Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

    Device Description

    A device intended for use during dental cleaning and periodontal (gum) therapy to remove adherent plaque and food debris from the teeth to reduce tooth decay, potential stains in the gingival groove, the back of the tooth and the tooth and associated gingival line. The electronic oscillation circuit generates ultrasonic frequency electric pulse wave, which is amplified to a certain intensity and conveyed to the transducer. The electrical energy is converted into mechanical energy, and the working head tool is stimulated to produce vibration of the same frequency, so as to achieve the purpose of crushing calculus and loosening tartar.

    AI/ML Overview

    The provided text is a 510(k) summary for an Ultrasonic Scaler (K240707). It describes the device's substantial equivalence to a predicate device and outlines performance data. However, it does not contain the acceptance criteria or a study that proves the device meets those criteria in the context of AI/ML performance.

    Instead, the document focuses on:

    • Comparison of technological characteristics between the proposed device and a predicate device (K163414). This comparison highlights similarities and differences in design, operation controls, power supply, operating conditions, storage conditions, and electrical/EMC/biocompatibility standards.
    • Performance data limited to biocompatibility testing and non-clinical testing of electrical safety and electromagnetic compatibility (EMC) according to various IEC standards.

    Therefore, I cannot provide the requested information regarding acceptance criteria for an AI/ML device, its reported performance, sample sizes, ground truth establishment, or specific AI/ML study details, because this information is not present in the provided document.

    The document discusses acceptance criteria and performance data for a physical medical device (ultrasonic scaler), not an AI/ML component.

    If this was a misunderstanding and you were expecting information related to a traditional medical device's performance, please clarify. Otherwise, the requested AI/ML specific information is absent from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230641

    Validate with FDA (Live)

    Date Cleared
    2023-11-21

    (258 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ultrasonic scaler tips are intended for use by dental professionals to:

    1. Remove supra and sub gingival calculus deposits and stain from the teeth;
    2. Clean and irrigate root canals.
    Device Description

    Ultrasonic scaler tips are an accessory to a piezoelectric ultrasonic handpiece and scaler unit. The devices are used during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of ultrasonic vibration and in various endodontic applications to prepare the teeth and/or root of the teeth for repair.
    The models are GD1, GD2, GD4, GD5, GD6, PD1, PD3, PD4, PD4D, ED1. The Ultrasonic scaler tips are made from Stainless Steel 30Cr13 and the endo files are made of Nickel Titanium Alloy. The tips will be available in M3x0.6 internal thread.
    GD series and PD series(except for PD4D) are used to remove calculus deposits from teeth; PD4D and ED1 are used for root canal indication.

    AI/ML Overview

    The provided text is a 510(k) summary for "Ultrasonic Scaler Tips" and primarily focuses on demonstrating substantial equivalence to predicate devices, rather than establishing acceptance criteria and proving performance through a clinical or algorithmic study with a defined test set, ground truth experts, and statistical outcomes.

    The document discusses:

    • Device Description and Intended Use: The device is an accessory for professional dental use to remove calculus and stain, and to clean/irrigate root canals.
    • Comparison to Predicate Devices: It details similarities and differences in product code, regulation, intended use, device description, operating principle, tip shapes, composition, coatings, interaction with other products, sterilization, and mechanism of treatment to establish substantial equivalence.
    • Performance Data: It lists bench performance testing (biocompatibility, reprocessing validation, conformity to ISO standards like ISO 18397 and ISO 3630-5) conducted to support substantial equivalence.
    • Absence of Clinical Data: Critically, it explicitly states, "The subject of this premarket submission, ultrasonic scaler tips, did not require clinical studies to support substantial equivalence."

    Therefore, the information required to answer your detailed questions about acceptance criteria, test set, data provenance, number/qualifications of experts, adjudication, MRMC studies, standalone performance, and ground truth for training/test sets is not available in this document. This document describes a 510(k) pathway, which often relies heavily on bench testing and comparison to legally marketed predicate devices rather than novel clinical performance studies, especially for devices like scaler tips.

    To directly answer your request based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide specific numerical acceptance criteria or clinical performance metrics for the device itself (e.g., how effectively it removes calculus). Instead, it states that tests were conducted to demonstrate conformity to applicable clauses of ISO standards and that biocompatibility and reprocessing validation were performed. The "performance" is implicitly deemed acceptable if it meets these bench test standards and allows for substantial equivalence to predicates.

