510(k) Data Aggregation

    K Number
    K212596
    Date Cleared
    2022-03-01

    (197 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Sterile Latex Surgical Gloves Powder Free is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The proposed device, Sterile Latex Surgical Gloves Powder Free with protein content labeling claim of 50 µg/ dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves. The classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device is Sterile Latex Surgical Gloves Powder Free, and variants of different sizes. All variants share the same natural color (No Color is added). The proposed device is sterilized using Gamma and ETO method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 5years.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for Sterile Latex Surgical Gloves Powder Free (K212596), intended to be worn in operating rooms. The study is a non-clinical performance (bench testing) study demonstrating substantial equivalence to a predicate device (K192328).

    Here's the breakdown of the information requested, based on the provided document:


    Acceptance Criteria and Device Performance Study for Sterile Latex Surgical Gloves Powder Free (K212596)

    The study performed is a non-clinical performance (bench testing) study, as explicitly stated under "SUMMARY OF NONCLINICAL PERFORMANCE DATA" and "SUMMARY OF CLINICAL PERFORMANCE DATA: Clinical data was not required for this submission." This type of study focuses on technical and physical characteristics rather than performance in a clinical setting with human participants.

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria are generally based on ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves, and other specific standards for each test.

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance (Average Value of Subject Device)Met Acceptance Criteria?
    DimensionsASTM D3577-2019
    Length (All sizes)Min 265 mmVaries by size (e.g., Size 7: 486 mm, Size 9: 488 mm)Yes
    Width (e.g., Size 7)89 ± 6 mm91 mmYes
    Finger Thickness (All sizes)Min 0.10 mm0.33 mmYes
    Palm Thickness (All sizes)Min 0.10 mm0.21 mmYes
    Cuff Thickness (All sizes)Min 0.10 mm0.16 mmYes
    Tensile StrengthASTM D3577-2019
    Before aging (All sizes)24 MPa minimum28.61 MPaYes
    After aging (All sizes)18 MPa minimum24.12 MPaYes
    Ultimate ElongationASTM D3577-2019
    Before aging (All sizes)750% minimum871%Yes
    After aging (All sizes)560% minimum732%Yes
    Stress at 500% before aging5.5 MPa Max5.1 MPaYes
    Pinhole AQLASTM D3577-2019, ASTM D5151-2019
    Before aging (All sizes)Max 1.51.0Yes
    After aging (All sizes)Max 1.51.0Yes
    Freedom from Holes AQLAQL 1.0 (ASTM D3577-2019, ASTM D5151-2019)1.0Yes
    Powder Residue< 2 mg/Glove; (ASTM D3577-2019, ASTM D6124-06)0.38 mg/GloveYes
    Protein Content< 50 µg/dm² (ASTM D3577-2019, ASTM D5712-15)46.40 µg/dm²Yes
    Biocompatibility
    Skin Irritation & SensitizationNon-irritant and Non-Sensitizer (ISO 10993-10)Non-irritant and Non-SensitizerYes
    In vitro CytotoxicityNon-cytotoxic (ISO 10993-5)Cytotoxic (Note: While reported as 'cytotoxic', this appears to be a misstatement or misinterpretation in the table, as the comparison table above it shows 'non-cytotoxic' for the predicate. However, in the detailed table, it is reported as 'cytotoxic', matching what is written for the predicate. This requires clarification to understand why a 'cytotoxic' result is deemed acceptable if the acceptance criteria is 'non-cytotoxic'.)Check with manufacturer/FDA guidance
    Material Mediated PyrogenicityNon-pyrogenic (ISO 10993-11, USP 41 <151>)Non-pyrogenicYes
    Systemic ToxicityDevice extracts do not pose a systemic toxicity concern (ISO 10993-11)Device extracts do not pose a systemic toxicity concernYes
    Bacterial Endotoxin<20 EU/pair of gloves (USP 42 <85>)<20 EU/pair of glovesYes

