510(k) Data Aggregation

    K Number
    K160588

    Validate with FDA (Live)

    Device Name
    XN CHECK BF
    Manufacturer
    Date Cleared
    2016-12-22

    (296 days)

    Product Code
    Regulation Number
    864.8625
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XN CHECK BF is used for control and calibration of Sysmex XN series (XN-10, XN-11, XN-20, XN-21, XN-L) analyzers. It is not, however, intended for actual calibration of these analyzers. Assayed parameters include:

    WBC-BF (10^3/μL), RBC-BF (10^6/μL), MN# (10^3/μL), PMN# (10^3/μL), MN%, PMN%, TC-BF# (10^3/uL)

    Device Description

    XN CHECK™ BF is an in-vitro diagnostic, two level, control. It contains the following: stabilized red blood cell component(s) and stabilized white blood cell component(s) in a preservative medium in polypropylene plastic vials with screw caps with a 3 mL fill. The vials will be packaged in (4) welled vacuum formed clamshell container with the Instructions for Use (IFU) / assay sheet. Product storage conditions are 2 - 8° C.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "XN CHECK BF." This device is a hematology quality control mixture, and the submission is for a new iteration of an already marketed product (K141957). The primary change in the new device is the addition of compatibility with the Sysmex XN-L analyzer.

    The document does not contain the information requested regarding acceptance criteria for an AI/ML device, the study to prove it meets those criteria, sample sizes, expert involvement, or MRMC studies. This is because XN CHECK BF is a quality control material for laboratory analyzers, not an AI/ML diagnostic or image analysis device.

    Therefore, I cannot extract the requested information from the provided text. The document focuses on performance testing relevant to a quality control material, such as precision and stability studies, which are different from the types of studies performed for AI/ML diagnostic tools.

    To answer your request, I would need a document describing the regulatory submission for an AI/ML-driven medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141957

    Validate with FDA (Live)

    Device Name
    XN CHECK BF
    Manufacturer
    Date Cleared
    2014-12-05

    (140 days)

    Product Code
    Regulation Number
    864.8625
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XN CHECK BF is used for control and calibration verification of Sysmex XN series (XN-10, XN-11, XN-20, XN-21) analyzers. It is not. however, intended for actual calibration of these analyzers. Assayed parameters include:

    WBC-BF (10^3/μl), RBC-BF (10^6/ μl), ΜΝ# (10^3/μl), ΜΝ% (%), ΡΜΝ% (%), ΡΜΝ% (%), ΤC-ΒΓ# (10°3/μl)

    Device Description

    XN CHECK™ BF is an in-vitro diagnostic product the following: stabilized red blood cell component(s) and stabilized white blood cell component(s) in a preservative medium. The product is packaged in polypropylene plastic vials with a 3 m fill. The vials will be packaged in (4) welled vacuum formed clamshell container with the Instructions for Use / assay sheet. The product storage conditions are at 2 - 8° C.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called XN CHECK™ BF. This document focuses on the substantial equivalence of the new device to a predicate device, rather than providing a detailed study design with specific acceptance criteria and performance metrics typically found for AI/ML-based medical devices or diagnostic tests.

    The XN CHECK™ BF is a hematology quality control mixture for Sysmex XN series analyzers, not an AI-powered diagnostic device. Therefore, many of the questions related to AI/ML device studies (such as sample size for test/training sets, ground truth establishment, expert adjudication, MRMC studies, and standalone performance) are not applicable to this submission.

    However, based on the information provided, here's a breakdown of what can be articulated regarding acceptance criteria and the study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria in a table format with specific thresholds (e.g., "accuracy > 90%"). Instead, the "acceptance criteria" are implied by the nature of a quality control product and its intended use, which is to demonstrate stability, reproducibility, and substantial equivalence to a predicate.

