Search Results
Found 34 results
510(k) Data Aggregation
(267 days)
The Bausch + Lomb Stellaris Elite™ vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/aspiration, bipolar coagulation, vitrectomy, viscous fluid injection/removal and air/fluid exchange operations.
The Bausch + Lomb Stellaris Elite™ vision enhancement system configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
The Bausch + Lomb Anterior Vitrectomy Cutter Packs are intended for use with a Bausch + Lomb Microsurgical System for vitrectomy during anterior segment surgery. The Posterior Vitrectomy Cutter Packs are intended for use with a Bausch + Lomb Microsurgical System for vitrectomy during posterior segment surgery. The Combined Vitrectomy Cutter Packs are intended for use with a Bausch + Lomb Microsurgical System for both vitrectomy during posterior segment surgery and for phacoemulsification of an opacified crystalline lens during anterior segment surgery.
The Bausch + Lomb Stellaris Elite™ vision enhancement system is comprised of an integrated ophthalmic microsurgical system designed for use in anterior and posterior segment surgery including phacofragmentation, vitreous aspirating, cutting, and endoillumination. Additionally, the Stellaris Elite™ vision enhancement system may be configured with a laser module intended for retinal photocoagulation and laser trabeculoplasty.
A selection of disposable single-use procedure packs is available for use with the system. These packs contain the necessary tubing to facilitate delivery and removal of air and fluids to and from the patient as well as a selection of components (cannulas, cutters, probes, drapes, etc.) that facilitate the surgical procedure. The items are arranged for physician convenience and may be presented as a group intended to support all the needs of a procedure or packaged singularly to allow the physician greatest flexibility.
The Bi-Blade™+ vitrectomy cutters are powered by a pneumatic pulse that operates the guillotine style cutter. The cutter has dual side ports and an aspiration channel that allows for aspiration during surgery. Each Bi-Blade™+ vitrectomy cutter is packaged and sterilized by gamma radiation.
N/A
Ask a specific question about this device
(30 days)
The Lumenis Stellar M22 system has connection capability with the following available treatment modules, for multi-application treatment options. All modules are designed for aesthetic and dermatological skin procedure application, as follows:
-
IPL with a spectrum of 400-1200nm (with different filters) is indicated for:
- Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma and ephelides (freckles).
- Cutaneous lesions, including warts, scars and striae.
- Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
- Removal of unwanted hair from all skin types, and to effect stable long term, or permanent hair reduction* in skin types I-V through selective targeting of melanin in hair follicles.
- Mild to moderate inflammatory Acne (Acne Vulgaris).
-
Multi-Spot Nd:YAG Laser, with a wavelength of 1064 nm is indicated for:
- The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1 - 4.0 mm diameter) of the leg.
- The non-ablative treatment of facial wrinkles.
- The removal of unwanted hair from all skin types, and to effect stable long term, or permanent hair reduction* in skin types I-V through selective targeting of melanin in hair follicles.
-
The 1565nm ResurFX module, with a wavelength of 1565 nm, is indicated for:
- Use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue.
-
Q-Switched Nd:YAG Laser, with a wavelength of 1064 nm, is indicated for:
- Treatment of pigmented lesions.
- Removal of dark tattoos.
(*) Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regimen.
The Lumenis Stellar M22 System is multi-application, multi-technology system intended for use in aesthetic dermatologic procedures.
The Stellar M22 System operates with four (4) treatment handpieces options (for connection capability), within serial operational mode:
- Intense Pulsed Light (IPL) handpiece;
- Multi-Spot Nd:YAG Laser handpiece;
- ResurFX Laser module and handpiece;
- Q-Switched Nd:YAG laser handpiece.
The following accessories are provided with and/or may be purchased independently for each of the four (4) available treatment handpieces of the Stellar M22 for Intense Pulsed Light (IPL) and Laser Systems:
- The Stellar IPL handpiece has nine (9) different filters available: Cut-off filters of 515, 560, 590,615, 640, 695 and 755 nm, Notch Filters of 400-600 & 800-1200 nm and 530650& 900-1200 nm. Further, the IPL handpiece has three (3) Sapphire Cool LightGuides available with sizes of: 15mm x 35mm, 8mm x 15 mm, 6 mm diameter.
- The Multi-Spot Nd:YAG handpiece has three (3) different LightGuides available in sizes of: 2mm x 4mm, 6 mm, and 9 mm.
