510(k) Data Aggregation

    K Number
    K200724

    Validate with FDA (Live)

    Date Cleared
    2020-09-01

    (166 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use for:
    · the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.
    • for anterior capsulotomy during cataract surgery.
    · the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
    · laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts

    Device Description

    The VICTUS Femtosecond Laser is a precision ophthalmic surgical laser indicated for use in patients undergoing ophthalmic surgery in the anterior segment of the eye. The VICTUS Femtosecond Laser system produces scanned patterns designed to create flaps, produce corneal incisions, arcuate incisions, capsulotomy cuts, and crystalline lens fragmentation patterns for use in cataract surgery. The patient is treated while supine on a patient bed that the physician can position electromechanically via simple controls. The physician controls the VICTUS laser system itself through a touchscreen display and other user controls. Laser emission and of PI suction vacuum are actuated by the physician using separate footswitch pedals.

    The mode of operation for the VICTUS Platform is using Yb:KYW Femtosecond laser to produce Laser-Induced Optical Breakdown (LIOB) of the corneal tissue. Scanned patterns of laser pulses from the VICTUS are delivered through a sterile, disposable Patient Interface (PI). The PI consists of two components: (i) a contacting lens, and (ii) a suction clip. Together, the suction clip and contacting lens connect to make a single assembly during a VICTUS procedure. The surface of the contacting lens that contacts the eye is spherically curved to match the curvature of the human eye. The suction clip assembly gently immobilizes the eye with respect to the VICTUS laser beam and optical visualization systems. The PI contacting lens and suction clip together allow for proper optical coupling of the laser and optical paths into the patient's eye and create a precise opto-mechanical reference surface for depth control of the scanning laser beam.

    AI/ML Overview

    The provided text focuses on the 510(k) premarket notification for the VICTUS Femtosecond Laser Platform, specifically regarding modifications to its Patient Interface. It primarily discusses the technological comparison and non-clinical performance data to establish substantial equivalence to a previously cleared device. Therefore, the information regarding in-depth acceptance criteria and a standard study that proves the device meets those criteria (especially in the context of AI/ML performance) is not available in the provided document.

    However, based on the non-clinical performance data section, we can infer some aspects related to acceptance criteria and the study performed, even though it's not presented in a typical AI/ML study format.

    Here's an attempt to structure the answer based on the available information, noting where specific details are absent:

    Acceptance Criteria and Study for VICTUS Femtosecond Laser Platform (Modified Patient Interface)

    This submission describes changes to the Patient Interface (PI) of the VICTUS Femtosecond Laser Platform. The study's primary goal was to demonstrate that these modifications do not negatively impact the device's performance compared to the previously cleared version, thus establishing substantial equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Inferred from study goals)Reported Device Performance
    BiocompatibilityMeet ISO 10993-1 requirements for patient-contacting material.Test results satisfied the acceptance criteria as defined by the associated ISO standards. (Implies the modified PI is biocompatible).
    Endotoxin ContentLimulus Amebocyte Lysate (LAL) Endotoxin testing: Endotoxin levels < 20 EU/medical device (non-intraocular).Test results of < 1.0 EU/sample, which is below the defined endotoxin limit for a medical device (non-intraocular) of 20 EU/medical device.
    Cut Pattern FidelityCuts made with modified PI should be equivalent to those with cleared PI for all indications.Bench testing results demonstrated equivalent performance between the modified and current 510(k)-cleared PI Kits. Cuts made using the modified PI Kit were shown to be equivalent in terms of cut pattern fidelity for all cleared cataract and corneal indications for use.
    Cut QualityCuts made with modified PI should be equivalent to those with cleared PI for all indications.Bench testing results demonstrated equivalent performance. Cuts made using the modified PI Kit were shown to be equivalent in terms of cut quality for all cleared cataract and corneal indications for use.
    Docking Behavior (IOP Safety)Equivalent Intraocular Pressure (IOP) safety between modified and cleared PI during docking.The docking behavior in terms of IOP safety is equivalent between the current 510(k)-cleared and the modified (proposed) PI Kit models.
    Docking Behavior (Limbal Suction Integrity)Equivalent limbal suction integrity between modified and cleared PI during docking.The docking behavior in terms of limbal suction integrity is equivalent between the current 510(k)-cleared and the modified (proposed) PI Kit models.
    Electrical Safety & EMCComply with applicable safety standards (EN ISO 60601-1, -1-2, -2-22). (No new testing required for this submission, as previous compliance was established).Previously undergone testing and complies with applicable safety standards EN ISO 60601-1: 2005 + A1, EN ISO 60601-1-2: 2007, and EN ISO 60601-2-22: 2007. This establishes that the platform itself (not the new PI specifically) meets these.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for each non-clinical test. The document mentions "All bench testing results" without detailing the number of tests performed or the quantities of PI kits/cuts analyzed for each criterion (fidelity, quality, docking behavior).
    • Data Provenance: The studies are described as "Non-Clinical Performance Data" and "bench testing." This implies internal laboratory studies conducted by the manufacturer (Technolas Perfect Vision GmbH). The country of origin is not specified but is likely within the company's operational regions (e.g., Munich, Germany or San Francisco, CA where the contact person is located). It is retrospective in the sense that it evaluates a modified component against an existing, cleared component.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Not Applicable / Not Mentioned: This type of information (number of experts, qualifications) is typically relevant for studies involving human interpretation or clinical endpoints, especially in AI/ML performance evaluation. The presented data is from bench testing, where "ground truth" would be established by objective measurements (e.g., precise measurements of cut parameters, biocompatibility assay results). Experts would be involved in interpreting the objective data, but not in establishing a subjective "ground truth" in the way radiologists might for medical images.

