Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K241230

    Validate with FDA (Live)

    Date Cleared
    2024-09-17

    (138 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Super Sheath Introducer Sheath is indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature of adult and pediatric patients of all ages.

    Device Description

    The Super Sheath Introducer Sheath (hereinafter referred to as the "Subject Device") is used to introduce diagnostic or interventional devices into a blood vessel. The Subject Device is available in two products: one consisting of a Sheath and Dilator (Super Sheath Introducer Sheath without Guidewire), and the other consisting of a Sheath, Dilator, and Guidewire (Super Sheath Introducer Sheath with Guidewire), similar to the Predicate Device K121504. These devices are provided sterile and intended for single use, and the shelf-life is three years.

    The main component of the Sheath consists of a Sheath hub, a Sheath shaft connected to the Sheath hub, and a three-way stopcock valve connected to the hub with a side port tube. The Sheath shaft is 3.3 French size and available in 5cm and 7cm lengths. The Sheath hub contains a hemostatic valve and has a suture wing. The shape and dimensions of the three-way stopcock luer connector are modified to comply with the new applicable standard ISO 80369-7:2021.

    The Dilator has an open, tapered plastic tube for Guidewire insertion. The Dilator tube is press-fit into the inner hub with a bushing. The length of the Dilator tube corresponds to the length of the Sheath shaft, which is longer and available in 10.5 cm and 12.5 cm.

    The 0.018" Guidewire is the smallest size Guidewire and a 40 cm long, thin, flexible wire with a inserter attached to the tip. This Guidewire is a legally marketed product which is cleared by the Food and Drug Administration (K021990).

    The upper cap and suture wing of the Sheath hub, and the Dilator hub are colored purple. The Dilator tube and the inserter are colored blue. And the unvented cap of the three-way stopcock is pink.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the "Super Sheath Introducer Sheath". This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria through a clinical study or performance data like an AI/ML diagnostic device would.

    Therefore, the provided text does not contain the information requested regarding acceptance criteria and a study to prove a device meets those criteria for an AI/ML-based device. The original request is formulated for a performance study of a diagnostic device (likely AI-powered), which requires metrics like sensitivity, specificity, or AUC, and details on clinical validation with ground truth.

    The document describes a physical medical device (an introducer sheath), and its substantial equivalence is proven through:

    • Bench Testing: Mechanical and physical properties (tensile strength, kink resistance, connection strength, pressure tests, lubricity, radiopacity, corrosion resistance).
    • Biocompatibility Testing: Cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, hemocompatibility, pyrogen test, LAL test.
    • Comparison of Technological Characteristics: To a predicate device (K121504) across features like French sizes, indications for use, regulation numbers, components, sterilizations, and dimensional comparisons.

    In summary, the provided document does not allow for a response to the specific questions regarding acceptance criteria and performance study details for an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200379

    Validate with FDA (Live)

    Device Name
    Super Sheath
    Date Cleared
    2020-03-19

    (30 days)

    Product Code
    Regulation Number
    870.1310
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in the introduction of diagnostic and interventional devices inserted into the peripheral and coronary vasculature.