    Acceptance Criteria (Implied)Reported Device Performance and Discussion of Tests
    Open-Vial Stability"product performance data for the following studies: Open-Vial Stability... The resultant data set established that XN CHECK BF is safe and effective for its intended use and that the product is stable for the entire product dating." The stated open-vial stability period is 30 days.
    Closed-Vial Stability"product performance data for the following studies: Closed-Vial Stability... The resultant data set established that XN CHECK BF is safe and effective for its intended use and that the product is stable for the entire product dating." The stated closed-vial stability period is 84 days.
    Precision Performance (Reproducibility)"product performance data for the following studies: ...Precision Performance. The resultant data set established that XN CHECK BF is safe and effective for its intended use and that the product is stable for the entire product dating." "Study results show XN CHECK BF to be consistently reproducible..."
    Substantial Equivalence to PredicateDemonstrated through comparison tables (Intended Use, Open Vial Stability, Closed Vial Stability, Reagents, Storage Conditions) which show mostly "Same" or minor "Differences" (addition of analyzer models) that don't alter fundamental performance. "Study results show XN CHECK BF to be ... substantially equivalent to the predicate products."
    Safety and Effectiveness"The resultant data set established that XN CHECK BF is safe and effective for its intended use..." and "XN CHECK BF is a safe and effective product, which fulfills its intended use when used as instructed in the Instructions for Use."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set in terms of the number of samples or runs for the stability and precision studies. It mentions "product performance data" was collected but does not provide details on the experimental design, number of replicates, or number of lots tested.

    • Data Provenance: Not explicitly stated, but the studies were conducted by Streck, Inc., implying internal company data.
    • Retrospective or Prospective: Given the nature of stability and precision studies for a quality control material, they would typically be prospective studies designed and executed specifically for this submission, following predefined protocols.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable. The device is a hematology quality control mixture, not a diagnostic device requiring expert interpretation for "ground truth." The "ground truth" for a control material would be its manufacturing specifications and the measured values on reference instruments.

    4. Adjudication Method for the Test Set

    This is not applicable for a quality control product study. Adjudication methods are typically used when subjective interpretations (e.g., image reads) are involved to resolve discrepancies among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC study was not done. This type of study is relevant for diagnostic devices (especially those involving human readers like radiologists) to assess the impact of AI assistance on human performance. The XN CHECK™ BF is a quality control material.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone performance study in the context of an "algorithm only" is not applicable. The device itself is a physical quality control material used with an analyzer, not an algorithm. The "performance" refers to the stability and reproducibility of the control material's measured parameters, not an algorithm's output.

    7. The Type of Ground Truth Used

    The "ground truth" for a quality control material like XN CHECK™ BF would be the expected target values and ranges for its assayed parameters, established through a rigorous manufacturing and assay process using reference methods and calibrated instruments. This is analogous to a reference standard, not ground truth derived from pathology, expert consensus, or outcomes data used for diagnostic algorithms.

    8. The Sample Size for the Training Set

    This is not applicable. The XN CHECK™ BF is a manufactured quality control material, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120744

    Validate with FDA (Live)

    Device Name
    XN CHECK BF
    Manufacturer
    Date Cleared
    2012-10-19

    (221 days)

    Product Code
    Regulation Number
    864.8625
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    XN CHECK BF is used for control and calibration verification of Sysmex XN series (XN-10, XN-20) analyzers. It is not, however, intended for actual calibration of these analyzers. Assayed parameters include:

    WBC-BF (103/μl), RBC-BF (10°/μl), MN# (103/μl), PMN# (103/μl), MN% (%), PMN% (%), TC-BF# (103/ul)

    Device Description

    XN CHECK™ BF is an in-vitro diagnostic product that contains the following: stabilized red blood cell component(s) and stabilized white blood cell component(s) in a preservative medium. The product is packaged in polypropylene plastic vials with screw caps containing 3 ml. The vials will be packaged in (4) welled vacuum formed clamshell container with the product information sheet / assay sheet. The product must be stored at 2 - 8° C.

    AI/ML Overview
    The provided document is a 510(k) summary for the Streck XN CHECK™ BF device, which is a hematology quality control mixture. It describes the device, its intended use, and compares it to a predicate device (Cell-Chex Auto-[K053362](https://510k.innolitics.com/device/K053362)). The document also includes information about the tests conducted to establish performance.
    