    Note on Cytotoxicity: The specific call out for "in vitro cytotoxicity" in the detailed results table (page 9) states "Under the conditions of the study cytotoxic." However, the acceptance criteria listed immediately above it is "Under the conditions of study non cytotoxic." And in the "Technological Characteristics Comparison" table (page 7), both the subject and predicate devices are listed as "cytotoxic" with an acceptance criterion of "cytotoxic" but then the standard states "non-cytotoxic." This is a discrepancy that would typically require clarification in a regulatory submission. Assuming the manufacturer has demonstrated that this level of cytotoxicity is acceptable for the intended use and does not pose a safety risk, or that the "acceptance criteria" presented in the detailed result is actually just reporting the finding, it passed. However, simply stating "cytotoxic" when the standard implies "non-cytotoxic" needs careful review. For the purpose of this summary, it's listed as reported.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not explicitly state the exact sample sizes (number of gloves) used for each bench test. Standards like ASTM D3577 typically specify sampling plans (e.g., AQL levels for defect testing).
    • Data Provenance: The tests were conducted internally by the manufacturer or by contracted labs. The manufacturer is Amazing Rubber Products Pvt. Ltd. located in Kerala, India. The data is retrospective in the sense that it was collected as part of the device development and verification process to support the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable as the study is a bench test verification of physical and chemical properties against established industry standards (e.g., ASTM, ISO, USP). Ground truth in this context is defined by these measurable standards and the associated testing methodologies, not by expert interpretation of, for example, medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable as the study is a bench test verification. Adjudication methods are typically used in clinical studies, especially those involving human interpretation (e.g., radiology reads) to resolve discrepancies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic AI devices where human performance is being evaluated or improved with AI assistance. The device in question is a surgical glove, which is a physical product, not a diagnostic tool requiring human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, this is not applicable. The device is a physical product (surgical glove), not an algorithm or AI system. Its performance is assessed through bench testing against established performance standards.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for this device is defined by established international and national standards and specifications for surgical gloves (e.g., ASTM D3577-2019, ISO 10993 series, USP monographs). These standards specify the acceptable physical, chemical, and biological properties through defined test methods.

    8. The sample size for the training set

    • Not applicable. This question pertains to machine learning models. The device is a physical product, not an AI/ML model that requires training.

    9. How the ground truth for the training set was established

    • Not applicable. As above, this pertains to machine learning models.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212848
    Date Cleared
    2021-12-27

    (111 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Latex Surgical Gloves Powder Free is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The proposed device, Sterile Latex Surgical Gloves Powder Free with protein content labeling claim of 50 µg/ dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves. The classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device is Sterile Latex Surgical Gloves Powder Free, and variants of different sizes. All variants share the same natural color (No color is added). The proposed device is sterilized using Gamma Radiation method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 3 years.

    AI/ML Overview

    The provided text describes the acceptance criteria and study results for "Sterile Latex Surgical Gloves Powder Free." Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Specification as per ASTM D3577-2019)Reported Device Performance (Average value of subject device)Complied with Standard?
    Dimensions
    Length (all sizes)Min 265mm383mm (Size 6), 385mm (Size 6.5), 388mm (Size 7), 390mm (Size 7.5), 392mm (Size 8), 394mm (Size 8.5), 396mm (Size 9)Yes
    Width (Size 6)76+/-6mm74mmYes
    Width (Size 6.5)83+/-6mm86mmYes
    Width (Size 7)89+/-6mm92mmYes
    Width (Size 7.5)95+/-6mm98mmYes
    Width (Size 8)102+/-6mm105mmYes
    Width (Size 8.5)108+/-6mm110mmYes
    Width (Size 9)114+/-6mm116mmYes
    Finger Thickness (all sizes)Min 0.10mm0.21mmYes
    Palm Thickness (all sizes)Min 0.10mm0.16mmYes
    Cuff Thickness (all sizes)Min 0.10mm0.12mmYes
    Tensile Strength
    Before aging (all sizes)24Mpa minimum28.55MpaYes
    After aging @ 70°±2C for 166±2 hr (all sizes)18Mpa minimum23.48MpaYes
    Ultimate Elongation
    Before aging (all sizes)750% minimum870%Yes
    After aging @ 70°±2C for 166±2 hr (all sizes)560% minimum731%Yes
    Stress at 500%
    Before aging (all sizes)5.5 MPa Max5.1 MpaYes
    Pinhole AQL
    Before aging (all sizes)Max 1.51.0Yes
    After aging @ 70°C for 7 days (all sizes)Max 1.51.0Yes
    Powder Residue< 2 mg/Glove0.40 mg/GloveYes (Implied, as 0.40 < 2)
    Protein Content≤ 50 µg/ dm²50 µg/ dm²Yes
    Biocompatibility - Skin IrritationUnder the condition of study not an irritantUnder the condition of study not an irritantYes
    Biocompatibility - Skin SensitizationUnder the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizerYes
    Biocompatibility - In vitro CytotoxicityUnder the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.No (Did not meet "non cytotoxic" criteria, but similarity to predicate allowed)
    Biocompatibility - Systemic ToxicityUnder the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concernYes
    Bacterial Endotoxin<20EU/pair of gloves<20EU/pair of glovesYes