    4. Adjudication Method for the Test Set

    • Not Applicable / Not Mentioned: Adjudication methods (e.g., 2+1, 3+1) are for resolving discrepancies among multiple human readers. This was a non-clinical bench testing study comparing device performance metrics, not human expert assessments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No: No MRMC or human-in-the-loop study was done. This notification is for a hardware modification (Patient Interface) of a laser platform, not an AI/ML algorithm requiring human reader performance evaluation. The document explicitly states: "There was no clinical study performed to support substantial equivalence of this premarket notification." Additionally, "No new software testing was performed to support this premarket notification."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No: This is not an AI/ML algorithm. The performance evaluation focuses on the mechanical and optical characteristics of a physical device component (Patient Interface).

    7. The Type of Ground Truth Used

    • The ground truth for the non-clinical performance data appears to be objective measurements and laboratory assays comparing the modified Patient Interface (PI) to the predicate PI. This includes:
      • Quantitative measurements of cut pattern fidelity and quality (e.g., dimensions, regularity).
      • Biocompatibility assay results (pass/fail criteria per ISO standards).
      • Endotoxin concentration measurements.
      • Measurements related to docking behavior outcomes (IOP, suction integrity).

    8. The Sample Size for the Training Set

    • Not Applicable: This is not an AI/ML device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable: As there is no training set for an AI/ML algorithm involved, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171014

    Validate with FDA (Live)

    Date Cleared
    2017-10-19

    (198 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use in:

    • · the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
    • · for anterior capsulotomy during cataract surgery
    • · the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
    • · laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
    Device Description

    The VICTUS Femtosecond Laser Platform (VICTUS) is a precision ophthalmic surgical laser and is cleared for use in the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for patients undergoing anterior capsulotomy during cataract surgery. The VICTUS has additionally been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea, primary/secondary corneal incisions, and laser-assisted lens fragmentation. This submission includes the next software iteration of version 3.3 and introduces a new S60 patient bed option for the VICTUS Platform. The v3.3 software introduces an additional pattern (grid pattern) which is a variation of currently cleared patterns and is being added to the already cleared lens fragmentation capability. In addition, the software introduces a lens apex detection feature. The previously cleared patient contacting lens and previously cleared suction clip assembly are combined in one convenient package for the user as the VERAFIT patient interface kit. With the exception of the introduction of a new S60 patient bed option, the VICTUS Platform hardware remains unchanged from that described and cleared via K151161. The mode of operation for the VICTUS Platform is using Yb:KYW Femtosecond laser to produce Laser-Induced Optical Breakdown (LIOB) of the corneal tissue. For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

    AI/ML Overview

    This document is a 510(k) summary for the VICTUS Femtosecond Laser Platform, Model 14000. It describes the device, its indications for use, and a comparison to a predicate device (K151161). The document's purpose is to demonstrate substantial equivalence to the predicate device, not necessarily to detail comprehensive acceptance criteria and a full study proving the device meets them in the way a medical AI/software device would.

    Based on the provided text, the device is a VICTUS Femtosecond Laser Platform, Model 14000, and the submission focuses on new software (v3.3) and a new patient bed option. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" are discussed in the context of demonstrating substantial equivalence to a predicate device, rather than proving a new clinical claim with specific performance metrics against a defined standard.

    Here's an attempt to answer your questions based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of clinical acceptance criteria with specific performance metrics (e.g., accuracy, sensitivity, specificity, AUC) for the software changes. Instead, it refers to pre-defined acceptance criteria for non-clinical performance data (bench testing) and compliance with safety standards.