    Acceptance Criteria CategoryReported Device Performance (as stated or implied)
    BiocompatibilityConducted in accordance with FDA Guidance (ISO 10993-1); included cytotoxicity, skin sensitization, irritation, acute systemic toxicity, material-mediated pyrogens tests. (Implicitly passed, as substantial equivalence was determined)
    Reprocessing ValidationSterilization validated to ISO 17665-1:2006. Cleaning validation to AAMI TIR30, AAMI TIR12, and FDA guidance. (Implicitly passed)
    Bench PerformanceConformity demonstrated to applicable clauses of: - ISO 18397 Dentistry Powered scaler - ISO 3630-5 Dentistry Endodontic instruments Part 5: Shaping and cleaning instruments (Implicitly passed)
    Clinical PerformanceNot Applicable / Not Required. "did not require clinical studies to support substantial equivalence."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any "test set" in the context of clinical performance data, as clinical studies were not required. For bench testing (biocompatibility, reprocessing, ISO conformity), sample sizes would be determined by the specific test protocols and standards, but are not detailed here.
    • Data Provenance: Not specified, as clinical data was not used. The document pertains to a Chinese manufacturer (Guilin Refine Medical Instrument Co., Ltd.) seeking FDA clearance in the US, so any bench testing would likely have been conducted by or for the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical test set requiring expert ground truth was performed or required for this 510(k) submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set requiring adjudication was performed or required.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is an ultrasonic scaler tip, not an AI-powered diagnostic or assistive tool. No MRMC study or AI assistance is mentioned or relevant to its function.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is an ultrasonic scaler tip, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable. No clinical ground truth was established or used, as no clinical studies were performed. Ground truth for the bench tests would refer to the validated methods and accepted reference standards of the specific physical, chemical, and mechanical tests conducted.

    8. The sample size for the training set

    • Not applicable. No training set for an algorithm was used or mentioned.

    9. How the ground truth for the training set was established

    • Not applicable. No training set for an algorithm was used or mentioned.

    In summary: The provided document is a 510(k) clearance letter and summary for an ultrasonic scaler tip, which falls under Class II medical devices. For such devices, particularly accessories, the FDA often relies on bench testing and substantial equivalence to existing predicate devices rather than requiring extensive clinical trials or AI-specific performance studies. The questions you've asked are highly relevant for AI/software as a medical device (SaMD) clearances, but not for this specific type of physical device and its associated 510(k) pathway.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163414

    Validate with FDA (Live)

    Date Cleared
    2017-11-21

    (351 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ultrasonic scaler is intended to:

    1. Remove supra and sub gingival calculus deposits and stain from the teeth;
    2. Carry out periodontal pocket lavage with simultaneous ultrasonic tip movement;
    3. Clean and irrigate root canals
    Device Description

    The Ultrasonic scaler employs piezoelectric technology to produce a steady power. It is composed of a control unit housing a generator that produces piezo-electric vibrations. The working instrument for the ultrasonic scaler is the handpiece, which is connected to the control unit via a handpiece cord. A scaling tip specific to a particular procedure is attached to the end of the handpiece. The ultrasonic scaler models VET-W3 and VST-W3 connect to an external water supply and delivers water via connected handpiece.
    Ultrasonic scaler consists of main generator, VE and VS handpiece, optional scaler tips (G series, P series, E series), power adapter and foot switch;
    The main part is as follows:

    1. Power adapter: DC 24V, input frequency: 3-20W
    2. Main engine: The main engine is the major part of the circuit control, including ultrasonic pulse circuit, modulator circuit and oscillatory circuit. It is used for converting the ultrasonic oscillatory signal to ultrasonic mechanical oscillatory signal by magnifying and improving the vibrating frequency.
    3. Handpiece: It mainly converts the signal from circuit,which will drive the ultrasonic scaler tips vibrating.
    4. Scaler tip: The scaler tip is used to remove the calculus and plague by vibrating on the teeth surface.
    AI/ML Overview

    This document describes a 510(k) premarket notification for an Ultrasonic Scaler, which is a dental device. The request asks for details related to acceptance criteria and a study proving the device meets those criteria, typically found in submissions for Artificial Intelligence (AI) or machine learning (ML) enabled medical devices.

