Found 3 results

510(k) Data Aggregation

    K Number
    K213009

    Validate with FDA (Live)

    Date Cleared
    2022-07-08

    (291 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system: 1) For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values; 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye. The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during: 1) Diagnostic exams; 2) Surgical interventions. The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field. This device is intended for use by qualified and appropriately trained healthcare professionals. It can be used in hospitals or medical clinics. Mode of Operation is Pulsed Wave (PW) Doppler mode

    Device Description

    The EMS-9M1 and EMS-9M2 Transcranial Doppler Ultrasound System is a Transcranial Doppler (TCD) system using non-invasive technique to obtain the information of blood flow velocities throughout the body. This method of measurement is particularly useful for examining the major arteries supplying blood to the brain. TCD is useful for evaluation of numerous neurological vascular diseases such as vasospasm and intracranial stenosis. TCD is also extremely valuable for intraoperative monitoring to help detect sudden changes in blood flow. Both the EMS-9Ml and EMS-9M2 are Track 1 devices. EMS-9M1 and EMS-9M2 should be used in hospitals or healthcare facilities by doctors or trained healthcare professionals.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a Transcranial Doppler Ultrasound System (models EMS-9M1/EMS-9M2). It focuses on demonstrating substantial equivalence to a predicate device (K173801).

    Crucially, the document explicitly states "No clinical study is included in this submission" (Page 5, Section 7). This means that the information requested about acceptance criteria and a study proving the device meets them (clinical performance) is not provided in this document. The submission relies solely on non-clinical tests to show that the proposed device meets design specifications and is substantially equivalent to the predicate device.

    Therefore, I cannot provide details about:

    • A table of acceptance criteria and reported device performance based on clinical data.
    • Sample size used for a clinical test set or its provenance (retrospective/prospective, country).
    • Number of experts, their qualifications, or adjudication methods for clinical ground truth.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Stand-alone (algorithm-only) performance studies.
    • Type of ground truth used for clinical validation.
    • Sample size for a training set or how ground truth was established for it, as no clinical studies were performed.

    However, based on the document, I can infer the "acceptance criteria" and "proof" in the context of this 510(k) submission as follows:

    Acceptance Criteria (as implied by a 510(k) submission aiming for Substantial Equivalence):

    The primary acceptance criteria for a 510(k) cleared device, particularly when no clinical studies are performed, are that the device:

    1. Has the same intended use as the predicate device.
    2. Has the same technological characteristics as the predicate device, or if there are differences, these differences do not raise new questions of safety or effectiveness.
    3. Meets recognized consensus standards for safety and performance.

    The Study Proving the Device Meets These Acceptance Criteria (Non-Clinical):

    The "study" proving the device meets these criteria is the non-clinical testing performed by the manufacturer, as detailed in Section 6 and 8 of the document.

    1. Table of Acceptance Criteria (Implied for Non-Clinical Equivalence) and Reported Device Performance:

    Since no clinical study was performed, the "performance" is largely demonstrated by meeting and adhering to established safety and performance standards.

    Acceptance Criterion (for Substantial Equivalence based on Non-Clinical Data)Reported Device Performance (from Non-Clinical Tests)
    Same Intended Use (21 CFR 892.1550/1570)Met: Proposed device's indications for use are identical to the predicate device, K173801. (Page 4, "Indication for use" table)
    Same Technological Characteristics, or differences do not raise new safety/effectiveness questions. (Key physical & functional parameters)Met: Most core technological characteristics (e.g., product code, regulation number, class, intended use, probe type, depth measurement, speed range, maximum error) are identical to the predicate. Minor differences (Connection modes, power supply mode, nominal voltage, internal power supply) were tested and deemed not to affect safety or effectiveness. (Pages 6-8, "Summary of Technological characteristics" tables and "Difference" explanations)
    Compliance with Recognized Consensus Standards (Electrical safety, EMC, Acoustic Output, Basic Safety & Essential Performance)Met: Non-clinical tests verified compliance with: - IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+AM1:2012 (General medical electrical equipment safety) - IEC 60601-2-37:2015 (Specific to ultrasonic medical diagnostic equipment) - IEC 60601-1-2: 2014 (EMC) - NEMA UD 2-2004 (R2009) Revision 3 (Acoustic Output) (Page 5, Section 6)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not applicable for clinical data: As stated, "No clinical study is included in this submission."
    • For non-clinical testing: The "test set" would refer to the physical devices and testing environments (e.g., test setups for electrical safety, EMC, acoustic measurements). The document implies that these tests were performed on the proposed device models (EMS-9M1/EMS-9M2). Data provenance would be from internal lab testing by the manufacturer (Shenzhen Delica Medical Equipment Co., Ltd., China). This is standard for pre-market submissions based on non-clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    • Not applicable for clinical data: No clinical study was performed.
    • For non-clinical testing: The "ground truth" for non-clinical testing is established by recognized standards and their accompanying test methodologies. The "experts" would be the engineers and technicians involved in performing these tests, following the procedures outlined in the standards. Their qualifications would be expertise in medical device testing, particularly in electrical safety, EMC, and ultrasound performance measurements according to the cited IEC and NEMA standards.