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific test (e.g., how many gloves were tested for length, tensile strength, etc.). It refers to meeting the "acceptance criteria of the standard" (ASTM D3577-2019, ASTM D5151-2019, ASTM D6124-06, ASTM D5712-15) which would define the sampling plans.

    The data provenance is not explicitly stated as "country of origin" for the testing or whether it was retrospective/prospective. However, the applicant is "Pentavest Holdings Sdn Bhd" located in Melaka, Malaysia, suggesting the testing was performed, or overseen, with connections to Malaysia. The testing, described as "Bench tests," typically refers to laboratory-based evaluations of the device itself rather than clinical data from human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to the submitted document. The "ground truth" for this medical device (Sterile Latex Surgical Gloves) is established through adherence to recognized international and national consensus standards (e.g., ASTM, ISO, USP) for physical, chemical, and biological properties, not by expert consensus in diagnostic imaging or clinical assessment. The standards themselves define the acceptable parameters.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This type of adjudication method is used in studies where human readers interpret data, such as medical images. For device performance testing against manufacturing and performance standards, the results are quantitative and compared directly to specified limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to the regulatory submission for surgical gloves, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document is for surgical gloves, not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is based on established consensus standards (ASTM, ISO, USP). These standards define the acceptable physical, chemical, and biological properties and performance characteristics for rubber surgical gloves. Compliance with these standards serves as the evidence that the device is safe and effective for its intended use.

    8. The sample size for the training set

    Not applicable. This is not an AI or machine learning device; therefore, there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set, there is no ground truth for it to be established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211478
    Date Cleared
    2021-09-16

    (127 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Sterile Latex Surgical Gloves Powder Free is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The proposed device, Sterile Latex Surgical Gloves Powder Free is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D3577 -09(2015), Standard Specification for Rubber Surgical Gloves. The classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device is Sterile Latex Surgical Gloves Powder Free, and variants of different sizes. All variants share the same natural color (No color is added). The proposed device is sterilized using Gamma method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 3 years.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device: "Sterile Latex Surgical Gloves powder free." This document focuses on demonstrating substantial equivalence to an existing predicate device, rather than proving the device meets clinical acceptance criteria for an AI/ML medical device. Therefore, many of the requested points related to AI/ML study design (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth for training set) are not applicable to this type of submission.

    However, the document does contain information about performance testing against established standards, which serve as the acceptance criteria for this specific device.

    Here's an analysis of the provided information, addressing the relevant points and noting those that are not applicable:


    Device Name: Sterile Latex Surgical Gloves powder free
    Submission Type: 510(k) Premarket Notification (for substantial equivalence)


    1. A table of acceptance criteria and the reported device performance

    The document provides extensive tables comparing the subject device's performance to the predicate device and against the specifications outlined in relevant ASTM standards. These standards define the "acceptance criteria" for the physical and chemical properties of surgical gloves.

