    K Number
    K121497

    Validate with FDA (Live)

    Device Name
    VENOFER PUMP
    Date Cleared
    2012-08-17

    (88 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Venofer Pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer® to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.

    Device Description

    The Venofer Pump is an optional module for use on Fresenius 2008 Series Hemodialysis Machines and is designed to administer Venofer® during dialysis treatments and consists of a control panel, vial holder, fluid detector, and a peristaltic pump. The module is a self contained microprocessor controlled device that receives its power from the Fresenius 2008 Hemodialysis machine.

    AI/ML Overview

    The provided text describes the Venofer Pump Version 1.3, an optional accessory for hemodialysis machines. This submission is a "Special 510(k)" for modifications to support the use of an additional blood tubing set. The document focuses on demonstrating substantial equivalence to a previously cleared device (K103564) and confirming that the modifications do not raise new questions of safety or effectiveness.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Verification Tests ConductedReported Device Performance/Outcome
    Venofer Pump PerformanceFunctional testing to demonstrate the device performs as designed and expected.The performance (verification) test results support the performance characteristics of the Venofer Pump. It was confirmed that the device does not perform in an unexpected and/or unsafe manner (system-level hazard analysis). Specific details of the performance metrics are not explicitly provided beyond this general statement.
    Venofer Pump SafetyElectromagnetic compatibility (EMC) testing (IEC 60601-1-2: 2007)Product safety testing demonstrates that the device performs per the FDA Consensus Standards. This specifically includes EMC testing.
    Biocompatibility testing (AAMI / ANSI / ISO 10993-1:2009)Biocompatibility testing was performed on all new materials that are patient-fluid contacting. The verification test results support the safety characteristics of the Venofer Pump. (No specific numerical results or pass/fail criteria are provided, only that testing was performed and supported safety).
    Venofer Pump SoftwareSoftware architecture SDD (Software Design Description) Unit test System and software requirements TraceabilitySoftware testing demonstrates the device software meets the design input requirements. The testing includes required documentation as described in the guidance titled "Guidance for the Content of Premarket Submissions for Software." (No specific metrics like bug rates, code coverage, or pass/fail percentages are provided, only a general statement of meeting design input requirements and review of documentation). The phrase "Software release history" is also listed as documentation reviewed, suggesting a process was followed.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the test set for any of the verification activities (functional, safety, or software).
    The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective. The studies described are non-clinical (verification) testing, not clinical trials on human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. Given that the studies are non-clinical verification tests, the "ground truth" would likely be established by engineering specifications, regulatory standards, or consensus standards, rather than expert clinical judgment in the traditional sense for medical imaging or diagnostic devices.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving human readers/interpreters, not for non-clinical device verification testing.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is a pump for administering medication, not an AI-powered diagnostic or assistive device for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    No standalone algorithm performance study was done in the context of AI. The "standalone" performance described would be the functional performance of the pump itself, which was tested during "Venofer Pump Performance Testing." The device is a physical pump, not a diagnostic algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical verification testing described:

    • Performance Testing: The "ground truth" would be the device's design specifications and expected functional behavior.
    • Safety Testing: The "ground truth" would be established by FDA Consensus Standards (e.g., IEC 60601-1-2 for EMC, AAMI / ANSI / ISO 10993-1 for biocompatibility).
    • Software Testing: The "ground truth" would be the software design input requirements and functional specifications.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. The Venofer Pump is a physical medical device, not an AI/machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided. As mentioned above, the device is not an AI/machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103564

    Validate with FDA (Live)

    Device Name
    VENOFER PUMP
    Date Cleared
    2011-02-10

    (66 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Venofer Pump 1.1 is identical to the unmodified device.

    The Venofer® Pump Is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.

    Fresenius 2008T is indicated for acute and chronic dialysis therapy.

    Device Description

    The Venofer Pump Module is an optional module for use on Fresenius 2008T Hemodialysis Machines. The module is designed to administer Venofer, an iron sucrose supplement, during dialysis treatments. The Venofer Pump Module is to be used in accordance with the approved Venofer Indications for Use and the physician's prescription.