- The ResurFX handpiece has two (2) different treatment tips available: SapphireCool and Precision tips
- The Q-Switched Nd:YAG handpiece has both disposable and gold-plated metal treatment tips available. The disposable treatment tips are available in four (4) different sizes of: 2, 2.5, 3.5, and 5 mm, while the metal treatment tips are available in seven (7) different sizes of: 2, 2.5, 3.5, 4, 5, 6 and 8 mm.
This FDA 510(k) clearance letter for the Stellar M22 device does not contain any information about acceptance criteria or a study proving the device meets said criteria for an AI/algorithm-driven component.
The document explicitly states:
- "No new clinical validation was required to support the Stellar M22 as the clinical validation of the Lumenis predicate device also applies to the subject device." This indicates that no new clinical study was performed for this specific submission to demonstrate performance.
- The modifications described are "mainly due to the addition of new Hardware to serve marketing, business and serviced purposes," and "to maintain Lumenis' high-level service capabilities to the professional end-users and their patients, through establishment of a secure transmission of technical information to the cloud for further analysis and rapid problem-solving." This strongly suggests the changes are related to connectivity and serviceability, not a new AI/algorithm feature requiring performance validation.
Therefore, I cannot provide the requested information from the provided text. The document describes a device modification for an existing system (Stellar M22) and asserts substantial equivalence to its predicate based on the fact that the changes are primarily hardware upgrades for service and marketing, and the clinical validation of the predicate device is considered applicable.
There is no mention of any AI component, performance criteria for such a component, or a study validating its performance in this submission.
Ask a specific question about this device
(69 days)
Stellar Imaging System is intended to provide real-time visible and near-infrared imaging in minimally invasive procedures.
Upon intravenous administration and use of an ICG agent consistent with its approved label, Stellar Imaging System is used to perform fluorescence imaging and visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts.
Upon interstitial administration and use of an ICG agent consistent with its approved label, the Stellar Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Stellar Imaging System comprises Image Processing Unit, Light Source, Camera Head, Laparoscope, Exoscope and relevant accessories.
During surgical procedures, Stellar Imaging System is used to provide real-time visible and fluorescence imaging similar to that provided by conventional imaging system used in endoscopic surgery and innovative 3D visible and fluorescence imaging in endoscopic surgery. Besides, Stellar Imaging System is also used to provide real-time fluorescence confirmation in open field.
The Stellar Imaging System including the laparoscope and the exoscope is designed to work with an approved infrared dye (Indocyanine green (ICG). ICG may be excited at excitation at either 785 or 805 nm. The System provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye using a camera head. Fluorescence imaging with ICG permits the system to visualize blood flow and related tissue circulation, of lymphatic flow, etc.
The System allows the capture of normal (white) light image in parallel with the fluorescence image and displays both to the surgeon to provide a view of the anatomy. In addition, the System permits recording surgical procedures, storing them on removable storage devices, and replaying the procedures.
The provided text is a 510(k) Summary for the Stellar Imaging System. It focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and safety standards. It does not contain information about the acceptance criteria for a study, the results of such a study, sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, or MRMC comparative effectiveness studies.
The section titled "VII. Performance data" only lists the non-clinical tests conducted to verify compliance with various safety and performance standards. It states: "Performance comparison testing are also conducted on the subject device and predicate/reference device, demonstrate that the proposed system performs according to specifications and functions as intended." However, no specific details or results of this performance comparison testing are provided within this document.
Therefore, I cannot fulfill your request for the specific information you asked for, as it is not present in the provided text.
Ask a specific question about this device
(186 days)
Ask a specific question about this device
(228 days)
The Bausch + Lomb Stellaris Elite™ vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/ aspiration, bipolar coagulation, vitrectomy, viscous fluid injection/removal and air/fluid exchange operations. The Bausch + Lomb Stellaris Elite™ vision enhancement system configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
The Bausch + Lomb Stellaris Elite™ vision enhancement system is comprised of an integrated ophthalmic microsurgical system designed for use in anterior and posterior segment surgery including phacofragmentation and vitreous aspirating and cutting as well as endoillumination. Additionally, the Stellaris Elite™ vision enhancement system may be configured with a 532 nm laser module for photocoagulation. The Stellaris Elite™ vision enhancement system was initially cleared under K 162342 which included the introduction of an Adaptive Fluidics feature as well as increase in vitrectomy cut rate. The Vitesse vitrectomy handpiece was introduced and cleared for use on the Stellaris Elite™ vision enhancement system under K170052.