    However, the provided text does NOT describe an AI/ML-enabled device or a study designed to evaluate its performance against specific acceptance criteria for AI/ML outputs.

    The document states:

    • "Clinical evaluation is not applicable for the proposed device." (Page 6)
    • The comparison is primarily focused on technological characteristics and bench testing against existing standards (IEC, ISO), not clinical performance metrics related to AI/ML.
    • The device is an "Ultrasonic Scaler" for removing calculus and stain, periodontal pocket lavage, and root canal cleaning. This is a mechanical device, not described as having any AI or ML components.

    Therefore, many of the requested points, such as "effect size of how much human readers improve with AI vs without AI assistance," "standalone (i.e. algorithm only without human-in-the loop performance)," "number of experts used to establish ground truth," and "sample size for training set," are not applicable to the content of this 510(k) submission.

    Based on the provided text, I can only address how the device's equivalence and safety were established, not how an AI/ML component meets specific performance criteria.


    Summary of Device Acceptance (for a non-AI/ML device based on this 510(k)):

    The acceptance of the "Ultrasonic Scaler" (Models VET-W3, VST-W3, VET-1, VST-1) is based on demonstrating substantial equivalence to a legally marketed predicate device (Nanning Baolai Medical Instruments Co., Ltd.'s Ultrasonic scaler P7, K140233) rather than proving performance against specific AI/ML acceptance criteria.

    The acceptance involves showing the proposed device has the same intended use, similar product design, same sterilization conditions, and comparable technological characteristics as the predicate device. Bench testing against recognized international standards was conducted to support safety and performance.

    1. Table of acceptance criteria and reported device performance:

    Acceptance Criteria (for Substantial Equivalence to Predicate)Reported Proposed Device Performance/CharacteristicsComments
    Intended Use (same as predicate)The ultrasonic scaler is intended to: 1. Remove supra and sub gingival calculus deposits and stain from the teeth; 2. Carry out periodontal pocket lavage with simultaneous ultrasonic tip movement; 3. Clean and irrigate root canals.Same as predicate device.
    Technological Characteristics (similar to predicate)Working Principle: Piezoelectric effect (same as predicate) Sterilization: Handpiece: autoclaved 135°C/0.22MPa; Scaler tip: autoclaved high temp/pressure (same as predicate) Operating Conditions: Temp 5-40℃, RH 30-80%, Pressure 50-106kPa (same as predicate) Storage Conditions: Temp -10-+50℃, RH ≤ 80%, Pressure 50-106kPa (same as predicate)Key technological aspects are similar.
    Power Supply (meet safety standards)Input voltage: 30VDC 50HZ; Output power: 3-20W. (Predicate: 24VDC 1.5A 50HZ, 3-20W)Slight difference in input voltage, but both meet IEC 60601-1.
    Safety and Performance Standards ComplianceIEC 60601-1:2006+A1:2012 (Electrical safety) IEC 60601-1-2:2007 (EMC) IEC 61205:2001 (Measurement & declaration of output characteristics) ISO 10993-5:2009 (Biocompatibility - cytotoxicity) ISO 10993-10:2010 (Biocompatibility - irritation/sensitization)All listed standards were used for non-clinical bench testing.
    Software Level of ConcernModerate (same as predicate)Addressed as "Moderate." (Details on software testing not provided in text).

    2. Sample size used for the test set and the data provenance:

    • Not applicable for AI/ML performance testing.
    • The device underwent bench testing against international standards (IEC, ISO). This typically involves testing a few representative units to confirm compliance with electrical safety, EMC, performance characteristics, and biocompatibility. The specific number of units tested is not detailed in this summary, but it would not be a "sample size" in the context of AI/ML validation with a large dataset.
    • Data provenance for such tests would be from the manufacturer's internal testing labs, likely in China (based on submitter's address). These are prospective tests performed to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device is not an AI/ML device requiring human expert annotation for ground truth. Ground truth for this device is based on physical measurements of device characteristics against engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. There was no human reader/adjudication involved in establishing ground truth for performance of this mechanical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, this was not done. This is a mechanical dental device, not an AI/ML tool designed to assist human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm. Device performance was assessed via bench tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this type of device, "ground truth" is typically defined by established engineering standards and specifications. For example, electrical safety is confirmed by measuring parameters against limits set in IEC 60601-1, and biocompatibility is assessed by testing against ISO 10993 standards. It's objective measurement against a known reference, not human expert consensus on diagnostic image interpretation.

