Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K153771

    Validate with FDA (Live)

    Date Cleared
    2016-09-13

    (258 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Guiding Sheath Introducer is intended for use in the hospital catheterization laboratory for the percutaneous introduction of various devices into veins and/or arteries in a variety of diagnostic and therapeutic procedures.

    Device Description

    The Guiding Sheath is intended for intravascular use for the introduction of interventional/diagnostic devices to the human vasculature. The Guiding Sheath is designed to work as a vessel introducer and a guiding catheter. It consists of a lubricous inner liner made from Teflon, and a stainless steel coil over the inner liner. The outer shaft consists of colored Polyether block amide. A Radiopaque Marker band is attached to the distal end of the shaft for radiopacity. The Distal end of the shaft is atraumatic. The Proximal end of the catheter is attached to a Hemostatic Valve. The device is available in three inner diameter sizes of 4Fr, 5Fr, and 6Fr; consisting of 0.063", 0.076" and 0.087" (throughout the shaft) respectably. A lubricious hydrophilic coating shall be applied to the outer diameter of the Catheter Sheath for improved trackability through the vasculature. The device contains a separate Dilator shaft made of two different materials depending on the size. For the 4Fr, Grilamid is used for the body and the Luer attached to the proximal end as well. For the 5Fr and 6F, high densiy polyethylene (HDPE) is used for the body and the Luer attached to the proximal end as well. The distal end of the Dilator is tapered for ease of access to the vessel. The device is available in lengths of 45cm and 90cm. The device is supplied sterile and is intended for single use.

    AI/ML Overview

    The document provided describes a medical device, the TXM Guiding Sheath, and its testing for substantial equivalence to a predicate device. It does not contain information about an AI/ML powered device, and therefore does not include details such as a multi-reader multi-case (MRMC) comparative effectiveness study, standalone algorithm performance, or training set specifics.

    Here is an analysis of the acceptance criteria and study that was conducted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a non-AI/ML medical device (catheter introducer), and its acceptance criteria are based on meeting performance standards for physical attributes, safety, and biocompatibility, rather than diagnostic accuracy metrics like sensitivity or specificity.

    Test NameApplicable Standard or Internal Test MethodAcceptance CriteriaReported Device Performance
    Dimensional & Physical AttributesISO 10555Meet specified dimensions and physical characteristicsT=0;T=2 Pass (Met)
    Lubricity of Hydrophilic CoatingInternal Test MethodMaintain sufficient lubricity for improved trackabilityT=0;T=2 Pass (Met)
    Dilator Guidewire and Sheath Catheter CompatibilityInternal Test MethodCompatible with specified guidewire and dilatorT=0;T=2 Pass (Met)
    Liquid LeakISO 10555No liquid leakage from the deviceT=0;T=2 Pass (Met)
    Air LeakISO 10555No air leakage from the deviceT=0;T=2 Pass (Met)
    Seal Strength (Pouch)ASTM F88/F88M-09Maintain sterile barrier integrityT=0;T=2 Pass (Met)
    Dye Penetration TestASTM-F1929-12No dye penetration into the sterile barrierT=0;T=2 Pass (Met)
    RadiopacityASTM-F640-12Sufficient visibility under fluoroscopyT=0;T=2 Pass (Met)
    In-Vitro Track ForceInternal Test MethodAcceptable track force through simulated vasculatureT=0;T=2 Pass (Met)
    Kink ResistanceInternal Test MethodResistance to kinking during useT=0;T=2 Pass (Met)
    Durability of Hydrophilic CoatingInternal Test MethodMaintain coating integrity and performance over timeT=0;T=2 Pass (Met)
    Tensile StrengthISO 10555Withstand specified tensile forces without breakageT=0;T=2 Pass (Met)
    Coating IntegrityInternal Test MethodMaintain integrity of the coatingT=0;T=2 Pass (Met)
    Corrosion ResistanceISO 10555Resistance to corrosionT=0;T=2 Pass (Met)
    Female Luer VerificationISO 594Conformance to luer connector standardsT=0;T=2 Pass (Met)
    Accelerated AgingASTM F1980-07Demonstrate performance equivalence after simulated agingT=0;T=2 Pass (Met)
    Biocompatibility TestsAll Met Predetermined Acceptance Criteria
    Irritation (Intracutaneous Injection)ISO10993-10No significant irritationPass (Met)
    Sensitization (Kligman Maximization / Murine Local Lymph Assay)ISO10993-10No significant sensitizationPass (Met)
    Systemic Toxicity (ISO Acute Systemic Toxicity)ISO10993-11No significant systemic toxicityPass (Met)
    Cytotoxicity (L929 Neutral Red Uptake)ISO 10993-5No significant cytotoxicityPass (Met)
    Pyrogenicity (Pyrogen Test in Rabbit)USP <151> / ISO10993-11No pyrogenic responsePass (Met)
    Hemocompatibility (Hemolysis)ISO10993-4Acceptable levels of hemolysisPass (Met)
    Hemocompatibility (Complement Activation)ISO10993-4No significant complement activationPass (Met)
    Hemocompatibility (In-Vivo Thrombogenicity)ISO10993-4No significant thrombogenicityPass (Met)
    Sterilization Test
    Ethylene Oxide Gas SterilizationN/A (Internal validation)Attain a Sterility Assurance Level (SAL) of 10^-6Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench testing and biocompatibility testing.

    • Bench Testing: The sample sizes for each specific bench test are not explicitly stated in the provided text. However, it indicates that testing was performed on "non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2)." This implies a test set composed of new devices and aged devices. The provenance is internal laboratory testing.
    • Biocompatibility Testing: The sample sizes (e.g., number of animals for in-vivo tests or replicates for in-vitro tests) are not explicitly stated. The provenance is internal laboratory testing following ISO standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is a medical device, not an AI/ML diagnostic or prognostic tool. The "ground truth" for these performance tests is defined by meeting physical specifications, regulatory standards (ISO, ASTM), and biological safety parameters, which are assessed by technical personnel and validated according to established test methodologies.

    4. Adjudication Method for the Test Set

    Not applicable. This is a medical device performance and safety evaluation against predefined standards, not a diagnostic assessment requiring expert adjudication of results. Each test outcome (e.g., pass/fail, quantitative measure) is evaluated against its specific acceptance criterion.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This type of study (comparing human readers with and without AI assistance) is relevant for AI/ML-powered diagnostic devices, not for a physical medical device like a guiding sheath.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This concept applies to AI/ML algorithms, not physical medical devices.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's evaluation is defined by:

    • Pre-established engineering specifications (e.g., dimensions, strength).
    • Compliance with recognized international standards (e.g., ISO 10555, ASTM F88, ISO 10993 for biocompatibility) and internal test methods that align with these standards.
    • Biological responses in validated in-vitro and in-vivo models (for biocompatibility).

    8. The Sample Size for the Training Set

    Not applicable. There is no AI/ML algorithm involved that would require a training set. The device itself is the product being tested.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1