    Based on the provided text, here's an analysis of the acceptance criteria and study information:
    
    **1. Table of Acceptance Criteria and Reported Device Performance**
    
    The document does not explicitly state quantitative acceptance criteria in a table format for performance metrics. Instead, it makes general statements about the study results. The primary "performance" assessed for this type of device (a quality control mixture) relates to its stability and consistency, and its substantial equivalence to a predicate device.
    
    | Performance Metric          | Acceptance Criteria (Implied)                                    | Reported Device Performance                                    |
    |-----------------------------|------------------------------------------------------------------|----------------------------------------------------------------|
    | Open-Vial Stability         | Stability for 30 days (matching predicate device)                | "Same" as predicate (30 days)                                  |
    | Closed-Vial Stability       | Stability for at least 75 days (matching predicate device)       | 84 days (exceeds predicate's 75 days)                          |
    | Precision Performance       | Consistency and reproducibility                                  | "consistently reproducible"                                    |
    | Substantial Equivalence     | Performance comparable to the predicate device (Cell-Chex Auto)  | "substantially equivalent to the predicate product"            |
    | Safety and Effectiveness    | Device is safe and effective for its intended use                | "XN CHECK™ BF is a safe and effective product, which fulfills its intended use" |
    
    **2. Sample Size Used for the Test Set and Data Provenance**
    
    The document does not specify the exact sample sizes used for the Open-Vial Stability, Closed-Vial Stability, or Precision Performance tests. It simply states that "The following studies were conducted to establish performance of XN CHECK™ BF."
    
    Data Provenance: Not explicitly stated, but these types of studies for in-vitro diagnostic controls are typically conducted internally by the manufacturer using fresh device samples, often in a laboratory setting. It is not mentioned whether the data is retrospective or prospective, but stability studies inherently involve prospective evaluation over time.
    
    **3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts**
    
    This information is not applicable and therefore not provided in the document. For a hematology quality control mixture, "ground truth" isn't established by human expert consensus on interpretations of images or clinical data. Instead, the "truth" is derived from the established values of the control material, measured on high-precision reference instruments or by defined methods. The stability and precision studies would involve comparing measurements over time and across different runs/lots, rather than expert interpretation.
    
    **4. Adjudication Method for the Test Set**
    
    Not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve discrepancies in human readings or interpretations, typically in studies involving imaging or clinical diagnosis. This device is a quality control material, and its performance is assessed through quantitative laboratory measurements, not through subjective human interpretation requiring adjudication.
    
    **5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance**
    
    Not applicable. This device is an in-vitro diagnostic control, not an AI-powered diagnostic tool for human readers. Therefore, an MRMC study and a human-in-the-loop comparison with AI assistance are not relevant to its evaluation.
    
    **6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done**
    
    Not applicable. This device is a physical quality control material for analyzers, not a software algorithm. Therefore, "standalone algorithm performance" is not a relevant concept for this product. Its performance is inherent to its chemical and biological composition and how it performs on specific analyzers.
    
    **7. The Type of Ground Truth Used**
    
    The concept of "ground truth" as typically applied in AI/diagnostic imaging studies (e.g., pathology, clinical outcomes, expert consensus) does not directly apply here. For a quality control material like XN CHECK™ BF, the "ground truth" is established by:
    
    *   **Reference Methodology:** The manufacturing process and characterization of the control material itself, which targets specific analyte concentrations (e.g., WBC-BF, RBC-BF counts).
    *   **Assayed Values:** The expected range of values for each parameter (WBC-BF, RBC-BF, etc.) determined through rigorous testing on reference instruments and methods, often by the manufacturer. These assayed values serve as the "true" or target values against which the control material's performance (stability, precision) is measured when run on user instruments.
    *   **Comparison to Predicate Device:** Substantial equivalence implies that the performance characteristics (including stability and precision of measurements) are comparable to an already legally marketed and accepted control material, the Cell-Chex™ Auto.
    
    **8. The Sample Size for the Training Set**
    
    Not applicable. This device is not an AI algorithm that requires a "training set" in the conventional machine learning sense. The "training" here would be the manufacturing process and initial characterization of the control material.
    
    **9. How the Ground Truth for the Training Set was Established**
    
    Not applicable, as there is no "training set" in the context of an AI algorithm. The "truth" for this control material is established through its manufacturing specifications and extensive testing to determine its assayed values and stability characteristics, often using validated laboratory methods and reference instruments.
    
    Ask a Question

    Ask a specific question about this device