    Intended Use

    SURGIS Staple Line Reinforcement II is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. The device may be used for buttressing and reinforcing staple lines during lung resection, (e.g., wedge resection, blebectomy, lobectomy, bullectomy, bronchial resection, segmentectomy, pneumonectomy/pneumectomy, pneumoreduction) and other incisions and excisions of the lung and bronchus. The device can be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding, and for reinforcement of staple lines during small bowel, mesentery, colon, and colorectal procedures. The device may be used with anastomotic staplers or non-anastomotic staplers. The device is supplied sterile and is intended for one-time use.

    Device Description

    SURGIS® Staple Line Reinforcement II is manufactured from porcine small intestinal submucosa (SIS) and is nominally supplied in strip sizes (unfolded) of 1.0 x 10.8 cm to 1.2 x 19.8 cm. The device is surface treated to allow it to temporarily adhere to the stapler arms, is packaged in a lyophilized (dried) state, and supplied sterile in a sealed double pouch system.

    AI/ML Overview

    The provided text describes the 510(k) Premarket Notification for the SURGISIS® Staple Line Reinforcement II device. It includes information about the device's intended use, predicate devices, and a summary of testing for substantial equivalence. However, it does not contain the specific details required to complete your request for acceptance criteria and a study proving those criteria are met.

    Here's why and what information is missing:

    • Acceptance Criteria and Reported Device Performance: The document states "Outcomes show the device to be biocompatible, manufacturing processes to adequately disinfect the material, and mechanical characteristics to be sufficient." This is a general statement, not a table of specific acceptance criteria (e.g., tensile strength > X MPa, burst pressure > Y mmHg) and the corresponding numerical results obtained from testing.
    • Sample Size (Test Set), Data Provenance: This information is not present. The document mentions "extensive biocompatibility testing, viral inactivation testing, and mechanical testing," but does not detail the sample sizes for these tests or the origin of any data (human or animal, retrospective/prospective).
    • Number of Experts, Qualifications, Adjudication Method: These points are relevant to studies involving human interpretation or consensus, which are not described here. The tests mentioned are laboratory-based (biocompatibility, viral inactivation, mechanical), not studies requiring human expert adjudication.
    • Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: This type of study would compare human performance with and without AI assistance. The SURGISIS® Staple Line Reinforcement II is a physical medical device (surgical mesh), not an AI algorithm. Therefore, an MRMC study is not applicable.
    • Standalone (Algorithm Only) Performance: As above, this device is a physical product, not an algorithm.
    • Type of Ground Truth Used: For physical device testing, "ground truth" typically refers to established standards and methods for measuring material properties. The document doesn't specify these, beyond general statements about "biocompatibility testing," "viral inactivation testing," and "mechanical testing." It doesn't mention expert consensus, pathology, or outcomes data in the context of establishing a benchmark for performance.
    • Sample Size for Training Set & How Ground Truth for Training Set was Established: These points are relevant to machine learning or AI development, which is not applicable to this physical device.

    In summary, the provided text describes a 510(k) submission for a physical surgical mesh. The "studies" mentioned are laboratory tests designed to demonstrate that the device is substantially equivalent to predicate devices based on its material properties and safety. It does not involve AI, human expert review, or clinical trials in the way your questions imply.

    Therefore, I cannot populate the table and answer the specific questions based only on the provided text. The document is a regulatory summary focusing on substantial equivalence, not a detailed scientific study report with specific performance metrics against defined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022044

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-08-23

    (60 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Surgisis Staple Line Reinforcement is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. The device may be used for buttressing and reinforcing staple lines during lung resection (e.g., wedge resection, blebectomy, bullectomy, bronchial resection, segmentectomy, pneumonectomy, pneumoreduction) and other incisions and excisions of the lung and bronchus. The device can be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding. The device can also be used for abdominal and thoracic wall repair, muscle flap reinforcement, trans-abdominal rectal and vaginal prolapse repair, trans-abdominal reconstruction of the pelvic floor, and repair of hernias (e.g., diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, scrotal, umbilical). The Surgisis Staple Line Reinforcement may be used with anastomotic staplers or with non-anastomotic staplers. The device is supplied sterile and is intended for one-time use.

    Device Description

    The Surgisis Stable Line Reinforcement is manufactured from porcine small intestinal submucosa and supplied in nominal strip sizes (unfolded) of 1 x 10.7 cm, 1.2 x 13.2 cm, and 1.2 x 17.3 cm. The device is packaged sterile, sealed double pouches.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text for the Cook Biotech Incorporated Surgisis® Staple Line Reinforcement:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityMeet ISO 10993-1 standardsMet test requirements, supporting substantial equivalence.
    DisinfectionEffectively disinfectedMet test requirements, supporting substantial equivalence.
    Performance CharacteristicsProvide reasonable assurance of device performance for intended use.Met test requirements, supporting substantial equivalence.
    Substantial EquivalenceSimilar to predicate devices in intended use, materials, and technological characteristics.Determined to be substantially equivalent to predicate devices.

    Note: The document does not explicitly state numerical acceptance criteria. Instead, it refers to meeting test requirements and demonstrating substantial equivalence to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify a sample size for a test set or data provenance for any clinical or performance study that would resemble a "test set" in the context of an AI/ML device. The document describes tests conducted on the material itself ("The material comprising the Surgisis Staple Line Reinforcement was subjected to a panel of tests...").

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this 510(k) submission. This device is a surgical mesh, not an AI/ML device that generates diagnoses or interpretations requiring expert-established ground truth. The "ground truth" for this medical device would relate to its physical and biological properties.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no mention of a test set requiring adjudication in the context of expert review.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done and is not applicable to this type of medical device (surgical mesh). This type of study is relevant for diagnostic devices, particularly those involving human interpretation, often with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study in the context of an algorithm or AI system was not done and is not applicable. This document describes a physical medical device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on:

    • Biocompatibility testing: Conforming to ISO 10993-1 standards.
    • Disinfection efficacy: Demonstrating effective disinfection.
    • Performance characteristics: Meeting mechanical and functional requirements relevant to its intended use as a surgical mesh.
    • Comparison to predicate devices: Demonstrating similar intended use, materials, and technological characteristics to legally marketed devices.

    These are established through laboratory testing and validated regulatory standards, rather than expert consensus, pathology, or outcomes data in the way an AI diagnostic device would use those terms.

    8. The Sample Size for the Training Set

    The provided text does not mention a training set sample size. A "training set" is typically associated with AI/ML model development, which is not relevant to this device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no mention of a training set or AI/ML model development.


    Summary of the Study that Proves the Device Meets the Acceptance Criteria:

    The study proving the Surgisis® Staple Line Reinforcement meets its (implied) acceptance criteria is described as a panel of tests conducted on the material itself. These tests assessed:

    • Biocompatibility: In accordance with ISO 10993-1 standards.
    • Disinfection: To ensure effective sterilization.
    • Performance Characteristics: Likely including mechanical properties and structural integrity relevant for its use as a soft tissue reinforcement.

    The document states that the "material met the test requirements," providing "reasonable assurance of device performance for its intended use and supporting substantial equivalence." The overall conclusion drawn from these tests was that the device is "substantially equivalent" to existing predicate devices (Surgisis® Soft Tissue Graft, Surgisis® Peripheral Vascular Patch, Peri-Strips® Staple Line Reinforcement-Strip, and Seamguard® Staple Line Reinforcement Material) with respect to intended use and technological characteristics. This substantial equivalence is the primary basis for its 510(k) clearance.