    Device Description

    Major components of the Super Sheath (Proposed) are a sheath and a dilator. In addition, a guidewire is covered under this premarket notification. The guidewire is placed in the guidewire case including the inserter. The sheath for the Super Sheath (Proposed) is named as the Super Sheath (Proposed) Introducer Sheath and is available in 4F- 9F diameters and in lengths ranging from 7 to 25 cm. Some types of sheaths have radiopaque markers. The Super Sheath (Proposed) are provided sterile and are intended for one procedure use only. The one-piece construction of the sheath shaft and hub allows smooth passage of concomitantly used interventional medical devices. The sheath shaft and hub are made from Ethylene Tetrafluoro ethylene (ETFE) and polyamide. The hubs contain a hemostatic valve to minimize blood leakage during a procedure. A side tube equipped with a three-way stopcock is attached to the sheath hub. The side tube extension may be used for fluid and medication administration, as well as blood sampling. Some types of sheaths contain a marker band that is embedded in the shaft approximately 2.0 mm from the distal tip. The dilator is an open tapered plastic tube with an integral luer hub for guidewire insertion. The dilator is inserted into the introducer sheath. The dilator is longer than the sheath with a rounded tapered distal tip. The dilator facilitates and supports the entry of the sheath into the patient's vasculature. Once the sheath is in place, the dilator is removed. There are two different dilator distal tip sizes, 0.038" and 0.035" guidewire compatible. The 4F dilators are only available with a 0.035"guidewire lumen dilator. The 5-9F dilators are available with a 0.035" or 0.038" guidewire lumen. The 4-8F dilator tubes are made from Polypropylene (PP). The 9F dilator tubes are made from Fluorinated Ethylene Propylene. The dilator tubes are attached to the inner hub with a metal bush. The dilator is locked into the sheath by rotating motion. The Super Sheath (Proposed) has a polypropylene suture wing which is color coded by French size. The suture wing is a small projection near the hub with a hole in it for correct placement of the sheath using sutures. The uncoated guidewire is made of a stainless steel coil wrapped tightly around an inner mandrel that tapers at the distal tip. The flexible tip is J-shaped available in diameters of 0.035" and 0.038". The guidewire, with inserter, is available in 45 cm and 80 cm lengths. The inserter is used strictly for guiding the guidewire into a cannula or introducer.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device called "Super Sheath" (Proposed), which is a vessel dilator for percutaneous catheterization. The submission aims to demonstrate substantial equivalence to a predicate device, also named "Super Sheath" (K141070).

    This document does not include information about AI/ML algorithms, human-in-the-loop performance, or the establishment of ground truth by expert consensus. Therefore, many of the requested elements for describing the acceptance criteria and study proving device performance will be inapplicable. The study's focus is on non-clinical performance testing to support changes in materials and an extended shelf-life for a physical medical device.

    However, I can extract the relevant information regarding the acceptance criteria and the performance testing conducted for this traditional medical device.

    1. A table of acceptance criteria and the reported device performance:

    The document details performance testing for the "Super Sheath" (Proposed) to support modifications from the predicate device, specifically for an extended shelf-life and changes to the 8F dilator. The acceptance criteria are described as "established methods and acceptance criteria in conformity with ISO 11070:2014 and the submission under the predicate device without deviation."

    Table: Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance Criteria (General)Reported Device Performance
    Biocompatibility Testing (for patient-contacting parts with new materials/additives)Adherence to ISO 10993-1 recommendations for: Cytotoxicity, Intracutaneous Reactivity, Sensitization, Acute Systemic Toxicity, Hemocompatibility (Hemolysis, Complement activation, Partial Thromboplastin Time, Platelet activation, in vivo Thromboresistance), Material-Mediated Pyrogen test.All "Biocompatibility evaluation... was conducted," meeting "recommendations."
    Residual Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) TestingConformity with ISO 10993-7 and predicate device submission exposure limits.Did not exceed exposure limit.
    Shelf-life Testing (Sheath & 8F Dilator)Maintenance of physical performance for three years, conforming to ISO 11070:2014 and predicate device submission without deviation for: Sheath shaft tensile testing, Sheath kink testing, Sheath hub to shaft tensile testing, Valve integrity/sheath pressure testing, Sheath lubricity testing, Sheath radiopacity testing, Cover tube durability testing, Hemostatic valve integrity/sheath pressure testing, Visual inspection on the dilator tube, Dilator shaft tensile test, Dilator hub to shaft tensile test.The sheath and the new 8F dilator maintain the physical performance for the three-year shelf-life.
    Performance Testing for 8F DilatorConformity with ISO 11070:2014 and predicate device submission acceptance criteria for: Visual inspection on the dilator tube, Dilator shaft tensile test, Dilator hub to shaft tensile test, Dimensional analysis of the tip of the dilator tube (within tolerance range for predicate device).Met the acceptance criteria. Dimensional analysis showed the 8F dilator tube was within the tolerance range for the predicate device.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify exact sample sizes for each test. It refers to "testing" performed.
    • Data Provenance: The testing was conducted by Togo Medikit Co., Ltd. in Japan. The data is likely from prospective testing, as it was performed to support design modifications and an extended shelf-life for the proposed device before its market introduction.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is a physical device with performance testing based on established engineering standards (e.g., ISO 11070:2014) and direct measurements, not expert human interpretation (like radiology image analysis).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. As mentioned above, this is non-clinical performance testing of a physical device against engineering standards, not a study involving human adjudication of clinical output.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI/ML device, nor does it involve human readers or image analysis.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used:

    • The "ground truth" for this device's performance is established by engineering standards, material specifications, and physical measurements (e.g., tensile strength, dimensional accuracy, leak integrity, biocompatibility test results). It is based on a comparison to the predicate device's performance and compliance with relevant ISO standards (ISO 10993 for biocompatibility, ISO 11070 for catheter/dilator performance).

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141070

    Validate with FDA (Live)

    Device Name
    SUPER SHEATH
    Date Cleared
    2015-01-15

    (265 days)

    Product Code
    Regulation Number
    870.1310
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use in the introduction of diagnostic and interventional devices inserted into the peripheral and coronary vasculature.

    Device Description

    Major components of the SUPER SHEATH (PROPOSED) are a sheath and a dilator. In addition, a guidewire is covered under this premarket notification. The guidewire is placed in the gudiewire case including inserter. The sheath for the SUPER SHEATH (PROPOSED) is named as the SUPER SHEATH (PROPOSED) Introducer Sheath and is available in 4F- 9F diameters and in lengths ranging from 7 to 25 cm. Some types of sheaths have radiopaque markers. The SUPER SHEATH (PROPOSED) are provided sterile and are intended for one procedure use only. The one-piece construction of the sheath shaft and hub allows smooth passage of concomitantly interventional medical devices. The sheath shaft and hub are made of Ethylene Tetrafluoro ethylene (ETFE) and polyamide. The hubs, color-coded by French size, contain a hemostatic valve to minimize blood leakage during a procedure. A side tube equipped with a three-way stopcock is attached to the sheath hub. The side tube extension may be used for fluid and medication administration, as well as blood sampling. Some types of sheaths contain a marker band that is embedded in the shaft approximately 2.0 mm from the distal tip. The dilator is an open tapered plastic tube with an integral luer hub for guidewire insertion. The dilator is inserted into the introducer sheath. The dilator is longer than the sheath with a rounded tapered distal tip. The dilator facilitates and supports the entry of the sheath into the patient's vasculature. Once the sheath is in place the dilator is removed. There are two different dilator distal tip sizes, 0.038" and 0.035" guidewire compatible. The 4F dilators are only available with a 0.035" guidewire lumen dilator. The 5-9F dilators are available with a 0.035" or 0.038" guidewire lumen. The 4bes are made from Polypropylene (PP). The 9F dilator tubes are made of Fluorinated Ethylene Propylene (FEP). The dilator tubes are attached to the inner hub with a metal bush. The sheath hub and the dilator hub lock using a rotating motion. The SUPER SHEATH (PROPOSED) has a polypropylene suture wing which is color coded by French size. The suture wing is a small projection near the hub with a hole in it for correct placement of the sheath using sutures. The uncoated guidewire is made of a stainless steel coil wrapped tightly around an inner mandrel that tapers at the distal tip. The flexible tip is J-shaped available in diameters of 0.035" and 0.038". The guidewire, with inserter, is available in 45 cm and 80 cm lengths. The inserter is used strictly for guiding the guidewire into a cannula or introducer.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called "SUPER SHEATH (PROPOSED)," a catheter introducer. The submission aims to establish substantial equivalence to previously marketed predicate devices, "SUPER SHEATH (K052557)" and "3.3F SUPER SHEATH (K121504)."