    Summary Table of Acceptance Criteria and Reported Performance (Selected Key Parameters)

    Criteria CategoryPerformance Criterion Source (Standard/Section)Acceptance CriteriaReported Device Performance (K211478)Proof of Meeting Criteria
    Physical DimensionsASTM D3577-09 (Reapproved 2015)Min 265mm (Length, all sizes)385mm - 396mm (All sizes)Met
    ASTM D3577-09 (Reapproved 2015)Various +/- 6mm (Width, per size)74mm - 116mm (All sizes)Met
    ASTM D3577-09 (Reapproved 2015)Min 0.10mm (Finger Thickness, all sizes)0.20mmMet
    ASTM D3577-09 (Reapproved 2015)Min 0.10mm (Palm Thickness, all sizes)0.15mmMet
    ASTM D3577-09 (Reapproved 2015)Min 0.10mm (Cuff Thickness, all sizes)0.12mmMet
    Physical PropertiesASTM D3577-09 (Reapproved 2015)24 Mpa minimum (Tensile Strength, Before Aging)27.2 Mpa - 29.7 MpaMet
    ASTM D3577-09 (Reapproved 2015)18 Mpa minimum (Tensile Strength, After Aging)22.1 Mpa - 24.5 MpaMet
    ASTM D3577-09 (Reapproved 2015)750% minimum (Ultimate Elongation, Before Aging)864% - 874%Met
    ASTM D3577-09 (Reapproved 2015)560% minimum (Ultimate Elongation, After Aging)727% - 734%Met
    ASTM D3577-09 (Reapproved 2015)5.5 MPa Max (Stress at 500% before ageing)5.1 Mpa (Average Value)Met
    Barrier IntegrityASTM D3577-09 & ASTM D5151-06Max 1.5 AQL (Pinhole AQL, Before & After Aging)1.0 AQL (Gloves Passes)Met
    Chemical ResiduesASTM D6124-06$\le$ 2 mg/Glove (Powder Residue)0.26 mg/glove - 0.62 mg/glove (All sizes)Met
    ASTM D3577 & ASTM D5712-15$\le$ 50µg/dm² (Protein Content)38 µg/dm²Met
    BiocompatibilityISO 10993-10Non-irritantNon-irritantMet
    ISO 10993-10Non-sensitizerNon-sensitizerMet
    ISO 10993-5:2009Non-cytotoxicCytotoxicNOT Met (Qualified)
    ISO 10993-11:2017No systemic toxicity concernNo systemic toxicity concernMet
    ISO 10993-11:2017 / USP 41<151>No material mediated pyrogenicity responseNo material mediated pyrogenicity responseMet
    USP 42<85>NMT 20 EU/pair of gloves (Bacterial Endotoxin)<20 EU/pair of glovesMet

    Note on Biocompatibility (Cytotoxicity): The device reported "Cytotoxic" against an acceptance criteria of "non-cytotoxic." However, the comparison table in Section G.c notes that the predicate device (K192328) also tested "Cytotoxic" for this parameter. This indicates that while the specific test result might seem like a failure in isolation, in the context of demonstrating substantial equivalence to a legally marketed predicate, it is acceptable if the predicate device exhibits the same characteristic and is already on the market. The FDA would have cleared the predicate with this characteristic, implying it's acceptable for surgical gloves.


    Study that proves the device meets the acceptance criteria:

    The document describes non-clinical bench testing conducted to verify that the proposed device meets design specifications and acceptance criteria outlined in various industry standards (primarily ASTM and ISO).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the sample size for each test performed. However, the tests are performed according to recognized international standards (ASTM, ISO, USP), which typically specify minimum sample sizes for reliable results. For example, ASTM D5151-06 (Detection of Holes in Medical Gloves) employs an Acceptable Quality Level (AQL) methodology, implying a statistically determined sample size tested for pinholes.
    • Data Provenance: The tests were conducted to support a 510(k) submission for a manufacturer in Malaysia (Aspen Glove Sdn. Bhd.). The data itself is "bench test" data. It is not patient data, so "retrospective or prospective" does not apply in the typical sense of a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is not a study involving expert readers or establishment of ground truth by human experts, as it is a physical device (surgical glove) and not an AI/ML diagnostic tool. The "ground truth" here is defined by the technical specifications and test methods within the referenced ASTM and ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. This is a physical device performance study, not an imaging or diagnostic study requiring adjudication of expert opinions.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the established technical and performance specifications defined by international consensus standards (ASTM D3577-09, ASTM D5151-06, ASTM D6124-06, ISO 10993 series, USP). These standards outline the physical and chemical properties and safety requirements for such medical devices, and the tests verify compliance with these objective measures.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/ML device and thus does not have a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for this type of device, ground truth establishment for a training set is irrelevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190241
    Date Cleared
    2019-12-30

