510(k) Data Aggregation

    K Number
    K180481

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-06-29

    (126 days)

    Product Code
    Regulation Number
    876.5880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Servator B SALF Solution is intended for the flushing and cold storage of kidney, liver and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.

    Device Description

    Servator B SALF Solution is a clear to light yellow, single use only, sterile (by steam sterilization) and non-pyrogenic for hypothermic flushing and preservation of organs.

    The formulation includes soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and phosphoric compounds to aid tissue viability by enabling regeneration of adenosine triphosphate (ATP).

    The primary containers used for the device are PVC free bags 1000ml and 2000ml; therefore, they are free of phthalates. The bags must be chilled to between 2° and 6° C (36° to 43° F) prior to use. The solution may be used without any point of use filtration.

    The Servator B SALF Solution is intended for perfusion and flushing donor kidneys, liver and pancreas immediately before or immediately after removal from a dead donor or immediately after removal from a living donor. The solution remains in the organ vessels during hypothermic storage and transportation. Servator B SALF Solution is meant for the cold storage of the organ and is not indicated for hypothermic preservation with continuous mechanical perfusion.

    AI/ML Overview

    The provided text describes the "Servator B SALF Solution," an organ preservation solution, and its substantial equivalence to a predicate device, CoStorSol® (Belzer UW). The document is a 510(k) summary for FDA clearance.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not present a formal table of "acceptance criteria" in the typical sense of numerical thresholds for clinical performance metrics (e.g., sensitivity, specificity for a diagnostic device). Instead, the acceptance criteria are implicitly the demonstration of substantial equivalence to the predicate device, CoStorSol®, across various technological characteristics and non-clinical performance aspects. The "reported device performance" is the result of tests confirming this equivalence.

    Here's a table based on the provided "Table of Comparison" and "Non-Clinical Performance Data" section, framed as a comparison to achieve substantial equivalence:

    Characteristic/TestAcceptance Criterion (Implicit: Substantial Equivalence to Predicate)Reported Servator B SALF Solution Performance
    Indications for UseSame as predicateSame as predicate (flushing and cold storage of kidney, liver and pancreas)
    Intended UseSame as predicateSame as predicate (Organ storage and preservation for transplantation)
    Composition (Key Chemicals)Identical to predicate formulationSame chemical composition as predicate
    Material Container/BagPVC Free / BagsPVC Free / Bags
    Protecting Overwrap BagYesYes
    Single Use OnlyYesYes
    Bag Connections1 flip off, 1 needle point1 flip off, 1 needle point
    Sodium Concentration29 mEq/L29 mEq/L
    Potassium Concentration125 mEq/L125 mEq/L
    pH7.4 at 20°C7.4 at 20°C
    Osmolality320 mOsm/Kg320 mOsm/Kg
    Sterilization MethodValidated according to ISO standardsSteam (ISO 17665-1 compliant)
    Shelf Life24 Months24 Months
    BiocompatibilityMeet ISO 10993 series standardsPassed Cytotoxicity, Irritation, System Toxicity, and Haemocompatibility (ISO 10993 series)
    Sterility ValidationMeet ISO 17665-1 and ASTM F1980-16Passed according to ISO 17655-1 and ASTM F1980-16
    Chemical ComparisonsDemonstrate equivalence to predicateDemonstrated substantial equivalence
    Leachables PerformanceDemonstrate equivalence to predicateDemonstrated substantial equivalence

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Performance Testing...was completed as a direct comparison between the subject and predicate device. The chemical comparisons and leachables performance testing demonstrated the substantial equivalence of this device to the predicate."

    However, it does not specify the sample size for these tests (e.g., how many batches were tested for chemical composition, how many samples for leachables) nor the data provenance (e.g., country of origin, retrospective or prospective). Given the nature of the device (a solution), the "test set" would refer to specific batches or formulations tested in a laboratory setting, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. For a 510(k) submission concerning a medical solution, ground truth is typically established through analytical chemical testing and standardized biological assays, not human expert consensus on interpretations of complex data like medical images.

    4. Adjudication method for the test set

    This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies or studies involving human readers interpreting data, especially in diagnostic imaging. For a chemical solution, the assessment relies on objective measurements and established scientific protocols.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done or is relevant for this device. This is a medical solution for organ preservation, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a solution, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for demonstrating substantial equivalence for this device relies on:

    • Analytical Chemistry and Physicochemical Properties: Direct measurement of concentrations (Na+, K+), pH, osmolality, and confirmation of chemical composition.
    • Sterility Testing: Conformance to ISO and ASTM standards.
    • Biocompatibility Testing: Conformance to ISO 10993 series standards.
    • Leachables Testing: To ensure no harmful substances migrate from the packaging into the solution.
    • Functional Equivalence: Implicitly, the solution's performance in preserving organs is assumed to be equivalent to the predicate due to identical composition and physical properties.

    8. The sample size for the training set

    Not applicable. This device is a medical solution, not based on an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.