Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071137

    Validate with FDA (Live)

    Date Cleared
    2007-08-01

    (99 days)

    Product Code
    Regulation Number
    890.5860
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Therapeutic Ultrasound:

    1. Relief of pain, muscle spasms and joint contractures:
    2. Relief of pain, muscle spasms and joint contractures that may be associated with:
    • Adhesive capsulitis
    • Bursitis with slight calcification
    • Myositis
    • Soft tissue injuries
    • Shortened tendons due to past injuries and scar tissues
    1. Relief of pain, muscle spasms and joint contractures resulting from:
    • Capsular tightness
    • Capsular tightening

    4-Pole Interferential, 2-Pole Interferential, TENS and Microcurrent waveforms:

    1. Symptomatic relief of chronic intractable pain
    2. Post-traumatic pain
    3. Post-surgical pain

    EMS, TENS, Hi Volt and Russian waveforms:

    1. Relaxation of muscle spasms
    2. Increase local blood circulation
    3. Prevention or retardation of disuse atrophy
    4. Muscle re-education
    5. Maintaining or increasing range of motion
    6. Immediate post surgical stimulation of calf muscles to prevent venous thrombosis

    DC (Direct Current):
    Relaxation of muscle spasms

    Device Description

    The Sonicator® Plus 940 , Model ME 940 is a four-channel combination unit for therapeutic ultrasound and muscle stimulation. The microprocessor controlled Sonicator Plus 940 provides premodulated medium frequency and symmetrical biphasic waveforms with enhanced reliability and ease of use. In addition the Sonicator Plus 940 offers 1 and 3 MHz ultrasound using a variety of interchangeable applicators.

    The four-channel Sonicator Plus 940 allows the clinician to utilize up to two different waveforms using four channels simultaneously. The clinician can choose between several different amplitude modulation options such as the surge, reciprocation and amplitude modulation (interferential only, vector rotation). The interferential and premodulated modes offer frequency modulation as well as a static frequency option.

    The membrane panel provides both tactile and audio feedback when buttons are pressed. Blinking LED's guide the operator through the easy setup routine.

    Large, soft-touch control knobs make adjusting power for ultrasound and stimulation easy to accomplish with no guesswork involved. A large LCD output display allows the clinician to monitor four channels simultaneously for four channel combination treatment protocols. It also allows the operator to adjust both channels of an interferential protocol simultaneously while monitoring the current.

    The Sonicator Plus 940 can provide electrical stimulation only. ultrasound only and combination therapy with the pre-modulated, biphasic and medium frequency waveforms

    AI/ML Overview

    The provided document is a 510(k) summary for the SONICATOR® PLUS 940, Model ME 940, an ultrasound and muscle stimulator device. The purpose of this summary is to demonstrate substantial equivalence to predicate devices, not to perform a clinical effectiveness study. As such, the document does not contain acceptance criteria for clinical performance, nor does it describe a study to prove the device meets such criteria.

    Instead, the document focuses on comparing the technological characteristics and intended uses of the proposed device to two predicate devices: Sonicator® Plus 994, Model ME 994 (K984142) and Chattanooga Vectra Genesis (K031077). The underlying assumption for 510(k) clearance is that if the new device is substantially equivalent in terms of technology, safety, and intended use to a legally marketed device, it does not require new clinical performance data to enter the market.

    Therefore, many of the requested elements for describing an acceptance criteria study (like sample size for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set information) are not applicable or present in this type of regulatory submission.

    However, I can extract the comparative information presented in the document, which serves as the basis for demonstrating equivalence in a 510(k) submission.

    Comparative Information between Proposed and Predicate Devices

    The "acceptance criteria" in this context are effectively the demonstrated characteristics and performance parameters that are comparable to the predicate devices, thereby establishing substantial equivalence. The "study" proving this is the detailed comparison provided in the 510(k) submission.

