Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K212396

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-10-22

    (81 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System Rigid Retrieval Kit Sterilization Tray is intended to enclose and protect surgical instruments optionally used for the UroLift System procedure for storage and sterilization in a prevacuum steam sterilizer. The following instruments are intended to be loaded into the UroLift System Rigid Retrieval Kit Sterilization Tray: · UL-SCOPE4-0 / UL-SCOPE4-0-FE* - Cystoscope, 4.0 mm diameter x 302 mm length, 0° angle of view, wide angle - · UL-GRASP-R Optical Graspers, double action jaws - · UL-SCI-R Optical Scissors, double action jaws * Factory Exchange: Devices with a part number ending in "FE" have been repaired to meet product specifications. The tray by itself is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, 510(k) cleared sterilization wrap to maintain sterility of the enclosed instruments when sterilized using the following prevacuum steam sterilization cycles: Prevacuum Steam Temperature: 132°C (270°F) Sterilization time: 4 minutes Minimum dry time: 30 minutes Maximum weight: 5 lbs (2.7 kg)

    Device Description

    The UroLift System Rigid Retrieval Kit Sterilization Tray is a rigid containment device consisting of a base with lid which enables reprocessing of surgical instruments that are optionally used in the UroLift System Procedure.

    AI/ML Overview

    The document describes the acceptance criteria and the results of non-clinical studies for the UroLift® System Rigid Retrieval Kit Sterilization Tray.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Title of TestPurpose of TestAcceptance Criteria / Source of ReferencesReported Device Performance
    Usability TestingAssured that the user could perform the steps of the Instructions for Use (IFU) to meet the intended use of the sterilization tray and that the tray performed as intended to securely store and sterilize the instruments.The user is able to load and unload the tray, perform the reprocessing workflow (Clean the tray, inspect the tray, clean and inspect the instruments, and load the tray and sterilize), and determine the end of the serviceable life of the tray. ANSI/AAMI ST77:2013 – Containment devices for reusable medical device sterilizationPass
    Cytotoxicity TestingEvaluate the cytotoxicity of a test article extract using an in vitro mammalian cell culture test.The test sample meets the requirements of the test if the biological response is less than or equal to grade 2 (mild). Cytotoxicity testing per ISO 10993-5:2009 – Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicityPass
    Sensitization TestingEvaluate the potential of the test article to cause delayed dermal contact sensitization in the guinea pig maximization test.Grades of 1 or greater observed in the test group generally indicate sensitization, provided that grades of less than 1 are observed on the control animals. If grades of 1 or greater are noted on control animals, then the reactions of test animals that exceeded the most severe control reaction will be considered to be due to sensitization. Sensitization testing per ISO 10993-10:2010 – Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitizationPass
    Intracutaneous Reactivity TestingEvaluate the local dermal irritation of a test article extract following intracutaneous injection in rabbits.The difference of overall mean of the test group to the control group on erythema and edema score must be less than 1. Intracutaneous Reactivity testing per ISO 10993-10:2010 – Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitizationPass
    Useful Life TestingDetermine the serviceable life span of the tray after being subject to 100 cleaning and sterilization cycles per the IFU reprocessing directions.The tray must pass the inspection criteria on the IFU after 100 reprocessing cycles. ANSI/AAMI ST77:2013 – Containment devices for reusable medical device sterilizationPass
    Cleaning ValidationValidate that the cleaning instructions listed in the IFU appropriately clean the tray to ensure the sterilization cycle will be effective.Per the protocol, there were three acceptance criteria: protein residual analysis, hemoglobin residual analysis, and visual inspection. The protein residual benchmark level was 6.4µg/cm² and the hemoglobin benchmark level was 2.2µg/cm². The tray must pass the visual inspection in the IFU. AAMI TIR30: 2011/(R)2016, A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices revised 15 December 2016 Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff issued 17 March 2015Pass
    Sterilization ValidationValidate that the sterilization instructions listed in the IFU appropriately sterilize the tray and contents.The minimum sterility assurance level (SAL) of 10-6 can be achieved if the sterilization instructions in the IFU were followed. ANSI/AAMI ST77:2013, Containment devices for reusable medical device sterilizationPass

    2. Sample Size Used for the Test Set and Data Provenance:
    The document does not explicitly state the sample sizes for each test in terms of the number of devices or data points beyond "100 cleaning and sterilization cycles" for Useful Life Testing. The studies are non-clinical, likely conducted in a laboratory setting by the manufacturer (NeoTract, Inc.), so the data provenance would be from internal testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
    This information is not provided in a way that is applicable to typical "ground truth" establishment for AI/ML devices. The tests described are engineering/performance validations, not clinical studies requiring expert interpretation of results for ground truth. The "acceptance criteria" themselves serve as the standard by which the device's performance is judged.

