Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242732

    Validate with FDA (Live)

    Device Name
    Tristel OPH
    Date Cleared
    2025-05-09

    (241 days)

    Product Code
    Regulation Number
    880.6886
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, A-scan and B-scan probes.

    Tristel OPH is a high level disinfectant when used in accordance with the Directions for Use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).

    Tristel OPH Foam must be applied on the surface of an ophthalmic device using Tristel OPH Wipes.

    Tristel OPH Wipes are intended for application of Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection.

    Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single use.

    The semi-critical ophthalmic devices processed by Tristel OPH must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions. Tristel OPH Foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.

    Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

    Device Description

    Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, and A-scan and B-scan probes.
    Tristel OPH Foam is packaged in a hand-held dispenser bottle. Tristel OPH Foam is made of two solutions, Tristel OPH Activator solution and Tristel OPH Base solution, which are mixed when dispensed to generate active ingredient chlorine dioxide. In-use concentration of chlorine dioxide (ClO2) is 0.032% (320ppm).
    Tristel OPH Foam Activator and Base solutions are contained in a dual compartment bottle with a dispenser pump and are ready to use.
    Tristel OPH Foam is supplied with Tristel OPH Wipes. Tristel OPH Wipes are dry wipes made of a non-woven and low-linting material. Tristel OPH Wipes are intended to apply Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection once the contact time of 2 minutes has elapsed.
    Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single-use.
    Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and its attachments describe the device Tristel OPH, a high-level disinfectant foam. However, the document focuses on demonstrating substantial equivalence to a predicate device (Tristel Duo ULT) through non-clinical performance testing. It meticulously compares the two devices based on their indications for use, technological characteristics, and performance data from various laboratory efficacy tests (bactericidal, fungicidal, virucidal, mycobactericidal, sporicidal, and additional clinically relevant efficacy) as well as simulated-use, in-use, compatibility, biocompatibility, residue, and inhalation studies.

    Crucially, the document does not describe a study that involves analyzing a test set of data with ground truth, nor does it refer to acceptance criteria in the context of diagnostic performance (e.g., sensitivity, specificity, AUC) for a medical image analysis or AI/ML device.

    The "acceptance criteria" referred to in the document are primarily the performance benchmarks met during chemical and microbiological efficacy testing, as well as safety and compatibility tests. The "study that proves the device meets the acceptance criteria" refers to a series of non-clinical, in-vitro, and simulated-use laboratory studies designed to demonstrate the disinfectant's potency, safety, and proper functioning.

    Therefore, many of the requested elements in your prompt, which are typical for studies evaluating AI/ML-based diagnostic devices, cannot be extracted from this document. I will focus on the information that is available.


    Here's an interpretation of the provided information, framed to address as many of your prompt's points as possible within the context of a disinfectant's clearance:

    Acceptance Criteria and Device Performance for Tristel OPH (High-Level Disinfectant)

