510(k) Data Aggregation

    K Number
    K170964

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-26

    (26 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medyssey Co, Ltd. Zenius Spinal System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft and /or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and / or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Zenius™ Spinal System, Internal Fixation Device for Spinal Surgery is comprised of: Rods, Pedicle Screw Assemblies, Compression Retaining Assemblies, and Transverse-Link Assemblies. Various forms and sizes of these implants are available so that adaptations can be utilized to take into account the unique pathology of individual patients.

    All implantable components are manufactured from Ti6Al4V ELI per ASTM F136 and wrought Co-Cr-Mo alloy per ASTM F1537. The subject of this submission is the addition of variable rods to the Zenius™ System.

    AI/ML Overview

    This document is a 510(k) premarket notification clearance letter from the FDA for the Zenius Spinal System. It does not contain information about the performance of a device based on a study with acceptance criteria, human readers, or ground truth.

    Specifically, there is no information in the provided text to address the following points in your request:

    1. A table of acceptance criteria and the reported device performance.
    2. Sample sizes used for the test set and data provenance.
    3. Number of experts used to establish ground truth and their qualifications.
    4. Adjudication method for the test set.
    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers' improvement with AI assistance.
    6. If a standalone (algorithm only) performance study was done.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    The document discusses the substantial equivalence of modifications (addition of variable rods) to an existing device (Zenius Spinal System) to its predicate device based on performance testing related to ASTM F1717. This likely refers to mechanical and fatigue testing of the device components, not a study involving human readers or AI performance on patient data.

    To clarify, the document states: "A rationale related to ASTM F1717 testing is provided showing the subject variable rod to be substantially equivalent to the predicate rods cleared in K131878." ASTM F1717 is a standard test method for both static and fatigue evaluation of pedicle screw systems. This indicates benchmark engineering performance, not clinical performance or AI algorithm performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110283

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-05-19

    (107 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medyssey Co. Ltd., Zenius™ Spinal System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

    Device Description

    The Zenius 146 Spinal System is a top-loading posterior spinal fixation system which consists of pedicle screws, rods, set screws, and a transverse (cross) linking mechanism. The Zenius ™ Spinal System implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available. Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Zenius "" Spinal System implants.

    AI/ML Overview

    The provided text describes a 510(k) summary for a spinal system and its modification, focusing on substantial equivalence to a predicate device. It does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria through clinical or even extensive non-clinical testing beyond basic mechanical and sterilization validation.

    This document specifically states: "Clinical tests: No clinical tests conducted on either the subject system nor the predicate system." The purpose of this submission was for the clearance of the redesign of the custom sterilization trays, not for assessing the clinical performance or efficacy of the spinal system itself beyond its substantial equivalence to the previously cleared predicate.

    Therefore, most of the requested information cannot be extracted from the provided text.

    Here is what can be inferred or explicitly stated based on the text, with clear limitations:


    1. Table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Mechanical Testing: Adherence to ASTM F 1717-04 standards for: - Static Compression Bending - Static Tension Bending - Static Torsional TestingAll testing performed per ASTM F 1717-04. (However, this particular submission did not include these tests for the redesigned sterilization trays, as they were not considered "worst case" for mechanical performance of the implant itself. The predicate device and previous submissions would have covered this.)
    Sterilization Validation: Achieve a Sterility Assurance Level (SAL) equal to at least 1 x 10⁻⁶ according to AAMI ST77 and AAMI ST79 guidelines for redesigned sterilization trays.Validation was satisfactorily performed on the re-designed sterilization trays, achieving the required SAL.
    Material: Fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136.Explicitly stated as conforming to ASTM F 136.
    Substantial Equivalence: To predicate device Zenius™ Spinal System (K093104) and Zenius™ Spinal System (K103272) for performance, safety & effectiveness requirements.Found substantially equivalent.

    Important Note: The document focuses on demonstrating substantial equivalence to a previously cleared device, particularly for a modification to sterilization trays. It does not present new performance data for the implant itself beyond the scope of that modification.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Mechanical Testing: Not specified for this submission, as the mechanical tests (ASTM F 1717-04) were already performed for the predicate device, and the current submission primarily concerned sterilization trays. For the sterilization validation, the sample size and provenance are not detailed in this summary; it just states it was "satisfactorily performed."
    • Clinical Data: No clinical tests were conducted for either the subject system or the predicate system (as stated in the document).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no clinical tests were performed, and thus no expert ground truth was established from clinical data for this submission. The "ground truth" for mechanical testing would be the ASTM standards, and for sterilization, the AAMI guidelines.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable as no clinical tests were performed and no human adjudication of results is mentioned for the non-clinical tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is a mechanical spinal implant, not an AI-assisted diagnostic or therapeutic system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a mechanical spinal implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For mechanical properties: ASTM F 1717-04 standards.
    • For material properties: ASTM F 136 standard for titanium alloy.
    • For sterility: AAMI ST77 and AAMI ST79 guidelines.
    • No clinical ground truth (expert consensus, pathology, outcomes data) was used or generated for this specific submission because no clinical studies were performed.

