510(k) Data Aggregation

    K Number
    K023953

    Validate with FDA (Live)

    Date Cleared
    2003-02-14

    (79 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a plastic disposable attachment for use in conjunction with a slow speed handpiece, used during the professional prophylaxis treatment of patient's teeth by a dentist or hygienist.

    Our disposable prophylaxis angle is intended as a single use device used by a Dental Professional, either a Dentist or a Dental Hygienist to assist in the cleaning process of patient's teeth. It is intended for a single patient and should be discarded after each patient.

    Device Description

    Plastic disposable item comprising a housing (body) that holds two plastic shafts aligned at 90 degrees (right angle) to each other that rotate a cup or brush.

    Plastic one-piece housing (outer body) holding 2 internal shafts aligned at 90 degree (right angle) to each other. Tip fitted with a rubber cup or brush to clean teeth.

    AI/ML Overview

    This document describes the St. George Technology, Inc. Perfection Plus Disposable Prophy Angles and their substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance Criteria (Predicate Device Performance)Reported Device Performance (St. George Technology Perfection Plus DPA)
    Indications for UseFor single use by Dental Professional in cleaning patient's teeth.Same (Intended as a single-use device during professional prophylaxis treatment of patient's teeth by a dentist or hygienist.)
    Target PopulationDentists and HygienistsSame
    DesignPlastic one-piece housing (outer body) holding 2 internal shafts aligned at 90 degree (right angle) to each other. Tip fitted with a rubber cup or brush to clean teeth.Same
    MaterialsTwo types of plastic with either a rubber cup or brush fitted at the cleaning tip.Same (Non-latex material used on all cups - Improvement over predicate)
    PerformanceTo be able to perform a full cleaning procedure.Satisfactorily performed when compared to the predicate device in all criteria (fit, wear, duration of performance, strength, and durability).
    SterilityNon SterileSame
    BiocompatibilityHave caused concern with latex allergies due to cup material.Non latex material used on all cups - Improvement
    Mechanical SafetyRobust construction to withstand forces involved in a cleaning.Same
    Energy Used/DeliveredRotation of the cleaning device cup/brush tip provided by a slow speed hand pieceSame
    Compatibility with other devicesDesigned to fit securely onto most available ISO fitting standard slow speed hand piecesSame
    Where UsedDental Office/OperatorySame

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in the way it would for AI/ML performance. Instead, the testing described appears to be for mechanical and functional evaluation of the device.

    • Sample Size: Not explicitly stated. The document mentions "extensively tested in in-vitro environments."
    • Data Provenance: The tests were conducted in "in-vitro environments," meaning in a laboratory setting, rather than on human subjects or with clinical data. The country of origin for the data is not specified beyond St. George Technology, Inc. being a US-based company. The data is retrospective in the sense that it evaluates the manufactured device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable. The study is a non-clinical performance evaluation of a medical device, not an AI/ML study requiring expert ground truth for classification or diagnosis.

    4. Adjudication Method for the Test Set

    This information is not applicable for this type of non-clinical device performance study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document describes a non-clinical performance study of a physical medical device (disposable prophy angles), not an AI/ML system. Therefore, an MRMC study and the concept of human readers improving with AI assistance are not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This document pertains to a physical medical device, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for this study is the observed performance characteristics of the predicate device and the St. George Technology device under defined in-vitro testing conditions. This includes measurements of fit, wear, duration of performance, strength, and durability. It is based on objective, quantifiable physical properties rather than expert consensus, pathology, or outcomes data in a clinical sense.

    8. The Sample Size for the Training Set

    This information is not applicable. This is a medical device performance study, not an AI/ML study that uses a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As stated above, there is no "training set" in the context of this device performance study.