    1. Table of "Acceptance Criteria" (Predicate Device Performance) and Reported Device Performance (Proposed Device)

    The document provides extensive tables comparing the technical specifications of the Sonicator® Plus 940 (Proposed Device) with the Chattanooga Vectra Genesis (Predicate Device K031077) for Neuromuscular Stimulation, Pain Management, Muscle Spasm, and Therapeutic Ultrasound. The "acceptance criteria" here are essentially the specifications of the predicate device. If the proposed device meets or is comparable to these specifications, it is deemed acceptable for marketing.

    Example Excerpts from Comparison Tables:

    Characteristic (Acceptance Criteria - Predicate K031077)Sonicator® Plus 940 (Proposed Device Performance)
    NEUROMUSCULAR STIMULATION
    Waveform
    EMS (Premod, Vectra): BiphasicBiphasic
    TENS (VMS, Vectra): BiphasicBiphasic
    Hi Volt: Pulsed MonophasicPulsed Monophasic and Biphasic
    Russian: BiphasicBiphasic
    Shape
    EMS (Premod, Vectra): SinusoidalSinusoidal
    TENS (VMS, Vectra): SquareSquare
    Hi Volt: Twin spikeTwin spike
    Russian: Gated SinusoidalGated Sinusoidal
    Max Output Voltage (V) ±20% @ 500 Ω
    EMS (Premod, Vectra): 5549
    Hi Volt: 544146
    PAIN MANAGEMENT
    Waveform
    Interferential, 4-pole: BiphasicBiphasic
    TENS: BiphasicBiphasic
    Max Output Current (mA) ±20% @ 500 Ω
    Interferential, 4-pole: 11496
    Microcurrent: 0.9600.760
    THERAPEUTIC ULTRASOUND
    Frequency: 1 MHz and 3.3 MHz, ± 5%1 MHz and 3 MHz, ± 5%
    Modes: Continuous and PulsedContinuous and Pulsed
    Maximum Output Power: 20 W for 10 cm² at 1 Mhz onlyN/A (12 W for ME 9401, 1.8 W for ME 9402)
    Maximum Intensity: 2.5 W/cm² for continuous mode2 W/cm² for continuous mode
    Effective Radiating Area: 5 cm² appl. ERA of 4 cm²5.5 cm² (1 MHz) / 6.0 cm² (3 MHz)

    Note: The tables clearly show differences in some parameters (e.g., maximum output voltage/current, maximum power/intensity for ultrasound). The 510(k) process accepts such differences as long as they do not raise new questions of safety and effectiveness and the device remains substantially equivalent for its intended use. The submission implicitly argues that these differences are not significant enough to warrant new clinical trials.

    2. Sample Size Used for the Test Set and the Data Provenance:
    Not applicable as this is a 510(k) premarket notification demonstrating substantial equivalence based on technical and performance comparisons to predicate devices, not a clinical study involving test sets of patients or data. The "data" here refers to the device's measured technical specifications. The provenance is internal testing performed by Mettler Electronics Corp. to establish these specifications.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
    Not applicable. No ground truth based on expert consensus for clinical outcomes was established as part of this 510(k) submission. The "ground truth" for the technical specifications is the output measured by the manufacturer against industry standards and the predicate device's published specifications.

    4. Adjudication Method for the Test Set:
    Not applicable. No clinical test set or adjudication process for clinical outcomes was part of this regulatory submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This device is an ultrasound and muscle stimulator, not an AI-powered diagnostic tool. Therefore, MRMC studies are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is a physical medical device for therapeutic purposes; it is not an algorithm, and the concept of "standalone performance" in this context is not relevant for a 510(k) based on substantial equivalence. Its performance is measured by its physical output (e.g., electrical stimulation parameters, ultrasound intensity).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The "ground truth" in this context refers to the technical specifications and performance characteristics of the device itself (e.g., voltage, current, frequency, power output, accuracy of timing). This "ground truth" is established through engineering and electrical testing, adhering to recognized standards (e.g., ISO, UL, CSA, IEC/EN, 21 CFR) and measurements against the predicate device's reported specifications.

    8. The sample size for the training set:
    Not applicable. This is not a machine learning or AI device.

    9. How the ground truth for the training set was established:
    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1