A selection of disposable single-use procedure packs is available for use with this system. These packs contain the necessary tubing to facilitate delivery and removal of air and fluids to/from the patient as well as a selection of components (cannulas, cutters, probes, etc.) that facilitate the surgical procedure. The items are arranged for physician convenience and may be presented as a group intended to support all the needs of a procedure or packaged singularly to allow the physician greatest flexibility. Additional Stellaris Elite™ vision enhancement system procedure packs have been made available and these packs contain components that have been used in previously available procedure packs. Some of these Stellaris Elite™ vision enhancement system procedure packs now include the BiBlade vitrectomy cutter produced by Medical Instruments Development Lab (separately cleared via K153168).
The previously cleared Stellaris Elite™ vision enhancement system introduced new feature sets that are not available on the legacy Stellaris PC systems previously cleared under K133242/K133486 respectively. A summary of these additional feature sets for the Stellaris Elite™ vision enhancement system are listed in Table 8-1. The various available system configurations allow for system feature flexibility and are made available for physician convenience.
The most recent clearance to the Stellaris Elite™ vision enhancement system introduced the Vitesse vitrectomy feature and was cleared via K170052 on April 19, 2017.
This traditional 510(k) incorporates updates to User Interface Computer System on Module (SOM) with an introduction of a new replacement Congatec SOM. In addition. the Stellaris Software platform is updated to Windows 10, since the current Stellaris software platform Windows XP is obsolete/incompatible with the new Congatec SOM.
The features remain the same as the previously approved Stellaris Elite™ vision enhancement system. The configuration matrix is listed in Table 8-1. The various available system configurations allow for system feature flexibility and are made available for physician convenience.
The provided document is a 510(k) Summary for the Bausch + Lomb Stellaris Elite™ vision enhancement system. It primarily focuses on demonstrating substantial equivalence to a predicate device after software and hardware updates, rather than presenting a study proving a device meets specific acceptance criteria for performance.
Therefore, many of the requested categories cannot be fully addressed in the traditional sense of a clinical or performance study aimed at defining and meeting specific analytical or clinical performance metrics. The document describes how substantial equivalence was demonstrated, which involves various types of testing, but not a study designed to prove new performance claims against defined acceptance criteria.
Below is the information extracted and interpreted based on the context of this 510(k) submission, highlighting what is available and what is not.
Acceptance Criteria and Device Performance Study for K232084
This 510(k) submission (K232084) is for updates to the Bausch + Lomb Stellaris Elite™ vision enhancement system, not for a new device making novel performance claims. The primary goal is to demonstrate "substantial equivalence" to a legally marketed predicate device (K170052 Stellaris Elite™ vision enhancement system) after changes to the User Interface Computer System on Module (SOM) and an update to the operating system from Windows XP to Windows 10. Consequently, the "acceptance criteria" and "device performance" are framed in terms of meeting regulatory standards and maintaining performance equivalent to the predicate device, rather than achieving specific performance thresholds for a novel diagnostic or therapeutic claim.
1. A table of acceptance criteria and the reported device performance
Given the nature of this 510(k) for system updates, the "acceptance criteria" are regulatory compliance and maintained functionality, rather than specific numerical performance metrics for a clinical task.
| Acceptance Criteria Category | Reported Device Performance / Assessment |
|---|---|
| Biocompatibility | Not required; device does not contain direct or indirect patient-contacting materials. |
| Electrical Safety | Complies with IEC 60601-1:2020, IEC 60601-1-2:2020, IEC 60601-1-6:2020, IEC 60601-2-2:2017, IEC 60601-2-22:2019, IEC 80601-2-58:2016. |
| Electromagnetic Compatibility (EMC) | Complies with IEC 60601-1-2:2020. |
| Software Verification & Validation | Functional, simulated use, environmental, and transport testing performed. Software changes verified and validated per Bausch + Lomb software quality procedures, complying with EN ISO IEC 623304:2006. All testing passed. |
| Mechanical Testing | No specific mechanical testing performed to support this premarket notification for substantial equivalence (implied that existing mechanical performance is assumed to be equivalent). |
| Acoustical Testing | No specific acoustical testing performed to support this premarket notification for substantial equivalence. |
| Overall Functional Equivalence | The Stellaris Elite™ vision enhancement system features remain the same as the previously approved system and are described as substantially equivalent to the predicate (K170052). |
| Maximum vacuum | Subject Device: 660 mmHg. Predicate (K170052): 600 mmHg (This is listed as an "incremental improvement" since the predicate of K170052, indicating a specific performance change that was assessed). |
| Operating System | Subject Device: Windows 10. Predicate (K170052): Windows XP. (Functional equivalence demonstrated via software V&V). |
| User Interface SOM | Subject Device: Congatec. Predicate (K170052): Kontron. (Functional equivalence demonstrated via software V&V and system testing). |
2. Sample size used for the test set and the data provenance
The document does not specify a "sample size" in the context of a dataset for AI or clinical study. Instead, it refers to "representative units" for hardware testing (electrical safety, EMC) and software verification/validation.