    4. Adjudication Method for the Test Set:
    Not applicable to these non-clinical performance and biocompatibility tests. The "Pass" results are based on meeting predefined quantitative and qualitative acceptance criteria specified by relevant standards (e.g., ISO, AAMI).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    Not applicable. This is a non-clinical device (sterilization tray) and does not involve AI or human readers for diagnostic purposes.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
    Not applicable. This is a non-AI medical device.

    7. Type of Ground Truth Used:
    The "ground truth" for these non-clinical tests is based on:

    • Predefined performance specifications and acceptance criteria derived from established industry standards (e.g., ANSI/AAMI ST77:2013, ISO 10993 series, AAMI TIR30: 2011/(R)2016).
    • Laboratory measurements (e.g., protein and hemoglobin residuals, biological responses in cell cultures/animals).
    • Visual inspection criteria.

    8. Sample Size for the Training Set:
    Not applicable. This is a non-AI medical device and does not involve machine learning models with training sets.

    9. How the Ground Truth for the Training Set Was Established:
    Not applicable. As noted above, this is a non-AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192781

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-14

    (197 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System Procedure Kit Sterilization Tray is intended to enclose and protect surgical instruments used for the UroLift System Procedure for storage and sterilization in a pre-vacuum steam sterilizer.

    The following instruments are intended to be loaded into the UroLift System Procedure Kit Sterilization Tray:

    • UL-SCOPE / UL-SCOPE-FE Cystoscope, 2.9 mm diameter x 365 mm length, 0° angle of view, 85° field of view
    • UL-SHEATH / UL-SHEATH-FE Sheath, 20 Fr., with 2 tube connectors (Luer lock and Luer lock with stopcock)
    • UL-VO / UL-VO-FE Visual obturator, 20 Fr.

    The tray by itself is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed. validated, 510(k) cleared sterilization wrap to maintain sterility of the enclosed instruments when sterilized using the following pre-vacuum steam sterilization cycles:

    132 °C Pre-vacuum Steam Cycle Exposure temperature: 270 °F (132 °C) Exposure time: 4 minutes Vacuum dry time: 30 minutes

    134 ℃ Pre-vacuum Steam Cycle Exposure temperature: 273 °F (134 °C) Exposure time: 3 minutes Vacuum dry time: 20 minutes

    The maximum weight of the tray is 1.8kg / 4lbs.

    Device Description

    The UroLift System Procedure Kit Sterilization Tray is a rigid containment device consisting of a base with lid which enables reprocessing of the surgical instruments used in the UroLift System Procedure.

    AI/ML Overview

    The UroLift® System Procedure Kit Sterilization Tray is a rigid containment device for surgical instruments, intended for storage and sterilization in a pre-vacuum steam sterilizer. The device was evaluated through non-clinical testing including usability, biocompatibility, cleaning, and sterilization testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Title of testPurpose of testAcceptance Criteria / Source of referencesResults
    Usability TestingAssured that the user could perform the steps of the Instructions for Use to meet the intended use of the sterilization tray and that the tray performed as intended to securely store and sterilize the instruments.The user is able to load and unload the tray, perform the reprocessing workflow (clean the tray, inspect the tray, clean and inspect the instruments, and load the tray and sterilize), and determine the end of the serviceable life of the tray. ANSI/AAMI ST77:2013 - Containment devices for reusable medical device sterilization.Pass
    Cytotoxicity TestingEvaluated the cytotoxicity of a test article extract using an in vitro mammalian cell culture test.The test sample meets the requirements of the test if the biological response is less than or equal to grade 2 (mild). Cytotoxicity testing per ISO 10993-5:2009 – Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity.Pass
    Sensitization TestingEvaluated the potential of the test article to cause delayed dermal contact sensitization in the guinea pig maximization test.Grades of 1 or greater observed in the test group generally indicate sensitization, provided that grades of less than 1 are observed on the control animals. If grades of 1 or greater are noted on control animals, then the reactions of test animals that exceeded the most severe control reaction will be considered to be due to sensitization. Sensitization testing per ISO 10993-10:2010 – Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.Pass
    Intracutaneous Reactivity TestingEvaluated the local dermal irritation of a test article extract following intracutaneous injection in rabbits.The difference of overall mean of the test group to the control group on erythema and edema score must be less than 1. Intracutaneous Reactivity testing per ISO 10993-10:2010 – Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.Pass
    Useful Life TestingDetermined the serviceable lifespan of the tray by subjecting it to 100 cleaning and sterilization cycles.The tray must pass the inspection criteria on the IFU after 100 reprocessing cycles. ANSI/AAMI ST77:2013 - Containment devices for reusable medical device sterilization.Pass
    Cleaning ValidationValidated that the cleaning instructions in the Instructions for Use appropriately clean the tray to ensure the sterilization cycle will be effective.Per the protocol, there were three acceptance criteria: protein residual analysis, hemoglobin residual analysis, and visual inspection. The protein residual benchmark level was 6.4µg/cm², and the hemoglobin benchmark level was 2.2µg/cm². The tray must pass the visual inspection in the IFU. AAMI TIR30: 2011/(R)2016 and Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff issued 17 March 2015.Pass
    Sterilization ValidationValidated that the sterilization instructions in the Instructions for Use appropriately sterilize the tray and contents.The minimum sterility assurance level (SAL) of 10-6 can be achieved if the sterilization instructions in the IFU were followed. ANSI/AAMI ST77:2013, Containment devices for reusable medical device sterilization.Pass

