510(k) Data Aggregation

    K Number
    K250087

    Validate with FDA (Live)

    Device Name
    Vscan Air
    Date Cleared
    2025-05-01

    (107 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vscan Air is a battery-operated software-based general-purpose ultrasound imaging system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The device is enabling visualization and measurement of anatomical structures and fluid including blood flow.

    Vscan Air's pocket-sized portability and simplified user interface enables integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and other environments as described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage.

    Vscan Air supports Black/ white (B-mode), Color flow (Color doppler), Pulsed wave Doppler mode, M-mode, combined (B + Color Doppler) and Harmonic Imaging modes with curved, linear and sector array transducers.

    With the curved array transducer of the dual headed probe solution, the specific clinical applications and exam types include: abdominal, fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult and pediatric, 40 kg and above), vascular/peripheral vascular, musculoskeletal (conventional), pediatrics, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

    With the linear array transducer of the dual headed probe solution, the specific clinical applications and exam types include: vascular/peripheral vascular, musculoskeletal (conventional and superficial), small organs, thoracic/lung, ophthalmic, pediatrics, neonatal cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block, vascular access and biopsy).

    With the sector array transducer of the dual headed probe solution, the specific clinical applications and exam types include: cardiac (adult and pediatric, 40 kg and above), abdominal, fetal/obstetrics, gynecological, urology, thoracic/ lung, pediatrics, adult cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

    Device Description

    Vscan Air™ is a battery-operated general-purpose diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. It enables ultrasound imaging guidance, visualization and measurement of anatomical structures and fluid.

    Vscan Air consists of an app which can be installed on Android™ or iOS devices, and 2 probes which use wireless technology for communication.

    Its pocket-sized portability and simplified user interface enable integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and in other environments. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage assessments for adult, pediatric and neonatal patients. Vscan Air can also be useful for interventional guidance.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Vscan Air's Auto Bladder Volume feature, extracted from the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Device Performance for Auto Bladder Volume

    Acceptance CriteriaReported Device Performance
    At least 90% success rate in automatic caliper placement for bladder volume measurements when bladder wall is entirely visualized.Automatic caliper placement success rate: 92.24% with a 90% confidence level. Consistent performance across key subgroups, e.g., BMI Overweight (>25): 92%, BMI Normal (<25): 95%.

    Study Details for Auto Bladder Volume (Test Set)

    1. Sample size used for the test set and data provenance:

      • Sample Size: 1,817 images from 142 individuals. Each individual was scanned in two views (Transverse and Longitudinal).
      • Data Provenance: The dataset included images from individuals of various ethnicities/countries, including USA, Germany, UK, Japan, and India. The study used retrospective data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the exact number of experts, but states that the verification dataset was assessed by "experts for accuracy." It does not provide their specific qualifications (e.g., years of experience as radiologists).
    3. Adjudication method for the test set:

      • The document describes "annotators" performing manual annotation on images to establish ground truth. It does not explicitly mention a formal adjudication method like "2+1" or "3+1" for resolving disagreements among multiple annotators.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

      • No, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was not done. This study solely focused on the standalone performance of the Auto Bladder Volume feature.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done for the Auto Bladder Volume feature. The performance reported is that of the algorithm's automatic caliper placement.
    6. The type of ground truth used:

      • Expert Consensus/Manual Annotation. Ground truth annotations were obtained through manual placement of landmarks (bladder edges) by annotators on images, representing where measurement calipers would be placed.
    7. The sample size for the training set:

      • The total dataset included 4,014 images from 301 individuals. Since 1,817 images from 142 individuals were used for verification, the remaining 2,197 images from 159 individuals were used for training/validation (tuning).
    8. How the ground truth for the training set was established:

      • Similar to the verification dataset, "annotators performed manual annotation on images converted from vscanet files." These annotations involved identifying 4-6 images representing different bladder volume statuses, and on each view (transverse and longitudinal), marking 4 distinct landmarks representing the bladder edges.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231301

    Validate with FDA (Live)

    Device Name
    Vscan Air
    Date Cleared
    2023-08-15

    (102 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vscan Air is a battery-operated software-based general-purpose ultrasound imaging system for use by qualified and trained healthcare professionals or practitioners that are legally authorized or licensed by law in the country, state or other local municipality in which he or she practices. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The device is enabling visualization and measurement of anatomical structures and fluid including blood flow.
    Vscan Air's pocket-sized portability and simplified user interface enables integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and other environments as described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage.
    Vscan Air supports black/white (B-mode), color flow (Color Doppler mode, M-mode, M-mode, combined (B + Color Doppler) and Harmonic Imaging modes with curved, linear and sector array transducers.
    With the curved array transducer of the dual headed probe solution, the specific clinical applications and exam types include: abdominal, fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult and pediatio, 40 kg and above), vascular/peripheral vascular, musculoskeletal (conventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).
    With the linear array transducer of the dual headed probe solution, the specific clinical applications and exam types include: vascular/peripheral vascular, musculoskeletal (conventional and superficial), small organs, thoraciclung, ophthalmic, pediatics, neonatal cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block, vascular access and biopsy).
    With the sector array transducer of the dual headed probe solution, the specific clinical applications and exam types include: cardiac (adult and pediatic, 40 kg and above), abdominal, fetal/obstetnes, gynecological, urology, thoracic/ lung, pediatics, adult cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

    Device Description

    Vscan Air™ is a battery-operated general-purpose diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. It enables ultrasound imaging guidance, visualization and measurement of anatomical structures and fluid.
    Vscan Air consists of an app which can be installed on Android™ or iOS devices, and 2 probes which use wireless technology for communication.
    Its pocket-sized portability and simplified user interface enable integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and in other environments. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage assessments for adult, pediatric and neonatal patients. Vscan Air can also be useful for interventional guidance.

