510(k) Data Aggregation

    K Number
    K191716

    Validate with FDA (Live)

    Device Name
    Somryst
    Date Cleared
    2020-03-23

    (271 days)

    Product Code
    Regulation Number
    882.5801
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Somryst is a prescription-only digital therapeutic intended to provide a neurobehavioral intervention (Cognitive Behavioral Therapy for Insomnia - CBT-I) in patients 22 years of age and older with chromic insomnia. Somryst treats chronic insomnia by improving a patient's insomnia symptoms.

    Device Description

    Somryst is a 9-week prescription digital therapeutic (computerized behavioral therapy) used in the treatment of chronic insomnia. Somryst is based on principles of Cognitive Behavioral Therapy (CBT) for Insomnia, Sleep Restriction, and other proven psychosocial treatment elements, which are delivered in a sequence of "cores" of patient education, training and skill building. The therapy is delivered via a mobile application intended to be used on a patient's mobile device and consists of text, video, animation and graphics. Clinicians, as part of a patient's general treatment program, have access to a clinician dashboard that shows patient utilization and engagement with the application.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the studies that prove Somryst meets these criteria, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance document for Somryst does not explicitly list "acceptance criteria" in a typical quantitative pass/fail format with predefined thresholds for performance metrics. Instead, it demonstrates the device's efficacy by comparing its performance to a control group in several clinical studies. The implicit acceptance criterion is that Somryst should significantly improve insomnia symptoms compared to usual care or a control.

    Below is a summary of the reported device performance from the two principal clinical studies, showing the statistically significant improvements.

    Performance MetricAcceptance Criteria (Implied)Somryst Performance (UC+SHUTi) - ANU Study (N=574)Somryst Performance (UC+SHUTi) - UVA Study (N=151)
    Insomnia Severity Index (ISI) Score ReductionSignificant reduction in ISI compared to control group.Week 9: LS Mean change of -8.63 vs. -2.85 for control (LS Mean Difference: -5.78; p<0.0001) Month 6: LS Mean change of -8.17 vs. -3.86 for control (LS Mean Difference: -4.31; p<0.0001) Month 12: LS Mean change of -8.21 vs. -4.63 for control (LS Mean Difference: -3.59; p<0.0001)Week 9: LS Mean change of -7.83 vs. -2.94 for control (LS Mean Difference: -4.89; p<0.0001) Month 6: LS Mean change of -8.52 vs. -5.36 for control (LS Mean Difference: -3.16; p<0.0001) Month 12: LS Mean change of -9.57 vs. -6.04 for control (LS Mean Difference: -3.52; p<0.0001)
    Proportion of ISI Responders (>=7-point reduction)Significantly higher proportion of responders than control.Week 9: 62.8% vs. 14.0% for control (p<0.0001) Month 6: 56.2% vs. 18.9% for control (p<0.0001) Month 12: 59.3% vs. 25.2% for control (p<0.0001)Week 9: 52.6% vs. 16.9% for control (p<0.0001) Month 6: 59.6% vs. 35.7% for control (p=0.0002) Month 12: 69.7% vs. 43.0% for control (p<0.0001)
    Proportion of ISI Remitters (ISI score < 8)Significantly higher proportion of remitters than control.Week 9: 61.6% vs. 14.9% for control (p<0.0001) Month 6: 63.7% vs. 20.4% for control (p<0.0001) Month 12: 63.0% vs. 25.7% for control (p<0.0001)Week 9: 40.6% vs. 11.3% for control (p<0.0001) Month 6: 49.1% vs. 24.0% for control (p<0.0001) Month 12: 56.6% vs. 27.3% for control (p<0.0001)
    Sleep Onset Latency (SOL) ReductionSignificant reduction in SOL compared to control group.Week 9: LS Mean change of -22.7 minutes vs. -0.46 minutes for control (LS Mean Difference: -22.4; p<0.0001) Month 6: LS Mean change of -22.6 minutes vs. -7.88 minutes for control (LS Mean Difference: -15.1; p<0.0001) Month 12: LS Mean change of -27.6 minutes vs. -10.8 minutes for control (LS Mean Difference: -17.2; p<0.0001)Week 9: LS Mean change of -21.5 minutes vs. -8.84 minutes for control (LS Mean Difference: -12.6; p<0.0001) Month 6: LS Mean change of -21.1 minutes vs. -13.9 minutes for control (LS Mean Difference: -7.25; p=0.0323) Month 12: LS Mean change of -23.7 minutes vs. -16.3 minutes for control (LS Mean Difference: -7.32; p=0.0255)
    Wake After Sleep Onset (WASO) ReductionSignificant reduction in WASO compared to control group.Week 9: LS Mean change of -28.8 minutes vs. -11.0 minutes for control (LS Mean Difference: -18.8; p<0.0001) Month 6: LS Mean change of -27.6 minutes vs. -12.5 minutes for control (LS Mean Difference: -15.6; p<0.0001) Month 12: LS Mean change of -25.1 minutes vs. -9.36 minutes for control (LS Mean Difference: -16.5; p<0.0001)Week 9: LS Mean change of -24.9 minutes vs. -8.46 minutes for control (LS Mean Difference: -17.2; p<0.0001) Month 6: LS Mean change of -23.9 minutes vs. -12.9 minutes for control (LS Mean Difference: -12.1; p=0.0006) Month 12: LS Mean change of -28.4 minutes vs. -16.8 minutes for control (LS Mean Difference: -12.7; p<0.0001)

    2. Sample Size for Test Set and Data Provenance

    The "test set" in this context refers to the participants in the two clinical trials where the effectiveness of Somryst (referred to as SHUTi in the studies) was evaluated.

