K Number
DEN130023
Device Name
UROLIFT SYSTEM
Manufacturer
Date Cleared
2013-09-13

(190 days)

Product Code
Regulation Number
876.5530
Type
Direct
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men age 50 and above.

Device Description

The UroLift® System is composed of two main components: the UroLift® Delivery Device and UroLift® Implant (Figure 1). Each Delivery Device comes pre-loaded with one UroLift® Implant. The insertion of the UroLift® Delivery Device into the male urethra is performed under direct visualization using standard surgical technique, using a standard cystoscopy sheath and a Karl Storz 10324AA telescope. The UroLift® Delivery Device is designed to access the prostatic urethra and deliver one UroLift Implant through a lateral lobe of the prostate. The UroLift Delivery Device is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. A UroLift® Implant is then deployed transversely through the prostatic tissue. The Implants secure the retracted position of the urethra thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLife® Implant (Figure 2).

AI/ML Overview

The UroLift System is an implantable transprostatic tissue retractor system intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men age 50 and above. The acceptance criteria and the study proving the device meets these criteria are detailed below.

1. Acceptance Criteria and Reported Device Performance

The clinical study for the UroLift System, known as the L.I.F.T. (Luminal Improvement Following Prostatic Tissue Approximation for the Treatment of Lower Urinary Tract Symptoms) Pivotal Study, established the following acceptance criteria and demonstrated performance:

Acceptance CriteriaReported Device Performance
Effectiveness Endpoints
Mean IPSS improvement from baseline at 3 months demonstrates a minimum statistical margin of 25% compared to cystoscopy alone (Control Group).UroLift System group demonstrated superiority to control at 3 months. IPSS was reduced from baseline by 50.0% at 3 months.
Lower bound of a one-sided 97.5% confidence interval of mean percent change (improvement) from baseline in IPSS at 12 months in the UroLift® group is ≥ 30%.Improvement of the treated group at 12 months was statistically significant compared to their baseline IPSS. IPSS was reduced from baseline by 45.5% at 12 months.
Safety Endpoints
Upper bound of a one-sided 97.5% confidence interval of the observed rate of post-operative urinary catheterization > 7 days is <10%.The probability of a harmful event (device or procedure related serious adverse event) was approximately 1.4%. There was a 1.4% chance of catheterization > 7 days. The risks of the device include a low risk of serious adverse events. Most events were mild or moderate, transient, and resolved without sequelae or intervention. The adverse events had an onset within 7 days of procedure and resolved within 30 days following the procedure.
Secondary Effectiveness Endpoints (Implicitly accepted if results show improvement)Most secondary endpoints showed clinical and statistical improvement. These included: - Qmax: 63.5% increase at 3 months over baseline and 54.8% at 12 months. - IPPS at 2 weeks: Symptom relief began as early as 2 weeks. - Quality of Life (QoL): Improved from 4.9 at baseline to 1.8 at 6 weeks and 2.4 at 12 months. - BPHII: Decreased 73% at 6 weeks and remained close out to 1 year (64.6 % and 66.7% respectively). - IIEF (International Index of Erectile Function): Showed preservation of function with no statistical change from baseline at 12 months.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: A total of 206 subjects were randomized in a 2:1 ratio (UroLift: 140; Control: 66).
  • Data Provenance: The study was a prospective, multinational clinical trial.
    • Countries of Origin: U.S. (14 investigational sites) and non-U.S. (5 sites: three in Australia and two in Canada).
    • Nature of Data: Prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the "number of experts used to establish the ground truth" in the context of diagnostic interpretation or image review for the test set. Instead, the ground truth for effectiveness was based on patient-reported outcomes (IPSS, QoL, BPHII) and objective measures (Qmax), and safety outcomes (adverse events, catheterization rates).

  • Clinical Trial Design: The effectiveness assessment was double-blinded in terms of both the subject and the assessor for the initial 3-month period. This implies that the assessors evaluating the outcomes were clinical professionals with the necessary qualifications to conduct these assessments.
  • Safety Event Adjudication: An independent reviewer adjudicated one event in the Australian Feasibility Study to be both a procedure and a device-related event. For the Pivotal Study, it states "Of 20, only 6 were adjudicated to be procedure or device related." This indicates that expert clinical review (adjudication) was performed for adverse events. The specific qualifications of these adjudicators are not detailed, but it can be inferred they were qualified medical professionals (e.g., urologists or clinical investigators).

4. Adjudication Method for the Test Set

  • Safety Event Adjudication: For the Feasibility Study, "An independent reviewer adjudicated one event to be both a procedure and a device related event." For the Pivotal Study, "Of 20, only 6 were adjudicated to be procedure or device related." This suggests that adverse events were subject to expert review and classification. The exact number of adjudicators per event (e.g., 2+1 or 3+1) is not specified, but it was not "none".
  • Effectiveness Assessment: The effectiveness assessment was "double blinded in terms of both the subject and the assessor" for the primary endpoints at 3 months. This implies a rigorous, standardized assessment without mentioning a specific adjudication method for primary outcome measures beyond this double-blinding.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study was a clinical trial comparing the UroLift System to a sham control, focusing on patient outcomes and device efficacy/safety, not on human reader performance with or without AI assistance. The device itself is a medical implant and delivery system, not an AI diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

  • Not applicable. The UroLift System is a medical device (implant and delivery system) for a surgical procedure, not an algorithm or AI system to be evaluated for standalone performance.