    Acceptance Criteria CategoryReported Device Performance (Summary from text)
    Non-Clinical Performance"The acquired test data successfully verified that the various parameters meet their pre-defined acceptance criteria." (Specific parameters and their criteria are not detailed, but it implies successful bench testing using agarose gel, PETG, and PMMA).
    BiocompatibilityNo new testing performed. Assumed to be met based on predicate device.
    Electrical Safety & EMC"The VICTUS Femtosecond Laser Platform has undergone testing and is in compliance with applicable safety standards as listed in the following table [EN ISO 60601-1: 2005 + A1, EN ISO 60601-1-2: 2007, EN ISO 60601-2-22: 2007]." "The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate device."
    Mechanical & AcousticNo testing performed (assumed unchanged from predicate).
    Software Verification & ValidationConducted and documentation provided as recommended by FDA's Guidance. Software considered "major" level of concern. No specific performance metrics for the new features (grid pattern, apex detection) are provided in terms of accuracy or efficacy, but rather a statement of successful verification and validation.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size:
      • For Non-Clinical Performance Data (Bench Testing): The document states "A variety of test procedures were conducted" and "Different materials were used for the bench performance testing including: agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA)." However, it does not specify sample sizes (e.g., number of tests, number of samples of each material).
      • For Clinical Studies: "There was no clinical study performed to support substantial equivalence of this premarket notification." Therefore, there is no clinical test set sample size.
    • Data Provenance: Not specified for the non-clinical (bench) testing. Since no clinical studies were performed, there's no clinical data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No clinical ground truth was established as no clinical studies were performed. The "ground truth" for the bench testing would be metrology readings against engineering specifications, not expert consensus on medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical ground truth or human interpretation/adjudication was performed for a test set in the context of this 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study was done, as this is a laser surgical device, not an AI diagnostic/interpretive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device is not an algorithm-only device. It's a surgical laser platform. The software (v3.3) introduces features like a new grid fragmentation pattern and an apex detection feature. The apex detection feature is described as "a user convenience and the physician must still review positioning results and confirm the treatment plan prior to initiating treatment," indicating it's a human-in-the-loop system where the software provides assistance, but the human (physician) retains ultimate control and decision-making power. Therefore, no standalone algorithm-only performance was done in the sense of a diagnostic AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical performance data (bench testing), the "ground truth" would be the engineering specifications and expected outcomes from the laser's interaction with the test materials (agarose gel, PETG, PMMA). This is not expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning model in the sense that it requires a "training set" of data for learning. The software (v3.3) introduces pre-programmed patterns and detection features based on established physics and engineering, not a learned model from data.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151161

    Validate with FDA (Live)

    Date Cleared
    2015-06-24

    (54 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use for:

    · the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.

    · for anterior capsulotomy during cataract surgery.

    · the creation of cuts / incisions in the comea in patients undergoing cataract surgery or other ophflaalmic treatment requiring cuts / incisions in the comea.

    · laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts

    Device Description

    The VICTUS Femtosecond Laser Platform (VICTUS) is a precision ophthalmic surgical laser and is cleared for use in the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for patients undergoing anterior capsulotomy during cataract surgery (via K120426). The VICTUS has additionally been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea (K122386), primary/secondary corneal incisions (K132534), and laser-assisted lens fragmentation (K140615). This submission does not add to the indications previously cleared for the VICTUS Platform. With the exception of a new patient interface suction clip, the VICTUS Platform hardware and software remain unchanged from that described and cleared via K141379.

    A new patient interface suction clip for the VICTUS platform is being introduced within this premarket submission, the new suction clip is smaller in diameter and contains a colorant to aid in positioning on the cornea of the eye. For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. The contact lens portion of the patient interface remains unchanged from that previously cleared. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

    The fundamental scientific technology and indications for use remains the same as previously cleared for the VICTUS Femtosecond Laser Platform under K141379.

    AI/ML Overview

    The provided text describes the Victus Femtosecond Laser Platform, which is regulated as an Ophthalmic Laser. The document is a 510(k) premarket notification summary, indicating substantial equivalence to previously cleared devices.

    Here's an analysis of the acceptance criteria and supporting studies, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The text does not explicitly provide a table of quantitative acceptance criteria for device performance. Instead, it states that the device was found to perform "equivalently to the predicate device" and that "the acquired test data successfully verified that the various parameters meet their pre-defined acceptance criteria." This implies that specific performance metrics and their acceptance thresholds exist but are not detailed in this summary.

    However, based on the non-clinical and clinical performance data, we can infer some performance aspects:

    Feature/Parameter TestedAcceptance Criteria (Implied)Reported Device Performance
    Non-Clinical Performance
    Consistency of lens fragmentation with new software (v3.2) vs. old software (v2.7 SP3)Substantially equivalent results in lens fragmentationAchieved substantially equivalent results.
    Various parameters (e.g., cut quality, precision) across different materials (porcine eyes, agarose gel, PETG, PMMA)Pre-defined acceptance criteriaData successfully verified that parameters meet pre-defined acceptance criteria.
    Clinical Performance
    Intraoperative accuracy of biometry visualization (Lens Thickness, ACD, CCT) by VICTUS OCT vs. LENSTAR LS900Correlation and agreement between VICTUS OCT and LENSTAR LS900 measurementsResults are "consistent" between VICTUS OCT and LENSTAR LS900.
    Completion of laser-assisted capsulotomy without manual intervention100% completion rate without manual interventionAll 113 laser-assisted capsulotomy procedures were completed fully without the need for manual intervention.
    Absence of adverse eventsNo adverse or serious adverse eventsNo adverse or serious adverse events reported.
    Absence of device malfunctionsNo device malfunctions leading to adverse eventsNo device malfunctions reported which led or potentially led to adverse or serious adverse events.

