Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K191719

    Validate with FDA (Live)

    Device Name
    X-MIND Trium
    Manufacturer
    Date Cleared
    2020-02-20

    (239 days)

    Product Code
    Regulation Number
    872.1800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    X-MIND trium is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, indicated for use in:

    • producing panoramic X-ray images for diagnostic examination (teeth), jaws and oral structures;
    • producing radiographs of maxillofacial region and parts of the skull for cephalometric examination, if equipped with CEPH arm;
    • producing radiographs of hands and wrists for carpus examination, if equipped with CEPH arm;
    • producing tomographic images of the oral and maxillofacial region, for diagnostic examination of dentition (teeth), jaws, oral structures and some cranial bones, if equipped with CBCT option.
      From a clinical point of view, X-MIND trium can be applied for the following medical indications:
    • Generic dentistry
    • Dental implantology
    • Dental surgery
    • Maxillo-facial surgery
    • Cephalometric analysis
    • Carpus radiology
    Device Description

    X-MIND trium is a digital panoramic, cephalometric and tomographic extra-oral X-ray system. The system consists of the X-MIND trium SCANNER and the OPERATOR'S WORKSTATION. The scanner includes a control panel, X-ray generator, sliding body, U-arm, detector sliding group, CBCT detector, PAN detector, TLD CASSETTE (optional), patient support, F group, and column. The workstation runs ACTEON IMAGING SUITE (AIS) + 2D diagnostic analysis and AIS 3D app for CBCT dataset display software. The device can be configured as PAN only, PAN / CBCT, PAN / CEPH, PAN / CBCT / CEPH (PAN Full), PAN / CBCT TLD version, and PAN / CBCT / CEPH TLD version. It can be wall mounted or self-standing. Main changes include the addition of a new accessory, the X-MIND trium TLD cassette, including a new model of CBCT sensor that can be used for both panoramic and CBCT examination, and the introduction of CEPH and PAN sensors with a wider admitted range of power supply value. A new movement of the cassette has been added to modify the SID, allowing for special exams for children with reduced SID. A new AUXCBCT board has been designed to support the new movement and cassette. A new functionality, the Low Dose functionality, allows for CBCT exams using Low Dose protocols as an alternative to Standard Dose protocols, ensuring a radiation dose saving of about 50%.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the X-MIND trium device, based on the provided document:

    Acceptance Criteria and Device Performance

    The document describes several changes to the X-MIND trium device, and the acceptance criteria and performance are tied to these specific changes, particularly related to the new PAN-CBCT detector and the Low Dose (LD) functionality.