    • "The GoodNight Study" (ANU Study):

      • Sample Size (Test Set):
        • UC+SHUTi group: N=574
        • UC+Control group: N=575
        • Total participants mentioned in some tables (e.g., ISI, Week 9): UC+SHUTi (250), UC+Control (342). These are "available patient data," possibly reflecting those who completed the assessments at that specific timepoint.
      • Data Provenance: Australia (Australian National University, Black Dog Institute, University of Sydney). Prospective randomized controlled trial.
    • Pivotal Trial (UVA Study - NCT01438697):

      • Sample Size (Test Set): Total N=303 adults with chronic insomnia.
        • UC+SHUTi group: N=151
        • UC+Control group: N=152
        • Total participants mentioned in some tables (e.g., ISI, Week 9): UC+SHUTi (133), UC+Control (142).
      • Data Provenance: United States (University of Virginia). Prospective randomized controlled trial.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not describe the use of experts to establish a "ground truth" for the test set in the traditional sense of image annotation or diagnostic consensus. Instead, the ground truth for measuring insomnia severity and outcomes was based on:

    • Patient-reported outcomes: Insomnia Severity Index (ISI) questionnaires. The ISI is a validated scale.
    • Patient-reported sleep diaries: These were administered online and collected for 10 days within a 2-week window at each assessment time point. These diaries were used to calculate objective-like sleep parameters such as Sleep Onset Latency (SOL) and Wake After Sleep Onset (WASO).

    The ISI itself incorporates a defined scoring system for severity and remission, which acts as a standardized "ground truth" established by clinical consensus (as referenced by published psychometric studies of the ISI).

    4. Adjudication Method for the Test Set

    No explicit adjudication method is mentioned. The primary outcome measures (ISI scores, sleep diary data for SOL and WASO) are quantitative and collected directly from patients. Clinical intervention studies like these typically rely on these validated instruments rather than expert adjudication of individual case outcomes after the trial.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. Somryst is a computerized behavioral therapy device designed to be used by patients, not an AI diagnostic tool that assists human readers in interpreting medical images or data. Therefore, the concept of assessing how much human readers improve with AI assistance is not applicable to this device.

    6. Standalone Performance

    Yes, a standalone performance study was done. The entire evaluation of Somryst (SHUTi in the studies) was based on its standalone performance in improving insomnia symptoms in patients directly. The comparator was either "Usual Care" plus an attention-matched digital control or an un-tailored digital education program, allowing for the isolation of Somryst's therapeutic effect. The device (SHUTi) itself delivered the intervention without human-in-the-loop assistance for the core therapy delivery. Clinicians had access to a dashboard, but this was for monitoring and management, not direct therapeutic intervention via the device.

    7. Type of Ground Truth Used

    The ground truth used was primarily patient-reported outcomes and validated psychometric scales:

    • Insomnia Severity Index (ISI): A validated questionnaire for assessing insomnia severity. The definitions of "responder" (>=7-point ISI reduction) and "remitter" (ISI score < 8) are based on established clinical criteria derived from psychometric studies of the ISI.
    • Sleep diaries: Patient-reported logs of sleep parameters like sleep onset latency (SOL) and wake after sleep onset (WASO). These provide quantitative, if subjective, measures of sleep behavior.

    8. Sample Size for the Training Set

    The document does not provide a specific "training set" for the Somryst algorithm in the context of machine learning. Somryst is described as a digital therapeutic based on "principles of Cognitive Behavioral Therapy (CBT) for Insomnia, Sleep Restriction, and other proven psychosocial treatment elements." This implies that the therapy logic and content are based on established medical and psychological knowledge (CBT-I protocols) rather than being developed through a data-driven machine learning training process.

    The clinical studies described (ANU and UVA studies) are effectiveness studies, essentially serving as a "test set" to validate the established therapeutic approach delivered by the device.

    9. How Ground Truth for the Training Set Was Established

    As noted above, there isn't a traditional "training set" with ground truth in the machine learning sense. The "ground truth" for the content and principles of Cognitive Behavioral Therapy for Insomnia (CBT-I) and other psychosocial treatment elements, which form the basis of Somryst, would have been established over decades of clinical research, expert consensus, and efficacy studies in the field of sleep medicine and psychology. It represents established medical knowledge and best practices for treating insomnia.