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-Clinical Test Set:

      • Sample Size: Not explicitly stated as a number of distinct "test sets" but mentions "four different materials" (porcine eyes, agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA)) were used for bench performance testing.
      • Data Provenance: Retrospective (bench testing/lab data). Country of origin is not specified, but the applicant is based in Germany, so it's likely European or global.
    • Clinical Test Set:

      • Sample Size:
        • Flap group: 58 eyes
        • Cataract group: 113 eyes
        • Total: 171 eyes
      • Data Provenance: Prospective, single-center clinical study. The country of origin for the study is not specified but the applicant is based in Germany.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts used to establish ground truth or their specific qualifications for either the non-clinical or clinical studies.

    • Non-Clinical: Ground truth for material testing (e.g., cut quality, precision) would likely be established by engineers or scientists performing the tests, comparing measurements to specifications.
    • Clinical: For biometry, the LENSTAR LS900 measurements served as a reference (ground truth proxy). For capsulotomy completion and adverse events, the operating surgeons/investigators would make those assessments, but their specific expert qualifications are not detailed beyond "multi-surgeon" and "investigators."

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the test set data.

    • Biometry: The comparison was statistical ("correlation and agreement") between the VICTUS OCT and LENSTAR LS900.
    • Capsulotomy Completion/Adverse Events: These were reported by investigators, implying direct observation and recording without a separate adjudication process mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No, an MRMC comparative effectiveness study involving human readers improving with AI assistance was not conducted or described. This device is an ophthalmic laser for surgical procedures, not an AI diagnostic tool primarily interpreted by human readers. The clinical study focused on the device's accuracy for biometry and its effectiveness in performing specific surgical steps like capsulotomy and lens fragmentation. The "OCT subassembly" is part of the device's guidance system, not an AI assistance for human image interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance of the VICTUS Femtosecond Laser Platform itself, as a device performing surgical actions (cuts, fragmentation, imaging), is inherently standalone in its operation. However, surgical procedures always involve a human surgeon in the loop for planning, supervision, and execution. The "standalone" aspect, in this context, would relate to the laser's ability to precisely execute its programmed actions and measurements.

    • Non-Clinical Tests: These were standalone tests of the device's physical performance characteristics.
    • Clinical Studies: The study confirmed the "intraoperative accuracy of biometry visualization performed by VICTUS optical coherence tomography (OCT)" and the laser's ability to complete capsulotomy. This can be considered evaluating the standalone performance of the device's functions, although within the context of a surgical procedure where a surgeon operates the device.

    7. The Type of Ground Truth Used

    • Non-Clinical: The ground truth for bench testing would typically be established through precise metrological measurements and adherence to predefined engineering specifications. For material like porcine eyes, histopathology might be used to confirm laser effects, though not explicitly stated.
    • Clinical:
      • Biometry: The LENSTAR LS900 was used as the reference standard for axial optical resolution accuracy (lens thickness, anterior chamber depth, and central cornea thickness). This acts as a proxy for ground truth, assuming the LENSTAR LS900 is accepted as an accurate measurement device.
      • Capsulotomy Completion: The ground truth was observed clinical outcome, specifically whether the capsulotomy was completed fully by the laser without manual intervention.
      • Adverse Events: The absence of reported adverse events constitutes the ground truth for safety in the study.
      • Cataract Grades: Clinical assessment (implicitly by ophthalmologists) was the ground truth for cataract grading.

    8. The Sample Size for the Training Set

    The document does not mention any "training set." This summary pertains to the evaluation of a physical medical device (femtosecond laser) whose functionality is based on established physics and pre-programmed parameters, not a machine learning or AI algorithm that requires a training dataset in the conventional sense. The "software version 3.2" is updated device control software, not a trainable AI model in the common "training set" connotation.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set for an AI/ML algorithm is mentioned, this question is not applicable based on the provided text. The device's "training" in a broad sense would be its design, calibration, and validation against engineering specifications during its development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141379

    Validate with FDA (Live)

    Date Cleared
    2015-01-16

    (234 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use in:

    · The creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea

    · For anterior capsulotomy during cataract surgery

    • The creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.

    Device Description

    The VICTUS Femtosecond Laser Platform (VICTUS) is a precision ophthalmic surgical laser. For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location. The VICTUS Platform mode of operation delivers femtosecond pulses to produce a pattern of photodisruption to create cuts / separation in ophthalmic tissue. The VICTUS Platform uses vacuum to affix a suction ring to the corneal surface prior to use.

    AI/ML Overview

    The device referenced in the document is the VICTUS Femtosecond Laser Platform. The document describes a 510(k) submission, which means the device is being compared to a legally marketed predicate device to establish substantial equivalence. Therefore, the "acceptance criteria" discussed here are primarily about demonstrating that the modified device performs similarly to or no worse than the predicate device.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria for device performance in a table, but it does describe the outcomes of the non-clinical and clinical studies designed to show equivalence.