    1. Table of Acceptance Criteria and Reported Device Performance

    Device Change / CharacteristicAcceptance CriteriaReported Device Performance
    New PAN Detector (1)Same technology. DQE values for new detector are in the range ±0.1 on the whole frequency range (1 – 4 LP/mm).PASS (DQE values are within range, same technology)
    New CEPH Detector (2)Same technology. DQE values for new detector are in the range ±0.1 on the whole frequency range (1 – 4 LP/mm).PASS (DQE values are within range, same technology)
    New PAN-CBCT Detector (3)Image Quality vs. Dose Trade-off (System Indicator - SI_DAP): Same or superior trade-off between dose and image quality (i.e. System Indicator (SI) in CBCT) $S_{DAP} ≥ S_{DAP ; predicate}$ where $S_{DAP} = \frac{CNR}{DAP \cdot (\frac{1}{2 \cdot V_{50%}})^2}$ (CNR = contrast to noise ratio, DAP = dose area product, V50% = resolution indicator at 50% modulation). Voxel Noise: Voxel Noise [a.u.] with TLD cassette including new PAN-CBCT sensor (VNTLD) ≤ Voxel Noise [a.u.] predicate (VN). Noise Power Spectrum (NPS): NPS0D [a.u.²] with TLD cassette including new PAN-CBCT sensor ≤ NPS0D [a.u.²] predicate. X-ray Field to Image Receptor Correspondence (IEC 60601-2-63): along each of the two axes of the IMAGE RECEPTION AREA, the edges of the X-RAY FIELD shall not exceed the corresponding edges of the EFFECTIVE IMAGE RECEPTION AREA by more than 2 cm or 3 % of the indicated FOCAL SPOT TO IMAGE RECEPTOR DISTANCE, whichever the larger; the sum of the discrepancies on both axes shall not exceed 3 cm or 4 % of the indicated FOCAL SPOT TO IMAGE RECEPTOR DISTANCE, whichever the larger. X-ray Field to Image Receptor Correspondence (21 CFR 1020.31 f) 4)): field does not exceed each dimension of the image receptor by more than 2 percent of the SID; align the center of the x-ray field with the center of the image receptor to within 2 percent of the SID.Image Quality vs. Dose Trade-off (SI_DAP): - Woman Medium Standard quality FOV 80X80-80X90: SDAP TLD > SDAP pred (PASS) - Woman Medium Standard quality FOV 110X80: SDAP TLD > SDAP pred (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Adult: SDAP TLD > SDAP pred (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Child: SDAP TLD > SDAP pred (PASS) Voxel Noise: - Woman Medium Standard quality FOV 80X90: VNTLD < VN (PASS) - Woman Medium Standard quality FOV 110X80: VNTLD < VN (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Adult: VNTLD < VN (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Child: VNTLD < VN (PASS) Noise Power Spectrum (NPS): - Woman Medium Standard quality FOV 80X90: NPS0D TLD < NPS0D (PASS) - Woman Medium Standard quality FOV 110X80: NPS0D TLD < NPS0D (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Adult: NPS0D TLD < NPS0D (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Child: NPS0D TLD < NPS0D (PASS) X-ray Field to Image Receptor Correspondence (IEC 60601-2-63): - SID Adult edge ≤ 20mm (PASS) - SID Child edge ≤ 20mm (PASS) - Sum of discrepancies: < 30mm (PASS) - Along scanning axis: ≤ 1mm (PASS) - Perpendicular to scanning axis: =0 (PASS) X-ray Field to Image Receptor Correspondence (21 CFR 1020.31 f) 4)): - PAN: ≤ 2% (PASS) - PAN reduced SID: ≤ 2% (PASS) - CBCT: ≤ 2% (PASS) - CBCT reduced SID: ≤ 2% (PASS) - Center alignment (PAN and CEPH only): PASS
    New Child Special Exam (4)Usability (IEC 60601-1-6): All comments are positive. Biocompatibility (ISO 10993-5: 2009 & ISO 10993-10: 2010 for new tool): Cytotoxicity: No cytotoxic. Hypersensitivity: Not sensitizing.Usability: PASS (all comments positive). Biocompatibility: PASS (no cytotoxicity, not sensitizing, same material and process as predicate)
    True Low Dose (TLD) / Reduced SID (4)Image Quality (PAN mode): Absence of artifacts. Equal high contrast resolution (Both 3.1 lp/mm) and equal low contrast resolution (Both 4 holes visible). Declared DAP (PAN mode): Declared DAP PAN 38.77 < 52. (image CHILD NEDIUM SID) < predicate. Image Quality vs. Dose Trade-off (CBCT mode - SI_DAP): Same or superior trade-off between dose and image quality (i.e. System Indicator (SI) in CBCT) $S_{DAP} ≥ S_{DAP ; predicate}$.Image Quality (PAN mode): Both verified PASS (Absence of artifacts, 3.1 lp/mm high contrast resolution, 4 holes visible low contrast resolution). Declared DAP (PAN mode): PASS (38.77 < predicate for child). Image Quality vs. Dose Trade-off (CBCT mode - SI_DAP): - Woman Medium Standard quality FOV 80X80-80X90: SDAP TLD > SDAP pred (PASS) - Woman Medium Standard quality FOV 110X80: TLD > pred (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Adult: SDAP TLD = > SDAP pred (PASS) - Child medium Standard quality FOV 80X80-80X90 SID Child: SDAP TLD > SDAP pred (PASS)
    Detectability Study (3 & 7)Same pathological or clinically relevant conditions identifiable both in the STD protocol scan and in the TLD protocol one (for TLD cassette / reduced SID) AND Same pathological or clinically relevant conditions identifiable both in the STD protocol scan and in the LD protocol one (for Low Dose functionality).TLD cassette / Reduced SID: PASS. All pathological or clinically relevant conditions identified with STD protocol scan have been identified with the LD protocol one. Note Pos. 45 "worn out dental neck" not identified nor in standard mode nor in TLD configuration (this implies both performed equally, neither identified this specific condition). Low Dose Functionality (LD): PASS. All pathological or clinically relevant conditions have been identified both in the STD protocol scan and in the LD protocol one.
    Low Dose Functionality (7)Image Quality vs. Dose Trade-off (SI_DAP): Same or superior trade-off between dose and image quality. $S_{DAP low dose} ≥ S_{DAP standard protocol}$. Voxel Noise (VN): $\Delta%VN \le 10%$, where $\Delta%VN = \frac{VNld - VNs}{100 + VNsd}$ (VNld = Voxel Noise [a.u.] low dose protocol, VNsd = Voxel Noise [a.u.] standard protocol). Noise Power Spectrum (NPS): $\Delta%NPS \le 30%$, where $\Delta%NPS = \frac{NPSld - NPSsd}{100 + NPSsd}$ (NPSld = NPS0D [a.u.²] low dose protocol, NPSsd = NPS0D [a.u.²] standard dose protocol).Image Quality vs. Dose Trade-off (SI_DAP): - Man: SDAP low dose > SDAP standard protocol (PASS) - Woman: SDAP low dose > SDAP standard protocol (PASS) - Child: SDAP low dose > SDAP standard protocol (PASS) Voxel Noise: - Man: $\Delta VN% < 10%$ (PASS) - Woman: $\Delta VN% < 10%$ (PASS) - Child: $\Delta VN% < 10%$ (PASS) Noise Power Spectrum (NPS): - Man: $\Delta NPS < 30%$ (PASS) - Woman: $\Delta NPS < 30%$ (PASS) - Child: $\Delta NPS < 30%$ (PASS)