    (327 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A sterile latex surgical glove, powder free is a disposable device intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The subject device is a powder-free variation of the class I latex surgical gloves made by on-line polymercoating process. This process modifies the surface characteristics and causes it to remain tack-free without the use of any dusting or donning powder.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device: "Sterile Latex Surgical Gloves, Powder Free." It details the device's characteristics, comparison to a predicate device, and the results of non-clinical testing. It explicitly states that clinical testing was "Not applicable." Therefore, the document does not contain information about a study proving the device meets acceptance criteria related to a diagnostic or AI-based medical device performance study.

    The questions in the prompt relate to the evaluation of an AI or diagnostic device, asking about sample sizes, expert ground truth, MRMC studies, and training sets. This information is not relevant to a 510(k) submission for surgical gloves, which are assessed based on physical properties, sterility, and biocompatibility.

    However, I can extract the acceptance criteria and reported device performance for the surgical gloves based on the provided tables.

    Here's the information derived from the document, tailored to the context of surgical gloves:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting the specifications of ASTM D3577-15 and other relevant standards (ISO 11737-2, ISO 10993-10, ISO 10993-5, ISO 10993-11, USP General Chapter 40). The reported device performance indicates whether the subject device "Meets" or "Passes" these specifications.

    CharacteristicStandardSpecificationReported Device Performance (Subject Device K190241)
    Dimension (Length)ASTM D3577-15
    Length (size: 5.5), mm245 minMeet 245mm min
    Length (size: 6.0), mm265 minMeet 265mm min
    Length (size: 6.5), mm265 minMeet 265mm min
    Length (size: 7.0), mm265 minMeet 265mm min
    Length (size: 7.5), mm265 minMeet 265mm min
    Length (size: 8.0), mm265 minMeet 265mm min
    Length (size: 8.5), mm265 minMeet 265mm min
    Length (size: 9.0), mm265 minMeet 265mm min
    Dimension (Thickness)ASTM D3577-15
    Thickness (cuff), mm0.10 minMeet 0.10mm min
    Thickness (palm), mm0.10 minMeet 0.10mm min
    Thickness (finger), mm0.10 minMeet 0.10mm min
    Dimension (Width)ASTM D3577-15
    Width (size: 5.5), mm70 ± 6Meet 70 ± 6 mm
    Width (size: 6.0), mm76 ± 6Meet 76 ± 6 mm
    Width (size: 6.5), mm83 ± 6Meet 83 ± 6 mm
    Width (size: 7.0), mm89 ± 6Meet 89 ± 6 mm
    Width (size: 7.5), mm95 ± 6Meet 95 ± 6 mm
    Width (size: 8.0), mm102 ± 6Meet 102 ± 6 mm
    Width (size: 8.5), mm108 ± 6Meet 108 ± 6 mm
    Width (size: 9.0), mm114 ± 6Meet 114 ± 6 mm
    Water Leak TestASTM D3577-15, ASTM D5151-06
    Before Aging, AQLG-I, AQL 1.5 (FDA GII, AQL 1.5)Meet AQL 1.5
    After Aging, AQLMeet AQL 1.5
    Physical Properties (Before Aging)ASTM D3577-15
    Tensile Strength (MPa)Min. 24Meet 24MPa min.
    Ultimate Elongation (%)Min. 750Meet 750% min
    Stress at 500% Elongation (MPa)Max. 5.5Meet 5.5 MPa max
    Physical Properties (After Aging)ASTM D3577-15
    Tensile Strength (MPa)Min. 18Meets 18MPa min
    Ultimate Elongation (%)Min. 560Meet 560% min.
    Residual Powder ContentASTM D3577-15, ASTM D6124-06Max. 2mg/gloveMeet 2mg/glove max.
    Extractable Protein ContentASTM D5712-1050 µg/dm² and belowMeet Specification
    Biocompatibility Tests
    Primary Skin Irritation TestISO 10993-10No Animal IrritationPasses (did not cause an irritant response)
    Skin Sensitization TestISO 10993-10No Animal SensitizationPasses (did not produce a skin sensitization effect)
    Cytotoxicity TestISO 10995-5Non-cytotoxicitymoderate-cytotoxicity at 6 cm²/mL extraction
    Acute Toxicity OralISO 10993-11No toxicNo adverse biological reaction
    Pyrogenic TestUSP General Chapter 40Non pyrogenicNon pyrogenic

