    K Number
    K191571

    Validate with FDA (Live)

    Date Cleared
    2020-02-06

    (237 days)

    Product Code
    Regulation Number
    878.4630
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UV Radiation Treatment System models YK-6000A-T, YK-6000B, andYK-6000B-T, are intended for use under the direction of a physician, for the treatment of vitilizo, psoriasis, and eczema. The devices are intended for treatment of these conditions on Fitzpatrick Skin Types (I-VI).

    Device Description

    The proposed device, UV Radiation Treatment System is a portable medical device that consists of tubes, irradiator, control circuit. It is a therapeutic product under the direction of a physician for individuals who require ultraviolet radiation for diagnosed skin disorders. It is for the partial treatment. It has four models: YK-6000A,YK-6000A-T,YK-6000B,YK-6000B-T,in which the model YK-6000A and YK-6000A-T emit UVA light and the model YK-6000B and YK-6000B-T emit UVB light.

    This product has the features are as follows:

    A. Use Philips special UV lamp as light source, with high stability and long service life:

    B. Small bulk, lightweight, operation simply, patients can take with themselves;

    C. There is glisten board inner radiation unit, which can raise radiant efficiency.

    The UV Radiation Treatment System only can be used in Hospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments under direction of physician.

    AI/ML Overview

    The provided document is a 510(k) summary for a UV Radiation Treatment System and specifies that no clinical study was performed to prove the device meets acceptance criteria. The submission relies solely on non-clinical (lab bench) testing and comparison to a predicate device. Therefore, many of the requested sections (2-5, 7-9) cannot be answered with information from this document because there was no study involving patient data, ground truth, or expert review.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of performance metrics against a defined standard (e.g., sensitivity, specificity for an AI device). Instead, it presents a comparison of the proposed device's characteristics against a predicate device, focusing on substantial equivalence. The "performance" is demonstrated through compliance with various safety and electrical standards via non-clinical bench testing.

    Characteristic / Acceptance Criterion (Implicitly "Substantially Equivalent")Proposed Device PerformancePredicate Device (K181805) Performance
    Product Code (FTC)FTCFTC
    Regulation No. (21 CFR 878.4630)21 CFR 878.463021 CFR 878.4630
    Class (II)IIII
    Intended UseTreatment of vitiligo, psoriasis, and eczema under direction of a physician, for Fitzpatrick Skin Types (I-VI).Treatment of psoriasis, vitiligo, and atopic dermatitis (eczema) under direction of a physician, for all skin types (I-VI). Can be used in hospitals, clinics, and households.
    Prescriptive (Yes)YesYes
    Mode of operationPortableHandheld
    Shell material (ABS)ABSABS
    Treatment AreaPartial treatment excluding eyesPartial treatment excluding eyes
    Treatment Time5~20 min0~30min
    UV spectral outputUVA or UVBUVA or UVB
    Lamp configuration1 (9W UV tube)1 or 2 (9W UV tube)
    Power SourceAC outletAC outlet or DC jack
    Wavelength range (nm)UVA: 320400; UVB: 300320UVA: 350400; UVB: 310315
    IPX – Rating / water resistanceIPX0IP22
    Application EnvironmentHospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments under direction of physicianHospital, Clinic, Medical Center, Private Medical Practice, or Other Professional Medical Environments, Household use under direction of physician.
    Electrical Safety/PerformanceComply with IEC60601-1 and IEC 60601-2-57Comply with IEC60601-1 and IEC 60601-2-57
    Home UseNot appliedComply with IEC60601-1-11
    SterileN/AN/A
    Single UseNoNo
    EMCComply with IEC60601-1-2Comply with IEC60601-1-2
    BiocompatibilityCytotoxicity (ISO 10993-5:2009), Sensitization (ISO 10993-10:2010), Irritation (ISO 10993-10:2010), Pyrogen (ISO 10993-11:2017), Acute Systemic Toxicity (ISO 10993-11:2017)Under the conditions of the study, Comply with the requirements
    Label and LabelingConforms to FDA Regulatory RequirementsConforms to FDA Regulatory Requirements

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • No test set was used. As explicitly stated in Section 8.0, "No clinical study is included in this submission." The evaluation for substantial equivalence was based on non-clinical bench testing and comparison to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical study was performed, and therefore no ground truth needed to be established by experts for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical study was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a UV radiation treatment system, not an AI diagnostic or assistance tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical UV treatment system, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable. No clinical study involving patient data and ground truth was conducted. The "ground truth" for the non-clinical tests would be the established performance specifications and regulatory standards (e.g., measured electrical safety, biocompatibility results).

    8. The sample size for the training set

    • Not applicable. As a physical medical device, there is no "training set" in the context of an AI/ML algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set for this device.