510(k) Data Aggregation

    K Number
    K141388

    Validate with FDA (Live)

    Date Cleared
    2014-07-11

    (45 days)

    Product Code
    Regulation Number
    862.1150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The System Calibrator is a calibration serum intended to be used for the calibration of ALP, ALT, AST, Amylase, CK-NAC, GGT and LDH enzymes on Beckman Coulter AU480, AU680 and AU5800 analysers. For In Vitro Diagnostic use only.

    Device Description

    The System calibrator kit is a lyophilized human serum based product intended to provide calibration for AU Enzyme reagents (ALP, ALT, AST, Amylase, CK-NAC, GGT and LDH) for the quantitative determination of the relevant analyte on Beckman Coulter AU analyzers. The following is the kit configuration for the system calibrator: 66300 kit has 20 vials with 5 ml contents for Level 1. The System calibrator is designed for optimal performance on Beckman Coulter AU analyzers. The System calibrator contains lyophilized human serum with chemical additives, preservatives, and enzymes of human, animal and plant origin.

    AI/ML Overview

    This document describes the Beckman Coulter System Calibrator (66300), a lyophilized human serum-based product intended for the calibration of ALP, ALT, AST, Amylase, CK-NAC, GGT, and LDH enzymes on Beckman Coulter AU480, AU680, and AU5800 analyzers.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Open Vial Stability≤ ±5% drift8 hours @ 2-8°C (meets criteria)
    Shelf Life Stability≤ ±5% drift36 months @ 2-8°C (meets criteria, tested for 36 months plus at least one month after expire date claim)

    2. Sample Size Used for the Test Set and Data Provenance

    • Open Vial Stability: 3 individual calibrator lots.
    • Shelf Life Stability: 3 individual lots of calibrator.
    • Value Assignment: Multiple reagent lots and calibrator vials were used. 10 replicates from 8 runs were performed, resulting in a total of 80 data points per individual enzyme. Distinct runs with minimum gaps of 2 hours were performed, and 8 calibration events were used.

    The data provenance is not explicitly stated in terms of country of origin. The studies appear to be internal validation studies conducted by Beckman Coulter Ireland, Inc. and may be considered prospective as they were specifically designed for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable for this type of device. The ground truth for this calibrator is based on analytical performance specifications and traceability to theoretical extinction coefficients, not expert consensus interpretation.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable. This device is a calibrator for in vitro diagnostic assays, and its performance is evaluated through quantitative analytical methods and stability testing, not through expert adjudication of qualitative outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC comparative effectiveness study is not relevant for this device. This product is a calibrator, not an imaging or diagnostic algorithm requiring human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    Yes, in the sense that the performance characteristics (stability, value assignment) are determined directly from the calibrator's analytical behavior on Beckman Coulter AU analyzers without human intervention in the result determination process. The "device" (calibrator) performs its function (calibrating assays) independently of human judgment on the results, although human operators perform the testing.

    7. The Type of Ground Truth Used

    The ground truth for the System Calibrator's values and performance is established through:

    • Analytical Performance Specifications: The device's stability (open vial and shelf life) is evaluated against a maximum allowable drift of ±5%.
    • Traceability to Theoretical Extinction Coefficients: The assigned values for each enzyme are traceable to specific theoretical extinction coefficients (e.g., Amylase OSR6x06: 11320, ALP OSR6x04: 17900), which are fundamental physicochemical properties.
    • Internal Test Procedures and Protocols: Value assignment was determined utilizing internal procedures and protocols that involved multiple reagent lots and calibrator vials.

    8. The Sample Size for the Training Set

    Not applicable. This document describes a calibrator, not a machine learning or AI model that requires a training set. The "training" in this context refers to the calibration of an analyzer using the calibrator, not the development of an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm. For the calibrator's value assignment, the ground truth was established by calculating the overall mean from 80 data points (10 replicates from 8 runs) for each individual enzyme, ensuring incorporation of variation from calibration and environmental sources, and tracing these values to theoretical extinction coefficients.