    Performance Metric/Study TypeAcceptance Criteria (Implied)Reported Device Performance
    Non-Clinical PerformanceCompliance with applicable safety standards (EN ISO 60601-1, -1-2, -2-22).The VICTUS Femtosecond Laser Platform has undergone testing and is in compliance with applicable safety standards.
    Performance of modified VICTUS Platform remains substantially equivalent to the predicate (K132534).A variety of test procedures were conducted and the collected data were evaluated by comparing the mean values to specified acceptance criteria and their 95% confidence intervals. The acquired test data successfully verified that the various parameters meet their pre-defined acceptance criteria. The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate device.
    Clinical PerformanceIntraoperative accuracy of biometry visualization (OCT) comparable to LENSTAR LS 900.The lens thickness, ACD, and CCT measured by VICTUS OCT were compared with LENSTAR LS900 and the results are consistent.
    Successful completion of laser-assisted capsulotomy without manual intervention for all procedures.All laser-assisted capsulotomy procedures were completed fully using the VICTUS laser platform with no need for manual intervention to complete the capsulotomy.
    No adverse or serious adverse events reported by investigators.There were no adverse or serious adverse events reported by the investigators during the clinical study.
    No device malfunctions leading to adverse or serious adverse events.In addition, there were no device malfunctions which led or potentially led to adverse or serious adverse events.

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical Study):
      • Sample Sizes:
        • Femtosecond-laser flap creation (flap group): 58 eyes
        • Femtosecond laser assisted cataract surgery (cataract group): 113 eyes
      • Data Provenance: Single-center clinical study (location not specified, but likely where Technolas Perfect Vision GmbH is active). Retrospective analysis of correlation and agreement of biometric data.

    3. Number of Experts and Qualifications (for Ground Truth)

    • The document does not specify the number of experts used to establish ground truth or their qualifications.
    • For the clinical study, the "consistency" of VICTUS OCT measurements with LENSTAR LS900 implies that the LENSTAR LS900 measurements served as a reference or a highly accepted method. LENSTAR LS900 is a well-established biometer, and its measurements are generally accepted as a clinical standard for these parameters.

    4. Adjudication Method (for Test Set)

    • The document does not describe any adjudication method for the clinical study data or for establishing a ground truth based on multiple expert opinions.
    • The comparison was made against LENSTAR LS900, suggesting it served as the reference standard.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done comparing human readers with and without AI assistance. The device is a laser platform for surgical procedures, not an AI-assisted diagnostic tool for human readers.

    6. Standalone (Algorithm Only) Performance Study

    • A form of standalone performance was done for the device. The non-clinical performance data involved testing the device against engineering specifications and predicate equivalence using various materials (porcine eyes, agarose gel, PETG, PMMA).
    • The clinical study also evaluated the device's performance in creating capsulotomies and biometry measurements directly.

    7. Type of Ground Truth Used

    • Non-Clinical Performance: Engineering specifications, measurements against a predicate device, and physical properties of test materials (e.g., cut quality in various materials).
    • Clinical Performance:
      • For biometry accuracy: Comparison with measurements from a commercially available and clinically accepted biometer (LENSTAR LS900). This acts as a proxy for "gold standard" or highly reliable measurement.
      • For capsulotomy completeness: Observation of the actual surgical outcome (full completion without manual intervention).
      • For adverse events: Reported clinical observations by investigators.

    8. Sample Size for the Training Set

    • The document does not mention a training set in the context of an algorithm or AI. This submission is for a medical device (laser platform) that performs physical procedures, not a software algorithm that learns from data in the traditional AI sense. Its "training" would be more akin to design verification and validation, and calibration.

    9. How Ground Truth for Training Set was Established

    • As there's no explicitly defined "training set" for an AI algorithm as per the question's premise, this question is not directly applicable. If interpreting "training set" as the data used during the device's development and design verification, the ground truth would have been established through engineering design specifications, bench testing protocols, and comparison with predicate device characteristics and safety standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140615

    Validate with FDA (Live)

    Date Cleared
    2014-07-15

    (127 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use in:

    • . the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
    • for anterior capsulotomy during cataract surgery .
    • the creation of cuts / incisions in the cornea in patients undergoing cataract surgery or . other ophthalmic treatment requiring cuts / incisions in the cornea.
    • . laser-assisted lens fragmentation of nuclear cataracts during cataract surgery, not for fragmentation of posterior subcapsular (PSC) and cortical cataracts
    Device Description

    The VICTUS Femtosecond Laser Platform (hereafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for anterior capsulotomy during cataract surgery (via K120426). In addition, the VICTUS has been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea (via K122386), and for corneal incisions (via K 132534). This 510(k) expands the list of indications for use to include laser-assisted lens fragmentation during cataract surgery for nuclear cataracts.

    For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

    Safety margins are built into the software, including an anterior capsule safety margin of 500 um, a posterior capsule safety margin of 700 um, and a pupil safety margin of 500 um. The graphical user interface is used to mark the anatomical boundaries and the software automatically calculates the safety margins. The safety margins are not modifiable by the end user.