    8. The sample size for the training set:

    • Not applicable. This device does not use an AI/ML training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This device does not have a training set or associated ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140233

    Validate with FDA (Live)

    Date Cleared
    2015-04-06

    (431 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ultrasonic scaler is intended to:

    1. Remove supra and sub gingival calculus deposits and stain from the teeth;
    2. Carry out periodontal pocket lavage with simultaneous ultrasonic tip movement;
    3. Clean and irrigate root canals.
    Device Description

    Ultrasonic scaler employs piezoelectric technology to produce a steady power. It is composed of a control unit housing a generator that produces piezo-electric vibrations. The working instrument for the ultrasonic scaler is the handpiece, which is connected to the control unit via a handpiece cord. A scaling tip specific to a particular procedure is attached to the end of the handpiece. The ultrasonic scaler connects to an external water supply and delivers water via connected handpiece.

    AI/ML Overview

    The provided text is related to the FDA 510(k) premarket notification for an Ultrasonic Scaler P7. This document is a regulatory submission demonstrating substantial equivalence to a predicate device, not a study proving device performance against specific acceptance criteria in a clinical context with human interaction.

    Therefore, many of the requested categories for a study proving acceptance criteria cannot be extracted from this document, as it focuses on demonstrating safety and efficacy through non-clinical testing and comparison to an already approved device.

    Here's the information that can be extracted and a clear indication of what cannot be found in this type of document:

    1. A table of acceptance criteria and the reported device performance

    This document does not present acceptance criteria in a tabulated format for clinical performance or a direct measure of human improvement with AI. Instead, it lists the non-clinical tests performed to demonstrate safety and performance equivalent to the predicate device.

    Acceptance Criteria CategoryReported Device Performance (as per document)
    Electrical SafetyTested according to IEC 60601-1: 2005+CORR. 1 (2006)+CORR. 2 (2007) and passed.
    Electromagnetic CompatibilityTested according to IEC 60601-1-2:2007 and passed.
    Output Characteristics (Bench Testing)Tested according to IEC 61205:1993 (Ultrasonics-Dental descaler systems-Measurement and declaration of the output characteristics) and performance is similar to the predicate device (both operate at 28kHz ± 3kHz).
    BiocompatibilityTested according to ISO 10993-5:2009 (in vitro cytotoxicity) and ISO 10993-10:2006 (irritation and skin sensitization) for the metal handpiece shell. Results passed.
    Moist Sterilization ValidationValidated according to AAMI / ANSI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013 (Consolidated Text) and passed.
    Intended Use Equivalence"has the same intended use" as the predicate device (removal of supra and sub gingival calculus deposits and stain, periodontal pocket lavage, clean and irrigate root canals).
    Technology Equivalence"both employ piezoelectric technology and produce ultrasonic vibrations at a frequency of 28kHz ± 3kHz".
    Component Equivalence"Have similar main components, i.e. control unit, handpiece cord, handpiece and detachable scaling tips." (Note: handpiece material changed, but biocompatibility validated).

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not applicable in this context. The testing described is non-clinical bench testing and biocompatibility. For these types of tests, "sample size" would refer to the number of units tested, which is not specified in the summary document. It's not a clinical study involving patients or a test set of data like images.
    • Data provenance (country of origin, retrospective/prospective): Not applicable for the non-clinical tests described. The tests were performed to international standards (IEC, ISO, AAMI/ANSI). The manufacturing country is China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This document describes non-clinical testing and regulatory equivalence, not a clinical study with expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not a study involving expert adjudication of clinical cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI/software device, and no MRMC study was conducted or reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. For the non-clinical tests, "ground truth" is established by the specifications and standards outlined (e.g., electrical safety standards, biocompatibility thresholds). The "ground truth" for showing equivalence is the performance characteristics of the predicate device.

    8. The sample size for the training set

    Not applicable. This is a physical device, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113717

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-03-05

    (442 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended only for dental clinic /dental office use. This device generates ultrasonic waves intended for use in dental applications such as scaling, root canal treatment, periodontal and cavity preparation.