    The acceptance criteria are not explicitly stated as numerical performance targets (e.g., "sensitivity of 90%"). Instead, the acceptance criteria are implicitly met by demonstrating that the "SUPER SHEATH (PROPOSED)" device is substantially equivalent to its predicate devices through various bench tests and biocompatibility testing. The study's goal was to ensure the proposed device's safety and effectiveness amidst material changes to the sheath shaft and packaging modifications.

    1. Table of acceptance criteria and the reported device performance:

    Since specific numerical acceptance criteria (e.g., target tensile strength, specific kink resistance values) are not provided in the document, the table below lists the types of tests performed and the general findings as the "reported device performance." The acceptance criteria, in this context, were that the proposed device would perform comparably to its predicate device or meet international/in-house requirements without raising new safety or effectiveness concerns.

    Test CategorySpecific Bench Test TypeImplicit Acceptance Criteria (based on predicate equivalence/safety)Reported Device Performance
    Mechanical Performance (Sheath)Sheath Shaft Tensile TestThe sheath shaft must maintain sufficient tensile strength to prevent breakage during use and be comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Sheath Kink TestThe sheath must resist kinking to ensure proper device insertion and functionality, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Sheath Hub to Shaft Tensile TestThe connection between the sheath hub and shaft must be strong enough to withstand forces during use, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Valve Integrity of SheathThe hemostatic valve must effectively minimize blood leakage during a procedure, performing comparably to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Sheath Lubricity TestThe sheath must exhibit adequate lubricity for smooth insertion and removal, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Sheath Radiopacity TestThe sheath should be sufficiently radiopaque for visualization under fluoroscopy, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Cover Tube Durability TestThe cover tube must be durable enough for its intended purpose.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Hemostatic Valve Integrity/Sheath Pressure TestThe hemostatic valve and sheath must withstand internal pressure and maintain integrity.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Sheath/Dilator Corrosion Resistance TestNo significant corrosion should occur that could compromise functionality or patient safety, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Mechanical Performance (Dilator)Dilator Shaft Tensile TestThe dilator shaft must possess adequate tensile strength, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Dilator Hub to Shaft Tensile TestThe connection between the dilator hub and shaft must be strong enough to withstand forces during use, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Mechanical Performance (Guidewire)Guidewire Tensile TestThe guidewire must exhibit sufficient tensile strength to prevent breakage during manipulation, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Guidewire Combined Load TestThe guidewire must perform reliably under combined loads experienced during procedures, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Guidewire Torqueability TestThe guidewire must have acceptable torqueability for precise positioning, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Guidewire Radiopacity TestThe guidewire should be sufficiently radiopaque for visualization under fluoroscopy, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    Guidewire Corrosion Resistance TestNo significant corrosion should occur that could compromise functionality or patient safety, comparable to the predicate device.Bench testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).
    BiocompatibilityCytotoxicityNo unacceptable cytotoxic effects should be observed.No biocompatibility concern was raised in the biocompatibility testing reports.
    Intracutaneous ReactivityNo unacceptable intracutaneous reactivity should be observed.No biocompatibility concern was raised in the biocompatibility testing reports.
    SensitizationNo unacceptable sensitization (allergic reaction) should be observed.No biocompatibility concern was raised in the biocompatibility testing reports.
    Acute Systemic ToxicityNo unacceptable acute systemic toxic effects should be observed.No biocompatibility concern was raised in the biocompatibility testing reports.
    HemocompatibilityThe device should not cause adverse blood-material interactions (e.g., hemolysis, thrombosis).No biocompatibility concern was raised in the biocompatibility testing reports.
    Pyrogen TestThe device should be non-pyrogenic to prevent fever reactions.No biocompatibility concern was raised in the biocompatibility testing reports.
    LAL Test (Endotoxin)Limits for endotoxins must be met.No biocompatibility concern was raised in the biocompatibility testing reports.
    Sterilization & PackagingPackaging Integrity TestThe packaging must maintain a sterile barrier throughout the shelf-life.Packaging integrity test and sterility test demonstrates the packaging works as a sterile barrier for the SUPER SHEATH (PROPOSED) and is substantial equivalent to the packaging of the 3.3F SUPER SHEATH (K121504).
    Sterility TestThe device must remain sterile until opened for use and meet requirements for ethylene oxide sterilization.Packaging integrity test and sterility test demonstrates the packaging works as a sterile barrier for the SUPER SHEATH (PROPOSED) and is substantial equivalent to the packaging of the 3.3F SUPER SHEATH (K121504).
    Shelf-Life TestingThe device components and sterile barrier must remain stable and functional for the stated shelf-life (2 years for the proposed device), comparable to the predicate devices (3 years for predicates, although proposed is 2 years, this difference in shelf-life is not highlighted as a concern).Bench testing including shelf-life testing demonstrated that the performance of the main body of the SUPER SHEATH (PROPOSED) was comparable to that of the SUPER SHEATH (K052557).