    2. Sample Size and Data Provenance

    The document mentions "bench tests" and a "clinical detectability analysis" using "radiographs of cadavers".

    • Test Set Sample Size: For the detectability analysis, the "radiographs of cadavers" constitute the test set for evaluating clinical performance. The number of cadavers scanned is not specified in the provided text. The number of specific conditions tested is also not explicitly stated, beyond the general statement of "All pathological or clinically relevant conditions".
    • Data Provenance: The data appears to be prospective as it involves new bench tests and a detectability study specifically designed to evaluate the changes in the device. The country of origin of the data is not specified, but the manufacturer is based in Italy.

    3. Number of Experts and their Qualifications

    • Number of Experts: For the detectability analysis, a "set of intended users" were used. The exact number is not specified.
    • Qualifications of Experts: The experts included "radiologist, dentist, oral surgeon." Specific years of experience are not mentioned.

    4. Adjudication Method

    • The document implies a consensus-based evaluation for the detectability study, stating that experts "evaluated the visibility of anatomical structures and features of interests in the different scanning conditions." However, a formal adjudication method (e.g., 2+1, 3+1, none) is not explicitly described. It states that "All pathological or clinically relevant conditions identified with STD protocol scan have been identified with the LD protocol one," suggesting an agreement on findings rather than a formal disagreement resolution process between multiple readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • A comparative effectiveness study in the sense of comparing human reader performance with and without AI assistance is not explicitly described or quantified.
    • The detectability study compares the device's performance (standard vs. low dose protocols) by assessing whether experts can identify pathological conditions in both, rather than assessing the uplift in human reader performance using AI. The changes described are enhancements to the imaging system itself (new detector, low dose protocols, reduced SID), not an AI-assisted diagnostic tool for readers.

    6. Standalone Performance Study

    • Yes, standalone performance studies were conducted for various technical aspects of the device. The document details extensive "bench testing" to evaluate parameters like:
      • DQE values for new PAN and CEPH detectors.
      • Trade-off between dose and image quality (SI_DAP) for the new PAN-CBCT detector and Low Dose functionality.
      • Voxel Noise and Noise Power Spectrum.
      • X-ray field to image receptor correspondence.
      • Mechanical stability (Self-standing version).
    • These tests were performed on the algorithm/device itself without human-in-the-loop intervention for the quantitative measurements, demonstrating the technical performance of the device's imaging capabilities.

    7. Type of Ground Truth Used

    • For the detectability analysis: The ground truth appears to be based on the identification of "pathological or clinically relevant conditions" by the experts in the standard protocol scan. The low dose scan was then compared against this baseline to ensure similar conditions could be identified. This suggests a form of expert consensus or established clinical findings from the standard protocol as the reference.
    • For the bench tests: The ground truth for technical parameters (DQE, CNR, DAP, Voxel Noise, NPS, field correspondence) is based on physical phantom measurements and engineering standards/specifications (e.g., IEC standards, physical dimensions, signal characteristics).

    8. Sample Size for the Training Set

    • The document describes changes to a medical imaging device (X-ray system), not an AI algorithm that requires a "training set" in the context of machine learning. Therefore, a training set sample size is not applicable in the conventional sense. The device's operation is based on established physics and engineering principles, and its performance is validated through bench tests and clinical detectability studies.

