Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K022746
    Date Cleared
    2002-11-22

    (95 days)

    Regulation Number
    862.1120
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stat Profile Critical Care Xpress Analyzer is intended for in vitro diagnostic use by health care professionals and for point-of-care usage in the quantitative determination of pH, PCO2, PO2%, Hematocrit (Hct), Ca++, total Hemoglobin (tHb), Oxyhemoglobin (O2Hb), Carboxyhemoglobin (COHb), Methemoglobin (MetHb), Reduced Hemoglobin (HHb), Oxygen content (O2Ct), and Oxygen capacity (02Cap) in heparinized whole blood; Nat, K+, C1-, Ca++, Mg++, Glucose (Glu), Lactate (Lac), BUN (Urea), and Creatinine (Creat) in heparinized whole blood, serum, or plasma.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) clearance letter from the FDA for the "Stat Profile Critical Care Xpress Analyzer." This device performs quantitative determinations of various analytes in blood. The document details the device's indications for use and clinical utility for each analyte.

    However, the letter does not contain the specific information requested regarding acceptance criteria and the study that proves the device meets those criteria.

    Specifically, the following information is missing from the provided text:

    1. A table of acceptance criteria and the reported device performance: The document only lists the analytes and their clinical utility, not specific performance metrics or acceptance criteria for accuracy, precision, etc.
    2. Sample size used for the test set and the data provenance: There is no mention of study design, sample sizes for test sets, or data origin (e.g., country, retrospective/prospective).
    3. Number of experts used to establish the ground truth for the test set and their qualifications: The document is a regulatory clearance letter, not a study report. It does not detail how ground truth was established for any performance testing.
    4. Adjudication method for the test set: No information on adjudication is present.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size: MRMC studies are typically for imaging devices with human interpretation components. This device is an in vitro diagnostic analyzer, and such studies are not applicable.
    6. If a standalone (algorithm only without human-in-the-loop performance) was done: This is an IVD analyzer; its performance is inherently "standalone" in its measurement function, but no specific performance data or study details are provided.
    7. The type of ground truth used: Given it's an IVD device, the "ground truth" would typically be established by a reference method or laboratory standard, but this is not discussed.
    8. The sample size for the training set: Training set information is relevant for machine learning algorithms. While the analyzer might use algorithms internally, this document doesn't detail any machine learning component or its training. However, if "training set" refers to calibration or method development, the information is not present.
    9. How the ground truth for the training set was established: See point 8.

    Conclusion:

    The provided FDA 510(k) clearance letter for the Stat Profile Critical Care Xpress Analyzer does not contain the detailed study results, acceptance criteria, or information about ground truth establishment for performance testing. This type of information would typically be found in the 510(k) submission summary or a separate clinical/technical study report, neither of which is part of the provided text. The letter primarily confirms substantial equivalence to a predicate device based on the information submitted by the manufacturer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020297
    Date Cleared
    2002-06-10

    (133 days)

    Product Code
    Regulation Number
    862.1120
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stat Profile Critical Care Xpress Analyzer is intended for in vitro diagnostic use by health care professionals for the quantitative determination of pH, pCO2, pO2, SO2%, Hct, Hb, Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac and Creat in heparinized whole blood, and Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac and Creat in serum and plasma.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter for the Stat Profile Critical Care Xpress Analyzer. It outlines the device's indications for use and states that the device is substantially equivalent to legally marketed predicate devices. However, it does not contain detailed information about specific acceptance criteria or the study that proves the device meets those criteria, beyond the general statement of substantial equivalence.

    Therefore, I cannot directly extract the specific acceptance criteria and detailed study information as requested.

    However, I can interpret what would typically be expected for a device like this based on the document's content and general FDA regulatory requirements for in vitro diagnostic devices.

    Based on the typical requirements for devices like the Stat Profile Critical Care Xpress Analyzer, and the provided document's limited information, here's what can be inferred or generally expected for such a submission:

    The document primarily focuses on establishing "substantial equivalence" to a predicate device, which means demonstrating that the new device is as safe and effective as a legally marketed device. This typically involves performance data, but the specifics are not included in this letter.


    Hypothetical Acceptance Criteria and Study Information (Based on typical performance studies for similar IVDs, NOT explicitly stated in the document):

    Given that this is an in vitro diagnostic (IVD) device for measuring various analytes in blood, serum, and plasma, the acceptance criteria would typically revolve around:

    • Accuracy/Bias: How close the device's measurements are to a reference method or true value.
    • Precision/Reproducibility: How consistent the device's measurements are when repeated under similar conditions.
    • Linearity/Measuring Range: The range over which the device can accurately measure the analytes.
    • Interference: How various substances in the sample (e.g., hemolysis, lipemia, common medications) affect the measurements.
    • Sample Matrix Equivalency: Ensuring comparable performance across different stated sample types (whole blood, serum, plasma).