7. The Type of Ground Truth Used

The ground truth for the clinical studies was established through:

  • Patient-Reported Outcomes: International Prostate Symptom Score (IPSS), Quality of Life (QoL) index, and Benign Prostatic Hyperplasia Impact Index (BPHII), International Index of Erectile Function (IIEF). These are subjective measures reported by the patients themselves.
  • Objective Clinical Measurements: Peak flow rate (Qmax), observations of adverse events (including catheterization duration), and various clinical assessments by investigators.
  • Adjudicated Adverse Events: As mentioned above, some adverse events were adjudicated by independent reviewers.

8. The Sample Size for the Training Set

  • Not applicable in the context of AI/ML. The UroLift System is a physical medical device. The "training" for its development involved non-clinical bench, animal, and cadaver studies, as well as an earlier "Feasibility Study" (64 subjects followed for 5 years in Australia) which refined methods and gathered initial safety/feasibility data before the larger Pivotal Study. This feasibility study could be considered analogous to early-stage development and refinement, but not a "training set" in the machine learning sense.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable as it's not an AI/ML device. For the "feasibility study" mentioned (which provided early clinical data used to refine the device and inform the pivotal study), the ground truth was established similarly to the pivotal study, through patient-reported outcomes (subject questionnaire responses) and safety events observed by investigators ("Site reported adverse events"). One event was independently adjudicated.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR NEOTRACT'S UROLIFT SYSTEM

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Implantable Transprostatic Tissue Retractor System. An implantable transprostatic tissue retractor system is a prescription use device that consists of a delivery device and implant. The delivery device is inserted transurethrally and deploys the implant through the prostate. It is designed to increase prostatic urethral patency by providing prostate lobe tissue retraction while preserving the potential for future prostate procedures and is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men.

NEW REGULATION NUMBER: 21 CFR 876.5530

CLASSIFICATION: II

PRODUCT CODE: PEW

BACKGROUND

DEVICE NAME: UROLIFT SYSTEM

SUBMISSION NUMBER: K130651

DATE OF DE NOVO: MARCH 7, 2013

  • CONTACT: NEOTRACT, INC., NANCY ISAAC, JD, MPH- VP, Clinical Affairs, Regulatory and Quality 4473 WILLOW RD STE 100 PLEASANTON, CA 94588

REQUESTER'S RECOMMENDED CLASSIFICATION: II

INDICATIONS FOR USE

The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men age 50 and above.

LIMITATIONS

    1. Caution: Federal Law restricts this device to sale by or on the order of a physician.

{1}------------------------------------------------

    1. The UroLift® System should not be used if the patient has:
    • Prostate volume of >80 cc
    • · An obstructive or protruding median lobe of the prostate
    • A urinary tract infection
    • · Urethra conditions that may prevent insertion of delivery system into bladder
    • · Urinary incontinence
    • Current gross hematuria
    • A known allergy to nickel
    1. The safety of the delivery system has not been evaluated in the MR environment, and therefore, the delivery system should not be used within the MR environment.
    1. The UroLift® Implant has been shown to be MR Conditional and can be scanned under the following conditions:
    • · Static magnetic field strength of 3 Tesla or less
    • · Maximum spatial gradient magnetic field of 720 Gauss/cm
    • · A maximum whole-body-averaged specific absorption rate (SAR) of 4 W/kg for 15 minutes of scanning

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

Device Name: UroLift® System

UL400 Device Model(s):

The UroLift® System is composed of two main components: the UroLift® Delivery Device and UroLift® Implant (Figure 1). Each Delivery Device comes pre-loaded with one UroLift® Implant. The insertion of the UroLift® Delivery Device into the male urethra is performed under direct visualization using standard surgical technique, using a standard cystoscopy sheath and a Karl Storz 10324AA telescope. The UroLift® Delivery Device is designed to access the prostatic urethra and deliver one UroLift Implant through a lateral lobe of the prostate. The UroLift Delivery Device is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. A UroLift® Implant is then deployed transversely through the prostatic tissue. The Implants secure the retracted position of the urethra thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLife® Implant (Figure 2).

{2}------------------------------------------------

Figure 1. The two main components of the UroLife® System, the UroLift® Delivery Device (Left) and the UroLife Implant (Right). Each Delivery Device comes pre-loaded with one UroLift® Implant.

Image /page/2/Figure/2 description: This image is titled "Figure 2: Male Prostatic Urethra - Pre/Post UroLift System Treatment". The image is split into two sections, pre-treatment and post-treatment. The pre-treatment section shows the bladder neck, capsule, urethra, and external urethral sphincter. The post-treatment section shows the UroLift implants.

Figure 2. Illustrations of coronal cross-sections of the male prostatic urethra, pre- and post-UroLift System treatment (Left and Right, respectively), demonstrating how the UroLift Implants can effect separation of the lobes and open the prostatic urethra.

