    K Number
    K201143

    Validate with FDA (Live)

    Device Name
    VICRYL Mesh
    Manufacturer
    Date Cleared
    2020-08-06

    (99 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The absorbable VICRYL™ Mesh may be used wherever temporary wound support is required. VICRYL™ Mesh may be cut to the shape or size desired for each specific application.

    Device Description

    VICRYL™ (polyglactin 910) Mesh is a synthetic absorbable sterile copolymer made from glycolide and L-lactide. The copolymer is identical in composition to that used in VICRYL™ (polyglactin 910) synthetic absorbable suture.

    Two weave configurations are available, knitted and woven. VICRYL™ Knitted Mesh is more porous than VICRYL™ Woven Mesh and may be used in instances in which compliant and stretchable support material is desired.

    AI/ML Overview

    This document, an FDA 510(k) Premarket Notification for VICRYL™ Mesh (K201143), does not describe a study involving an AI/image analysis device. It is a submission for a surgical mesh, and the basis for its clearance is substantial equivalence to a previously cleared device (K191373), with the only change being a revision to the Instructions for Use labeling.

    Therefore, many of the requested elements for describing the acceptance criteria and study proving device performance (e.g., sample size for test/training sets, expert adjudication, MRMC studies, standalone performance, ground truth establishment) are not applicable to this specific FDA submission.

    However, I can extract the general acceptance criteria and "performance" of this device within the context of a 510(k) clearance, which is based on demonstrating substantial equivalence to a predicate device.


    Acceptance Criteria and Device Performance (in the context of Substantial Equivalence for a Medical Device)

    Since this document is a 510(k) clearance for a surgical mesh and not an AI/image analysis device, the concept of "acceptance criteria" and "device performance" is framed in terms of demonstrating substantial equivalence to a legally marketed predicate device. The primary "acceptance criteria" for a 510(k) submission like this is that the new device is as safe and effective as a predicate device and does not raise new questions of safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for Substantial Equivalence)Reported Device Performance (as stated in the 510(k) Summary)
    Fundamental Scientific Technology: Be of the same type of technology.Both the subject device (VICRYL™ Mesh) and the predicate device (VICRYL™ Mesh, K191373) are synthetic absorbable sterile copolymers.
    Intended Use: Have the same intended use.Both devices are intended for use wherever temporary wound support is required.
    Materials: Be composed of the same materials.Both are made from glycolide and L-lactide copolymer, identical to that used in VICRYL™ (polyglactin 910) synthetic absorbable suture.
    Design: Have similar design characteristics (e.g., two weave configurations available: knitted and woven).Both devices have two weave configurations available (knitted and woven). ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Construction: Be constructed using the same manufacturing processes.The subject mesh is manufactured within the existing manufacturing processes for the predicate device. No changes to manufacturing, packaging, sterilization processes, or shelf life.
    Performance Characteristics: Exhibit comparable performance.The devices function in the same manner. No changes to performance characteristics were noted or implied by the change only in labeling. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Impact of differences: Any differences do not raise new questions of safety or effectiveness.The only stated difference is a revision to the Instructions for Use (labeling) to provide additional clarity to the Indications statement. This change does not alter the fundamental characteristics or function of the device and thus does not raise new questions of safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable. This 510(k) submission did not involve a "test set" in the sense of a dataset for evaluating an AI algorithm's performance. The clearance was based on substantial equivalence, meaning the device itself (VICRYL™ Mesh) was deemed equivalent to a previously cleared version.
    • Data Provenance: Not applicable as no new clinical or non-clinical testing data were generated or relied upon for this specific submission, other than the updated labeling. The core "data" supporting this clearance is the demonstration of sameness to the predicate device and the predicate's existing clearance history.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not Applicable. No ground truth establishment by experts was required for this 510(k) given that no new performance data was submitted or required. The review process is primarily regulatory and technical comparison to a predicate, not clinical validation of performance with a study.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No test set requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • Not Applicable. This is not an AI/image analysis device. Therefore, no MRMC study comparing human readers with and without AI assistance was performed.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Performance:

    • Not Applicable. This is not an AI/image analysis device. Therefore, no standalone algorithm performance was assessed.

