510(k) Data Aggregation

    K Number
    K972436

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-09-22

    (84 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Siemens T1 LINE dental handpieces are intended for use in the dental operatory to prepare dental cavities for restorations such as fillings, and for cleaning teeth.

    Device Description

    The Siemens T1 LINE Dental Handpieces are intended for use in the dental operatory to prepare dental cavities for restorations, such as fillings, and for dental cleanings. The devices can be either AC-powered or air-powered. They are reusable, ergonomically shaped, and are provided both with and without a fiber optic light system. Water delivery has optional one, two, and three spray outlets, and includes a one-way retraction valve to prevent the ingress of external air or liquid when the spray system is not in use. The device can be sterilized by the steam autoclave method.

    AI/ML Overview

    The provided 510(k) summary for the Siemens T1 LINE Dental Handpieces describes performance testing rather than acceptance criteria and study results in the typical sense of a clinical trial for a diagnostic device. Dental handpieces are Class I devices, which generally have lower regulatory hurdles and typically rely on adherence to established consensus standards and mechanical performance testing rather than clinical effectiveness studies with human subjects.

    Here's a breakdown of the information based on the provided text, addressing your questions where applicable and noting when information is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a Class I mechanical device, the "acceptance criteria" are compliance with relevant ISO standards and demonstrated durability/sterilization effectiveness.

    Acceptance CriteriaReported Device Performance
    Compliance with ISO Standard 1797-1: Dental Rotary Instruments - Shanks made of metal"The Siemens T1 LINE Dental Handpieces comply with ISO Standard 1797-1"
    Compliance with ISO Standard 3964: Dental Handpieces - Coupling dimensions"The Siemens T1 LINE Dental Handpieces comply with ... ISO Standard 3964"
    Compliance with ISO 7785-2: Straight and geared angle handpieces"The Siemens T1 LINE Dental Handpieces comply with ... ISO 7785-2"
    Sterilization Process Efficacy: Device effectiveness and integrity after repeated sterilization cycles"Additional performance testing was conducted on the device to validate the sterilization process and device effectiveness following 2000 sterilization cycles." (No specific quantitative results provided in this summary, just that it was validated).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "sample size" in the context of clinical testing with patients or data. The testing mentioned refers to mechanical and sterilization validation.

    • Test set sample size: Not applicable/not specified for clinical data. For the mechanical and sterilization testing, the sample size of handpieces tested is not listed.
    • Data provenance: Not applicable for clinical data. The performance testing was conducted by Siemens AG, a German company (Bensheim, Germany).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. The device is a mechanical tool, and its performance is assessed against engineering standards and functional tests, not by expert interpretation of clinical data.

    4. Adjudication Method

    Not applicable. There is no expert adjudication process described for the mechanical performance testing of the handpiece.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. A MRMC study is relevant for diagnostic devices where human readers interpret medical images or data. This is a mechanical dental tool, and as such, an MRMC study was not performed.

    6. Standalone (Algorithm Only) Performance

    No. The concept of "standalone performance" typically applies to AI algorithms or diagnostic tools operating without human intervention. This is a physical dental handpiece, so this concept is not applicable.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Adherence to international engineering standards (ISO 1797-1, 3964, 7785-2).
    • Demonstrated functionality and safety (sterilization effectiveness) through laboratory performance testing.

    8. Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this device does not involve machine learning or AI algorithms that require training data.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.