{3}------------------------------------------------

UroLift® Delivery Device

The UroLift® Delivery Device, shown in greater detail in Figures 3 and 4, is a single-use device that delivers one permanent UroLift Implant. The key design elements of the Delivery Device are: i) the hollow nitinol Needle, which penetrates the urethra and prostatic lobe to access the prostatic capsule and then places the Capsular Tab; ii) the stainless steel Shaft Assembly with Distal Tip, which guides the hollow nitinol needle and stores the Urethral End-Piece; and iii) the Handle Assembly, which contains a Needle Safety Lock and three actuators (the Needle Trigger, the Retraction Lever, and the Urethral Release) that allow the user to deploy the UroLift Implant.

Figure 3. Details of the UroLife® Delivery Device, identifying the Distal Tip, The Shaft Assembly and, in a magnified view, the Hollow Nitinol Needle as it appears when protruding from the Distal Tip.

The Handle Assembly of the UroLift Delivery Device is the primary user interface and contains all of the actuators that are used to deploy the UroLift Implant. The Handle geometry is a pistolstyle grip with actuators located on the front and rear of the handgrip portion. The pistol-style grip design of the Handle Assembly provides a user interface familiar to urologists and incorporates several safeguards for the user and the patient.

The Delivery Device user interface comprises seven elements, all part of the Handle Assembly, as shown in Figure 4: 1) the Handle, 2) the Scope Lock, 3) the Sheath Lock, 4) the Needle Safety Lock, 5) the Needle Trigger, 6) the Retraction Lever, and 7) the Urethral Release. These elements are used to connect the UroLift Delivery Device to the ancillary equipment and complete the implantation process. Additional aspects of the user interface are provided via the ancillary equipment. The telescope, light source, and endoscopy camera provide a video image of the urethra. The video is used to select the proper locations to place the UroLift Implants.

{4}------------------------------------------------

Image /page/4/Figure/0 description: The image is titled "Figure 4: UroLift Delivery Device - User Interface Elements". The image shows a UroLift delivery device with labels pointing to different parts of the device. The labels include: Sheath Lock, Scope Lock, Needle Trigger, Needle Safety Lock, Retraction Lever, Urethral Release, and Handle.

Figure 4. Details of the UroLift® Delivery Device, identifying the Handle, Scope Lock, Sheath Lock, Needle Safety Lock, Needle Trigger, Retraction Lever, and Urethral Release.

UroLift® Implant

The UroLift® Implant is a permanent implant delivered transurethrally into the prostate (Figure 2). The UroLift Implant comprises a nitinol Capsular Tab, a stainless steel Urethral End-Piece, and a monofilament Poly Ethylene Terephthalate (PET) suture. All of the materials and methods of fabrication utilized to produce the UroLift ® Implant are commonly used in the construction of other medical implants that have been cleared/approved for use in various markets, including the United States.

Refer to the Instructions for Use for additional details.

SUMMARY OF NONCLINICAL/BENCH STUDIES

The sponsor conducted a series of non-clinical performance testing to demonstrate that the UroLift System would perform as anticipated. Non-clinical testing included biocompatibility, shelf-life, toxicological, sterility, package integrity, transit, MR compatibility, bench performance, implant corrosion, animal and cadaveric testing, as summarized in Table 1, below.