    9. How the Ground Truth for the Training Set Was Established

    • As explained above, the concept of a "training set" in this context is not applicable for an X-ray imaging system. The system's functionalities are developed based on engineering design, and its performance is validated against physical and clinical criteria, not by training on a dataset with pre-established ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160166

    Validate with FDA (Live)

    Device Name
    X-MIND trium
    Manufacturer
    Date Cleared
    2016-11-15

    (293 days)

    Product Code
    Regulation Number
    892.1750
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    X-MIND trium is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, indicated for use in:

    • producing panoramic X-ray images for diagnostic examination of dentition (teeth), jaws and oral structures;
    • producing radiographs of maxillofacial region and parts of the skull for cephalometric examination, if equipped with CEPH arm;
    • producing radiographs of hands and wrists for carpus examination, if equipped with CEPH arm;
    • producing tomographic images of the oral and maxillofacial region, for diagnostic examination of dentition (teeth), jaws, oral structures and some cranial bones, if equipped with CBCT option.

    From a clinical point of view, X-MIND trium can be applied for the following medical indications :

    • Generic dentistry
    • Dental implantology
    • Dental surgery
    • Maxillo-facial surgery
    • Cephalometric analysis
    • Carpus radiology
    Device Description

    X-MIND trium is a digital panoramic, cephalometric and tomographic extra-oral Xray system, which consists of:

    1. Operator's workstation
    2. X-MIND trium remote control and X-MIND trium light
    3. X-MIND Trium medical device (Control panel, X-ray generator, Sliding body, U-arm, Detector sliding group, CBCT detector, PAN detector, Patient support, F group, Column)
      Cephalometric extension (CEPH arm extension, CEPH control panel, CEPH patient support, CEPH detector sliding group, CEPH secondary collimator, CEPH detector)
      Accessories (Temple rest, Chin rest, Sliding bite block and chin rest assembly, TMJ nose support, Calibration tray and geometric calibration phantom, Quality phantom support for Germany)
    AI/ML Overview

    The provided text is a 510(k) summary for the X-MIND trium device, which is a digital panoramic, cephalometric, and tomographic extra-oral X-ray system. The document compares the proposed device to a predicate device, the Hyperion X9, to establish substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document implicitly uses the performance of the predicate device, Hyperion X9, as the acceptance criteria for substantial equivalence, particularly for image quality and radiation dose, while also assessing compliance with relevant safety standards. The table below summarizes the key performance comparisons presented as part of the substantial equivalence claim. Since specific numerical acceptance criteria (e.g., "must meet X value") are not explicitly stated as distinct criteria but rather as comparison outcomes, this table presents the comparative performance as reported.