- Test Set Sample Size: Not applicable in the context of an AI-driven clinical test set. Testing was performed on the device itself and its software.
- Data Provenance: Not applicable as no patient data or image data was used for a test set. This submission is for hardware/software updates to an ophthalmic surgical system. The standards cited (IEC standards for electrical safety, EMC, software lifecycle) are international regulatory standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. No "ground truth" for a test set (e.g., diagnostic labels, disease states) was established by experts in this submission, as it's not a diagnostic AI device.
4. Adjudication method for the test set
Not applicable. There was no test set requiring expert adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an ophthalmic surgical system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable in the context of a clinical performance study. For software and hardware validation, the "ground truth" would be the functional requirements and established performance specifications of the predicate device and relevant international standards.
8. The sample size for the training set
Not applicable. This device does not use a "training set" in the sense of machine learning for interpretation or diagnosis.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(104 days)
STELLAR Knee is a stereotaxic system including an intraoperative software as a medical device and surgical instruments. STELLAR Knee is indicated for primary Total Knee Replacement, to assist the surgeon in determining reference alignment axes in relation to anatomical landmarks, in order to position the cutting guide regarding computed mechanical axis. The STELLAR Knee is installed on a Head Mounted Device (HMD) for displaying information to the user intraoperatively. The HMD should not be relied upon solely and should always be used in conjunction with traditional methods.
The main purpose of the STELLAR Knee is to assist the surgeon during the primary Total Knee Replacement (TKR) intervention. STELLAR Knee includes software and surgical instruments.
STELLAR Knee uses established surgical navigation techniques to provide information to help track patient bony landmarks in real time to assist the surgeon in determining resection angles and measurements as required in knee replacement surgery. STELLAR Knee allows the surgeon to adjust the cutting plane orientation and the resection level. This includes means for the surgeon to collect anatomical references during the TKR intervention using the surgical instruments. The software locates in a 3D reference frame the instruments which include marker arrays. All collected coordinates are treated by software algorithms to provide the surgeon with the relevant orientation of the tracked cutting quide. STELLAR Knee software is installed on a wearable Head Mounted Device (HMD) which includes an embedded camera and displays intraoperative information to the user. This near-eye display allows the surgeon to look at the HMD screen or the field of view when needed.
The provided text describes a 510(k) premarket notification for the STELLAR Knee device, which is a stereotaxic system for Total Knee Replacement. However, the document does not contain the detailed acceptance criteria or the study data that proves the device meets those criteria.
The "Non-clinical performance testing" section briefly mentions that "Bench testing was conducted in order to demonstrate that STELLAR Knee performs according to its requirements and specifications when installed on the Head Mounted Device. In particular, overall system repeatability and accuracy were tested." It also states, "The results demonstrated that the STELLAR Knee performs according to its specifications and functions as intended."
However, information regarding specific quantitative acceptance criteria (e.g., accuracy thresholds in mm or degrees), the reported device performance against these criteria, sample sizes used, data provenance, ground truth establishment, or any comparative effectiveness studies (MRMC) is not provided in this document.
Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance, nor the details of any study that proves the device meets specific acceptance criteria based on the given input. The document focuses on establishing substantial equivalence with a predicate device based on similar indications for use, operating principles, and general performance testing, without detailing the results of those performance tests in a quantitative manner against predefined acceptance criteria.
Ask a specific question about this device
(112 days)
Stellaris 2D: Extraoral source dental X-ray system intended to perform panoramic exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry.
Stellaris 2D Ceph: Extraoral source dental X-ray system intended to perform panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial regions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
Stellaris 3D: Extraoral source dental X-ray system intended to perform 3D and panoramic exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry.
Stellaris 3D Ceph: Extraoral source dental X-ray system intended to perform 3D, panoramic and cephalometric exams with production of diagnostic images in the dento-maxillo-facial regions, for general and pediatric dentistry, as well as carpal images for dental clinical investigations.