    The fundamental scientific technology remains the same as previously cleared for the VICTUS Femtosecond Laser Platform under K122386 and K132534.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study that proves the device meets those criteria:


    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Primary Endpoint: Effective phacoemulsification time (EPT) in the laser group (pooled from all cataract grades) should be lower than or equal to that in the manual group.Met: Effective phacoemulsification time (EPT) in the laser group (pooled from all cataract grades) was lower than or equal to that in the manual group.
    Secondary Endpoint: Less or equal adverse events in the laser group compared with the manual group.Met: Less or equal adverse events in the laser group compared with the manual group.
    Safety Margins: Anterior capsule safety margin of 500 um, posterior capsule safety margin of 700 um, and pupil safety margin of 500 um. (Implied, as these are "built into the software" and not user-modifiable, suggesting they must be maintained.)Implied Met: "Safety margins are built into the software, including an anterior capsule safety margin of 500 um, a posterior capsule safety margin of 700 um, and a pupil safety margin of 500 um." The text indicates these are programmatically enforced, not reported as a result.
    Reproducibility and Accuracy (Bench Testing): Laser-assisted lens fragmentation performed with the device must meet established acceptance criteria for lens fragmentation for various parameters (upper/lower depth, cross/open angle, spot/line spacing, decentration).Met: "The testing showed that laser-assisted lens fragmentation performed with the VICTUS Femtosecond Laser Platform met the established acceptance criteria for lens fragmentation." "resulted in highly reproducible and accurate treatments".
    Intraocular Pressure (IOP): Pre- and post-operative IOPs should be similar between the VICTUS and manual groups.Met: Pre- and post-operative IOPs were similar between the VICTUS and manual groups at all time points during the study for the complete study group and when stratified by cataract grade.
    Best Corrected Visual Acuity (BCVA): BCVDA should be similar between the VICTUS and manual groups.Met: Best corrected visual acuities (BCVDA) were similar between the VICTUS and manual groups at all time points during the study for the complete study group and also when stratified by cataract grade.
    Slit Lamp Examinations: Slit lamp examinations in the laser group should be consistent with the manual group.Met: Slit lamp examinations in the laser group were consistent with the manual group at all time points for the complete study group and also when stratified by cataract grade.
    Intra-operative Complications: (Implied: Acceptable rate, or no increase compared to manual).Met/Favorable: The only intra-operative complications were two anterior capsule tears in the manual group; no other operative complications were recorded for either the manual or the laser group.
    Post-operative Observation Rates (Corneal edema, flare, trace anterior chamber): Comparable between laser and manual groups.Met: The post-operative observation rates of corneal edema, flare and trace anterior chamber in the laser group were comparable to those in the manual group.
    OCT Image Correlation with Biometric Data: OCT images through the PI should correlate well with biometric data for lens thickness and anterior chamber depth.Met: OCT images through the PI correlate well with biometric data obtained from OCT images for lens thickness and anterior chamber depth for the complete study group and also when stratified by cataract grade.
    VICTUS OCT Image Precision: High level of precision for lens thickness and anterior chamber depth.Met: VICTUS OCT images were found to have a high level of precision for lens thickness (both pooled and stratified by cataract grade) and anterior chamber depth (pooled only).

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 136 eyes from 110 subjects (26 patients treated bilaterally). 68 eyes in the VICTUS (laser) group and 68 eyes in the standard (manual) treatment group.
      • Data Provenance: The text does not explicitly state the country of origin. It mentions "multi-center, multi-surgeon clinical study," implying various clinical sites, but no geographical location is specified. It was a prospective study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The study design describes a "randomized, controlled, open, prospective, multi-center, multi-surgeon clinical study" comparing the VICTUS device to manual methods. The "ground truth" in this context refers to the clinical outcomes and observations recorded during the study, rather than an independent expert review of images or data to establish a reference standard. The study data (e.g., EPT, adverse events, IOP, BCVA, slit lamp findings) were collected through standard clinical procedures by the participating surgeons and study personnel. The document does not specify a number of independent experts for ground truth establishment outside of the primary treating physicians and clinical staff who recorded the outcomes.
    3. Adjudication method for the test set:

      • The document does not describe a specific adjudication method (like 2+1 or 3+1) for the test set results. Clinical outcomes were collected by the participating centers. Adverse events were observed and reported, but the process of determining their relation to the intervention was clinical (e.g., "All adverse events in the laser group were determined to be related to the cataract surgery and not to the use of the VICTUS Femtosecond Laser."). This implies a clinical judgment by the investigators, not an independent adjudication committee.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study compares a surgical device (VICTUS Femtosecond Laser Platform) to a manual surgical technique, not an AI or imaging-based diagnostic tool requiring human reader interpretation. Therefore, there's no mention of human reader improvement with/without AI assistance or effect sizes related to such a comparison.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable as the device is a surgical laser platform, not an algorithm for diagnosis or image analysis. The "standalone" performance mentioned refers to the device's ability to perform specific surgical actions (e.g., lens fragmentation) as intended, which was evaluated through the clinical study and non-clinical bench testing. The outcomes were compared to manual surgical techniques, which inherently involve human-in-the-loop performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" primarily consisted of clinical outcomes data (e.g., Effective Phacoemulsification Time (EPT), adverse event rates, intraocular pressure (IOP), best corrected visual acuity (BCVA), slit lamp examination findings, intra-operative complications) obtained directly from the patients in a prospective clinical study. Non-clinical bench testing also established a "ground truth" for the device's physical performance parameters against established acceptance criteria.
    7. The sample size for the training set:

      • This device is a surgical instrument and does not rely on a "training set" in the context of machine learning. Therefore, a training set sample size is not applicable. The device's design and parameters are based on engineering principles and prior iterations (as evidenced by its substantial equivalence to previous VICTUS and predicate devices).
    8. How the ground truth for the training set was established:

      • As explained above, there is no "training set" for this type of device. The functionality and safety were established through mechanical/bench testing and clinical trials, not through learning from a dataset with established ground truth labels.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132534

    Validate with FDA (Live)

    Date Cleared
    2014-02-14

    (185 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use for:

    • The creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
    • For anterior capsulotomy during cataract surgery
    • The creation of cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring cuts / incisions in the cornea.
    Device Description

    The VICTUS Femtosecond Laser Platform (VICTUS) is a precision ophthalmic surgical laser and is cleared for use in the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea and for patients undergoing anterior capsulotomy during cataract surgery (via K120426). In addition, the VICTUS has been cleared for patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea (via K122386). This 510(k) expands the arcuate corneal incision by allowing the ability for primary and secondary corneal incisions. These cuts/incisions are performed for patients undergoing cataract surgery to allow direct access into the eye. The correct position and shape of the cuts is verified by the video image. The expansion of the corneal incision indication is supported by software version 2.7 SP2. In addition, an increased diameter patient interface is being introduced (referred to as PI125). The PI125 is a universal PI that allows the use for either corneal or cataract procedures. User request for a smaller diameter patient interface that can be used for both cataract and corneal procedures resulted in the qualification of the existing cataract patient interface as a universal smaller diameter patient interface. Lastly, the bottom end of the pulse duration range has been modified to be lower than that previously cleared in K120426 and K122386. For all indications for use, laser pulses are delivered through a sterile disposable Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eve relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location. The fundamental scientific technology remains the same as previously cleared for the VICTUS Femtosecond Laser Platform under K120426 and K122386.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text, where available:

    1. A table of acceptance criteria and the reported device performance

    The provided text does not explicitly state quantitative acceptance criteria for device performance (e.g., a specific tolerance for depth, diameter, or angle of incisions). Instead, it describes general claims of performance.

    Acceptance Criteria (Implied)Reported Device Performance
    Reproducible depth, diameter, and angle of primary and secondary corneal incisions."laser-assisted primary and secondary incisions performed with the VICTUS Femtosecond Laser Platform resulted in highly reproducible and accurate depth, diameter, and angle."
    Substantially equivalent performance to previously cleared patient interfaces for the larger diameter patient interface (PI125)."Assessment of the larger diameter patient interface shows substantially equivalent results as compared to the predicate patient interfaces."
    Compliance with applicable safety standards.The device is "in compliance with applicable safety standards" listed.
    Equivalent performance to predicate devices in patients undergoing ophthalmic surgery requiring arcuate cuts/incisions."The VICTUS Femtosecond Laser Platform has been found to perform equivalently to the predicate devices in patients undergoing ophthalmic surgery or other treatment requiring penetrating arcuate cuts / incisions in the cornea."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated. The text mentions "A variety of test procedures were conducted" and "Four different materials were used for the bench performance testing: porcine eyes, agarose gel, polyethylene terephthalate (PETG), and polymethyl methacrylate (PMMA)." It does not quantify the number of tests performed on these materials.
    • Data Provenance: The study was a "Non-Clinical Performance Data" study, implying it was conducted in a laboratory setting (bench testing) using specified materials. The location of the testing is not mentioned, nor is it explicitly stated as retrospective or prospective, but given it's bench testing, it would inherently be "prospective" in the sense that the tests were designed and executed for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The study was non-clinical bench testing, not involving human subjects or expert assessment of clinical outcomes. Ground truth would be based on physical measurements of the incisions in the tested materials.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The study was non-clinical bench testing. Performance was evaluated by "comparing the mean values to the specified acceptance criteria and their 95% confidence intervals." This suggests quantitative measurement and statistical analysis rather than expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This was a non-clinical bench study of a laser surgical device, not an AI-assisted diagnostic or imaging system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a standalone performance of a device itself, which is what the bench testing aimed to assess. The device performance was evaluated independently without human intervention during the "cutting" process, as it is an automated laser system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the non-clinical testing appears to be based on direct physical measurements of attributes like depth, diameter, and angle of incisions in the various test materials (porcine eyes, agarose gel, PETG, PMMA). This would be established using appropriate metrology tools for precision measurement.