    2. Sample Size Used for the Test Set and Data Provenance:
    The provided document does not specify the exact sample sizes for each test in the test set (e.g., number of units tested for usability, number of samples for biocompatibility). The testing is described as non-clinical in nature. The provenance of the data is manufacturer-generated (NeoTract, Inc.) through non-clinical design verification testing. It is retrospective in the sense that the testing was conducted on the device to demonstrate conformance to pre-defined criteria.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
    This information is not explicitly provided. Non-clinical studies like these generally rely on established scientific protocols and standards (e.g., ISO, AAMI) rather than expert consensus on interpretation for ground truth. For "Usability Testing," the evaluation would typically involve trained users or human factors engineers, but their number and specific qualifications are not detailed.

    4. Adjudication Method for the Test Set:
    The document does not detail specific adjudication methods as would be typical for clinical studies involving multiple reviewers or subjective assessments. The results are reported as "Pass" or "Fail" based on whether the acceptance criteria were met according to established laboratory methods and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
    No, an MRMC comparative effectiveness study was not done. This device is a sterilization tray, not a diagnostic or therapeutic device evaluated for human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:
    Not applicable. This device is a physical medical device (sterilization tray), not an algorithm or AI system.

    7. The Type of Ground Truth Used:
    The ground truth for the non-clinical tests was based on:

    • Established scientific and regulatory standards: e.g., ISO 10993 for biocompatibility, AAMI ST77 and TIR30 for sterilization and cleaning.
    • Predefined acceptance criteria: Specific quantitative (e.g., protein residual levels, SAL) and qualitative (e.g., visual inspection, user performance steps) benchmarks derived from these standards and industry best practices.
    • Performance of the device itself through repeated cycles (for useful life testing).

    8. The Sample Size for the Training Set:
    Not applicable. This device is a physical medical device, not an AI or algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:
    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193269

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-20

    (24 days)

    Product Code
    Regulation Number
    876.5530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

    Device Description

    The UroLift System is designed to access the prostatic urethra and deliver one UroLift Implant through a lobe of the prostate. The UroLift System is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. Multiple implants are deployed in the UroLift System procedure. The implants secure the retracted position of the urethra, thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant. The procedure typically requires 2-6 implants to retract the obstruction. The UL400 (most recently cleared in K173087), consists of two main components, the UroLift Delivery Device (single use), and the UroLift Implants (one implant per delivery device). Each Delivery Device comes pre-loaded with one UroLift Implant.

    AI/ML Overview

    This document is a 510(k) summary for the UroLift System (UL400) by NeoTract, Inc. It describes a submission concerning a modification to a contraindication, specifically increasing the maximum prostate volume from >80 cc to >100 cc for patients indicated for treatment with the UroLift System.