    Regarding the study proving the device meets the acceptance criteria:

    The study described is a series of bench tests and biocompatibility tests designed to demonstrate that the "SUPER SHEATH (PROPOSED)" is substantially equivalent to its predicate devices, despite a material change in the sheath shaft and changes in packaging.

    • 2. Sample size used for the test set and the data provenance:

      • Sample Size: The document does not specify the exact number of devices or samples used for each bench test or biocompatibility test. It simply states that "bench testing" and "biocompatibility testing" were performed.
      • Data Provenance: The tests were conducted by TOGO MEDIKIT CO., LTD., a Japanese company. The data would therefore originate from their internal testing processes. The document does not specify if the tests were retrospective or prospective, but typically such verification tests are performed prospectively on newly manufactured devices or components.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable to this type of submission. This premarket notification is for a physical medical device (catheter introducer) and its performance is evaluated through objective physical and chemical testing (bench tests, biocompatibility tests). There is no "ground truth" established by human experts in the way that an AI diagnostic device would require (e.g., radiologist consensus on images). The "ground truth" for these tests are objective measurements against established engineering standards or comparative performance against predicate devices.
    • 4. Adjudication method for the test set:

      • This question is not applicable for the reasons stated above. There is no human interpretation or adjudication of "cases" as would be seen in an AI study. The results of the physical and chemical tests are objectively measured.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This question is not applicable. This is not an AI/software device that involves human interpretation of diagnostic data. It is a physical medical device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable. This is not an AI/software device. The performance is of the physical device components.
    • 7. The type of ground truth used:

      • The "ground truth" in this context refers to the objective performance criteria derived from engineering specifications, international standards, and the established performance of the predicate devices. For biocompatibility, the ground truth is the absence of adverse biological reactions/toxicity as defined by ISO standards for biocompatibility.
    • 8. The sample size for the training set:

      • This question is not applicable. This submission does not involve an AI algorithm; therefore, there is no "training set." The device is a physical product.
    • 9. How the ground truth for the training set was established:

      • This question is not applicable. As there is no AI algorithm, there is no training set or associated ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K121504

    Validate with FDA (Live)

    Date Cleared
    2012-09-27

    (129 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Sets are indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature of adult and pediatric patients of all ages.

    Device Description

    The Super Sheath Introducer Sheaths and Super Sheath Sets are similar to the 4F-9F with the addition of the 3.3 French sheath, compatible dilator and guidewire to the sheath line. The Super Sheath Introducer Sheath is available in 3.3F and lengths of 5 cm and 7 cm. The device is provided sterile and intended for one procedure use only.

    The Super Sheath Introducer Sheath is packaged with one dilator. The Super Sheath Introducer Sheath Set consists of one sheath, one dilator, and one guidewire with inserter. The sheath shaft and hub are manufactured of polyamide and ethylene tetrafluoro ethylene; one-piece construction of the sheath shaft and hub allows smooth passage of medical devices. The hub, color-coded by French size, contains a hemostatic valve to prevent blood leakage during a procedure. A side tube equipped with a three-way stopcock is attached to the sheath hub. The side tube extension may be used for fluid and medication administration, as well as blood sampling.