    Device Description

    The Varios 370 is a compact, portable control unit powered by the iPiezo® engine. The product comes with a wide range of tip inserts, which can be attached at the distal end of the Varios 2 Handpiece transducer and vibrates at ultrasonic frequencies of 28 to 32 KHz. The Varios 370 LUX features twin LED lights, assuring generally clearer vision and easier identification of the treatment area.

    AI/ML Overview

    This report does not contain information about the acceptance criteria and the study proving the device meets those criteria from an AI/ML perspective. The document is a 510(k) summary for an ultrasonic scaler (Varios 370 / Varios 370 Lux) and focuses on regulatory compliance, outlining the device's description, intended use, technological characteristics, and a summary of testing against applicable industry standards (e.g., electrical safety, biocompatibility, sterilization).

    Here's a breakdown of what is provided, and why it doesn't align with the requested AI/ML-centric information:

    What is provided in the document:

    • Device: Varios 370 / Varios 370 Lux (Ultrasonic Scaler)
    • Intended Use: Dental clinic/office use for applications like scaling, root canal treatment, periodontal and cavity preparation using ultrasonic waves.
    • Summary of Testing: The document states the device underwent "design validation, including software validation" as required by 21 CFR 820.30(g) and was tested in accordance with several standards:
      • IEC 60601-1 (Electrical Safety)
      • UL 60601-1 (Electrical Safety)
      • IEC 60601-1-2 (Electromagnetic Compatibility)
      • ISO 10993-1 (Biocompatibility)
      • ISO 22374 (Dental handpieces Electrical-powered scalers and scaler tips)
      • AAMI/ANSI/ISO 17665-1 (Sterilization)
    • Predicate Device: K031421 – Nakanishi Varios 350 / Varios 350 Lux, to which the current device is considered "substantially equivalent."
    • Conclusion: Substantial equivalence is based on similarities in primary intended use, principles of operation, design rationale, test results, and performance.

    Why this doesn't fit the requested AI/ML acceptance criteria study:

    The provided document describes a traditional medical device (an ultrasonic scaler) and its regulatory submission. It does not mention any AI or Machine Learning components. Therefore, there are no AI/ML acceptance criteria, no studies on algorithm performance, no discussions of training or test sets, no data provenance related to AI, no ground truth establishment for AI, and no MRMC studies or standalone algorithm performance metrics.

    To answer your specific questions based on the absence of AI/ML in this document:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria (for regulatory submission): The device successfully met the requirements of the standards listed (IEC 60601-1, UL 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 22374, AAMI/ANSI/ISO 17665-1) and successfully underwent design validation, including software validation, as per 21 CFR 820.30(g). Performance was also deemed similar to the predicate device.
      • Reported Device Performance (for AI/ML): Not applicable, as there is no AI/ML component.
    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable for AI/ML. The device's performance was likely evaluated through engineering tests (e.g., electrical safety, EMC, vibration frequency, biocompatibility tests) rather than a clinical "test set" in the context of data analysis for AI.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable for AI/ML. Ground truth, in this context, would relate to the physical and functional specifications of the device meeting its design intent and safety standards, rather than expert interpretation of data.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable for AI/ML.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as there is no AI component.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as there is no AI component.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): The "ground truth" for this device would be defined by the technical specifications, performance standards, and safety requirements outlined in the referenced IEC/ISO/UL standards. For example, for electrical safety, the ground truth is whether the device adheres to leakage current limits. For biocompatibility, it's whether the materials are non-toxic.

    8. The sample size for the training set: Not applicable, as there is no AI/ML component.

    9. How the ground truth for the training set was established: Not applicable, as there is no AI/ML component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113530

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-31

    (428 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device, Varios 970 / Varios 970 Lux, an electronic scaler, is intended for use with an appropriate tip for following use: Scaling, Perio, Implant Maintenance, Endodontic, Retrograde Endo, Restorative (for ・ Minimal Intervention/ Finishing/Trimming/ Polishing/Caries of Dentin), Prosthetics (Condensation / Loosening / Plugging)

    Device Description

    Varios 970 is a stand-alone ultrasonic scaler device powered by the iPiezo® engine. The Varios 970 consists of the Varios 970 Control Unit, two independent 400 mL irrigation bottles, handpiece, and a wide range of tip inserts. The tip inserts, when attached at the distal end of the handpiece transducer, resonate at ultrasonic frequencies of 28 -32 kHz. The 400 mL bottles each can carry different solutions with independent pumps. The Varios 970 LUX features twin LED lights that last longer, are more durable, and generate less heat than halogen bulbs.