    Based on the information provided in the document, here's a description of the acceptance criteria and the study that proves the device meets them, specifically regarding the change in contraindication:

    Acceptance Criteria and Study for Modified Contraindication (Increased Prostate Volume)

    The core of this submission is not about proving a new device's performance from scratch, but rather demonstrating that a modification to a contraindication (increasing the prostate volume limit) does not negatively impact the device's established safety and effectiveness. Therefore, the "acceptance criteria" and "study" described herein are focused on demonstrating non-inferiority or equivalence of outcomes in the expanded patient population.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (for expanded prostate volume)Reported Device Performance (as demonstrated by clinical literature review)
    Device Performance Equivalence (Safety & Effectiveness):Outcomes (from sponsored and independent clinical studies):
    - Symptom response in patients with prostate volumes >80cc is equivalent to those with prostate volumes <80cc."These studies show that the symptom response... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Quality of life in patients with prostate volumes >80cc is equivalent to those with prostate volumes <80cc."These studies show that the... quality of life... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Uroflowmetry results in patients with prostate volumes >80cc are equivalent to those with prostate volumes <80cc."These studies show that the... uroflowmetry... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Adverse events in patients with prostate volumes >80cc are equivalent to those with prostate volumes <80cc."These studies show that the... adverse events... are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    - Catheterization rates in patients with prostate volumes >80cc are equivalent to those with prostate volumes <80cc."These studies show that the... catheterization rates are equivalent to the outcomes of patients with prostate volumes less than 80cc." (page 5)
    Biocompatibility Equivalence:"The modification of the contraindication does not impact the biocompatibility of the existing UroLift System. As such, the data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication." (page 5)
    Sterilization & Shelf-Life Equivalence:"The modification of the contraindication does not impact the sterilization or shelf life of the existing UroLift System. As such, the data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication." (page 5)
    Technological Characteristics & Principles of Operation Equivalence:"The UroLift System (UL400) described in this submission is substantially equivalent to the previously cleared generations of the device. The UL400 was previously cleared in K173087. This submission concerns the modification of a contraindication of the device and does not impact the device itself." (page 4) "The data provided demonstrated the NeoTract UroLift System with a modified contraindication is as safe and effective, has the same intended use, technological characteristics and principles of operation as the predicate device." (page 5)

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Description: The "test set" for this submission is not a newly collected dataset for the modified contraindication. Instead, it relies on a "clinical literature review" of existing studies.
    • Sample Size: The document states "clinical literature review of both sponsored and independent clinical studies that included men with prostate volumes greater than 80cc." (page 5). The specific number of patients or studies reviewed is not provided in this summary.
    • Data Provenance: The document does not specify the country of origin. It indicates the data comes from "sponsored and independent clinical studies," implying a mix of retrospective and prospective data collection, but does not explicitly state the design. For example, the predicate device (K173087) would have included prospective study data to support its original clearance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • This submission relies on a clinical literature review to support the change in contraindication. Therefore, the "ground truth" for the equivalence of outcomes for patients with prostate volumes >80cc is established by the aggregate findings and conclusions of the respective clinical studies included in that review.
    • The document does not specify the number of experts, their qualifications, or their role in reviewing the literature for this specific submission. The "review" itself would have been conducted by the sponsor's clinical/regulatory team, and ultimately evaluated by the FDA.

    4. Adjudication Method for the Test Set

    • Given that the evidence is based on a "clinical literature review" rather than a de novo test set with expert reads, there is no explicit adjudication method described for this submission. The "adjudication" effectively comes from the scientific rigor and peer review (if published) of the underlying studies themselves, and the FDA's assessment of the presented literature review.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done for this specific 510(k) submission. This submission is focused on a change to a contraindication based on existing clinical evidence, not on evaluating the performance of an AI model or the improvement of human readers with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. The UroLift System is a medical device (implantable system), not an AI algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for the claim that outcomes (symptom response, quality of life, uroflowmetry, adverse events, catheterization rates) are equivalent for patients with prostate volumes >80cc compared to those <80cc is based on clinical study outcomes data (clinical endpoints) as reported in sponsored and independent clinical literature.

    8. The Sample Size for the Training Set

    • Not applicable. This submission is for a medical device with a modified contraindication, not an AI/ML algorithm that requires a training set. The device's original clearance (K173087) would have involved clinical trials, which could be considered the "training data" in a broader sense for establishing the device's original safety and efficacy.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. See point 8. The ground truth for the predicate device's efficacy and safety would have been established through well-defined clinical trial endpoints and data collection, usually with a panel of clinical investigators and potentially independent data monitoring committees.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190377

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-21

    (30 days)

    Product Code
    Regulation Number
    876.5530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

    Device Description

    The UroLift System is designed to access the prostatic urethra and deliver one UroLift Implant through a lobe of the prostate. The UroLift System is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. Multiple implants are deployed in the UroLift System procedure. The implants secure the retracted position of the urethra, thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant. The procedure typically requires 2-6 implants to retract the obstruction. The UL400 (most recently cleared in K173087), consists of two main components, the UroLift Delivery Device (single use), and the UroLift Implants (one implant per delivery device). Each Delivery Device comes pre-loaded with one UroLift Implant.