    The dilator is an open, tapered plastic tube with an integral luer hub for guidewire insertion. The guidewire is inserted into the introducer sheath to facilitate and support entry of the sheath into the patient's vasculature. The dilator is longer than the sheath with a rounded tapered distal tip. The 3.3F Super Sheath Introducer Sheaths and provided dilator are compatible with a maximum recommended 0.025" guidewire. The dilator tubes are manufactured of polypropylene. Dilator tubes are press-fit into the dilator hub with a bushing. The sheath hub and dilator hub lock using a rotating motion.

    AI/ML Overview

    The provided text describes a medical device (Super Sheath Introducer Sheaths and Sets) and its equivalence to a predicate device, rather than the acceptance criteria and study results for a new AI or diagnostic device. As such, most of the requested information (sample sizes, expert qualifications, MRMC studies, ground truth types, training data, etc.) is not applicable or present in this document.

    However, I can extract the information related to the functional testing conducted to establish the performance of this medical device.

    Acceptance Criteria and Device Performance (Functional Testing)

    The document states that "Functional testing was conducted to verify the 3.3F Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Sets met product specification." While specific numerical acceptance criteria are not provided, the types of tests performed indicate the performance characteristics that were evaluated and met.

    Acceptance Criteria (Implied)Reported Device Performance
    Sheath Shaft Tensile Strength: Must withstand tensile forces.Met product specification through testing.
    Sheath Kink Resistance: Must resist kinking during use.Met product specification through testing.
    Connection Strength: Connections must be secure.Met product specification through testing.
    Hemostatic Valve Pressure Integrity: Valve must prevent blood leakage under pressure.Met product specification through testing.
    Sheath Valve Integrity / Sheath Pressure: Valve and sheath must maintain integrity under pressure.Met product specification through testing.
    Sheath Lubricity: Sheath must allow smooth passage.Met product specification through testing.
    Sheath Radiopacity: Sheath must be visible under fluoroscopy.(Note: Predicate had a marker, new device states "Not applicable" for marker. This implies the sheath material itself might be radiopaque or it's not a P.criterion for the 3.3F based on its intended use/size.)Met product specification through testing.
    Sheath / Dilator Corrosion Resistance: Must resist corrosion.Met product specification through testing.
    Dilator Tensile Strength: Dilator must withstand tensile forces.Met product specification through testing.
    Dilator Connection Strength: Dilator connections must be secure.Met product specification through testing.
    Guidewire Tensile Strength: Guidewire must withstand tensile forces.Met product specification through testing.
    Guidewire Radiopacity: Guidewire must be visible under fluoroscopy.Met product specification through testing.
    Guidewire Corrosion Resistance: Guidewire must resist corrosion.Met product specification through testing.
    Guidewire Torque Load: Guidewire must withstand torque.Met product specification through testing.
    Guidewire Torqueability: Guidewire must be steerable/torqueable.Met product specification through testing.
    Sterilization Efficacy: Device must be sterile.Validated in accordance with ISO 11135-1:2007.
    Biocompatibility: Device materials must be biocompatible.Demonstrated by similarity to predicate in K052557.
    Packaging Integrity: Packaging must maintain sterility and product integrity.Demonstrated by similarity to predicate in K052557.
    Product Shelf Life: Device must maintain performance over time.Demonstrated by similarity to predicate in K052557 (3 years).

    Study Details (for a medical device, not AI)

    1. Sample size used for the test set and the data provenance: Not explicitly stated. The testing was functional/engineering testing of the device itself, not patient data. Therefore, "data provenance" in terms of country of origin or retrospective/prospective is not applicable.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These were non-clinical, functional tests performed according to protocols. "Ground truth" in this context would refer to engineering specifications and international standards, not expert medical opinion.
    3. Adjudication method for the test set: Not applicable. Results would be compared against established product specifications and standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, not applicable. This is not an AI or diagnostic imaging device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No, not applicable.
    6. The type of ground truth used: Product Specifications, International Standards (e.g., ISO 11135-1:2007 for sterilization), and in-house requirements.
    7. The sample size for the training set: Not applicable. This document describes the testing of a physical medical device, not a machine learning model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K060190

    Validate with FDA (Live)

    Date Cleared
    2006-04-04

    (70 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medikit Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Sets are indicated for use in the introduction of diagnostic and interventional devices inserted into the human vasculature.