Stellaris 3D is a dental X-ray system for Panoramic and Cone Beam Computed Tomography (CBCT) which allows to perform all the radiographic projections, both 2D and 3D, of most interest for the dentist, the surgeon and the maxilla-facial radiologist.
In addition to the functions of Panoramic and 3D radiography, Stellaris 3D Ceph also allows to perform One-Shot Cephalometric radiographs.
Stellaris 2D is a dental X-ray system for Panoramic which allows to perform all the radiographic projections of most interest for the dentist, the surgeon and the maxilla-facial radiologist.
In addition to the functions of Panoramic, Stellaris 2D Ceph also allows to perform One-Shot Cephalometric radiographs.
Except for the panoramic acquisition sensor, the remaining mechanical, electrical and software characteristics of Stellaris 2D are exactly identical to those of Stellaris 3D.
The control panel interface on the unit provides a complete control of the operation and for the setting of the desired technique factors.
Class I LASER aiming lights support positioning of patient's head, which is stabilized through the use of bite blocks, chin rest, and, if required, temple supports.
The user controls the exposure using a manual hand-switch, implementing the dead man functionality.
An Ethernet connection cable allows the FONA Stellaris devices to interface with a computer for image acquisition, processing and storage.
The provided document is a 510(k) Premarket Notification from FONA S.r.l. to the FDA for their Stellaris dental X-ray systems. It primarily focuses on demonstrating substantial equivalence to predicate devices based on technical specifications and adherence to relevant standards.
The document does NOT describe the acceptance criteria for an AI/ML powered device, nor does it detail a study proving such a device meets acceptance criteria.
The Stellaris devices are dental X-ray systems (2D and 3D, with and without cephalometric capabilities). The document explicitly states:
- "The Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices can be used with the OrisWin DG Suite software for image acquisition and database (not part of this submission)..." (page 8)
- "...the 3D sensor has been replaced with a 2D sensor." (page 10)
- "The principle of operation of the Stellaris 2D, Stellaris 2D Ceph, Stellaris 3D, Stellaris 3D Ceph devices remains the same as the predicates. The difference is the type of sensors used which work on similar principles as the sensors found in the predicate devices." (page 7)
There is no mention of any AI or machine learning components within the device itself for diagnostic or image processing purposes that would require specific performance criteria or a study as outlined in the request. The "reconstruction algorithm" mentioned is standard for CT and panoramic imaging, not an AI/ML algorithm requiring separate clinical performance validation from the device's image acquisition capabilities.
Therefore, I cannot extract the requested information as it is not present in the provided text. The document describes a traditional 510(k) submission for an imaging device, demonstrating substantial equivalence through technical characteristics, safety standards (IEC, EN), software standards (EN 62304), and biocompatibility (ISO 10993-1).
In summary, none of the requested information regarding AI/ML acceptance criteria or a comparative study proving device performance against such criteria can be found in this document.
Ask a specific question about this device
(30 days)
The subject Stellar M22 has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
- The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for:
- Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and o ephelides (freckles)
- Cutaneous lesions, including warts, scars and striae O
- Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, O truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
- Removal of unwanted hair and to effect stable long term, or permanent* hair O reduction in skin types I-V through selective targeting of melanin in hair follicles
- Mild to moderate inflammatory Acne (Acne vulgaris) o
- The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is ● indicated for:
- The coagulation and hemostasis of vascular lesions and soft tissue, including the O treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg
- The removal of unwanted hair and to effect table long term, or permanent* hair O reduction in skin types I-V through selective targeting of melanin in hair follicles
- The non-ablative treatment of facial wrinkles o
- . ResurFX module and handpiece, with wavelength of 1565 nm, is indicated for:
- Use in dermatological procedures requiring fractional skin resurfacing and O
coagulation of soft tissue
- Use in dermatological procedures requiring fractional skin resurfacing and O
The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated ● for:
- Removal of dark tattoos O
- o Treatment of pigmented lesions
*Note
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of treatment regime.
The subject Stellar M22 for Intense Pulsed Light (IPL) and Laser System is a multi-application, multitechnology platform with four (4) available treatment handpieces:
- Stellar Intense Pulsed Light (IPL) handpiece; ●
- . Multi-Spot Nd:YAG laser handpiece;
- ResurFX non-ablative laser handpiece: ●
- Q-Switched Nd:YAG laser handpiece. ●
The following accessories are provided with and/or may be purchased independently for each of the four (4) available treatment handpieces of the Stellar M22 for Intense Pulsed Light (IPL) and Laser Systems:
- The Stellar IPL handpiece has nine (9) different filters available: Cut-off filters of 515, . 560, 590,615, 640, 695 and 755 nm, Notch Filters of 400-600 & 800-1200 nm and 530-650& 900-1200 nm. Further, the IPL handpiece has three (3) Sapphire Cool LightGuides available with sizes of: 15mm x 35mm, 8mm x 15 mm, 6 mm diameter.
