Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K052637

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-03-08

    (163 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The infusion catheter is intended to assist in the controlled infusion of diagnostic or therapeutic solutions into the peripheral or coronary vasculature. The target vessel effective diameter is ≥ 2.0 mm. The catheter is placed by a physician using a guide catheter and over a guidewire that extends beyond the final position of the infusion catheter.

    Device Description

    The TherOx® Infusion Catheter is a sterile, single-use, 4.6 French (F) intravascular catheter with a 127 cm usable length. The catheter is comprised of three sections; 1) a distal atraumatic soft tip 2) a semi-rigid but flexible shaft; and 3) a proximal luer hub for connection of solution-delivery devices. A radiopaque band at the distal tip allows fluoroscopic visualization of the catheter's location in the vessel.

    AI/ML Overview

    TherOx, Inc. sought 510(k) clearance for their TherOx® Infusion Catheter, indicating substantial equivalence to the Boston Scientific/Target Therapeutics Tracker® - 38 Infusion Catheter. The study performed was a non-clinical test data summary comparing the TherOx® Infusion Catheter to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicAcceptance Criteria (based on predicate device and standards)TherOx® Infusion Catheter Performance
    Physical/Material
    Outer DiameterComparable to predicate to allow intended use4.6 F (0.060 in) overall (smaller than predicate's 5.3F/5F)
    Inner DiameterSufficient for recommended guidewire0.046 in overall, 0.037 in min at marker band (slightly smaller than predicate at marker band)
    Usable LengthComparable to predicate for physician manipulation127 cm (longer than predicate's 115cm/117cm)
    MaterialsBiocompatible and suitable for intravascular useHigh density polyethylene (HDPE) shaft, rigid HDPE luer, LDPE plasticized tip (minor differences from predicate, but compliant)
    RadiopacityVisible under fluoroscopyRadiopaque band at distal tip allows visualization
    SterileSterileSterilized with ethylene oxide
    Single-UseSingle-use onlySingle-use only
    Shelf LifeEstablished shelf life3-year shelf life established
    Functional/Performance
    StrengthMeets standards (ISO 10555 series) for intravascular cathetersTest results met acceptance criteria
    FlexibilityMeets standards (ISO 10555 series) for intravascular cathetersTest results met acceptance criteria
    PressureMeets standards (ISO 10555 series) for intravascular cathetersTest results met acceptance criteria
    FlowMeets standards (ISO 10555 series) for intravascular cathetersTest results met acceptance criteria
    BiocompatibilityMeets ISO 10993-1Satisfies ISO 10993-1 requirements
    SterilizationValidated per ISO 11135Validated per ISO 11135
    PackagingDesigned and tested per ISO 11607Designed and tested per ISO 11607
    Joint StrengthWithin specified requirementsTest results met acceptance criteria
    Dimensional ConformanceWithin specified requirementsTest results met acceptance criteria

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not explicitly state the exact sample size (number of catheters) used for each individual test. It mentions "production samples that were sterilized twice in a validated sterilization cycle" and "samples that were sterilized twice and then thermally aged to simulate an expected shelf life." This implies a sufficient number of devices were tested to validate the performance and shelf life.
    • Data Provenance: The data is from non-clinical bench testing conducted by TherOx, Inc. explicitly for this 510(k) submission. It is prospective in the sense that the tests were performed to demonstrate compliance of the TherOx® Infusion Catheter. The country of origin of the data is implied to be where TherOx, Inc. is located (Irvine, CA, USA) or where their testing facilities are.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This was a non-clinical bench study, not a clinical study involving human expert interpretation. Therefore, there were no experts used to establish a "ground truth" in the sense of clinical diagnosis or interpretation. The "ground truth" or acceptance criteria were derived from international standards (ISO 10555 series, ISO 10993-1, ISO 11135, ISO 11607) and internal device specifications, as well as comparison to the predicate device. The engineers and scientists conducting the tests would have been qualified in their respective fields of materials science, biomechanics, and medical device testing.

    4. Adjudication method for the test set:

    • Not applicable. As a non-clinical bench study, there was no adjudication method involving multiple human readers or experts to determine a consensus "ground truth." Test results were compared directly to the predetermined acceptance criteria outlined in the relevant standards and internal specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not performed. This submission is for a physical medical device (an infusion catheter) and evaluates its mechanical and material properties, not an AI or imaging diagnostic tool that would involve human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm or software. No standalone algorithm performance was evaluated.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the non-clinical tests was established by predetermined engineering specifications and compliance with recognized international and national standards for intravascular catheters and medical device testing. These standards dictate acceptable ranges for physical dimensions, material properties, strength, flexibility, pressure resistance, flow rates, biocompatibility, sterilization efficacy, and packaging integrity. The predicate device's characteristics also served as a reference point.

    8. The sample size for the training set:

    • Not applicable. This is a non-clinical bench study for a physical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable, as there was no training set for a machine learning model.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1