510(k) Data Aggregation

    K Number
    K062702

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-11-29

    (79 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insulin Syringes (0.3 to 1.0 cc) are intended for subcutaneous injection of U-100 insulin.

    The UltiCare disposable syringes and needles (0.3 to 3.0 cc) are used to draw a quantity of pharmaceutical from its container and allow administration of the pharmaceur icul directly to the patient. As an alternative use the product will allow administration of the pharmacenty to to the patient via an intravenous port, heparin lock, or saline lock.

    Device Description

    The UltiCare Disposable Syringes and Needles and Insulin Syringes are standard piston type syringes with permanently attached (unibody) needles. They are sterile, single-use, disposable piston syringes consisting of a syringe barrel, a plunger rod, and a hypodermic single lumen needle. The UltiCare Disposable Syringes and Insulin Syringes are non-toxic and non-pyrogenic, and are available in a variety of combinations of syringe sizes (0.3 to 3.0 cc (ml)), needle sizes (31 to 22 gauge), and needle lengths (5/16" to 1.5").

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria and a study proving a device meets these criteria in the way typically found for AI/ML-based medical devices or diagnostic tools. The document is a 510(k) summary for a traditional medical device (syringes and needles), which focuses on "substantial equivalence" to predicate devices rather than novel performance metrics and validation studies common for AI.

    Therefore, I cannot populate the requested table and answer the questions related to AI/ML device performance validation. The document describes the device, its intended use, and lists predicate devices, but it does not include data on acceptance criteria, a test set, expert ground truth, adjudication methods, MRMC studies, or standalone performance for an algorithm.

    Here's why the information isn't available in the provided text:

    • Device Type: The device is a traditional disposable syringe and needle, not an AI/ML-based diagnostic or imaging device. Its clearance hinges on demonstrating substantial equivalence to existing devices in terms of materials, design, and intended use, rather than demonstrating specific accuracy or performance metrics against a ground truth dataset.
    • 510(k) Summary Nature: A 510(k) summary (especially for traditional devices) typically outlines the device description, intended use, predicate devices, and a comparison demonstrating substantial equivalence. It generally does not include extensive clinical study data with detailed performance metrics, especially not in the context of AI/ML validation.
    • Lack of Performance Data: There are no sections discussing performance metrics like sensitivity, specificity, AUC, or a "reported device performance" that would align with acceptance criteria for a diagnostic algorithm. The "performance" for these devices would typically relate to sterility, material compatibility, and physical dimensions, which are not outlined in such a summary.

    Therefore, I must state that the requested information cannot be extracted from the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033794

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-12-30

    (25 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UltiCare Disposable Syringe is used to draw a quantity of pharmaccutical from its container and allow administration of the pharmaceutical directly to the patient. As an alternative use the product will allow administration of the pharmaceutical to the patient via an intravenous port, heparin lock, or saline lock.

    Device Description

    The UltiCare 3 ml Disposable Syringe is a standard piston type hypodermic syringe, with a black colored protective needle cap. This device is a sterile, single-use, disposable piston svringe with permanently affixed hypodermic single lumen needle. This device is a sterile, single use, disposable piston syringe with permanently affixed hypodermic single lumen needle. The UltiCare 3 ml Disposable Syringe consists of a syringe barrel, a plunger rod, and a hypodermic single lumen needle permanently bonded to the tip of the syringe with epoxy. The UltiCare 3 ml Disposable Syringe is available in a 3.0 cc syringe capacity with a 22 GAUGE x 1½" hypodermic single lumen needle.

    AI/ML Overview

    This document is a 510(k) summary for a disposable medical device (syringe), not an AI/ML medical device. Therefore, many of the requested categories related to AI/ML device performance and testing (e.g., sample size for test sets, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) are not applicable.

    The submission focuses on demonstrating substantial equivalence to predicate devices based on design, materials, and compliance with established performance standards for syringes.