    4. Adjudication Method for the Test Set:

    • Not applicable for clinical data: No clinical study was performed.
    • For non-clinical testing: Adjudication is not typically performed in the same way as in clinical studies. Test results from the non-clinical studies are compared directly against predefined acceptance limits specified by the recognized consensus standards. Any deviations would lead to failure and require re-design or re-testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable: The device is a Transcranial Doppler Ultrasound System, not an AI-powered image analysis tool for human readers. No MRMC study was conducted as no clinical study was included.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable: This is a hardware diagnostic device that measures blood flow velocities; it is not an algorithm for standalone diagnostic interpretation.

    7. The type of ground truth used:

    • Not applicable for clinical data: No clinical study was performed to establish a clinical ground truth (e.g., expert consensus, pathology, or outcomes data).
    • For non-clinical testing: The "ground truth" is based on the specifications and limits defined within recognized national and international consensus standards (IEC 60601 series, NEMA UD 2). For example, the accuracy of velocity measurement (±15%) is compared against physical phantoms or established reference measurements rather than clinical patient outcomes.

    8. The sample size for the training set:

    • Not applicable: No clinical study was performed, and the device does not involve machine learning algorithms that would require a "training set" in the traditional sense of AI/ML development.

    9. How the ground truth for the training set was established:

    • Not applicable: See #8.

    In summary, the document details a 510(k) submission based purely on substantial equivalence through non-clinical performance data and adherence to recognized standards, not on a clinical study demonstrating efficacy or specific performance metrics in a patient population.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190228

    Validate with FDA (Live)

    Date Cleared
    2019-06-13

    (128 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    While the Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system, it can be:

    1. For the measurement of cerebral artery blood velocities to determine the presence of hemo-dynamically significant deviations from normal values;
    2. To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and internal carotid artery via the eye. The robotic probe headband facilitates monitoring use by its ability to track the Doppler signal.
      While the Transcranial Doppler Ultrasound System to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The system provides a noninyasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The noninvasive blood pressure waveform is measured on the subject's finger.
      Using the blood pressure calibration module, the system can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
      The system is intended for use during:
    3. Diagnostic exams;
    4. Surgical interventions.
      The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
      The EMS-9D EXP and EMS-9D PRO is intended to be used for subjects above 18 years of age. The MS-9D EXP and EMS-9D PRO is intended for use in hospitals, clinics and research centers.
    Device Description

    EMS-9D PRO and EMS-9D EXP are Transcranial Doppler (TCD) ultrasound system, which is developed based on the EMS-9D, it inherits all the EMS-9D's hardware and mechanical structure design. EMS-9D PRO and EMS-9D EXP are not only a Transcranial Doppler ultrasound system which used non-invasive technique to obtain the information of blood flow velocities throughout the body, but also a blood pressure monitor. This method of measurement is particularly useful for examining the major arteries supplying blood to the brain, and continuously measuring blood pressure.
    TCD is useful for evaluation of numerous neurological vascular diseases such as vasospasm and intracranial stenosis. TCD is also extremely valuable for intraoperative monitoring to help detect sudden changes in blood flow.
    The EMS-9D PRO and EMS-9D EXP integrate main unit, TCD probes, Nano Core module, upper arm blood pressure module and analog output module (EMS-9D EXP only), dedicated software, remote control, headframe, upper arm cuff, finger cuff and optional probes.
    EMS-9D EXP/EMS-9D PRO should be used in hospitals or healthcare facilities by doctors or trained healthcare professionals.

    AI/ML Overview

    The provided text describes a Transcranial Doppler Ultrasound System (EMS-9D EXP/PRO) and references several non-clinical tests to assert its substantial equivalence to predicate devices, particularly regarding its Transcranial Doppler Ultrasound part, Nano Core part (Finger Arterial Blood Pressure), and Upper Arm Cuff Measurement part.

    However, the document is a 510(k) Summary for regulatory submission, not a detailed clinical study report. It focuses on demonstrating equivalence to previously cleared devices rather than presenting novel clinical efficacy data with defined acceptance criteria and study results in the typical sense of a diagnostic performance study.

    Therefore, many of the requested details about acceptance criteria, specific study design (e.g., sample size for test sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance), and ground truth establishment for new clinical performance evaluation are not explicitly present or applicable in this regulatory summary. The document primarily relies on engineering performance specifications and comparison to predicate device specifications.