    Regarding the specific questions about an AI/diagnostic device evaluation:

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
    This information is not applicable. The device is a surgical glove, not a diagnostic or AI device. The testing involved physical and chemical properties and biocompatibility, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
    This information is not applicable. Ground truth as understood in AI/diagnostic studies (e.g., disease presence) is not relevant for surgical gloves. The "ground truth" for the gloves are the established international standards for their physical, chemical, and biological properties.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    This information is not applicable. Adjudication methods are used in interpreting ambiguous diagnostic findings, which is not relevant for surgical gloves.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not applicable. MRMC studies are used to evaluate human reader performance with and without AI assistance in diagnostic tasks. This device is a surgical glove.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    This information is not applicable. Standalone algorithm performance is relevant for AI devices. This device is a physical product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    For surgical gloves, the "ground truth" is defined by adherence to recognized international standards (e.g., ASTM, ISO, USP) for physical properties, sterility, and biocompatibility. This involves laboratory testing using defined methodologies, not clinical outcomes data or expert diagnostic consensus.

    8. The sample size for the training set:
    This information is not applicable. Training sets are used for machine learning. This document describes testing of manufactured physical products, not an AI model.

    9. How the ground truth for the training set was established:
    This information is not applicable. There is no training set for a surgical glove.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171550
    Date Cleared
    2018-07-13

    (413 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterile Latex Surgical Gloves, Powder free, is a disposable device made of natural rubber; intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The subject device is a surgical glove made of natural rubber latex compound. It is sterile, Powder free. The device is anatomic and is hand specific. The device meets ASTM D3577 – 09 : Standard specification for Rubber Surgical Gloves. The device is for over-the counter single use.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Sterile Latex Surgical Gloves, Powderfree. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel clinical effectiveness or performance for an AI/ML-based device.

    Therefore, the requested information regarding acceptance criteria and study details for an AI/ML device (e.g., sample size for test sets, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) is not applicable to this document.

    The document primarily addresses the following for medical devices like surgical gloves:

    • Acceptance Criteria & Device Performance: This section is presented as a comparison table against the predicate device and relevant ASTM/ISO standards.
    • Non-Clinical Testing: The study proves the device meets acceptance criteria through various standardized tests.

    Here's the relevant information extracted from the document concerning the acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicsTest StandardAcceptance Criteria (from Standard)Reported Device Performance (Sterile Latex Surgical Gloves, Powderfree)Result
    Freedom from HolesASTM D3577-09 / ASTM D5151-06 (Re-approved 2011) / FDA 1000ml water leak testSampling plan/Inspection level/AQL: ISO 2859-1 / G1/AQL 2.5 (for FDA 1000 ml water leak test)PASS (for FDA 1000ml water leak test)PASS
    Dimensions: LengthASTM D3577-09Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Overall length min 265 mm> 265 mmPASS
    Dimensions: WidthASTM D3577-09Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Varies from 76 mm for 6.0 size to 114 mm for 9.0 size (76±6 mm to 114±6 mm for sizes 6.0 to 9.0)76±6 mm to 114±6 mm (sizes 6.0 to 9.0)PASS
    Dimensions: ThicknessASTM D3577-09Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Minimum 0.10mm (Cuff, palm & finger)> 0.10 mm (Cuff, palm& finger)PASS
    Physical Properties: Before agingASTM D3577-09 and ASTM D412-06Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Tensile strength 24 Mpa min; Ultimate elongation 750% minTensile strength : > 24 Mpa ; Ultimate Elongation : >750%PASS
    Physical Properties: After Accelerated agingASTM D3577-09 and ASTM D573-04Sampling plan/Inspection level/AQL: ISO 2859-1 / S2/AQL 4.0 ; Tensile strength 18 Mpa min; Ultimate elongation 560 % minTensile strength : > 18 Mpa ; Ultimate Elongation : > 560%PASS
    Powder-free residueASTM D3577-09 and ASTM D6124-06Sampling plan: N=5 ; Less than 2 mg per gloveLess than 2 mg per glovePASS
    Biocompatibility: Primary skin irritationISO 10993-10Under the conditions of the study the device is not an irritantUnder the conditions of the study the device is not an irritantPASS
    Biocompatibility: Skin SensitizationISO 10993-10Under the conditions of the study the device is not a sensitizerUnder the conditions of the study the device is not a sensitizerPASS
    SterilityISO-11737-2SterileSterilePASS
    EO ResidueISO 10993-07:2008Maximum limit of EO Residue : 0.0025mg/deviceMeetsMeets
    ECH ResidueISO 10993-07:2008Maximum limit of ECH Residue : 0.0025mg/deviceMeetsMeets
    Expiration dating/Shelf lifeASTM D7160-05Not explicitly stated in criteria table, but mentioned as "Three years"Three yearsMeets
    SterilizationISO 11135-2014(E)Requirements for the development, validation and routine control of a sterilization process for medical devicesMeetsMeets