    The "acceptance criteria" for Tristel OPH are not diagnostic performance metrics, but rather the successful demonstration of its high-level disinfection capabilities, safety, and functional equivalence to its predicate device under various laboratory and simulated-use conditions.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implied from Testing)Reported Device Performance (Tristel OPH)
    Potency - BactericidalBactericidal against Pseudomonas aeruginosa, Staphylococcus aureus, Salmonella enterica in 2.0 minutes exposure at 20°C (AOAC 955.15, 964.02, 955.14).Bactericidal against Pseudomonas aeruginosa, Staphylococcus aureus, and Salmonella enterica in 2.0 minutes exposure at 20°C.
    Potency - FungicidalFungicidal against Trichophyton interdigitale in 1.0 minute exposure at 20°C (AOAC 955.17).Fungicidal against Trichophyton interdigitale in 1.0 minute exposure at 20°C.
    Potency - VirucidalVirucidal against Poliovirus type 1, Herpes simplex virus type 1, Adenovirus type 5, and Feline calicivirus in 30 seconds exposure at 20°C (ASTM E1053).Virucidal against Poliovirus type 1, Herpes simplex virus type 1, Adenovirus type 5, and Feline calicivirus in 30 seconds exposure at 20°C.
    Potency - MycobactericidalMycobactericidal against Mycobacterium terrae in 1.0 minute exposure at 20°C (Ascenzi et al., 1987).Mycobactericidal against Mycobacterium terrae in 1.0 minute exposure at 20°C.
    Potency - SporicidalSporicidal against Clostridium sporogenes and Bacillus subtilis in 12 hours exposure at 20°C (AOAC 966.04).Sporicidal against Clostridium sporogenes and Bacillus subtilis in 12 hours exposure at 20°C.
    Potency - Additional Clinically Relevant EfficacyPassing AOAC and ASTM1053 against specified clinically relevant bacteria, fungi, and viruses, and inactivation of HPV types 16 & 18.Passed AOAC and ASTM1053 against Candida albicans, Aspergillus brasiliensis, Streptococcus agalacticae, Carbapenem-resistant Klebsiella pneumoniae, Extended Spectrum Beta-Lactamase (ESBL) Escherichia coli, Haemophilus influenza, Vancomycin-Resistant Enterococcus faecalis (VRE), Multi-Drug Resistant Streptococcus pneumoniae, Staphylococcus epidermidis, Human Coronovirus SARS-CoV-2, Human hepatitis B virus surrogate (duck hepatitis B virus), Human immunodeficiency virus type 1, Influenza A Virus. Inactivates Human papillomavirus (HPV) type 16 and type 18.
    Simulated-Use>6 log reduction mycobactericidal efficacy on worst-case inoculated devices in <2 minutes exposure at 20°C at end of shelf-life, end of use-life, and MRC.Provides >6 log reduction mycobactericidal efficacy on worst-case inoculated devices in <2 minutes exposure at 20°C at end of shelf-life, end of use-life and MRC.
    In-UseNo evidence of growth of wild type bacteria or fungi recovered from clinically used instruments following disinfection.No evidence of growth of wild type bacteria or fungi recovered from devices following disinfection on clinically used instruments.
    Device CompatibilityNo corrosion or other visible damage to tested devices/materials; inclusion in compatible disinfectant lists.Tested and approved by several device manufacturers and included in the List of Compatible disinfectants. Does not produce any corrosion or other visible damage in the devices and materials tested.
    Biocompatibility ToxicityNon-mutagenic, non-clastogenic, non-oral toxic, not a skin sensitizer, negligible irritant to skin, no cytotoxic effect (in use with wipe).Tristel OPH foam is non-mutagenic, non-clastogenic in maturing erythrocytes, and non-oral toxic. It is not a skin sensitizer. It is classified as a slight irritant to the skin and an eye irritant in direct contact. In use with the Tristel OPH Wipe, Tristel OPH foam is not considered to have a cytotoxic effect, is negligible irritating to the skin and has no sensitizing potential.
    ResiduesPost-reprocessing residuals quantified and pose no additional risk to patient/user; non-irritating to dermal and eye tissues.Residue testing showed that Tristel OPH germicide poses no additional risk to the patient or user of the device. Any potential residues remaining post reprocessing are non-irritating to dermal and eye tissues.
    Inhalation ExposureAverage 8-hour TWA readings do not exceed U.S. OSHA PEL of 0.1ppm for chlorine dioxide; average 15-minute STEL readings do not exceed California Division of Occupational Safety and Health PEL of 0.3ppm.In a study simulating 51 disinfection procedures, the average 8-hour TWA readings did not exceed the U.S. OSHA permissible exposure limit of 0.1ppm for chlorine dioxide, and the results of the average 15-minute STEL readings did not exceed the California Division of Occupational Safety and Health PEL of 0.3ppm.
    Human Factors/UsabilityEvaluated for intended users, uses, and use environments.Tristel OPH has been evaluated for the intended users, uses, and use environments.

    2. Sample sizes used for the test set and the data provenance

    • Test Set (for Efficacy/Safety/Compatibility): The document does not provide explicit sample sizes in terms of "number of devices" or "number of tests" for each specific study (e.g., number of inoculated items for simulated-use, number of materials for compatibility). However, it implies standardized testing (e.g., AOAC, ASTM, ISO standards) which would inherently include appropriate sample sizes for replicates and controls.
    • Data Provenance: The studies are non-clinical, laboratory-based tests. The country of origin for the data is not explicitly stated but is assumed to be from the manufacturer's labs or certified contract research organizations. These are prospective experiments conducted to demonstrate the device's performance against pre-defined standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For a disinfectant, "ground truth" is established through standardized microbiological assays (e.g., presence or absence of viable microorganisms after exposure to the disinfectant) or chemical analyses, not through expert consensus on diagnostic images. The performance is objectively measured based on established scientific protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in human observer studies, particularly in medical image interpretation. The studies for Tristel OPH are laboratory-based and involve objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. MRMC studies are relevant for evaluating the impact of AI on human diagnostic performance. Tristel OPH is a chemical disinfectant, not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is also not applicable in the context of a disinfectant. The device (disinfectant) inherently performs its function (killing microbes) in a "standalone" chemical process, and its efficacy is tested in laboratory settings without human intervention in the disinfection process itself, though human compliance with the instructions for use is necessary.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance of this disinfectant is based on:

    • Microbiological Culture Results: Absence or significant reduction (e.g., >6 log reduction) of specific viable microorganisms after exposure to the disinfectant.
    • Chemical Analysis: Measurement of residual chemicals, concentration of active ingredients.
    • Physical Observation: Assessment of material integrity (corrosion, damage).
    • Environmental Monitoring: Measurement of airborne chemical concentrations.
      These are all objective, laboratory-derived measurements based on established scientific methodologies and standards.

    8. The sample size for the training set

    Not applicable. This device is a chemical disinfectant, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable (as above).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1