    Here's a breakdown based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are implicitly substantial equivalence to the predicate devices and adherence to relevant safety and performance standards. The "reported device performance" refers to the specifications of the proposed device, which are shown to be equivalent to the predicate.

    CharacteristicAcceptance Criteria (Predicate Device K173801)Reported Device Performance (Proposed Device)Comparison
    Transcranial Doppler Ultrasound Part
    Product CodeIYN, ITX, OQQIYN, ITX, OQQNo difference
    Regulation No.21 CFR 892.1550, 21 CFR 892.157021 CFR 892.1550, 21 CFR 892.1570No difference
    ClassIIIINo difference
    Intended UseAs described for K173801 (long text)As described for proposed device (long text)No difference
    ConfigurationEMS-9D: Main Unit + TransducerEMS-9D EXP/PRO: Main Unit + Transducer + NIBP ModuleYes, added NIBP module (see Analysis 1)
    Input Power80VA120VAYes, higher (see Analysis 2)
    1.6MHz PW Probe
    Depthnot less than 150 mmnot less than 150 mmNo difference
    Speed range10cm/s~300cm/s10cm/s~300cm/sNo difference
    Maximum Error±15%±15%No difference
    2MHz PW Probe
    Depthnot less than 150 mmnot less than 150 mmNo difference
    Speed range10cm/s~300cm/s10cm/s~300cm/sNo difference
    Maximum Error±15%±15%No difference
    4MHz CW Probe
    Depthnot less than 85 mmnot less than 85 mmNo difference
    Speed range10cm/s~300cm/s10cm/s~300cm/sNo difference
    Maximum Error±15%±15%No difference
    8MHz CW Probe
    Depthnot less than 35 mmnot less than 35 mmNo difference
    Speed range10cm/s~200cm/s10cm/s~200cm/sNo difference
    Maximum Error±15%±15%No difference
    16MHz PW Probe
    Depthnot less than 8 mmnot less than 8 mmNo difference
    Speed range10cm/s~120cm/s10cm/s~120cm/sNo difference
    Maximum Error±15%±15%No difference
    Nano Core Module (Finger Arterial Blood Pressure)
    Product CodeDXQ (K173916 Finapres NOVA)DXQNo difference
    Regulation No.21 CFR 870.113021 CFR 870.1130No difference
    ClassificationClass IIClass IINo difference
    Upper Arm Cuff Measurement Part
    Regulation No.21CRF 870.1120 (K182433)21CRF 870.1120No difference
    ClassificationClass IIClass IINo difference
    Product CodeDXQDXQNo difference
    Intended UseReusable blood pressure cuff for NIBP systemsReusable blood pressure cuff for NIBP systemsNo difference
    Patient PopulationsAdultAdultNo difference
    Circumference25-35cm25-35cmNo difference
    Pressure Range0-300 mmHg0-300 mmHgNo difference
    SterilityNon-sterileNon-sterileNo difference
    Max. Leakage< 4mm Hg/min.< 4mm Hg/min.No difference
    MaterialCuff (Patient contacted): PUCuff (Patient contacted): PUNo difference
    BiocompatibilityComply with ISO 10993-5; ISO 10993-10Comply with ISO 10993-5; ISO 10993-10No difference

    2. Sample size used for the test set and the data provenance

    • Transcranial Doppler Ultrasound System: This document does not specify a sample size for a clinical test set or details on data provenance (e.g., country of origin, retrospective/prospective). The performance data cited are technical specifications compared to a predicate device.
    • Nano Core Module & Upper Arm Cuff: Similarly, no specific sample sizes for clinical test sets or data provenance are mentioned for these components, as they are being compared to cleared predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not provided. This document demonstrates technical equivalence and adherence to standards rather than a clinical performance study requiring expert ground truth establishment for a diagnostic output.

    4. Adjudication method for the test set

    • Not applicable/Not provided, as this is not a study focused on diagnostic accuracy requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. The described device is a Transcranial Doppler Ultrasound System, not an AI-assisted diagnostic tool for image interpretation that would typically be evaluated with MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable, as this device is not an algorithm for standalone diagnostic interpretation. It is a diagnostic ultrasound system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the technical specifications, the "ground truth" is derived from engineering measurements, adherence to recognized standards (e.g., IEC 60601 series, NEMA UD 2-2004), and comparison against the specifications of the predicate devices. Clinical ground truth like pathology or outcomes data is not mentioned in this summary.

    8. The sample size for the training set

    • Not applicable. This document does not describe a machine learning or AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no mention of a training set for an AI algorithm.