TestPurposeResults
Biocompatibility - Implant
Cytotoxicity (ISO 10993-5) -DeployedTo evaluate the biocompatibility of the test article (deployed and undeployedUroLift Implant) extract using an in vitro mammalian cell culture test byPass
Cytotoxicity (ISO 10993-5) -Undeployeddetermining whether leachables extracted from the test article would causecytotoxicity.Pass
Sensitization (ISO 10993-10) -To evaluate the potential of the test article (undeployed UroLift Implant) toPass
Undeployedcause delayed contact sensitization in the guinea pig maximization test.Pass
Intracutaneous Reactivity (ISO10993-10) - UndeployedTo evaluate the local dermal irritant effects of leachables extracted from thetest article (non-deployed UroLift Implant) following intracutaneous injectioninto rabbits.Pass
Systemic Toxicity (ISO 10993-11) - UndeployedTo evaluate the acute systemic toxicity of leachables extracted from the testarticle (UroLift Implant) following a single intravenous or intraperitonealinjection in mice.Pass
Genotoxicity (ISO 10993-3) -Ames Bacterial ReverseMutation - UndeployedTo evaluate whether an extract of the test material (UroLift Implant) willcause mutagenic changes in a tryptophan-dependent strain of Escherichia colior one or more strains of histidine-dependent Salmonella typhimurium in thepresence or absence of S9 metabolic activation.Pass
Genotoxicity (ISO 10993-3) -Mouse Lymphoma Assay -ImplantTo evaluate the mutagenic potential of a test material extract using the mouselymphoma forward mutations assay procedure.Pass
Genotoxicity (ISO 10993-3) -Mouse Peripheral BloodMicronucleus - ImplantTo evaluate the potential for a test article extract to cause damage tochromosomes or the mitotic apparatus of murine erythroblasts by measuringthe frequency of micronucleated reticulocytes (MN-RET) of treated mice.Pass
Implantation – 4 week (ISO10993-6) - UndeployedTo evaluate the local tissue response to the test article implanted in muscletissue. This test was performed twice, once with a four week endpoint andPass
Implantation - 12 week (ISO10993-6) - Undeployedonce with a twelve week endpoint.Pass
Biocompatibility - Delivery Device
Cytotoxicity (ISO 10993-5)To evaluate the biocompatibility of the test article (patient-contacting portionof the UroLift Delivery Device) extract using an in vitro mammalian cellculture test by determining whether leachables extracted from the test articlewould cause cytotoxicity.Pass
Sensitization (ISO 10993-10)To evaluate the potential of the test article (patient-contacting portion of theUroLift Delivery Device) to cause delayed contact sensitization in the guineapig maximization test.Pass
Intracutaneous Reactivity (ISO10993-10)To evaluate the local dermal irritant effects of leachables extracted from thetest article (patient-contacting portion of the UroLift Delivery Device)following intracutaneous injection into rabbits.Pass
Shelf LifeAfter aging conditions, test articles were subjected to Joint and MaterialsTesting, System Compatibility and Simulated Use Testing and PackageIntegrity Testing.Pass
Implant Extractables andImmersionComprehensive toxicological assessment of UroLift Implant to determine thedegradation products on the sterilized and deployed UroLift Implant.NA
Exhaustive Extraction Testing- UroLift ImplantThis supplemental testing was conducted for information purposes only. Thisexhaustive extraction testing evaluated extractables and leachables frommaterials in the UroLift suture.NA
Exaggerated Extraction (60Day Immersion) Testing -UroLift ImplantDefine the method for identifying and quantifying ions leached from finishedNeoTract Urolift ®Suture Implants under simulated in vivo conditions.Provide a standardized method based on ISO 10993-15 "BiologicalEvaluation of Medical Devices- Part 15: Identification and Quantification ofDegradation Products from Metals and Alloys" to assess ion release from thedevices.Pass
Total Digestion - Suture OnlyIn order to evaluate the overall toxicity of the Neotract implant, this testidentifies and quantifies the residual metallic content that may have been usedduring the production of the suture.NA
Extractions Study - SutureOnlyThe purpose of this test was to repeat extraction studies completed on thePET suture with test methods demonstrated to be capable of measuring tracequantities of PET extractables and degradation products in biorelevantextraction media.NA
Toxicological AssessmentTo evaluate the risks to patients associated with exposure to substancesidentified as potential leachables from components of the UroLift® Implant.Pass
Sterility
Sterilization ValidationTo ensure that 25 kGy gamma irradiation is an adequate minimum dose toachieve an SAL of 10-6 (validation). To determine the appropriate dose mapand ensure reproducibility of same.ANSI/AAMI/ISO 11137-1ANSI/AAMI/ISO 11137-2ANSI/AAMI/ISO 11137-3Pass
PyrogenicityEvaluate whether a material extraction of the UroLift Implant induced apyrogenic response following intravenous injection in rabbits.Pass
PackagingSee Package Integrity testing.
Package Integrity
Bubble EmissionSterile Barrier Testing via bubble emission, dye penetration and lid peelstrength.Pass
Dye PenetrationPass
Lid Peel strengthPass
LabelsPass
Transit TestingClimatic conditioning and transit simulation per ASTM D4169-09. Testarticles were subjected to Joint and Materials Testing, System Compatibilityand Simulated Use Testing and Package Integrity Testing.Pass
Magnetic Resonance (MR)CompatibilityMR compatibility per ASTM F2052, ASTM F2213, ASTM F2182, ASTMF2119Pass
Performance Testing - Bench
Joint and Materials StrengthImplant strengths, critical device Joint and Material strength.
• Capsular Tab-SutureJoint StrengthPass
• Urethral End-Piece-Suture Joint StrengthPass
• Suture MaterialStrengthPass
• Suture and Ferrule toSuture SpoolAssemblyPass
• Suture to FerruleAssemblyPass
• Needle to Needle EndOvermoldPass
• Shaft to ShaftOvermoldPass
• Lever Lock and TapePass
• Distal Tip Lateral LoadPass
• Distal Tip Axial TensileLoadPass
• Atraumatic Tape toShaft Bond FunctionPass
System Compatibility andSimulated UseSystem Compatibility, Simulated Use Testing, Needle Depth, Suture TagLength.
• Compatible withcommerciallyavailable 20F StorzSheath minimumPass
• Compatible withcommerciallyPass
available 2.9mmscope
• Deploy one UroLift®ImplantPass
• Suture Tag Length onUrethral End-PiecePass
• Needle depthPass
Implant Corrosion
SEM/Auger - Capsular TabTo determine the oxide layer thickness and composition of NeoTractCapsular Tab component.Pass
SEM/Auger – Urethral End-PieceTo document the oxide layer composition and thickness of the SS UrethralEnd-Piece.Pass
Potentiodynamic - CapsularTabTo determine the corrosion behavior of the NeoTract UroLift NitinolCapsular Tab.Pass
Potentiodynamic - UrethralEnd-PieceTo determine the corrosion behavior of the NeoTract UroLift stainless steelUrethral End-Piece components.Pass
Potentiostatic – Capsular TabTo determine the resistance to crevice corrosion of overlapped NeoTractCapsular Tab to suture assembly.NA
Potentiostatic - Urethral End-PieceTo determine the resistance to crevice corrosion of overlapped NeotractStainless Steel Urethral End-Piece components.NA
Performance Testing - AnimalCanineLong term safety of the UroLift Implant in mammalian prostate.Pass
Chronic Animal Study (6months)Initial safety and feasibility of luminal restoration of the prostatic urethraprocedure and the UroLift Implant (1, 3, 6 months)
Chronic Animal Study (1month)Feasibility and safety of the UroLift Implant at 1 month
Chronic Animal Study (Dog,GLP, 90 and 180 Days)Systemic and local tissue safety evaluation of the UroLift Implant at 3 and 6months
Chronic Animal Study (Dog.GLP, 12 months)To confirm the safety of long-term effect on the prostate and surroundingurinary tract tissues presence of the UroLift Implant.NA
Performance Testing - CadaverAssessment of procedural success andpotential device safetyPass

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

Table 1. A summary of non-clinical testing conducted on the UroLift System.