    AI/ML Overview

    The provided text is a 510(k) Summary for the GE Medical Systems Ultrasound and Primary Care Diagnostics Vscan Air, outlining its substantial equivalence to predicate devices. It does not contain information on specific acceptance criteria and the detailed study results that prove the device meets these criteria in terms of clinical performance (e.g., accuracy, sensitivity, specificity) for a particular AI function.

    The document primarily focuses on:

    • Regulatory information: Device name, regulation, product codes, predicate devices.
    • Intended Use/Indications for Use: Describes the general applications of the ultrasound system.
    • Technological comparison: Explains how the Vscan Air is similar to its predicate and reference devices in terms of technology, modes, transducers, and software features.
    • Non-Clinical Tests: Mentions evaluation for acoustic output, biocompatibility, cleaning/disinfection, and electrical/electromagnetic/mechanical safety, and adherence to various safety standards.
    • Absence of Clinical Studies: Explicitly states, "The subject of this premarket submission, Vscan Air, did not require clinical studies to support substantial equivalence."

    Therefore, based on the provided input, I cannot fulfill the request for:

    1. A table of acceptance criteria and reported device performance (for clinical/AI performance).
    2. Sample size used for the test set and data provenance.
    3. Number of experts and their qualifications for ground truth.
    4. Adjudication method for the test set.
    5. MRMC study details or effect size.
    6. Standalone performance details.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    The document confirms that this particular submission for Vscan Air did not require clinical studies, implying that the performance evaluation relied on non-clinical testing and comparison to predicate devices, rather than a prospective clinical trial with a defined test set and established ground truth for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202035

    Validate with FDA (Live)

    Device Name
    Vscan Air
    Date Cleared
    2020-11-20

    (120 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vscan Air is a battery-operated software-based general-purpose ultrasound imaging system for use by qualified and trained healthcare professionals or practitioners that are legally authorized by law in the country, state or other local municipality in which he or she practices. The users may or may not be working under supervision or authority of a physician. Users may also include medical students working under the supervision or authority of a physician during their education / training. The device is enabling visualization and measurement of anatomical structures and fluid including blood flow.

    Vscan Air's pocket-sized portability and simplified user interface enables integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and other environments as described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage.

    Vscan Air supports Black/ white (B-mode), Color flow (Color doppler), Combined (B + Color Doppler) and Harmonic imaging modes with both the curved and linear array transducers.

    With the curved array transducer of the dual headed probe solution, the specific clinical applications and exam types include: abdominal, fetal/obstetrics, gynecological, urology, thoracic/lung, cardiac (adult and pediatric, 40 kg and above), vascular/peripheral vascular, musculoskeletal (conventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block and biopsy).

    With the linear array transducer of the dual headed probe solution, the specific clinical applications and exam types include: vascular/peripheral vascular, musculoskeletal (conventional and superficial), small organs, thoraciclung, ophthalmic, pediatrics, neonatal cephalic, interventional guidance (includes free hand needle/catheter placement, fluid drainage, nerve block, vascular access and biopsy).

    Device Description

    Vscan Air™ is a battery-operated general-purpose diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals. It enables ultrasound imaging guidance, visualization and measurement of anatomical structures and fluid.
    Vscan Air consists of an app which can be installed on Android™ or iOS devices, and a probe which uses wireless technology for communication.

    Its pocket-sized portability and simplified user interface enable integration into training sessions and examinations in professional healthcare facilities (ex. Hospital, clinic, medical office), home environment, road/air ambulance and in other environments. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic follow-up, and triage assessments for adult, pediatric and neonatal patients. Vscan Air can also be useful for interventional guidance.

    AI/ML Overview

    The provided document indicates that the Vscan Air device did not require clinical studies to support substantial equivalence. Therefore, there is no detailed information regarding acceptance criteria, device performance, sample sizes for test or training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for a study proving the device meets acceptance criteria.

    The document primarily focuses on non-clinical tests to demonstrate substantial equivalence to predicate devices. These non-clinical tests pertain to acoustic output, biocompatibility, cleaning and disinfection, as well as thermal, electrical, electromagnetic, mechanical safety, and wireless communication. The device was found to conform to applicable medical device safety standards.

    Summary of Non-Clinical Tests and Conformance:

    Test CategoryStandards/Requirements Conformed To
    Acoustic OutputIEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017. Acoustic power levels are below FDA applicable limits.
    BiocompatibilityISO 10993-1, Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing- Third Edition, 2009. Probe materials are biocompatible.
    Cleaning & DisinfectionEvaluated
    Electrical SafetyAAMI/ANSI ES60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance, 2005/A2:2012. IEC 60601-2-37, Medical Electrical Equipment Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015. IEC 60601-1-11 and IEC 60601-1-12 for home and EMS environments.
    Electromagnetic SafetyIEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2014.
    Mechanical SafetyEvaluated
    WirelessEvaluated (uses Wi-Fi direct)
    Risk ManagementISO 14971, Application of risk management to medical devices, 2012.

    Key takeaway regarding clinical studies:

    • Clinical Studies: "The subject of this premarket submission, Vscan Air, did not require clinical studies to support substantial equivalence."

    Therefore, the requested information regarding a study proving acceptance criteria for clinical performance is not available in the provided text. The submission relied on non-clinical testing and comparison to predicate devices to establish substantial equivalence.