    AI/ML Overview

    The provided document describes a 510(k) submission for the UroLift System (UL400) by NeoTract, Inc. This submission focuses on minor device modifications and asserts substantial equivalence to a previously cleared predicate device (K173087).

    Crucially, this document focuses on non-clinical performance testing and does not describe a study involving patient data, ground truth establishment, or clinical outcomes that would typically be associated with AI/ML device performance. The acceptance criteria and "device performance" discussed here pertain to engineering and material testing, not diagnostic or predictive accuracy.

    Therefore, many of the requested sections (sample size for test/training sets, data provenance, expert qualifications, adjudication, MRMC study, standalone performance, type of ground truth) are not applicable to the information provided in this regulatory submission.

    Here is the information that can be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
    Mechanical PerformanceDeployment TestingNot explicitly stated, but implied to ensure safe and effective deployment."All acceptance criteria were met."
    Needle Depth TestingNot explicitly stated, but implied to ensure correct tissue penetration."All acceptance criteria were met."
    Urethral End-piece to Suture Joint StrengthNot explicitly stated, but implied to ensure structural integrity."All acceptance criteria were met."
    Needle Spool to Right Case wall gap testingNot explicitly stated, but implied to ensure proper assembly and function."All acceptance criteria were met."
    BiocompatibilityGeneral Biocompatibility TestingCompliance with ISO 10993-1: Biological evaluation of medical devices."Passed the relevant tests according to ISO 10993-1."
    SterilitySterilization ValidationMinimum gamma irradiation dose to ensure a 10⁻⁶ Sterility Assurance Level (SAL)."Confirmed that the modification did not impact product sterility."
    Aging/TransitImpact of sterilization, accelerated aging, and transitNot explicitly stated, but implied that all mechanical tests were performed on devices subjected to these conditions, and acceptance criteria were met after these conditions."All acceptance criteria were met" even after worst case sterilization, accelerated aging, and transit testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. The described tests are non-clinical engineering and material tests, not clinical studies with patient data. Therefore, there is no "test set" in the context of diagnostic performance or patient data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. Ground truth in the context of expert consensus for diagnostic performance is not relevant to this type of device modification submission. The "ground truth" here would be the physical properties and performance metrics defined in engineering standards and specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are used in clinical studies for resolving discrepancies in expert interpretations, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This document pertains to a medical device (UroLift System) for treating BPH, not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI assistance metrics are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. The UroLift System is a physical medical device, not an algorithm. Standalone algorithm performance is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the tests performed would be defined by engineering specifications, international standards (e.g., ISO 10993-1 for biocompatibility), and established test methods (e.g., for mechanical strength, sterility assurance). This is not related to clinical "ground truth" derived from patient data.

    8. The sample size for the training set

    • Not Applicable. This submission does not involve an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This submission does not involve an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173087

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-28

    (90 days)

    Product Code
    Regulation Number
    876.5530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

    Device Description

    The UroLift System is designed to access the prostatic urethra and deliver one UroLift Implant through a lobe of the prostate. The UroLift System is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. Multiple implants are deployed in the UroLift System procedure. The implants secure the retracted position of the urethra, thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant. The procedure typically requires 2-6 implants to retract the obstruction.

    The UroLift System includes two generations of the device, the UL400 and the UL500. Both generations use the same UroLift Implant. The only differences are in the delivery device.

    The UL400 (cleared in K133281), consists of two main components, the UroLift® Delivery Device (single use), and the UroLift Implants (one implant per delivery device). Each Delivery Device comes pre-loaded with one UroLift Implant.

    The UL500 (UroLift 2 System, cleared in K172359) is comprised of the UroLift® Delivery Handle (single patient reusable), the UroLift Implant Cartridges (single-use) and the UroLift Implants (one implant per cartridge). Each cartridge is pre-loaded with one UroLift Implant. The cartridges fit into the delivery handle. Each patient procedure will use one dedicated handle and the number of cartridges/implants necessary to perform a typical procedure (estimated 2-6 implants). For the UL500, users are also provided with an optional Scope Seal which enables them to examine the anatomy between implant deployments without removal of the telescope from the Delivery Handle.