BIOCOMPATIBILITY/MATERIALS

The UroLift System is comprised of two main components, UroLift Implant and the UroLift Delivery Device. The biocompatibility and materials testing included those tests recommended by both FDA memorandum G95-1 entitled "Required Biocompatibility Training and Testing Profiles for Evaluation of Medical Devices" and the current revision of the International Standard ISO 10993, Biological Evaluation of Medical Devices. In addition, UroLift Implant leachables were evaluated through immersion testing.

Biocompatibility

The materials selected for use in the UroLift System are well characterized, used for similar medical device applications, and appropriate for gamma sterilization.

Leachables and Extractables

A review was conducted of the results of various studies in which the amounts of extracted metals and polymer-related organic substances were determined following extractions conducted

{8}------------------------------------------------

with the UroLift Implant and its PET suture component. These studies were performed for different durations of time and with various extraction solvents (hexane, isopropyl alcohol, purified water, and/or artificial urine) to provide a range of solubility conditions. Some of these conditions mimicked the physiological environment in which the UroLift Implant is implanted in male patients while others represented unrealistically aggressive worst-case conditions. As summarized in Table 1, the results of the studies indicate that none of the metals or polymerrelated organic substances would be expected to leach from the implant or the PET suture in amounts that would be of toxicological concern to the patient, even in the event that 10 UroLift Implants were to be implanted on a single occasion. The available data do not suggest a toxicological risk to the patient due to the possible leaching of metals or polymer-related organic substances from the UroLift Implant.

Conclusions

In summary, the UroLift System has been evaluated to determine the potential for toxicity resulting from contact of the device component materials with the body. The results of this testing demonstrated the UroLift System is biocompatible when used as intended.

SHELF LIFE, PACKAGE INTEGRITY AND TRANSIT

The shelf life/stability of the UroLift System and its packaging has been demonstrated via real time aging at multiple time points including 18 months. The shelf life testing consisted of demonstrating that the UroLift System and its packaging meet all key design requirements, including joint and materials strength testing, system compatibility and simulated use, and package integrity, after gamma sterilization and aging for a specified time period. The successful completion of these tests demonstrated that the product meets the applicable design requirements and supports the device bearing a shelf life of 18 months.

Package integrity testing was performed on the UroLift System after gamma sterilization, climatic conditioning and transit simulation and demonstrated that the sterile barrier is maintained after these challenge conditions. Label integrity was visually inspected to confirm that labels are intact and legible. The UroLift System packaging test demonstrated that the packaging components met all design requirements.

Transit testing was performed on the UroLift System after gamma sterilization, climatic conditioning and transit simulation in accordance with the appropriate Assurance Levels of ASTM D4169-09 (Standard Practice for Performance Testing of Shipping Containers and Systems.) The test battery comprehensively assessed the key design requirements with the exception of the shelf life and MR conditional requirements and sterility). The UroLift System met all design requirements after the transit simulation.

STERILITY

The UroLift System is gamma sterilized via a validated minimum dose of halidation was conducted in accordance with the ISO11137-2: Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose, and resulted in a sterility assurance level (SAL) of 10-6. The packaging configuration of a tray, lid, sealed to a Tyvek cover has been shown to maintain the sterile barrier. In addition, the UroLift System has been found to be nonpyrogenic in accordance with ISO 10993-11. Based on this, the UroLift System will be labeled

{9}------------------------------------------------

and provided as "sterile".

MAGNETIC RESONANCE (MR) COMPATIBILITY

Testing was conducted by an independent laboratory to evaluate the magnetic resonance (MR) safety and compatibility of the UroLift Implant. Three (3) UroLift Implants were used to test for implant displacement, implant torque, magnetic resonance imaging (MRI) related heating, and image artifacts. In magnetic field interaction tests, the UroLift Implant showed a translational attraction of 200 and 1200 torque. The MRI-related heating test resulted in a physiologically inconsequential implant temperature rise of 2000 °C. The Implant displayed an artifact that appeared moderate in size in relation to the size and shape of this device. Based on all the test results, the UroLift Implant was assigned a status of MR Conditional. The safety of the delivery system has not been evaluated in the MR environment and therefore the delivery system should not be used within the MR environment.

PERFORMANCE TESTING - BENCH

The UroLift System, comprising the UroLift Delivery Device and the UroLift Implant, as packaged, labeled and gamma sterilized, has been tested in accordance with a comprehensive test plan that was derived in accordance with the FDA 2010 Guidance for the Non-Clinical and Clinical Investigation of Devices Used for the Treatment of Benign Prostatic Hyperplasia (BPH) and applicable ISO, ASTM and USP standards.