    The Venofer Pump Module fits into the module compartment of Fresenius 2008T Machines and consists of a control panel, vial holder, fluld detector, and a peristaltic pump.

    AI/ML Overview

    This document is a 510(k) Summary for the Venofer Pump Version 1.1, a device intended to administer Venofer (iron sucrose) during hemodialysis. This filing is a "Special 510(k)" for a modification to an existing device (Venofer Pump K093964).

    Crucially, this document focuses on demonstrating substantial equivalence through non-clinical verification testing (performance, safety, and software) and does NOT describe a study involving human patients or a process for establishing ground truth from expert consensus for diagnostic accuracy.

    Therefore, most of the requested information regarding clinical studies, sample sizes, expert involvement, and ground truth establishment will not be applicable or available in this specific document.

    Here's an analysis based on the provided text:


    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria CategoryReported Device Performance (Summary from this document)
    Venofer Pump Performance TestingFunctional testing demonstrated that the device performs as designed and expected. This included specific verification tests with defined objectives, test articles, test methods and procedures, and pre-defined acceptance or pass/fail criteria. A system-level hazard analysis confirmed the device does not perform in an unexpected and/or unsafe manner. (Specific quantitative performance metrics are not detailed in this summary.)
    Venofer Pump Safety TestingProduct safety testing demonstrated that the device performs per FDA Consensus Standards. This included: - Electromagnetic Compatibility (EMC): Tested to IEC 60601-1-2: 2007.- Biocompatibility: Tested to AAMI / ANSI / ISO 10993-1:2009 for all new patient-fluid contacting materials. (Specific quantitative results or pass/fail criteria for these tests are not detailed in this summary.)
    Venofer Pump Software TestingSoftware testing demonstrated the device software meets the design input requirements. This included: - Structure chart/flow chart of software architecture.- Summary of software development procedures, including changes.- Software requirements specification with traceability to hazard analysis.- Verification and validation test plans, including pass/fail criteria and traceability to requirements.- System-level test results.- Current software version number and date of latest revision. (Specific quantitative test results or detailed pass/fail criteria are not detailed in this summary.)

    Study Proving Device Meets Acceptance Criteria:

    The study referenced is the "510(k) Verification Testing Non-Clinical Summary" which includes "Performance, Safety, and Software Testing." The document emphasizes that "The conclusions drawn from this testing demonstrates that the Venofer Pump is as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices to which it was compared."


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: This refers to the number of devices or components tested, not patient data. The document does not specify the exact number of units or test iterations for the performance, safety, or software verification tests. It generally refers to "the device" implying tests were conducted on one or more representative units.
    • Data Provenance: Not applicable. This is not a study involving patient data. The testing was non-clinical (laboratory/engineering) and would have been conducted by the manufacturer, Renal Solutions, Inc. (USA).
    • Retrospective or Prospective: Not applicable. This applies to clinical data collection, not non-clinical device verification.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This device is a pump for administering medication, not an AI/diagnostic device that requires expert-established ground truth from medical imaging or patient data. The "ground truth" for this device's performance is its adherence to engineering specifications, safety standards, and software requirements, which are verified through the non-clinical testing described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods like 2+1 or 3+1 refer to how discrepancies in expert opinions are resolved to establish a definitive ground truth in diagnostic studies. This is a non-clinical device verification, so such adjudication is not performed. Acceptance was based on pre-defined pass/fail criteria for engineering and safety tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. An MRMC study is relevant for diagnostic devices that assist human readers (e.g., radiologists interpreting images). This device is a pump, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable in the traditional sense of an AI algorithm. This is a hardware device with embedded software. The "standalone" performance was effectively what the document refers to as "Venofer Pump Performance Testing" and "Venofer Pump Software Testing" – verifying the device and its embedded software performed according to specifications and safety requirements without human intervention during the test cycles, but it's not an AI algorithm performing a diagnostic task.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable/Defined Differently. For this type of device, "ground truth" is defined by:
      • Engineering Specifications: The pre-defined functional requirements and performance parameters of the pump (e.g., flow rate accuracy, pressure limits).
      • Regulatory Standards: Compliance with recognized FDA Consensus Standards (e.g., IEC 60601-1-2 for EMC, ISO 10993-1 for Biocompatibility).
      • Software Requirements: The predefined design input requirements for the device's embedded software.
      • Hazard Analysis: Confirmation that the device does not perform in an unexpected and/or unsafe manner.