    8. The sample size for the training set:

    • Not applicable. This is a mechanical device, not a machine learning algorithm.

    9. How the ground truth for the training set was established:

    • Not applicable. This is a mechanical device, not a machine learning algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103272

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-12-29

    (54 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medyssey Co. Ltd., Zenius™ Spinal System, a posterior spinal fixation device, indicated for skeletally mature patients receiving fusion by autogenous bone graft with removal of the implants after the attainment of a solid fusion and is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

    Device Description

    The Zenius "M Spinal System is a top-loading posterior spinal fixation system which consists of pedicle screws, rods, set screws, and a transverse (cross) linking mechanism. The Zenius ™ Spinal System implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136. Various sizes of these implants are available. Specialized instruments made from surgical instrument grade stainless steel are available for the application and removal of the Zenius™ Spinal System implants.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Zenius™ Spinal System.

    It's important to note that this document is a 510(k) Special Premarket Notification for a modification to an existing device. This type of submission often relies heavily on proving substantial equivalence to the predicate device, rather than entirely new, extensive studies.


    Acceptance Criteria and Device Performance for Zenius™ Spinal System (K103272)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    Mechanical PerformanceStatic Compression Bending (ASTM F 1717-04)The predicate device underwent this testing. The subject device (modified Zenius™ Spinal System) contains "dimensionally modified components (not worst case)" and therefore was "not subject to ASTM F 1717-04 additional testing." This implies the modification was minor enough not to negatively impact the predicate's established performance or it was shown through analysis not to compromise the predicate's performance.
    Static Tension Bending (ASTM F 1717-04)Same as above.
    Static Torsional Testing (ASTM F 1717-04)Same as above.
    Dynamic Compression Bending (ASTM F 1717-04)Same as above.
    Biocompatibility/MaterialsConformance to ASTM F 136 for Titanium Alloy (Ti-6AI-4V ELI)The implant components are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136.
    Quality System ConformanceAdherence to 21 CFR part 820, ISO 9001, ISO 13485The design and development process of the manufacturer (both subject and predicate system) conforms to 21 CFR part 820, ISO 9001, and ISO 13485 quality systems.
    Substantial Equivalence (Overall)Same indications for use, contraindications, warnings, precautions, principles of operation.The Modified Zenius™ Spinal System is stated to be "substantially equivalent" to the currently marketed Zenius™ Spinal System (K093104). Both systems use the same universal housing, pre-assembled pedicle screw, set screw, and vertical rods. The principles of operation are the same, employing the same indications for use, contraindications for use, warnings, and precautions within labeling. The design and development processes conform to relevant quality standards. The subject device "contains dimensionally modified components (not worst case)" and was therefore not subject to additional mechanical testing under ASTM F 1717-04 because the predicate device had already undergone such testing. No clinical tests were conducted on either the subject or predicate system.

    Regarding the Absence of Specific Information for AI-Related Questions:

    It is crucial to understand that the provided text is for a spinal fixation system (a physical medical device), not an AI/software-as-a-medical-device (SaMD). Therefore, many of the questions related to AI performance, sample sizes for test/training sets, experts for ground truth, adjudication methods, and MRMC studies are not applicable to this type of device submission. The study is mechanical/physical testing, not an algorithmic performance evaluation.


    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated as a numerical sample size in the context of device testing. Mechanical tests specified by ASTM F 1717-04 would involve a certain number of test units, but this detail is not provided.
    • Data Provenance: The mechanical testing was performed by Medyssey Co., Ltd., which is a Korean company (Gyeonggi-do, Korea). The ASTM standards are international standards. This is lab-based physical testing, not a clinical data set.
    • Retrospective/Prospective: Neither. This refers to laboratory mechanical testing, not clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. For a mechanical spinal implant, "ground truth" is established through engineering specifications, material properties, and adherence to validated testing standards (like ASTM F 1717-04). It does not involve expert consensus in the way an AI diagnostic tool would.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. Since there's no diagnostic or interpretative output from the device being evaluated by human experts, there's no adjudication method in play.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a hardware device; AI assistance for human readers is irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Engineering Specifications and Standardized Mechanical Test Results: The "ground truth" for this device revolves around its ability to meet the defined mechanical performance thresholds and material specifications outlined in standards like ASTM F 1717-04 and ASTM F 136. The predicate device's existing clearance (K093104) serves as the basis for substantial equivalence for similar material and design modifications.

    8. The sample size for the training set:

    • Not Applicable. This is not an AI device that requires a training set. The "training" for a physical device involves its design, manufacturing processes, and material selection, all adhering to established quality systems.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set in the AI sense. The "ground truth" for the device's design and manufacturing is established through adherence to engineering principles, material science, and quality management systems (21 CFR part 820, ISO 9001, ISO 13485).