    Device Description

    The Medikit Super Sheath Introducer XL Sheaths and Super Sheath XL Introducer Sheath Sets are similar to the 4-9F sheaths with the addition of four French sizes to the sheath line. The Super Sheath XL Introducer Sheaths are available in 10F, 11F, 12F and 14F sizes and available in lengths ranging from 11 to 25 cm. The devices are provided sterile and are intended for one procedure use only. The Medikit Super Sheath XL Introducer Sheath is packaged with a dilator. The Super Sheath Introducer XL Sheath Set consists of a sheath, a dilator and a mini guidewire, with inserter. The mini guidewires with inserter were cleared to market in K052557 (January 17, 2006). The construction of the sheath shaft and hub allows smooth passage of medical devices. The sheath shaft and hub are made of polyamide and fluorinated ethylene propylene FEP (Teflon). The hubs, color-coded by French size, contain a hemostatic valve to prevent blood leakage during a procedure. A side tube equipped with a three-way stopcock is attached to the sheath hub. The side tube extension may be used for fluid and medication administration as well as blood sampling. The dilator is an open tapered plastic tube with an integral luer hub for guidewire insertion. The guidewire is inserted into the introducer sheath and facilitates and supports the entry of the sheath into the patient's vasculature. The dilator is longer than the sheath with a rounded tapered distal tip. Once the sheath is in place the dilator is removed. The dilators are 0.038" and a 0.035" guidewire compatible. The dilator tubes are made of fluorinated ethylene propylene (FEP). The dilator tubes are attached to the dilator hub using an epoxy resin. The sheath hub and the dilator hub lock using a rotating motion.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (introducer sheaths) and does not describe acceptance criteria or a study that proves a device meets those criteria in the context of an AI/ML clinical validation study. Instead, it details the non-clinical (bench) testing performed to demonstrate substantial equivalence to a predicate device.

    Therefore, many of the requested categories are not applicable to the provided document.

    Here's a breakdown of what can be extracted and what cannot:

    1. Table of acceptance criteria and the reported device performance:

    The document describes several non-clinical tests but does not explicitly state acceptance criteria in a quantitative format or report performance against specific thresholds. It generally states that "Test results verified that the Super Sheath XL Introducer Sheaths and Super Sheath XL Introducer Sheath Sets are adequate for their intended use."

    Acceptance Criteria (Explicitly Stated)Reported Device Performance
    Not explicitly stated in the document for each test. General goal: "meet product specification" and be "adequate for their intended use."Verified as "adequate for their intended use."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for individual non-clinical tests.
    • Data Provenance: N/A (non-clinical bench testing, not human data).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    N/A. Ground truth in this context would be test standards/specifications, not expert interpretation of data.

    4. Adjudication method for the test set:

    N/A. This is relevant for diagnostic studies, not bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    N/A. This document describes a medical device, not an AI/ML diagnostic system, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    N/A. This is not an algorithm.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" would be the established engineering specifications and performance standards for introducer sheaths, which are not detailed in this summary but are referred to as "product specification."

    8. The sample size for the training set:

    N/A. No training set is applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established:

    N/A. No training set.


    Summary of Device-Specific Information from the Text:

    The document describes the Super Sheath XL Introducer Sheaths and Super Sheath XL Introducer Sheath Sets, which are indicated for the introduction of diagnostic and interventional devices into the human vasculature.

    Study Type: Non-clinical (bench) testing was performed to verify that the device met product specifications and to establish substantial equivalence to a predicate device (Medikit Super Sheath Introducer Sheaths and Super Sheath Introducer Sheath Sets, K052557).