    8. The sample size for the training set

    • Not Applicable. This device does not use machine learning or AI that requires a "training set" of data. Its functionality is based on deterministic programming and electro-mechanical design, not learned patterns from data.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an AI algorithm, there is no ground truth to establish for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K093964

    Validate with FDA (Live)

    Device Name
    VENOFER PUMP
    Date Cleared
    2010-07-09

    (197 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fresenius Model 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy.

    The Venofer® Pump Module is an optional accessory for use on Fresenius 2008T Hemodialysis Machines and is intended to administer Venofer®, an iron sucrose, to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer®, in conjunction with hemodialysis, is prescribed by a physician.

    Device Description

    The Venofer Pump Module is an optional module for use on Fresenius 2008T Hemodialysis Machines. The module is designed to administer Venofer, an iron sucrose supplement, during dialysis treatments. The Venofer Pump Module is to be used in accordance with the approved Venofer Indications for Use and the physician's prescription.

    The Venofer Pump Module is intended to provide ease and consistent delivery of Venofer, an iron sucrose supplement, indicated for the treatment of iron deficiency anemia for Hemodialysis dependent-chronic kidney disease patients. Anemia commonly occurs in patients with chronic kidney disease undergoing Hemodialysis and effective anemia management is recognized as an important factor in improving the outcomes of these patients. Iron deficiency is a frequent contributing factor which complicates the treatment of anemia in chronic kidney disease patients. Iron sucrose is commonly delivered to improve iron status in patients who have chronic kidney disease. Venofer is an injectable iron preparation drug approved for treatment of iron deficiency and is the most commonly used drug to treat iron deficiency anemia in dialysis patients. Renal Solutions has developed the Venofer Pump Module that provides the means to deliver Venofer when undergoing Hemodialysis on a Fresenius 2008T Dialysis System.

    The Venofer Pump Module fits into the module compartment of existing Fresenius 2008T Machines and consists of a control panel, vial holder, fluid detector, and a peristaltic pump.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Fresenius Model 2008T Hemodialysis Machine with Venofer Pump Module." This document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study with acceptance criteria and performance data for a novel device. Therefore, much of the requested information regarding acceptance criteria, sample sizes, expert ground truth, adjudication methods, and comparative effectiveness studies is not available in the provided text.

    Based on the information given, here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The submission is for a medical device accessory and states that non-clinical testing was performed, but it doesn't detail specific acceptance criteria or quantitative performance results in a table format. It generally claims the device is "as safe, as effective, and performs at least as safely and effectively as the legally marketed devices identified as predicate devices."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in the document. The text mentions "non-clinical testing" but does not specify sample sizes or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided in the document. The nature of the device (a pump for administering medication during hemodialysis) suggests that ground truth would likely be established through engineering and medical performance standards, rather than expert-derived ground truth in the way a diagnostic imaging AI might.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and not provided in the document. The Venofer Pump Module is an accessory for a hemodialysis machine, administering medication. It is not an AI-assisted diagnostic tool that would typically involve human "readers" or an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This information is not applicable in the context of an AI algorithm. The device is a physical pump module. The document states, "Like the Heparin Pump, the Venofer Pump Module is not a stand alone device. The Venofer Pump Module cannot operate unless it is connected to the 2008T Hemodialysis Machine." This refers to its operational dependency on the larger hemodialysis machine, not a standalone AI performance evaluation.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    This information is not explicitly stated in the document. Given the device's function, ground truth for its performance would likely be established through engineering specifications, flow rate accuracy, absence of adverse events, and efficacy of drug delivery as measured by standard medical parameters (e.g., successful iron administration without pump-related complications), rather than categorical labels like pathology or expert consensus on image interpretation. The submission relies on "verification, safety, performance and software testing."

    8. The sample size for the training set:

    This information is not provided in the document. The device is a physical pump, not an AI algorithm that undergoes a "training set" in the machine learning sense.

    9. How the ground truth for the training set was established:

    This information is not provided in the document, as it's not an AI device with a training set.