    Here's the information that is extractable and relevant from the provided document, presented as closely as possible to the requested format:

    Acceptance Criteria and Device Performance for UltiCare™ Disposable Syringe (3 ml/cc)

    This 510(k) submission primarily relies on demonstrating substantial equivalence to two predicate devices and compliance with recognized international standards for syringes. The "acceptance criteria" are implicitly these standards, and the "reported device performance" is the claim of meeting these standards and being substantially equivalent.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria / StandardReported Device Performance
    Design & FunctionSubstantial Equivalence: - Intended Use - Function - Basic CompositionThe UltiCare 3 ml Disposable Syringe is substantially equivalent in intended use, function, and basic composition to the Ulti Med UltiCare 1 ml Disposable Syringe (K994230) and the Becton Dickinson Medsaver Syringe (K941095). The only design change from the 1ml UltiCare syringe is syringe capacity and corresponding dimensional changes, and minimal material component changes. All other aspects are identical.
    Material & SterilityISO 8537: Sterile, Single-Use Syringes, with or without Needle, for InsulinUltiCare Disposable Syringes "meet the following standards." (Implies compliance)
    Needle MaterialISO 9626: Stainless Steel Needle Tubing for Manufacture of Medical DevicesUltiCare Disposable Syringes "meet the following standards." (Implies compliance)
    Intended UseDraw and administer pharmaceuticals via direct injection or intravenous port/heparin lock/saline lock.Device is used to draw and administer pharmaceuticals directly to the patient or via an intravenous port, heparin lock, or saline lock.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This submission is for a conventional medical device (syringe), not an AI/ML device. Performance is demonstrated through substantial equivalence to predicate devices and compliance with ISO standards, rather than through a clinical test set with specific sample sizes and data provenance as would be relevant for an AI/ML algorithm. Testing would have typically involved laboratory-based verification and validation of engineering design and manufacturing against the ISO standards. The document does not detail this testing (which is common for 510(k) summaries of such devices).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. See point 2. Ground truth establishment by experts is not a concept applied to the evaluation of a disposable syringe in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 2.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a non-AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a non-AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. For this device, "ground truth" equates to the physical and functional characteristics specified in the ISO standards (e.g., sterility, dimensional accuracy, material properties, fluid delivery pressure/flow) and the functionality of predicate devices. These are not established by expert consensus, pathology, or outcomes data in the way an AI algorithm's performance is evaluated.

    8. The sample size for the training set

    • Not Applicable. This is a non-AI device; there is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is a non-AI device; there is no training set or associated ground truth establishment process.
    Ask a Question

    Ask a specific question about this device

    K Number
    K994230

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-03-24

    (99 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The product is a standard piston type hypodermic syringe. Such device will be used to draw a quantity of pharmaceutical from its container and allow administration of the pharmaceutical directly to the patient. As an alternative use, and which the product will be specifically marketed for, the product will allow administration of the pharmaceutical to the patient via an intravenous port, heparin lock, or saline lock.

    Device Description

    The Ulti Care Low Dead Space syringe is substantially equivalent to the Becton Dicksnson MedSaver The syringe. The device is intended to minimize the consumption of expensive pharmacenticals by reducing the distance between the calibrated barrel of the syringe and the needle. The devise has a permanently attached needle that eliminates the tip and hub-of a normal needle/syringe combination. This results in less waste of pharmaceutical and greater efficiency. The intended device has the same technological characteristics as the presently marketed Becton Dickinson MedSaver using a manually operated piston type hypodemic syringe, injected molded plastic parts, a synthetic latex gasket and permanently attached stainless steel needle.

    AI/ML Overview

    The provided text describes a 510(k) submission for the UltiCare Disposable Syringe. It focuses on establishing substantial equivalence to a predicate device (Becton Dickinson MedSaver™) and does not contain information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert involvement as requested in your prompt.

    The document is a regulatory submission for market clearance, outlining the device's intended use, technological characteristics, and comparison to a legally marketed predicate. It does not detail specific performance studies or acceptance criteria beyond stating substantial equivalence.

    Therefore, I cannot provide the requested information based on the given text.