains to expert review in diagnostic studies, not functional device testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (stent), not an AI/diagnostic software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a medical device (stent), not an AI/diagnostic software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the functional tests described, the "ground truth" would be the engineering specifications and established test methodologies (e.g., measuring stent diameter, deployment accuracy against a defined target, tensile strength against a known material property standard). It's based on physical measurements and material science, not clinical ground truth.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device. The design and manufacturing processes are likely informed by engineering principles and previous device designs, but there isn't a "training set" in the AI sense.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1