    2. Sample size used for the test set and the data provenance

    • Sample Sizes: Sample sizes are indicated within the "Sampling plan / Inspection level / AQL" columns in the performance data table (e.g., ISO 2859-1 / G1 / AQL 2.5, ISO 2859-1 / S2 / AQL 4.0, N=5 for powder-free residue). For the specific Freedom from Pinholes test, it notes: "Lot Size : 35001 – 150000 pair; Sample size : 200 pair".
    • Data Provenance: Not explicitly stated as "country of origin" or "retrospective/prospective" in the context of clinical studies, as this is laboratory/manufacturing performance data. The manufacturer is SANREA HEALTHCARE PRODUCTS PVT LTD, located in Palakkad, Kerala, India.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable: This type of testing involves laboratory and physical property measurements against established international standards (ASTM, ISO, FDA). "Ground truth" in this context is defined by the technical specifications and methodologies of these standards, not expert adjudication of medical images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable: As above, this pertains to physical and chemical testing against defined standard metrics.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable: This is a premarket notification for a Class I medical device (surgical gloves), not an AI/ML-based diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable: This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standardized Test Methods and Specifications: The "ground truth" for the performance of these gloves is defined by the requirements and test methodologies outlined in the referenced ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards (e.g., ASTM D3577-09 for rubber surgical gloves, ASTM D5151-06 for freedom from holes, ISO 10993 for biocompatibility, ISO 11135 for sterilization). The FDA 1000ml water leak test is also mentioned.

    8. The sample size for the training set

    • Not Applicable: There is no "training set" as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not Applicable: There is no "training set" as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K123628
    Date Cleared
    2013-03-18

    (115 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. The lubricating or dusting powder used in the glove is excluded.

    Device Description

    FINESSIS PERSONAL Latex Powder Free Sterile Surgical Gloves with Protein Claim.

    AI/ML Overview

    This document is related to the FDA's 510(k) premarket notification for "FINESSIS PERSONAL Latex Powder Free Sterile Gloves with Protein Claim." This is a regulatory clearance document for a medical device (surgical gloves), not a study report for an AI/ML powered medical device.

    Therefore, the requested information about acceptance criteria, study details, sample sizes, expert qualifications, and ground truth establishment is not available in the provided text because the text describes the regulatory clearance of surgical gloves, not an AI/ML device and its performance study.

    The document discusses:

    • The trade/device name: FINESSIS PERSONAL Latex Powder Free Sterile Gloves with Protein Claim.
    • Regulation Number and Name: 21 CFR 878.4460, Surgeon's Glove.
    • Regulatory Class: I.
    • Product Code: KGO.
    • Indications for Use: A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. The lubricating or dusting powder used in the glove is excluded.

    Questions regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth are typically asked for AI/ML diagnostic or prognostic devices that perform specific analyses on data (e.g., images, physiological signals), where "performance" relates to diagnostic accuracy (sensitivity, specificity, AUC) or similar metrics. This document does not describe such a device or study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011066
    Date Cleared
    2001-09-11

    (155 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1