Joint and materials strength testing was performed on the UroLift System after gamma sterilization, climatic conditioning and transit simulation and demonstrates that these design requirements were met after challenge conditions. The UroLift System components and subsystems met all design requirements for joint and materials strength.

Simulated use testing was performed on the UroLift System after gamma sterilization, climatic conditioning and transit simulation and demonstrates that these design requirements are met after challenge conditions. These tests required demonstrating fit and mechanical connection to a cystoscope sheath and telescope and deploying the delivery device in a simulated use environment. The UroLift System met all design requirements for system compatibility and simulated use.

PERFORMANCE TESTING - ANIMAL

NeoTract has conducted multiple animal studies to demonstrate the safety of the UroLift System when implanted into mammalian prostates. Several studies were conducted prior to 2012 (Prior Studies) and were reviewed by FDA as part of the Investigation Device Exemption (IDE). The progression of the Prior Studies served to refine critical study methods, such as surgical approach and urine collections techniques, as well as obtain on-going safety data for the UroLift System.

NeoTract conducted an additional study (New GLP Study) to confirm the safety of the UroLift Implant with regard to the long-term effect on the prostate and surrounding tissue and with regard to chronic systemic toxicity. The control arm of the study used animals that received the full surgical procedure including deployment of the Needle from the UroLift Delivery Device, though no UroLift® Implant was delivered. The New GLP Study included five animals in each

{10}------------------------------------------------

group (test and control) at each timepoint: 1, 6, and 12 months (30 animals total).

The results show that the local tissue reaction attributed to the UroLift® Implant up to 6 months post-implantation was characterized by minimal chronic inflammation and minimal to mild fibrosis and was considered to be a steady-state end-stage response typical for a biocompatible chronic implant (New GLP Study). No proliferative changes including urothelial or glandular hyperplasia or neoplasia in the prostate attributable to UroLift® Implant were observed (New GLP Study). In Summary, the results from these studies demonstrated the following:

  • · Minimal chronic inflammation associated with the UroLift® Implant.
  • · Prostate tissue responds to physical pressure via local atrophy that relieves the applied pressure over time. The New GLP Study demonstrated a stable, end-stage atrophy at 6 months with no evidence of any proliferative changes. Prior Studies demonstrated no meaningful difference in induced proliferation or apoptosis between UroLift® Implanttreated animals and untreated control animals as previously presented by NeoTract.
  • · Minimal chronic inflammation associated with the UroLift® Implant. This was demonstrated in the New GLP Study and in Case Reports of clinical subjects who progressed to transurethral resection of the prostate (TURP).

PERFORMANCE TESTING - CADAVER

As part of the design development of the UroLift® System, a series of cadaver studies were performed. Cadaveric evaluations allowed early assessments of procedural success to be made. All devices performed as intended and met the established product specifications. A cumulative total of thirteen UroLift® Implants were deployed within the prostatic tissue resulting in an observed increase in luminal area in all three cadaver studies. The Delivery Device was easily inserted and withdrawn. The overall time per deployment was approximately 5 minutes. All deployments were performed by a board-certified urologists made observations that the System was simple to use. As confirmed by cystoscopy, there was no evidence of damage to the urethra. All devices performed as intended and no safety issues were observed. The results from this cadaver study further confirm the safety and ability to deploy the UroLift Implant.

SUMMARY OF CLINICAL INFORMATION

Two clinical studies were conducted to confirm the safety, feasibility and effectiveness of the UroLift® System.

Feasibility Study - Australia, 2005

In 2005, a feasibility study was initiated in Australia with the objective of determining whether the placement of UroLift® transprostatic implants was feasible and safe for the effective treatment of lower urinary tract symptoms (LUTS) secondary to benign prostatic hyperplasia (BPH). Safety and feasibility of the device was evaluated on the 64 subjects enrolled and followed out to 5 years. All subjects have reached their two year follow-up. For the intent to treat (ITT) population, all procedures were completed successfully and symptomatic improvement was seen at follow up intervals compared to baseline based on subject questionnaire responses. No unanticipated adverse events have been recorded for this study. Site reported adverse events reported early in the study were typical postoperative symptoms. An independent reviewer adjudicated one event to be both a

{11}------------------------------------------------

procedure and a device related event. No other events were adjudicated as device related in this study. Twenty-four (24) month results from this study demonstrated that placement of the UroLife implants into the prostate was feasible and that it was sufficiently safe to support continuation onto a broader randomized controlled trial.

Pivotal Study - US/OUS, 2010

The L.I.F.T. (Luminal Improvement Following Prostatic Tissue Approximation for the Treatment of Lower Urinary Tract Symptoms) Pivotal Study was an FDA approved Investigational Device Exemption (IDE) Study (G090012), designed as a prospective, multinational, 2:1 randomized, single-blinded controlled clinical trial of the UroLife® System. The study had two phases, a randomized single-blind period followed by a non-randomized open-label period. The blinded randomized trial portion of the study started at the time of the procedure and ended at the subject's three-month visit. Once the 3-month follow-up was completed, the subjects were unblinded. After unblinding, if their symptoms returned and treatment was required, subjects were allowed to receive treatment (either de novo treatment if originally in the Control group or as a retreatment if originally treated with UroLift) with the UroLift® System or any other approved BPH treatment.