    AI/ML Overview

    The provided text is a 510(k) summary for the NSK Varios 970 / Varios 970 Lux Ultrasonic Scaler. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and expert-adjudicated ground truth.

    Therefore, an exhaustive response to all specific points regarding acceptance criteria, study details, expert involvement, and ground truth establishment, as typically expected for studies proving AI device performance, cannot be fully provided based solely on the given extract. This document describes a medical device's regulatory submission, not the results of a specific clinical performance study for an AI algorithm.

    However, I can extract and infer information relevant to what would be the "acceptance criteria" and the "study" in the context of a 510(k) submission for a non-AI medical device, which relies heavily on demonstrating equivalence to an already legally marketed product.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of device (ultrasonic scaler), the "acceptance criteria" revolve around demonstrating that the new device operates similarly and safely to the predicate device, especially in terms of fundamental performance characteristics and compliance with electrical safety and EMC standards.

    Acceptance Criteria CategoryReported Device Performance
    Intended Use EquivalenceVarios 970 has the same intended use: Scaling, Perio, Implant Maintenance, Endodontic, Retrograde Endo, Restorative (MI/Finishing/Trimming/Polishing/Caries), Prosthetics (Condensation/Loosening/Plugging).
    Principle of Operation EquivalenceBoth Varios 970 and Varios 75 utilize the iPiezo® engine to generate ultrasonic frequencies of 28-32 kHz. Both convert electrical signals into mechanical vibrations propagated to the handpiece.
    Technological Characteristics EquivalenceBoth have the same operating modes, irrigation modes, and operating time. Varios 970 uses LED lights, which are longer-lasting, more durable, and generate less heat than the predicate, considered an improvement but not a change in fundamental performance.
    Electrical Safety Standards ComplianceThe Varios 970 has been tested in accordance with applicable standards for medical device electrical safety.
    Electromagnetic Compatibility (EMC) ComplianceThe Varios 970 has been tested in accordance with applicable standards for electromagnetic compatibility.
    Sterilization Requirements ComplianceThe Varios 970 has been tested in accordance with the particular requirements for sterilization.
    Design Rationale EquivalenceBasis for the conclusion of substantial equivalence.

    2. Sample Size for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify a "sample size" in terms of patient data or clinical cases for a performance study. For a traditional medical device like an ultrasonic scaler demonstrating substantial equivalence, the "test set" primarily refers to the device itself being subjected to engineering verification and validation tests rather than a cohort of patients.
    • Data Provenance: The 'data' mentioned refers to the results of engineering tests (electrical safety, EMC, sterilization). This data would be generated in a lab setting by the manufacturer (Nakanishi, Inc. in Japan). It is not retrospective or prospective in the clinical sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This is not applicable in the context of this 510(k) submission for an ultrasonic scaler. Ground truth establishment by clinical experts is typically required for diagnostic or AI-driven devices where interpretation of medical data is central. For an ultrasonic scaler, performance is evaluated against engineering specifications and safety standards, not against "ground truth" established by radiologists or other clinical experts for comparative diagnostic accuracy.

    4. Adjudication Method for the Test Set

    • This is not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve discrepancies in expert interpretations to establish a definitive ground truth in a clinical study. As explained above, this 510(k) relies on engineering and performance testing against objective standards and comparison to a predicate device, not on interpreting medical images or data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    • No, an MRMC comparative effectiveness study was not done. Such studies are relevant for evaluating the impact of an AI diagnostic aid on human reader performance, typically in interpreting medical images. This 510(k) is for a physical medical device (an ultrasonic scaler), not a diagnostic AI.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, this is not an AI device, so the concept of a "standalone" or "algorithm only" performance study is not applicable. The device itself is a standalone physical instrument.