    AI/ML Overview

    The provided text describes a 510(k) submission for the UroLift System (UL400 and UL500) and includes information about a clinical study conducted to support its safety and effectiveness, specifically for patients with an enlarged median lobe.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Effectiveness Endpoint (at 6 months): The 95% lower confidence limit of the mean percent improvement in subject's International Prostatic Symptom Score (IPSS) over baseline for the UroLift System must be $\geq$ 25%.Results: At 6 months, the 95% lower confidence limit of the mean percent improvement in subject's International Prostatic Symptom Score (IPSS) over baseline for the UroLift System was 50.8%, which is substantially above the goal of $\geq$ 25%. The mean percent change at 6 months was 57.7%.
    Safety Endpoint (at 3 months): The composite observed rate of post-procedure device-related serious complications ≤ 15%. Composite device-related serious complications include: de Novo (new) severe urinary retention lasting more than 21 consecutive days post procedure, device-related formation of fistula between the rectum and urethra, perforation of the rectum or GI tract, damage to ureter or ureteral orifices, damage to the trigone requiring surgical repair, or de novo, sustained erectile dysfunction.Results: The safety endpoint was achieved with 0% (5.7% CI upper limit) meeting criteria within the composite. From 3 to 6 months, there were no reported events that would meet the endpoint criteria.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 45 subjects were enrolled in the study.
    • Data Provenance: The study was a prospective, multicenter, non-blinded, single-arm study conducted at 9 investigational sites located in the US. It gathered data through 1, 3, 6, and 12 months of follow-up.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The information provided does not specify the number of experts or their qualifications used to establish ground truth for the test set in the conventional sense of image or diagnosis review. Instead, a Clinical Events Committee (CEC) was used for adjudication related to the safety endpoint.

    4. Adjudication method for the test set

    • Adjudication Method: A Clinical Events Committee (CEC) was responsible for adjudicating adverse events and relevant subject questionnaires to evaluate against the safety endpoint. The specific composition or method (e.g., 2+1, 3+1) of the CEC is not detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focused on the performance of the UroLift System directly on patients, not on human reader performance with or without AI assistance. This device is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. The UroLift System is a medical device (implantable system) that is used in a procedure performed by a clinician. Its "performance" is measured by its impact on patient symptoms and safety, not as a standalone algorithm.

    7. The type of ground truth used

    • Ground Truth: The ground truth for the effectiveness endpoint was based on patient-reported outcomes, specifically the International Prostatic Symptom Score (IPSS).
    • The ground truth for the safety endpoint was based on clinical events adjudicated by a Clinical Events Committee against predefined criteria for serious complications.

    8. The sample size for the training set

    • Training Set Sample Size: The document does not mention a "training set" as this is a clinical trial for a medical device and not an AI/machine learning model that would typically have a training set. The 45 enrolled subjects constitute the study population for evaluating the device's safety and effectiveness.

    9. How the ground truth for the training set was established

    • This question is not applicable as there was no explicit "training set" in the context of an AI/ML model for this medical device study. The ground truth for evaluating the device's performance was established through the objective measurement of IPSS scores and the adjudication of adverse events by a CEC.
    Ask a Question

    Ask a specific question about this device

    K Number
    K172359

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-08-18

    (14 days)

    Product Code
    Regulation Number
    876.5530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift 2 System (UL500) is intended for treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men 50 years of age or older.

    Device Description

    The UroLift 2 System (UL500) is comprised of the UroLift® Delivery Handle (single patient reusable), the UroLift Implant Cartridges (single-use) and the UroLift Implants (one implant per cartridge). Each patient procedure will use one dedicated sterile handle and the number of cartridges/implants necessary to perform a successful procedure (estimated 2-6 implants). The cartridges fit into the delivery handle.

    The UroLift 2 System (UL500) is designed to access the prostatic urethra and deliver one UroLift Implant through a lateral lobe of the prostate. The UroLift 2 System (UL500) is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. The implants secure the retracted position of the urethra thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving LUTS. This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant.

    Minor modifications have been made to the UroLift 2 System (UL500) to improve the device. The overall design, as well as the patient contacting materials, is substantially equivalent to the predicate device.

    Users are now provided with an optional Scope Seal which enables them to examine the anatomy between implant deployments without removal of the telescope from the device handle. In addition, a Suture Support Tube has been incorporated into the device to provide additional support to the suture during Implant Cartridge assembly at NeoTract's manufacturing facility.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the UroLift 2 System (UL500), which is a medical device and not an AI/ML powered device. Therefore, the information requested in your prompt about an AI device's acceptance criteria, study details, and related metrics (such as sample sizes for test and training sets, expert ground truth, adjudication methods, MRMC studies, and standalone performance) cannot be found in this document.