    CharacteristicAcceptance Criteria (Predicate Device - Hyperion X9)Reported Device Performance (X-MIND trium)Outcome
    Cephalometric Image QualityTest pattern images of certain qualityTest pattern images of equivalent qualitySubstantial equivalence confirmed by metrics
    Cephalometric Radiation DoseUnder DRL for Dental lateral cephalogram adult (32 mGycm²) according to NCRP Report No. 172 (2016)Under DRL for Dental lateral cephalogram adult (32 mGycm²) according to NCRP Report No. 172 (2016)Both Hyperion and X-MIND trium are under the DRL
    Panoramic Image QualityTest pattern images of certain qualityTest pattern images of equivalent qualitySubstantial equivalence confirmed by metrics
    Panoramic Radiation DoseUnder DRL for Dental panoramic (100 mGycm²) according to NCRP Report No. 172 (2016)Under DRL for Dental panoramic (100 mGycm²) according to NCRP Report No. 172 (2016)Both Hyperion and X-MIND trium are under the DRL
    CBCT Image Quality (Standard Quality Protocols)Homogeneity: baseline as Hyperion X9Homogeneity: 2.8-10.9 times lower than Hyperion X9Homogeneity for Hyperion is 2.8-10.9 times higher than X-MIND trium
    CNR: baseline as Hyperion X9CNR: 2.12 ÷ 2.37 times higher than Hyperion X9CNR for Trium is 2.12 ÷ 2.37 times higher than the CNR for Hyperion
    MTF: baseline as Hyperion X9MTF plots show Trium better maintains high frequencies between 1.25 lp/mm and 1.5 lp/mmTrium better maintains high frequencies
    Nyquist frequency: baseline as Hyperion X9EquivalentEquivalent
    Absence of artifacts: baseline as Hyperion X9EquivalentEquivalent
    CBCT Radiation Dose (Standard Quality Protocols)Similar to Hyperion X9, within reasonable variationLower than Hyperion X9, but X-MIND trium dose can be reduced by using other scan protocols (implicitly aiming for equivalence or improvement)Hyperion is lower but the dose for X-MIND trium can be reduced by using other scan protocols (suggesting overall acceptable dose, though initially higher for some protocols if not adjusted)
    CBCT System Indicator ($S_{DAP}$)Substantial equivalence to Hyperion X9Calculated $S_{DAP}$ value (not explicitly reported but claimed to establish equivalence)Substantial equivalence is confirmed by $S_{DAP}$
    CBCT Image Quality (High Quality Protocols)Noise: baseline as Hyperion X9Voxel noise for Hyperion is 43.5% ÷ 53.8% higher than TriumVoxel noise for Hyperion is higher than Trium (implying Trium has lower noise)
    NPS: baseline as Hyperion X9Noise content very similar at low frequencies, Hyperion shows higher noise at medium and high frequencyHyperion shows higher noise at medium and high frequencies (implying Trium has lower noise)
    CBCT Radiation Dose (High Quality Protocols)Substantial equivalence to Hyperion X9Metrics confirmed substantial equivalenceSubstantial equivalence is confirmed by metrics
    Electrical Safety & EMCCompliance with IEC 60601-1 3rd ed., IEC 60601-1-2: 2007, etc. (as per predicate)Compliance with same standards, plus US and Canada deviations (IEC 60601-1 deviations for US and Canada), IEC 60601-1: 2007, ANSI/AAMI ES60601-1: 2005, CAN/CSA C22.2 No. 60601-1:08, IEC 60601-1-3: 2008, IEC 60601-2-63: 2012, IEC 60601-1-6: 2010, IEC 62366: 2007Met and exceeded standards, with formal certifications from NRTLs.
    Software Verification & ValidationAs recommended by FDA GuidanceDocumentation provided as recommended by FDA Guidance for software in medical devicesSoftware considered "moderate" level of concern, V&V testing conducted. General tools and means made available by both software are equivalent.

    2. Sample size used for the test set and the data provenance:

    The document mentions that "bench tests" were performed for image quality and radiation dose. It refers to "test pattern images" and "clinical images obtained with X-MIND trium and predicate device." However, the sample size for the test set (number of images/cases) is not explicitly stated.

    The data provenance is likely retrospective, as the comparison involves existing images from the predicate device and newly generated images from the X-MIND trium under controlled bench test conditions. The country of origin of the data is not specified but given the manufacturer's location in Italy and the use of a Nationally Recognized Testing Laboratory (NRTL) (Intertek Semko, Nemko S.p.A.), it is likely that the testing was performed in Europe.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    The document does not mention the use of human experts to establish ground truth for the test set. The image quality assessments appear to be based on objective metrics derived from phantom studies (e.g., homogeneity, CNR, MTF, Nyquist frequency, artifacts, NPS), rather than subjective evaluations by clinical experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as human expert evaluation for ground truth determination is not described. The assessment was based on objective metrics from bench tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This device is an X-ray imaging system, not an AI-assisted diagnostic tool. The focus of the submission is on hardware and software equivalence for image acquisition and display, not on reader performance improvement with AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, standalone performance was done, but not in the context of an "algorithm" as typically conceived in AI/ML performance. The "device performance" described (image quality metrics, radiation dose, electrical safety, EMC) is the performance of the X-MIND trium system itself in a standalone capacity, i.e., without human intervention in the image generation part of the test. The "software" component also underwent verification and validation, but this refers to its functionality and reliability in image management and processing features available to the user, not an AI algorithm performing diagnostic tasks.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the performance evaluation mostly relies on objective physical measurements and phantom studies. For image quality, metrics like homogeneity, CNR, MTF, Nyquist frequency, and NPS derived from test patterns or phantoms serve as the "truth" against which the device's imaging capabilities are measured. For radiation dose, established Dosimetry Reference Levels (DRLs) from NCRP REPORT No. 172 (2016) serve as the standard. For electrical safety and EMC, recognized international standards (e.g., IEC 60601 series) are the ground truth for compliance.

    8. The sample size for the training set:

    Not applicable. The device is an X-ray imaging system, not an AI/ML diagnostic algorithm that requires a training set of data.

    9. How the ground truth for the training set was established:

    Not applicable. As stated above, this device does not involve a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1