- The Multi-Spot Nd:YAG handpiece has three (3) different LightGuides available in sizes . of: 2mm x 4mm, 6 mm, and 9 mm.
- The ResurFX handpiece has two (2) different treatment tips available: SapphireCool and Precision tips
- . The Q-Switched Nd:YAG handpiece has both disposable and gold plated metal treatment tips available. The disposable treatment tips and metal treatment tips are available in seven (7) different sizes of: 2, 2.5, 3.5, 4, 5, 6 and 8 mm.
The provided text is a 510(k) summary for the Stellar M22 for Intense Pulsed Light (IPL) and Laser system. This document outlines the device's intended use, technological characteristics, and compares it to a predicate device (M22 System, K170060) to establish substantial equivalence.
However, the summary does not include specific acceptance criteria or details of a study that proves the device meets those criteria in a quantitative sense as often seen in clinical performance studies with specific endpoints. Instead, it describes "Performance Data" in the context of design verification, compliance with standards, and software validation to demonstrate that modifications to the device do not raise new questions of safety or effectiveness compared to the predicate device.
Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) are not detailed in this type of regulatory submission, as it focuses on demonstrating substantial equivalence rather than proving a specific diagnostic accuracy or treatment efficacy de novo.
Below is a summary based on the information provided, with "N/A" for information not present in the document.
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present acceptance criteria in a quantitative table format that measures specific performance metrics against a defined threshold. Instead, it focuses on demonstrating that the Stellar M22 device's modifications do not introduce new safety or effectiveness concerns compared to its predicate. The "reported device performance" is framed in terms of functional verification and compliance with standards.
| Acceptance Criteria Category | Reported Device Performance/Compliance |
|---|---|
| Design Verification & Validation | Tests performed to validate modifications to the device (specifics not detailed) |
| Risk Analysis | Compliance with ISO 14971 |
| Electrical Safety | Compliance with IEC 60601-1 |
| Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 |
| IPL Compatibility | Compliance with IEC 60601-2-57 |
| Software Verification & Validation | Tests performed (specifics not detailed) |
| Technological Characteristics Equivalence | The device's technological characteristics (wavelength, pulse duration, maximum fluence, etc.) are substantially equivalent to the predicate, with minor modifications not raising new safety/effectiveness questions (see detailed comparison table in the source text). |
2. Sample size used for the test set and the data provenance
N/A - The document describes general performance data related to engineering verification and validation, and compliance with standards. It does not refer to a clinical "test set" with a specific sample size of patients or images for diagnostic or therapeutic performance evaluation. The data provenance is also not specified in terms of country of origin or retrospective/prospective nature, as this is not a clinical study report.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A - No clinical "test set" requiring ground truth established by experts is described in this regulatory submission for substantial equivalence.
4. Adjudication method for the test set
N/A - No clinical "test set" or adjudication method is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A - This document does not describe an MRMC comparative effectiveness study or involve AI assistance for human readers. The device is a laser and IPL system for dermatological procedures, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
N/A - This device is a medical instrument (laser and IPL system), not an algorithm or AI system. Therefore, standalone algorithm performance is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
N/A - As there is no clinical "test set" for performance evaluation in the context of diagnostic accuracy or treatment efficacy against a reference standard, the concept of ground truth in that sense is not detailed. The performance data focuses on engineering and safety compliance.
8. The sample size for the training set
N/A - The document does not describe a "training set" in the context of machine learning or AI algorithms.
9. How the ground truth for the training set was established
N/A - Not applicable, as there is no mention of an AI training set or related ground truth.
Ask a specific question about this device
(202 days)
The Stellar LEAP Manual Tilt Wheelchair is intended to provide mobility to persons restricted to a seated position.