    Tests Performed:

    • Sheath Shaft Tensile test
    • Sheath Kink Test
    • Sheath Hub to Shaft Tensile Test
    • Hemostatic Valve Integrity/Sheath Pressure Test
    • Sheath Valve Integrity/Sheath Pressure Test
    • Dilator Shaft Tensile Test
    • Dilator Hub to Shaft Tensile Test
    • Real Time Product Shelf Lift Testing
    • (Also referenced from predicate device testing: Lubricity, radiopacity, sheath/dilator corrosion resistance testing, biocompatibility, and package test reports).

    Conclusion of Testing: "Test results verified that the Super Sheath XL Introducer Sheaths and Super Sheath XL Introducer Sheath Sets are adequate for their intended use. Based on a comparison of intended use, design and the results of bench testing, Medikit 10F, 11F, 12F, and 14F Super Sheath XL Introducer Sheaths and Super Sheath XL Introducer Sheath Sets have been shown to be substantially equivalent to the predicate (K052557) device."

    Ask a Question

    Ask a specific question about this device

    K Number
    K052557

    Validate with FDA (Live)

    Date Cleared
    2006-01-17

    (123 days)

    Product Code
    Regulation Number
    870.1310
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Introducer Sheaths and Introducer Sheath Sets are intended for use in the introduction of diagnostic and interventional devices into the human vasculature.

    Device Description

    The Super Sheath Introducer Sheaths are available in-4F-Device Description 9F diameters and available in lengths ranging from 7 to 25 cm. Some sheaths may be ordered with a suture wing and some sheaths have radiopaque markers. The devices are provided sterile and are intended for one procedure use only.

    The Medikit Super Sheath Introducer Sheath is packaged with a dilator. The Super Introducer Sheath Sets consists of a sheath, a dilator and a mini guidewire, with Inserter, as described in this submission or a US commercially available mini guidewire meeting Medikit specifications. The one-piece construction of the sheath shaft and hub allows smooth passage of medical devices. The sheath shaft and hub are made of polyamide and Ethylene shart and mac mac the (Teflon). Some sheaths have radiopaque markers that are embedded into the shaft 1.5 mm from the distal tip. The hubs, color-coded by French size, contain a hemostatic valve to prevent blood leakage during a procedure. A side tube equipped with a three-way stopcock is attached to the sheath hub. The sidetube extension is used for site management.

    The dilator is an open tapered plastic tube with an integral luer hub for guidewire insertion. It is inserted into the introducer sheath and facilitates and supports the entry of the sheath into the patient's vasculature. The dilator is longer than the sheath with a rounded tapered distal tip. Once the sheath is in place the dilator is removed. There are two different dilators, 0.038" and a 0.035" guidewire compatible. The 4-8F-dilator tubes are made from Polypropylene (PP). The 9F dilator tubes are made of Fluorinated Ethylene Propylene (FEP). The dilator tubes are pressed into the dilator hub with a bushing. The sheath hub and the dilator hub lock using a rotating motion.

    Super Sheaths can be ordered with a Polypropylene suture wing color coded by French size. The suture wing is a small projection near the hub with a hole in it for correct placement of the sheath using sutures.

    The uncoated mini guidewire is made of a stainless steel coil wrapped tightly around an inner mandrel that tapers at the distal tip. The flexible tip is J-shaped available in diameters of 0.035" and 0.038". The mini guidewire (with Inserter) is available in 45cm and 80cm lengths. The Inserter does not contact blood and is used strictly for guiding the guidewire into a cannula or introducer. A US commercially available mini guidewire (meeting Medikit specifications) may be used in the sets or the mini guidewire that is described in this submission.

    AI/ML Overview

    This is a medical device submission for a "Super Sheath Introducer Sheath" which is a physical device, not an AI/ML powered device. Therefore, the questions regarding acceptance criteria and study details for an AI/ML device are not applicable to this document. The document describes functional testing of the physical device to demonstrate substantial equivalence to a predicate device, as opposed to performance metrics for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1