    7. The Type of Ground Truth Used

    • Not applicable in the AI context. For this device, the "ground truth" for performance is effectively defined by:
      • Compliance with specific engineering and safety standards (e.g., ISO, IEC for electrical safety, EMC, sterilization).
      • Demonstrated functionality (generating ultrasonic frequencies, irrigation, operating modes) matching the predicate device.
      • The predicate device itself (Varios 75) serving as the benchmark for "acceptable" performance and safety.

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI algorithm, no ground truth needed to be established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062731

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-05-01

    (230 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device, Varios 560, an electric ultrasonic scaler, is intended for use with an appropriate tip for following use.

    • Scaling, Perio, Implant Maintenance, Endodontic, Retrograde Endo, Restorative (for Minimal Scameg, Forto, Inprintin Mailier, Balishing / Caries of Dentin), Prosthetics (Condensation / Loosening / Plugging)
    Device Description

    electric ultrasonic scaler

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for an ultrasonic scaler. It does not contain information about acceptance criteria or a study proving the device meets those criteria. The letter primarily confirms that the device is substantially equivalent to legally marketed predicate devices and outlines regulatory compliance requirements.

    Therefore, I cannot provide the requested information based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053555

    Validate with FDA (Live)

    Date Cleared
    2006-08-21

    (244 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Model :UDS-J/K/N1/N2 :

    1. Removing supra and sub gingival calculus deposits and stains from the teeth

    2. Periodontal pocket lavage with simultaneous ultrasonic tip movement

    Model: UDS-L :

      1. Removing supra and sub gingival calculus deposits and stains from the teeth
        AND/OR
    1. Periodontal pocket lavage with simultaneous ultrasonic tip movement

    2. Preparing, cleaning, and irrigating root canals

    3. Retrograde preparation of root canals

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text only contains a letter from the FDA regarding a 510(k) premarket notification for an ultrasonic scaler. It confirms that the device is substantially equivalent to legally marketed predicate devices and outlines regulatory requirements.

    The document does not contain any information regarding:

    • Acceptance criteria for device performance.
    • A study that demonstrates the device meets acceptance criteria.
    • Performance metrics of the device.
    • Sample sizes for test or training sets.
    • Data provenance.
    • Number or qualifications of experts for ground truth.
    • Adjudication methods.
    • MRMC comparativeeffectiveness studies.
    • Standalone performance.
    • Type of ground truth used.
    • How ground truth for training was established.

    Therefore, I cannot fulfill your request for this specific information based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040396

    Validate with FDA (Live)

    Date Cleared
    2004-04-22

    (65 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    THE ELEGANCE MS-10 (25K) ULTRASONIC SCALER IS DESIGNED FOR USE IN PROPHYLAXIS TREATMENTS PERIODONTIA , AND OTHER AREAS OF OPERATIVE DENTISTRY.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for the Elegance MS-10 (25k) Ultrasonic Scaler. It does not contain information about acceptance criteria, specific study designs, or performance data for a device meeting acceptance criteria.

    The letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976..."

    This means the device was cleared based on its substantial equivalence to a predicate device, not necessarily on a study demonstrating it met specific acceptance criteria. A 510(k) submission typically includes a comparison to a legally marketed predicate device, addressing aspects like technological characteristics and performance, but it often does not involve the kind of detailed clinical study with acceptance criteria, ground truth, and expert evaluation described in your request for AI/medical imaging devices.

    Therefore, I cannot provide the requested information from this document. The document confirms the device's clearance for marketing and its intended use, but it does not detail the specific performance metrics or studies you are asking for.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031421

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-03-26

    (326 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These devices are ultrasonic scalers, intended for use with an appropriate tip for scaling, crown removal, amalgam condensation, excavating, periodontal, and endodontic use.

    Device Description

    Not Found

    AI/ML Overview

    This FDA 510(k) clearance letter dated March 26, 2004, for the Nakanishi, Inc. Varios ultrasonic scalers (Models 350, 350 Lux, 150, 150 Lux, and 550) does not contain information about acceptance criteria, device performance, or any studies as requested.

    The letter explicitly states that the FDA has determined the device is "substantially equivalent" to legally marketed predicate devices. This means that the FDA's decision is based on a comparison to existing devices already on the market, rather than requiring the submission of new clinical or performance study data with specific acceptance criteria.

    Therefore, I cannot provide the requested information based on the provided document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1