The Stellar LEAP manual tilt wheelchair is an indoor / outdoor, manually operated, Tilt-in-Space wheelchair. Its intended function and use is to provide mobility to persons ages 16 and up (adolescents and adults). The Stellar LEAP manual tilt wheelchair consists of metal frame architecture with a backrest, seat, and rear wheels, which allow for occupant or attendant propelling of the device. Smaller caster wheels are also mounted on the front of the frame to facilitate steering and turning. The Stellar LEAP consists of three (3) metal frame subassemblies: a lower frame, upper seat frame, and sub-frame assembly. The manual Tilt-in-Space operation is achieved using a gas strut mechanism connected between the upper seat frame and sub-frame. The Stellar LEAP Manual Tilt Wheelchair is available in a standard configuration, which allows 20° Tilt-in-Space (posterior) tilt, with a maximum weight capacity of 250 lbs.
PDG offers optional features / options to be added to the Stellar LEAP wheelchair, including the two below features:
- Stellar LEAP Anterior Tilt Feature (0° 30°): Functional reach extension & transfer . assist
- Stellar LEAP Dynamic Recline Backrest Tilt Feature: 30° range of dynamic recline for the backrest combined with the anterior tilt
Other available options / accessories to complete the Stellar LEAP configuration include: seat width / depth / height, armrests, casters, frog leg suspension forks, rear wheel / tire, hand rim, wheel locks, anti-tippers, back post style, back options, back upholstery, headrest, tilt control, recline control, front rigging, foot plates, footrest / leg rest accessories, position straps, IV pole, 02 holder.
The provided text describes the 510(k) summary for the Stellar LEAP manual tilt wheelchair (K163432). It details the device, its intended use, comparison to predicate devices, and performance data. However, the requested information focuses on performance criteria and studies for AI/ML-driven devices, which is not applicable to this submission for a mechanical wheelchair.
Therefore, most of the specific points in the request cannot be answered from the provided document.
Here's a breakdown of what can be extracted and what cannot:
1. Table of acceptance criteria and the reported device performance:
The acceptance criteria for the Stellar LEAP wheelchair are defined by its compliance with various ISO 7176 standards. The reported performance is that the device "meets the requirements" or "performed compliant" to these standards.
| Acceptance Criteria (ISO Standard Requirement) | Reported Device Performance |
|---|---|
| ISO 7176 – Part 1 – Static stability | Meets requirements |
| ISO 7176 – Part 3 – Effectiveness of brakes | Meets requirements |
| ISO 7176 Part 5 – Overall dimensions, mass, turning space | Meets requirements |
| ISO 7176 Part 7 – Measurement of seating and wheel dimensions | Meets requirements |
| ISO 7176 Part 8 – Static impact and fatigue strengths | Meets requirements |
| ISO 7176 – Part 11 – Test dummies | Meets requirements |
| ISO 7176 – Part 13 – Coefficient of friction of test surfaces | Meets requirements |
| ISO 7176 – Part 15 – Information disclosure, documentation, labeling | Meets requirements |
| ISO 7176 – Part 16 – Resistance to ignition of postural support devices | Meets requirements |
| ISO 7176 – Part 19 – Wheeled mobility devices for use as seat in motor vehicles | Meets requirements |
| Biocompatibility (ISO 10993-1, FDA Guidance) | Biocompatibility assessment carried out and deemed safe |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable/not specified for a mechanical device. The testing described is based on engineering verification and validation according to international standards, typically involving a set number of units or prototypes to test specific characteristics.
- Data Provenance: Not applicable for this type of device. The provenance is the testing conducted by the manufacturer against established ISO standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. "Ground truth" in the context of AI/ML is not relevant here. Compliance with ISO standards is verified through defined test methods and measurements, not expert consensus on data interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This concept is for diagnostic performance assessment, not mechanical device testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is for AI/ML diagnostic devices.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for AI/ML systems.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable. The "ground truth" for a mechanical device is its adherence to specified engineering, safety, and performance standards as determined by standardized tests and measurements, as well as risk analysis and biocompatibility assessment.
8. The sample size for the training set
Not applicable. The Stellar LEAP is a mechanical medical device, not an AI/ML product that requires a training set for algorithm development.
9. How the ground truth for the training set was established
Not applicable. No training set is used for this device.
Ask a specific question about this device
(104 days)
The Bausch + Lomb Stellaris Elite vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/aspiration, bipolar coagulation, vitrectomy, viscous fluid injection/removal, and air/fluid exchange operations. The Stellaris Elite Vision Enhancement System configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
The Bausch + Lomb Stellaris Elite Vision Enhancement System is comprised of an integrated ophthalmic microsurgical system designed for use in anterior and posterior segment surgery including phacofragmentation and vitreous aspirating and cutting as well as endoillumination. Additionally, the Stellaris Elite Vision Enhancement System may be configured with a 532 nm laser module for photocoagulation.
