Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K222856

    Validate with FDA (Live)

    Date Cleared
    2022-10-19

    (28 days)

    Product Code
    Regulation Number
    878.4780
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SOMAVAC® 100 Sustained Vacuum System (SOMAVAC® 100) is a portable battery powered vacuum source / waste container intended for the removal of surgical and bodily fluids from a closed wound following plastic surgery and other general surgery forming large flaps for hematoma and seroma prophylaxis. It is intended for use in homecare and healthcare environments.

    Device Description

    The SOMAVAC® 100 Sustained Vacuum System (SOMAVAC® 100) is a portable battery-powered vacuum pump/waste container intended for the removal of surgical and bodily fluids from a closed wound following plastic surgery and other general surgery forming large flaps for hematoma and seroma prophylaxis. It is intended for homecare and healthcare environments. The SOMAVAC® 100 may be used for most instances a surgeon determines a closed suction drain device is applicable. The SOMAVAC® 100 is compatible with drains commonly used after surgeries and is intended to be used by a single patient. The SOMAVAC® 100 is to be installed by trained medical personnel. Post-installation (in patient setting or at home), the SOMAVAC® 100 is intended to be operated by patients or their caregivers.

    AI/ML Overview

    This document details a 510(k) premarket notification for the SOMAVAC® 100 Sustained Vacuum System. It establishes substantial equivalence to a predicate device, K180606.

    Here’s an breakdown of the acceptance criteria and study information provided:

    Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly present a table of "acceptance criteria" against which a clinical performance study would be evaluated in terms of sensitivity, specificity, accuracy, or similar metrics. Instead, the document focuses on demonstrating substantial equivalence to a predicate device (SOMAVAC® Device, K180606) through non-clinical performance testing and direct comparison of specifications.

    The "performance" described refers to the device's functional characteristics and compliance with standards, rather than diagnostic or treatment efficacy from a clinical study. The table below summarizes the key comparisons made to demonstrate substantial equivalence, which serves as the "acceptance criteria" for this type of regulatory submission in the absence of a new clinical claim.

    CharacteristicAcceptance Criteria (Predicate K180606)Reported Device Performance (Subject K222856)Assessment of Differences
    Regulatory statusClass IIClass IIIdentical
    Classification namePump, portable, aspiration.Pump, portable, aspiration.Identical
    Regulation878.4780878.4780Identical
    Product CodeBTABTAIdentical
    Vacuum pressure-60 to -350 mmHg-60 to -350 mmHgIdentical
    Use settingHome and health care environmentHome and health care environmentIdentical
    Indication for useRemoval of surgical and bodily fluids from a closed wound following plastic surgery and other general surgery forming large flaps for hematoma and seroma prophylaxis. Intended for use in homecare and healthcare environments.Removal of surgical and bodily fluids from a closed wound following plastic surgery and other general surgery forming large flaps for hematoma and seroma prophylaxis. Intended for use in homecare and healthcare environments.Identical
    Waste Collection Capacity50mL/drain, 100 mL total (disposable waste collection units)50mL/drain, 100 mL total (disposable waste collection units)Identical
    Flow @ Max Vacuum1.0 mL/min1.0 mL/minIdentical
    Electrical3VDC, 2xAA alkaline batteries (series connection)3VDC, 2xAA batteries (series connection)Substantially Equivalent (Alkaline or lithium-ion batteries are appropriate)
    Weight300 gram w/batteries300 gram w/batteriesIdentical
    Electrical Equipment ClassificationClass II, Type BF, IP (22)Class II, Type BF, IP (22)Identical
    TransportPortable, wearable; delivered in a belt to be worn by the patientPortable, wearable; delivered in a belt to be worn by the patientIdentical
    Mode of operationNon-continuous, intermittent, controlled by vacuum pressure and amount of exudate collectedNon-continuous, intermittent, controlled by vacuum pressure and amount of exudate collectedIdentical
    Accessoriesdrain connectors, non-sterile waste collection units belt AA batteriesdrain connectors, sterile waste collection units belt AA batteriesSubstantially Equivalent (Offering of sterile drain connectors does not affect substantial equivalence)

    Study Information

    The document describes non-clinical testing rather than a traditional “study” with patient data.

    1. Sample size used for the test set and the data provenance: Not applicable. This submission relies on non-clinical performance verification and validation. The "test set" would refer to the physical devices and components tested. The provenance is not explicitly stated in terms of country of origin of "data," but it implicitly comes from the manufacturer's internal testing. The testing is prospective in the sense that the new device was built and then tested.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For non-clinical performance testing of a physical device like this, "ground truth" is established by engineering specifications, validated measurement equipment, and industry standards (e.g., electrical safety, EMC). Expert review would be part of the design and verification processes but not as a formal "ground truth" for a test set in the clinical sense.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This type of adjudication is typically used in clinical studies for interpretation of imaging or clinical outcomes, not for engineering performance testing.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a powered suction pump, not an AI-assisted diagnostic or therapeutic tool that would involve human readers or cases in an MRMC study.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a hardware product. While it contains software, the performance listed is for the complete system.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the non-clinical testing is based on:

      • Engineering specifications: The defined parameters for vacuum, flow, electrical characteristics, and physical properties.
      • Regulatory standards: Compliance with voluntary standards for electrical safety (e.g., IEC 60601-1), electromagnetic compatibility (e.g., IEC 60601-1-2), and powered suction pumps (e.g., ISO 10079-1).
      • Previous device performance: The performance of the predicate device (K180606) served as a benchmark for equivalence.
    7. The sample size for the training set: Not applicable. This device is not an AI/ML algorithm that requires a training set. The software changes were validated in accordance with FDA guidances for software in medical devices, but this is distinct from training an AI model.

    8. How the ground truth for the training set was established: Not applicable. (See answer to point 7).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1