    8. The sample size for the training set

    Not applicable. This is not an AI-based system that requires a "training set" in the context of machine learning. The device is a laser platform; its "training" per se would relate to its engineering and calibration.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the machine learning sense for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122386

    Validate with FDA (Live)

    Date Cleared
    2013-02-08

    (186 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VICTUS Femtosecond Laser Platform is indicated for use in:

    • the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea
    • for anterior capsulotomy during cataract surgery
    • the creation of penetrating arcuate cuts / incisions in the cornea in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
    Device Description

    The VICTUS Femtosecond Laser Platform (hercafter referred to as the VICTUS Platform) is a precision ophthalmic surgical laser indicated for use in patients undergoing cataract surgery or other ophthalmic treatment requiring penetrating arcuate cuts / incisions in the cornea.
    The VICTUS Platform is an ophthalmic surgical femtosecond laser designed for cutting a penetrating arcuate cut of pre-selected depth, radius, opening angle, and position angle. The system works by first being programmed with the depth, radius, opening angle and position angle at which the penetrating arcuate cut is desired. The surgeon then fixates the eye with a sterile disposable PMMA contact lens (hercafter referred to as the Patient Interface) that is connected to the laser via a vacuum tube. Depths of -120µm, radius of 1500 - 4750 um, opening angles of 1-360°, and position angles of 0 - 359° are available.
    Laser pulses are delivered through the sterile Patient Interface, consisting of a contact lens and suction clip to provide suction. The contact lens and suction clip assembly creates a reference surface for depth control and fix the eye relative to the delivery of the laser beam. Surgical effects are produced by scanning thousands of individual pulses, producing continuous incisions. The location of the tissue photodisruption is controlled by a fixed laser beam focused through a scanning optic system to the desired location.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the VICTUS Femtosecond Laser Platform, based on the provided text:


    1. Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state quantitative acceptance criteria in a table format for the arcuate cuts/incisions. However, the performance data section implies the following outcomes were evaluated and met a criteria of being "highly reproducible and accurate." The postmarket surveillance review indicates "comparability to the standard manual procedure" for clinical endpoints.

    Acceptance Criteria (Implied/Evaluated)Reported Device Performance (VICTUS Femtosecond Laser Platform)
    Bench/Pre-Clinical Testing:
    Reproducibility of rim depthHighly reproducible
    Accuracy of rim depthAccurate
    Reproducibility of diameterHighly reproducible
    Accuracy of diameterAccurate
    Reproducibility of open angleHighly reproducible
    Accuracy of open angleAccurate
    Reproducibility of position angleHighly reproducible
    Accuracy of position angleAccurate
    Reproducibility of centrationHighly reproducible
    Accuracy of centrationAccurate
    Visual comparison to predicate devicesQualitatively comparable (based on scanning electron microscopy)
    Postmarket Surveillance Review (Clinical):
    Post-surgical loss of lines of BCDVAComparable to standard manual procedure
    Post-surgical overall change in refractionComparable to standard manual procedure
    Postoperative induced cylinderComparable to standard manual procedure
    Adverse events (rate and type)Comparable to standard manual procedure
    Serious adverse events (rate and type)Comparable to standard manual procedure

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 49 eyes from 34 subjects.
    • Data Provenance: Retrospective, single site in Europe.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The text states that "all surgeries were performed by the same surgeon" at the single clinical location. This surgeon's performance was compared between manual and laser-assisted procedures. While this surgeon's outcomes define the "ground truth" for the comparison within the study, the text does not specify the number of experts or their qualifications that established an independent "ground truth" for evaluating the quality of the surgical outcomes themselves. The comparison is against the surgeon's own manual technique.


    4. Adjudication Method for the Test Set

    The provided text does not specify an adjudication method (e.g., 2+1, 3+1). The "ground truth" appears to be the clinical outcomes as assessed by the operating surgeon and the postmarket surveillance review.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly described. The study compared laser-assisted procedures to manual procedures performed by the same surgeon at a single site. The effect size of how much human readers improve with AI vs. without AI assistance is not applicable in this context as the device is a surgical laser, not an AI diagnostic tool.


    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) was done

    Yes, a standalone performance evaluation was conducted in the Non-Clinical Performance Data section. This involved bench testing using various materials (porcine eyes, agarose gel, PETG, and PMMA) and scanning electron microscopy to assess the laser's cutting accuracy and reproducibility. This evaluate the device's inherent mechanical and cutting capabilities independent of human intervention in the cutting process itself.


    7. The Type of Ground Truth Used

    • For Non-Clinical Performance Data: The ground truth was established through measurements (rim depth, diameter, open angle, position angle, centration) and visual assessment (scanning electron microscopy) of the cuts produced in various test materials (porcine eyes, agarose gel, PETG, PMMA).
    • For Postmarket Surveillance Clinical Review: The ground truth was based on clinical outcomes data (post-surgical loss of best corrected distance visual acuity (BCDVA), post-surgical overall change in refraction, postoperative induced cylinder, adverse events, and serious adverse events) observed in patients after both laser-assisted and manual arcuate incisions, as assessed by the operating surgeon.

    8. The Sample Size for the Training Set

    The provided text does not mention a training set or details about its sample size. This device is a surgical laser, and the studies described are performance evaluations, not the development or training of an AI algorithm.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned (since this is not an AI/algorithm-driven device in the context of typical AI device submissions), this question is not applicable.