    Disclaimer: The following table and sections contain assumed information based on general IVD performance claims, as the specific data is not present in the provided document.

    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter (Analyte)Acceptance Criteria (Hypothetical)Reported Device Performance (Hypothetical)
    pHBias ≤ ±0.02 pH units (vs. reference)Bias: ±0.015 pH units
    pCO2Bias ≤ ±2 mmHg or 5% (whichever is greater)Bias: ±1.5 mmHg
    pO2Bias ≤ ±3 mmHg or 5% (whichever is greater)Bias: ±2.0 mmHg
    Na+Bias ≤ ±2.0 mmol/L or 2% (whichever is greater)Bias: ±1.5 mmol/L
    K+Bias ≤ ±0.15 mmol/L or 4% (whichever is greater)Bias: ±0.10 mmol/L
    Cl-Bias ≤ ±2.0 mmol/L or 2% (whichever is greater)Bias: ±1.8 mmol/L
    GlucoseBias ≤ ±6 mg/dL or 10% (whichever is greater)Bias: ±5 mg/dL
    Precision (CV%)≤ 2.0% for most analytes at medical decision levels< 1.5% for most analytes
    Measuring RangeFor example, pH: 6.8 - 7.8For example, pH: 6.75 - 7.85

    Note: These are hypothetical acceptance criteria and performance values, typically established against a recognized standard or predicate device in a comparative study.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size or data provenance for any test set. For an IVD device, a typical test set for a 510(k) submission would involve:

    • Sample Size: Hundreds to thousands of patient samples would typically be used across various studies (accuracy, precision, linearity, interference). For accuracy studies, 40-100 patient samples spanning the measuring range are often used per analyte, compared against a reference method or predicate. For interference, numerous potential interferents would be tested.
    • Data Provenance: Studies would typically be conducted at multiple clinical sites (e.g., hospitals, reference labs) within the country where regulatory approval is sought (e.g., USA). The data would be prospective (samples collected and analyzed during the study) or retrospective clinical samples with known characteristics, sometimes spiked to cover specific ranges or interference conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    For in vitro diagnostic devices like the Stat Profile Critical Care Xpress Analyzer, "ground truth" for the test set is established by:

    • Reference Methods: Using highly accurate and precise laboratory reference methods (e.g., a gas chromatograph for blood gases, atomic absorption spectroscopy for electrolytes) or a well-validated predicate device.
    • Not human expert consensus related to image interpretation: This section typically applies to diagnostic imaging AI, not directly to quantitative chemical analyzers. The "ground truth" is analytical.

    Thus, there are no human experts establishing ground truth in the sense of adjudication for this type of device documentation. Instead, highly trained clinical laboratory scientists and specialists operate the reference instrumentation.

    4. Adjudication Method for the Test Set

    Not applicable for this type of IVD device. "Adjudication" typically refers to resolving disagreements among human reviewers of images or clinical assessments. For quantitative laboratory analyzers, the ground truth is determined by the output of a reference method or predicate device, and statistical agreement/difference is calculated.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    Not applicable. MRMC studies are relevant for diagnostic imaging or subjective clinical assessments where multiple human readers interpret cases. This device is a quantitative in vitro diagnostic analyzer, not an imaging or interpretive AI system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is also not directly applicable in the same way it would be for an AI algorithm. The device is a standalone instrument that provides quantitative measurements. Its "performance" is measured by its analytical accuracy, precision, etc., without direct human interpretation being part of the primary output. Human operators initiate the tests and interpret the results, but the measurement itself is standalone.

    7. The Type of Ground Truth Used

    The ground truth used for performance validation of this IVD device would be:

    • Reference Method Analysis: Measurements obtained from highly accurate and well-established laboratory reference methods for each analyte.
    • Predicate Device Comparison: Measurements obtained from the legally marketed predicate device to which substantial equivalence is claimed.
    • Certified Reference Materials/Calibrators: For linearity, accuracy, and calibration verification.

    8. The Sample Size for the Training Set

    The document does not specify a "training set" in the context of an AI model.

    For traditional IVDs, "training" is more about:

    • Calibration: Using a small set of known-value calibrator materials to establish the instrument's measurement curve.
    • Algorithm Development/Optimization: This is proprietary but would involve internal R&D with a dataset to fine-tune the sensor readings to final concentration values. This is not typically disclosed as a "training set" in regulatory filings but is part of product development.

    9. How the Ground Truth for the Training Set Was Established

    Again, "training set" refers more to AI/machine learning. For an IVD, the ground truth for calibration and internal development would be established through:

    • Certified Reference Materials: Materials with precisely known analyte concentrations.
    • Traceability to International Standards: Ensuring accuracy is traceable to recognized international measurement standards.
    • Comparative Analysis: Running samples alongside established reference methods or predicate devices during the development phase.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1