    Summary of Study Type:

    The document describes a 510(k) premarket notification, which is a process to demonstrate that the new device is substantially equivalent to a legally marketed predicate device. The "study" referenced in the document is primarily non-clinical testing to ensure safety, electrical performance, acoustic output, and electromagnetic compatibility according to recognized international and national standards (IEC 60601-1, IEC 60601-2-37, IEC 80601-2-30, IEC 60601-1-2, NEMA UD 2-2004). The clinical performance is inferred through the substantial equivalence claim based on identical intended use and equivalent technical specifications to predicate devices. The modifications (addition of NIBP module, higher input power) are addressed by stating they do not impact safety and effectiveness or by referencing compliance with additional relevant standards for those components.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173801

    Validate with FDA (Live)

    Date Cleared
    2018-07-11

    (209 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system: 1) For the measurement of cerebral artery blood velocities to determine the presence of hemo dynamically significant deviations from normal values; 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye. The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. Transcranial Doppler is intended for use during: 1) Diagnostic exams; 2) Surgical interventions. The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.

    Device Description

    The EMS-9D/EMS-9F Transcranial Doppler Ultrasound System is a Transcranial Doppler (TCD) system using non-invasive technique to obtain the information of blood flow velocities throughout the body. This method of measurement is particularly useful for examining the major arteries supplying blood to the brain. TCD is useful for evaluation of numerous neurological vascular diseases such as vasospasm and intracranial stenosis. TCD is also extremely valuable for intraoperative monitoring to help detect sudden changes in blood flow. EMS-9D/EMS-9F should be used in hospitals or healthcare facilities by doctors or trained healthcare professionals.

    AI/ML Overview

    This document is a 510(k) Summary for the Shenzhen Delica Medical Equipment Co., Ltd.'s Transcranial Doppler Ultrasound System (models EMS-9D and EMS-9F). The document aims to demonstrate substantial equivalence to a predicate device (K122710).

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the performance of the predicate device (K122710). The reported device performance for the proposed device (EMS-9D and EMS-9F) is compared against these values. It's noted that the "performance of proposed device is fully better than that of predicate device - the proposed device has longer measurement depth, larger speed range and less error."

    ParameterPredicate Device (K122710) Acceptance CriteriaProposed Device (EMS-9D/EMS-9F) Reported Performance
    1.6MHz PW
    Depthnot less 134 mmnot less 150 mm
    Speed range20cm/s~200cm/s10cm/s~300cm/s
    Maximum Error±20%±15%
    2MHz PW
    Depthnot less 134 mmnot less 150 mm
    Speed range20cm/s~200cm/s10cm/s~300cm/s
    Maximum Error±20%±15%
    4MHz CW
    Depthnot less 50 mm (EMS-9UA) / not less 85 mm (EMS-9PB)not less 85 mm
    Speed range10cm/s100cm/s (EMS-9UA) / 10cm/s300cm/s (EMS-9PB)10cm/s~300cm/s
    Maximum Error±20%±15%
    8MHz CW
    Depthnot less 20 mm (EMS-9UA) / not less 35 mm (EMS-9PB)not less 35 mm
    Speed range10cm/s50cm/s (EMS-9UA) / 10cm/s200cm/s (EMS-9PB)10cm/s~200cm/s
    Maximum Error±20%±15%
    16MHz PWN/A (not present in predicate)Depth: not less 8 mm, Speed range: 10cm/s~120cm/s, Max Error: ±15%
    All ProbesCompliance with Track 3 Acoustic Output LimitAcoustic output test demonstrated compliance

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the "sample size used for the test set" in terms of patient data or clinical cases. The non-clinical tests mentioned are related to engineering standards (IEC, NEMA) and device performance specifications. It also does not explicitly state the "country of origin of the data" or if the data was "retrospective or prospective". The testing appears to be primarily laboratory-based non-clinical verification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The testing described is non-clinical performance verification against engineering standards rather than diagnostic accuracy studies requiring expert-established ground truth.

    4. Adjudication Method

    This information is not provided as it is not relevant to the non-clinical performance verification described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission focuses on non-clinical performance testing and substantial equivalence based on technical specifications and adherence to standards.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device is a Transcranial Doppler Ultrasound System, which is a diagnostic tool that provides data for human interpretation. The concept of "standalone performance" for an algorithm without human-in-the-loop is not directly applicable in the AI/algorithm sense to this type of medical device as described in the document. The document describes the technical performance specifications of the ultrasound system itself.

    7. The Type of Ground Truth Used

    For the non-clinical performance tests mentioned (depth, speed range, maximum error, acoustic output), the ground truth would be based on physical phantom measurements and engineering standards/calibrations. For example, a calibrated flow phantom would be used to measure speed range and error, and specialized equipment would measure acoustic output.

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable. The Transcranial Doppler Ultrasound System is a hardware diagnostic device, not an AI/machine learning algorithm that requires a training set in the conventional sense.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided and is not applicable, as there is no "training set" for an AI/machine learning algorithm mentioned for this device.