    The document primarily focuses on establishing substantial equivalence to a predicate device based on:

    • Intended Use: Treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men 50 years of age or older.
    • Technological Characteristics: Minor modifications (optional Scope Seal, Suture Support Tube) to improve the device, but the overall design and patient-contacting materials are substantially equivalent to the predicate. The UroLift Implant is identical.
    • Performance Testing: Testing demonstrated the modified device meets the same performance requirements as the predicate, plus additional requirements for the Scope Seal. The changes do not affect the implant deployment procedure.

    Without additional information about an AI/ML component or a separate study describing such, I cannot fulfill your request as it pertains to an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162345

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-09-21

    (30 days)

    Product Code
    Regulation Number
    876.5530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UroLift System is intended for treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men 50 years of age or older.

    Device Description

    The UroLift System UL500 is comprised of the UroLift® Delivery Handle (single patient resusable), the UroLift Implant Cartridges (single-use) and the UroLift Implants (one implant per cartridge). Each patient procedure will use one dedicated sterile handle and the number of cartridges/implants necessary to perform a successful procedure (estimated 2-6 implants). The cartridges fit into the delivery handle.

    The UroLift System UL500 is designed to access the prostatic urethra and deliver one UroLift Implant through a lateral lobe of the prostate. The UroLift System UL500 is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. The implants secure the retracted position of the urethra thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving LUTS. This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant.

    AI/ML Overview

    I am sorry, but the provided text does not contain the specific information required to answer your request about acceptance criteria and a study proving device performance. The document is a 510(k) summary for the UroLift® System UL500, which focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a comprehensive study for initial device efficacy.

    While it mentions "Performance Testing" was conducted, it only states that "the device meets the same performance requirements of the predicate device" and that "the changes made to the device do not affect the implant deployment procedure." It does not provide:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes, data provenance, or details about the test set.
    • Information about experts, ground truth establishment, or adjudication methods.
    • Details about multi-reader multi-case (MRMC) comparative effectiveness studies or standalone algorithm performance.
    • Sample size or ground truth details for a training set.

    The document primarily focuses on establishing that modifications to the UroLift System UL500 do not introduce new questions of safety or effectiveness compared to its predicate device, rather than providing a detailed study of its primary efficacy or safety from scratch.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN130023

    Validate with FDA (Live)

    Device Name
    UROLIFT SYSTEM
    Manufacturer
    Date Cleared
    2013-09-13

    (190 days)

    Product Code
    Regulation Number
    876.5530
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men age 50 and above.

    Device Description

    The UroLift® System is composed of two main components: the UroLift® Delivery Device and UroLift® Implant (Figure 1). Each Delivery Device comes pre-loaded with one UroLift® Implant. The insertion of the UroLift® Delivery Device into the male urethra is performed under direct visualization using standard surgical technique, using a standard cystoscopy sheath and a Karl Storz 10324AA telescope. The UroLift® Delivery Device is designed to access the prostatic urethra and deliver one UroLift Implant through a lateral lobe of the prostate. The UroLift Delivery Device is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. A UroLift® Implant is then deployed transversely through the prostatic tissue. The Implants secure the retracted position of the urethra thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLife® Implant (Figure 2).

    AI/ML Overview

    The UroLift System is an implantable transprostatic tissue retractor system intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men age 50 and above. The acceptance criteria and the study proving the device meets these criteria are detailed below.

    1. Acceptance Criteria and Reported Device Performance

    The clinical study for the UroLift System, known as the L.I.F.T. (Luminal Improvement Following Prostatic Tissue Approximation for the Treatment of Lower Urinary Tract Symptoms) Pivotal Study, established the following acceptance criteria and demonstrated performance:

    Acceptance CriteriaReported Device Performance
    Effectiveness Endpoints
    Mean IPSS improvement from baseline at 3 months demonstrates a minimum statistical margin of 25% compared to cystoscopy alone (Control Group).UroLift System group demonstrated superiority to control at 3 months. IPSS was reduced from baseline by 50.0% at 3 months.
    Lower bound of a one-sided 97.5% confidence interval of mean percent change (improvement) from baseline in IPSS at 12 months in the UroLift® group is ≥ 30%.Improvement of the treated group at 12 months was statistically significant compared to their baseline IPSS. IPSS was reduced from baseline by 45.5% at 12 months.
    Safety Endpoints
    Upper bound of a one-sided 97.5% confidence interval of the observed rate of post-operative urinary catheterization > 7 days is <10%.The probability of a harmful event (device or procedure related serious adverse event) was approximately 1.4%. There was a 1.4% chance of catheterization > 7 days. The risks of the device include a low risk of serious adverse events. Most events were mild or moderate, transient, and resolved without sequelae or intervention. The adverse events had an onset within 7 days of procedure and resolved within 30 days following the procedure.
    Secondary Effectiveness Endpoints (Implicitly accepted if results show improvement)Most secondary endpoints showed clinical and statistical improvement. These included: - Qmax: 63.5% increase at 3 months over baseline and 54.8% at 12 months. - IPPS at 2 weeks: Symptom relief began as early as 2 weeks. - Quality of Life (QoL): Improved from 4.9 at baseline to 1.8 at 6 weeks and 2.4 at 12 months. - BPHII: Decreased 73% at 6 weeks and remained close out to 1 year (64.6 % and 66.7% respectively). - IIEF (International Index of Erectile Function): Showed preservation of function with no statistical change from baseline at 12 months.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: A total of 206 subjects were randomized in a 2:1 ratio (UroLift: 140; Control: 66).
    • Data Provenance: The study was a prospective, multinational clinical trial.
      • Countries of Origin: U.S. (14 investigational sites) and non-U.S. (5 sites: three in Australia and two in Canada).
      • Nature of Data: Prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the "number of experts used to establish the ground truth" in the context of diagnostic interpretation or image review for the test set. Instead, the ground truth for effectiveness was based on patient-reported outcomes (IPSS, QoL, BPHII) and objective measures (Qmax), and safety outcomes (adverse events, catheterization rates).

    • Clinical Trial Design: The effectiveness assessment was double-blinded in terms of both the subject and the assessor for the initial 3-month period. This implies that the assessors evaluating the outcomes were clinical professionals with the necessary qualifications to conduct these assessments.
    • Safety Event Adjudication: An independent reviewer adjudicated one event in the Australian Feasibility Study to be both a procedure and a device-related event. For the Pivotal Study, it states "Of 20, only 6 were adjudicated to be procedure or device related." This indicates that expert clinical review (adjudication) was performed for adverse events. The specific qualifications of these adjudicators are not detailed, but it can be inferred they were qualified medical professionals (e.g., urologists or clinical investigators).

    4. Adjudication Method for the Test Set

    • Safety Event Adjudication: For the Feasibility Study, "An independent reviewer adjudicated one event to be both a procedure and a device related event." For the Pivotal Study, "Of 20, only 6 were adjudicated to be procedure or device related." This suggests that adverse events were subject to expert review and classification. The exact number of adjudicators per event (e.g., 2+1 or 3+1) is not specified, but it was not "none".
    • Effectiveness Assessment: The effectiveness assessment was "double blinded in terms of both the subject and the assessor" for the primary endpoints at 3 months. This implies a rigorous, standardized assessment without mentioning a specific adjudication method for primary outcome measures beyond this double-blinding.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study was a clinical trial comparing the UroLift System to a sham control, focusing on patient outcomes and device efficacy/safety, not on human reader performance with or without AI assistance. The device itself is a medical implant and delivery system, not an AI diagnostic tool.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Not applicable. The UroLift System is a medical device (implant and delivery system) for a surgical procedure, not an algorithm or AI system to be evaluated for standalone performance.

    7. The Type of Ground Truth Used

    The ground truth for the clinical studies was established through:

    • Patient-Reported Outcomes: International Prostate Symptom Score (IPSS), Quality of Life (QoL) index, and Benign Prostatic Hyperplasia Impact Index (BPHII), International Index of Erectile Function (IIEF). These are subjective measures reported by the patients themselves.
    • Objective Clinical Measurements: Peak flow rate (Qmax), observations of adverse events (including catheterization duration), and various clinical assessments by investigators.
    • Adjudicated Adverse Events: As mentioned above, some adverse events were adjudicated by independent reviewers.

    8. The Sample Size for the Training Set

    • Not applicable in the context of AI/ML. The UroLift System is a physical medical device. The "training" for its development involved non-clinical bench, animal, and cadaver studies, as well as an earlier "Feasibility Study" (64 subjects followed for 5 years in Australia) which refined methods and gathered initial safety/feasibility data before the larger Pivotal Study. This feasibility study could be considered analogous to early-stage development and refinement, but not a "training set" in the machine learning sense.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable as it's not an AI/ML device. For the "feasibility study" mentioned (which provided early clinical data used to refine the device and inform the pivotal study), the ground truth was established similarly to the pivotal study, through patient-reported outcomes (subject questionnaire responses) and safety events observed by investigators ("Site reported adverse events"). One event was independently adjudicated.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1