The Stellaris Elite Vision Enhancement System is a rebranding of the currently cleared Stellaris (K133242) Vision Enhancement System and Stellaris PC (K133486) Vision Enhancement System. The system is based on the technology and the performance of the existing Stellaris and Stellaris PC Vision Enhancement Systems, and this traditional 510(k) incorporates software and hardware revisions to support the introduction of the new Vitesse vitrectomy feature (including 2 new Vitesse handpieces and supporting accessories). The new Vitesse vitrectomy feature is available on the Stellaris Elite Vision Enhancement System based on the configuration matrix listed in Table 1. The various available system configurations allow for system feature flexibility and are made available for physician convenience.
This document describes the Bausch + Lomb Stellaris Elite Vision Enhancement System, a device intended for surgical procedures related to cataracts and vitrectomy, and its substantial equivalence to previously cleared predicate devices.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Stellaris Elite Vision Enhancement System are based on compliance with various international electrical and medical device standards, and successful functional, simulated use, environmental, and transport testing. The reported device performance is that it met all these criteria.
| Acceptance Criteria (Standards and Testing Goals) | Reported Device Performance |
|---|---|
| Electrical Safety Standards: | Compliance Demonstrated: |
| IEC 60601-1:2005 + C1(2006) + C2(2007) + AM1(2012) or IEC 60601-1:2012 (General requirements for basic safety and essential performance) | Passed all referenced testing. |
| IEC 60601-1-2 ed3.0 (2007) (Electromagnetic compatibility) | Passed all referenced testing. |
| IEC 60601-1-6:2010 (Usability) | Passed all referenced testing. |
| IEC 60601-2-2:2009 (Particular requirements for high frequency surgical equipment) | Passed all referenced testing. |
| IEC 60601-2-22:2007 (Particular requirements for diagnostic and therapeutic laser equipment) | Passed all referenced testing. |
| IEC 80601-2-58:2008 (Particular requirements for lens removal devices and vitrectomy devices) | Passed all referenced testing. |
| Functional, Simulated Use, Environmental & Transport Testing: Ensure the device performs as intended in various conditions and scenarios. | Successfully completed on representative units, demonstrating compliance and substantial equivalence to predicate devices. Functional requirements have been met. |
| Software Verification and Validation: Compliance with EN ISO IEC 62304:2006 (Medical device software life cycle processes) | Software changes were verified and validated in accordance with Bausch + Lomb's software quality procedures, which comply with EN ISO IEC 62304:2006. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in terms of patient data or
clinical cases. The testing conducted was primarily hardware and software performance verification. The non-clinical tests involved "representative units" of the device. There is no information regarding country of origin of data or whether it was retrospective or prospective, as the testing described is not clinical in nature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable. The listed tests are engineering and compliance standards, not clinical studies requiring expert ground truth establishment for patient data. Expert involvement would be in the form of engineers, quality assurance personnel, and regulatory specialists ensuring compliance with the standards.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods are typically used in clinical studies when there's uncertainty about ground truth or disagreement among readers (e.g., radiologists). The testing described here is based on objective measurements against engineering and safety standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. The submission focuses on demonstrating substantial equivalence through technical characteristics and compliance with standards, not on comparing human reader performance with and without AI assistance. This device is not an AI-powered diagnostic tool, but a surgical system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This concept is not directly applicable to the Stellaris Elite Vision Enhancement System. As a surgical system, it is inherently designed for human-in-the-loop operation. The "standalone" performance being assessed relates to the system's compliance with engineering and safety standards as an independent medical device, not as an algorithm performing a task without human intervention. The software verification and validation would be considered "standalone" in the sense of the software itself being tested, but in the context of contributing to the overall device's safety and functionality.
7. The Type of Ground Truth Used
The "ground truth" for the device's performance is established by its adherence and successful execution to various international engineering and medical device safety standards (e.g., IEC 60601 series, EN ISO IEC 62304). These standards define the acceptable performance, safety, and operational parameters for such medical equipment.
8. The Sample Size for the Training Set
This information is not applicable. This document describes a medical device, not an AI model that relies on training sets of data. The software within the device undergoes verification and validation, but this is distinct from machine learning model training.
9. How the Ground Truth for the Training Set was Established
This information is not applicable as there is no mention of a "training set" for an AI model in the context of this device's submission.
Ask a specific question about this device
Page 1 of 4