Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132542

    Validate with FDA (Live)

    Date Cleared
    2013-11-15

    (94 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TransGate™ Guide Catheters are intended to facilitate the placement of interventional devices into the coronary, peripheral, and neuro vasculature.

    Device Description

    Stryker Neurovascular's TransGate Guide Catheters are constructed with polytetrafluoroethylene (PTFE) on the inner lumen for lubricity, stainless steel wire reinforcement within the wall for trackability. torque transmission and strength, and polymer materials along the length of the catheter for support and flexibility. The catheter has an atraumatic tip, a hub for device connectivity and device handling, and a strain relief at the hub for kink resistance. The distal segment of the device shaft is radiopacified and. in addition, a radiopaque marker is embedded at the distal tip of the device to aid visualization under fluoroscopy. .

    The Stryker Neurovascular TransGate Guide Catheter has straight (ST) and pre-shaped distal tips, available in 40 (40), and Multipurpose (MP) shapes. The TransGate Guide Catheters are offered in both 90 and 100 cm lengths and 5 and 6 French diameters.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Stryker Neurovascular TransGate™ Guide Catheter.

    Important Note: The provided document is a 510(k) Summary for a medical device – a guide catheter specifically. These types of devices do not typically involve AI components or software algorithms that would require the typical "AI/ML acceptance criteria" found in software as a medical device (SaMD) clearances. Therefore, many of your requested points regarding AI/ML studies (like standalone algorithm performance, MRMC studies, training set details, ground truth for training) are not applicable to this particular device and document.

    The "acceptance criteria" here refer to meeting performance standards for a physical medical device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Performance TestAcceptance Criteria (Implicit)Reported Device Performance
    Kink DistanceEstablished criteriaMet established criteria
    Torque Degrees Rotation to KinkEstablished criteriaMet established criteria
    Tip Shape RetentionEstablished criteriaMet established criteria
    Distal Shaft FlexibilityEstablished criteriaMet established criteria
    PushabilityEstablished criteriaMet established criteria
    Tensile StrengthEstablished criteriaMet established criteria
    Chemical CompatibilityEstablished criteriaMet established criteria
    Corrosion ResistanceEstablished criteriaMet established criteria
    ParticulateEstablished criteriaMet established criteria
    Liquid LeakageEstablished criteriaMet established criteria
    Air LeakageEstablished criteriaMet established criteria
    Unscrewing TorqueEstablished criteriaMet established criteria
    Resistance to OverridingEstablished criteriaMet established criteria
    Stress CrackingEstablished criteriaMet established criteria
    Ease of AssemblyEstablished criteriaMet established criteria
    Luer GaugingEstablished criteriaMet established criteria
    Luer Separation ForceEstablished criteriaMet established criteria
    Freedom from leakage and damage under high static pressureEstablished criteriaMet established criteria
    RadiopacityEstablished criteriaMet established criteria
    Shelf Life Testing (Product and Packaging)Established criteriaMet established criteria
    Distribution / Shipping Challenge Conditioning and TestingEstablished criteriaMet established criteria
    Packaging Verification (sterile barrier, protection)EN ISO 11607-1 and -2 requirementsComplies and met established criteria
    Biocompatibility (Hemolysis, Sensitization, etc.)EN ISO 10993-1, relevant sub-partsSuccessfully passed all tests
    Sterilization (EO and ECH residuals)EN ISO 10993-7:2008 limitsMeets limits
    Sterility Assurance Level (SAL)10^-6Demonstrated 10^-6

    Note on "Established Criteria": For physical medical devices, "established criteria" typically refer to internal design specifications, industry standards (like ISO standards), and recognized test methods that demonstrate the device is safe and effective for its intended use. The document explicitly mentions adherence to international standards like EN ISO 10555-1, EN ISO 10555-2, EN 1707, EN ISO 11607-1 and -2, EN ISO 10993-1, EN ISO 10993-7, EN ISO 11135-1, and EN 556-1. These standards define many of the specific acceptance criteria for these types of tests.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific performance test (e.g., how many catheters were tested for kink distance, tensile strength, etc.). It generally states that "Testing was conducted" and "Test samples consisted of 2X EtO sterilized devices (entire device)" for the SAL measurement.

    Data Provenance: This is a physical device testing, not a clinical data study. The "provenance" would be from laboratory testing conducted by Stryker Neurovascular. There is no mention of country of origin for data beyond the company's location (Fremont, CA, USA) and adherence to international (EN ISO) standards. The testing is prospective for the device's design verification and validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    N/A. This is not an AI/ML device that generates diagnostic outputs requiring expert ground truth for a test set. The "ground truth" for these tests is defined by the physical or chemical properties measured against predefined scientific and engineering standards and specifications.

    4. Adjudication Method for the Test Set

    N/A. Adjudication methods like 2+1 or 3+1 are used for clinical image interpretation or diagnostic performance evaluation, which is not relevant to the design verification and validation of a guide catheter.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    N/A. MRMC studies are specific to evaluating diagnostic technologies, often AI-powered, where human readers interpret cases. This is not applicable to a physical medical device like a guide catheter. There is no AI component in this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    N/A. This device does not involve an algorithm or AI.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's safety and performance studies is based on:

    • Engineering and Design Specifications: The device must meet its own internal specifications for physical properties.
    • International Standards: Adherence to recognized international standards (e.g., EN ISO 10555 for catheters, EN ISO 10993 for biocompatibility, EN ISO 11135 for sterilization). These standards define acceptable ranges and methodologies.
    • Scientific Principles: Basic principles of material science, mechanics, and biology for biocompatibility.

    8. The Sample Size for the Training Set

    N/A. This is a physical medical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    N/A. As there is no training set, there is no ground truth to establish for it.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1