    7. Type of Ground Truth Used:

    • Not Applicable. No clinical "ground truth" (e.g., pathology, outcomes data) was used or generated for this submission as it was based on substantial equivalence and a labeling change for an existing device. The "ground truth" in a regulatory sense here is the established safety and effectiveness of the predicate device.

    8. Sample Size for the Training Set:

    • Not Applicable. As this is not an AI/image analysis device, there was no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. As there was no "training set," there was no ground truth to establish for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K191373

    Validate with FDA (Live)

    Device Name
    VICRYL Mesh
    Manufacturer
    Date Cleared
    2019-10-22

    (152 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VICRYL™ Mesh may be used wherever temporary wound or organ support is required (kidney, liver, spleen). VICRYL™ Mesh may be cut to the shape or size desired for each specific application.

    Device Description

    VICRYL™ (polyglactin 910) Mesh is a synthetic absorbable sterile copolymer made from glycolide and L-lactide. The copolymer is identical in composition to that used in VICRYL™ (polyglactin 910) synthetic absorbable suture.

    VICRYL™ Knitted Mesh, which is more porous than VICRYL™ Woven Mesh, may be used in instances in which compliant and stretchable support material is desired.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the VICRYL™ Mesh device. It asserts substantial equivalence to predicate devices rather than proving the device meets specific acceptance criteria through a study with performance metrics.

    Therefore, many of the requested categories regarding acceptance criteria, study design, ground truth, and statistical analysis are not applicable (N/A) because this type of submission focuses on demonstrating equivalence to existing devices, not on presenting novel performance data from a clinical or analytical study with defined acceptance criteria.

    Here's a breakdown based on the information available:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (N/A for 510(k) Substantial Equivalence)Reported Device Performance (N/A for 510(k) Substantial Equivalence)
    No explicit acceptance criteria based on specific performance metrics are defined in this 510(k). The submission relies on substantial equivalence to predicate devices.No specific performance metrics (e.g., sensitivity, specificity, accuracy) are reported for the VICRYL™ Mesh itself, as the submission relies on substantial equivalence.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • N/A. This 510(k) submission does not describe a test set or data provenance for a performance study. It's a regulatory submission demonstrating substantial equivalence based on technological characteristics and intended use, not clinical performance data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • N/A. As no test set performance study is described, there's no mention of experts establishing ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • N/A. No test set performance study is described, therefore no adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is a medical device (surgical mesh), not an AI-based diagnostic tool. No MRMC study was conducted or described.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • N/A. This is a physical medical device (surgical mesh), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. No new performance study requiring ground truth is described. The basis for clearance is substantial equivalence to legally marketed predicate devices, meaning their known safety and effectiveness serve as the "ground truth" for the category.

    8. The sample size for the training set

    • N/A. This 510(k) submission pertains to a physical medical device (surgical mesh) and does not involve an algorithm or an AI model with a training set.

    9. How the ground truth for the training set was established

    • N/A. As there is no training set for an algorithm, there's no ground truth establishment for it.

    Summary of the 510(k) Submission:

    The submission for VICRYL™ Mesh (K191373) focuses on demonstrating substantial equivalence to two predicate devices: VICRYL™ (Polyglactin 910) Mesh (K810428) and VICRYL™ Mesh Bag (K051701).

    The core argument for equivalence is:

    • Technological Characteristics: Both the subject and predicate devices are synthetic absorbable sterile copolymers made from glycolide and L-lactide.
    • Intended Use: All devices are designed for temporary wound or organ support (kidney, liver, spleen) and can be cut to size.
    • Manufacturing and Sterilization: The subject mesh is manufactured within existing processes for the predicate device, with no changes to manufacturing, packaging, sterilization, or shelf life.
    • Differences: The only notable difference for the subject device is a revision to the Instructions for Use (IFU) to clarify indications and align with evolving regulatory expectations.

    Therefore, Ethicon, Inc. did not conduct a performance study with acceptance criteria and a test set to "prove the device meets the acceptance criteria" in the manner typically expected for a novel device or AI algorithm. Instead, they demonstrated that their device is fundamentally the same as already cleared devices, making it "as safe and effective as the predicates."