The primary endpoints of this study were to establish safety of the device. test the superiority of UroLift® system effectiveness when compared at 3 months to the randomized Control group, and to test UroLift® system durability at 1 year. At 3 months, the effectiveness of the UroLife System was demonstrated by comparison of the change from baseline of the International Prostate Symptom Score (IPSS) of the treated group to the Control group. At 12 months, the long-term effectiveness for subjects in the UroLift® group was demonstrated by comparison of the IPSS at 12 months to baseline. The effectiveness assessment was double blinded in terms of both the subject and the assessor. The primary safety endpoint was an assessment of the proportion of subjects requiring extended post-operative urinary catheterization, defined as an occurrence of a subject requiring catheterization within the first 3 days as part of a postoperative management for inability to void and requiring the catheter for more than 7 days.

The UroLife System would be considered effective if the mean IPSS improvement from baseline at 3 months demonstrates a minimum statistical margin of 25% compared to cystoscopy alone, and the lower bound of a one-sided 97.5% confidence interval of mean percent change (improvement) from baseline in IPSS at 12 months in the UroLift® group is ≥ 30%. The primary safety endpoint was to demonstrate that the upper bound of a one-sided 97.5% confidence interval of the observed rate of post-operative urinary catheterization > 7 days is <10%. The secondary effectiveness endpoints were: 1) peak flow rate (Qmax) at 3 and 12 months; 2) IPPS at 2 weeks; 3) Quality of Life (QoL) at 12 months; and 4) BPHII at 12 months.

A total of 206 subiects were randomized in a 2:1 ratio. (UroLift: 140: Control: 66) at 19 investigational sites. Fourteen U.S. sites and five non-U.S. sites (three in Australia and two in Canada) participated in the investigation. Subjects in the UroLift® group underwent the UroLift® System procedure. Subjects in the Control group underwent a sham procedure, which included standard cystoscopy with perioperative sounds and verbal comments that mimicked the UroLift® group procedure.

{12}------------------------------------------------

All study co-primary effectiveness and safety endpoints were achieved. The UroLift® System group was superior to control at 3 months and improvement of the treated group at 12 months was statistically significant as compared to their baseline IPSS. All secondary endpoints were clinically and statistically improved. International Index of Erectile Function (IIEF) showed preservation of function with no statistical change from baseline at 12 months. Symptom relief began as early as 2 weeks and subjects returned to normal activity by 8.6 days. IPSS was reduced from baseline by 50.0% at 3 months and by 45.5% at 12 months. QoL improved from 4.9 at baseline to 1.8 at 6 weeks and 2.4 at 12 months. BPHII decreased 73% at 6 weeks and remained close out to 1 year (64.6 % and 66.7% respectively). Urine Outflow demonstrated a 63.5% increase at 3 months over baseline and 54.8% was achieved at 12 months.

The majority of subjects experienced a greater than 83% and 87% responder rate (>3 point IPSS) at 12 months in the Intent to Treat (ITT) and Per Protocol (PP) groups, respectively. There were no appreciable differences in the treatment effect noted in co-primary effectiveness endpoint 1 based on geographical region. Based on sub-group analyses, it was determined that younger patients (< 66 years) allowed for greater improvement in IPSS score. Also, higher baseline IPSS scores allowed for greater improvement.

The risks of the device include a less than 20% chance that a patient will experience no benefit and a low risk of serious adverse events, though mild, transient, adverse events may be expected. All reported adverse events occurred within the first 30 days post procedure. Of 20, only 6 were adjudicated to be procedure or device related. No unanticipated adverse events have been recorded for this study. The adverse reactions associated with the UroLift® were comparable to other minimally invasive surgical therapies as well as standard cystoscopy, as shown here: Dysuria (35.0 %); Hematuria (26.4 %); Urgency (9.3 %); Incontinence (7.9 %); Calculus Urinary (6.4 %); Retention (5.7 %); Constipation (5.0 %); Nocturia (5.0 %); Bladder spasms (4.3 %); and PSA increase (4.3 %).

The probability of a harmful event was approximately 1.4% for a device or procedure related serious adverse event. There was a 1.4% chance of catheterization > 7 days and a 10% chance of return of lower urinary tract symptoms (LUTS) significant enough to seek medications or additional surgical intervention. Most events were mild or moderate, transient, and resolved without sequelae or intervention. The adverse events had an onset within 7 days of procedure and resolved within 30 days following the procedure.

The L.I.F.T. pivotal clinical study demonstrated a 12 months improvement of the primary and secondary endpoints with a high compliance rate, few significant deviations and low loss to followup. The results of this study corroborate the Australian Feasibility results.

LABELING

The labeling for the UroLift® System is consistent with the clinical data and covers all the hazards and other clinically relevant information that may impact safe and effective use of the device. The labeling is sufficient and satisfies the requirements of 21 CFR Part 801.109 Prescription devices.

{13}------------------------------------------------

The following summarizes how the UroLift® System labeling addresses the special controls related to labeling:

    1. To address failure to deploy the device or misdeployment, the labeling includes an explanation of how proper implant depth is achieved during deployment will be included in the Instructions for Use (IFU) document and other product labeling.
    1. To address infection due to presence of foreign body, an expiration date is included in the IFU document.
    1. To address failure to deploy the device or misdeployment, failure of the implanted device, improperly placed implants, the occurrence of genito-urinary adverse events, and the presence of implants adversely affecting subsequent interventions, a detailed description of the training required prior to use of the device is included in the IFU document.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with the use of an implantable transprostatic tissue retractor system and the measures necessary to mitigate these risks.

Identified RiskMitigation Measure
Adverse Tissue Reaction to the Device1. Biocompatibility Testing
2. In Vivo Testing
Infection Due to Presence of Foreign Body1. Sterilization Validation
2. Labeling (including expiration dating)
3. Shelf life testing
Migration of Implanted Device1. In Vivo Testing
2. MR Compatibility Testing
Failure to Deploy Device or Misdeployment1. Non-clinical Testing
2. In Vivo Testing
3. Labeling
Failure of Implanted Device1. Non-clinical Testing (Mechanical)
2. Non-clinical Testing (Resistance to Degradation)
3. Shelf life testing
4. In Vivo Testing
5. Labeling
Improperly Placed Implants1. In Vivo Testing
2. Labeling
Occurrence of Genito-Urinary Adverse Events1. In Vivo Testing
2. Labeling
Presence of Implants Adversely Affects Subsequent1. Non-clinical Testing
Interventions2. In Vivo Testing
3. Labeling

Table 2. The identified risks that may be associated with the use of an implantable transprostatic tissue retractor system and the measures necessary to mitigate these risks.

SPECIAL CONTROLS:

In combination with the general controls of the FD&C Act, the UroLift® System is subject to the following special controls:

{14}------------------------------------------------

  • (1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
  • (2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
  • (3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity and device functionality over the requested shelf life.
  • (4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
    • (A) Deployment testing must be conducted
    • (B) Mechanical strength must be conducted
    • (C) Resistance-to-degradation testing must be conducted
  • (5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance (MR) environment
  • (6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of
    • use. The following performance characteristics must be tested:
    • (A) Deployment testing must be conducted
    • (B) Implant migration must be conducted
  • (7) Labeling must bear all information required for safe and effective use of the device, and must include:
    • (A) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device
    • (B) Information on the patient population for which the device has been demonstrated to be effective
    • (C) A detailed summary of the device technical parameters
    • (D) Information on how the device operates and the typical course of treatment
    • (E) An expiration date/shelf life
    • (F) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.

BENEFIT/RISK DETERMINATION

The results of the L.I.F.T. Study showed that the UroLite® System is effective in improving symptoms associated with LUTS as shown by the significant improvement in mean IPSS score at three months when compared to Control, and at 12 months when compared to baseline IPSS value. Corresponding, significant improvements in other relevant clinical parameters, Qmax, BPHII, and IPSS QoL support the conclusion. Further, the study demonstrates a favorable safety profile associated with the UroLift® System.

The low rate of extended catheterization, and of catheterization rates in general, as well as the demonstration of no occurrences of de novo sustained erectile dysfunction or anejaculation are cited as evidence for the conclusion. The low rate of SAEs associate with the UroLift® procedure, and the rates and duration of the observed adverse events are deemed acceptable. In fact, the observed procedure-related adverse events are the same as those commonly associated with transurethral

{15}------------------------------------------------

procedures, such as dysuria, hematuria, pelvic pain, urinary retention, and elevated prostate specific antigen (PSA).

There are significant benefits provided by the UroLift® System including improvement in lower urinary tract symptoms (LUTS) as measured by IPSS, low catheterization rate, improvement in Qmax, Quality of Life, and BPHII with a correspondingly low risk of complications. In addition patients are interested in the preservation of sexual function provided by this technology. It offers a less invasive procedure that can be performed under local anesthesia in an outpatient setting with rapid recovery and relief. Finally, the risks associated with the device are not significant and are lower than those for other BPH interventions. The risks can be mitigated by limiting the device to prescription use only, through the Instructions for Use (IFU) which provides instructions, contraindications, warnings and precautions, and through a training program which is available for qualified physicians (e.g. urologists).

In conclusion, given the available information above, the data support that, for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men age 50 and above, the probable benefits outweigh the probable risks for the UroLift® System. The device provides substantial benefits and the risks can be mitigated by the use of general and the identified special controls.

CONCLUSION

The de novo for the UroLift® System is granted and the device is classified under the following:

Product Code:PEW
Device Type:Implantable Transprostatic Tissue Retractor System
Class:II
Regulation:21 CFR 876.5530

§ 876.5530 Implantable transprostatic tissue retractor system.

(a)
Identification. An implantable transprostatic tissue retractor system is a prescription use device that consists of a delivery device and implant. The delivery device is inserted transurethrally and deploys the implant through the prostate. It is designed to increase prostatic urethral patency by providing prostate lobe tissue retraction while preserving the potential for future prostate procedures and is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia in men.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity, and device functionality over the requested shelf life.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Mechanical strength must be conducted.
(iii) Resistance-to-degradation testing must be conducted.
(5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance environment.
(6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Implant migration must be conducted.
(7) Labeling must bear all information required for safe and effective use of the device, and must include:
(i) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device.
(ii) Information on the patient population for which the device has been demonstrated to be effective.
(iii) A detailed summary of the device technical parameters.
(iv) Information on how the device operates and the typical course of treatment.
(v) An expiration date/shelf life.
(vi) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.