Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K991403

    Validate with FDA (Live)

    Date Cleared
    1999-07-16

    (85 days)

    Product Code
    Regulation Number
    886.5928
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SOLO-Care™ Brand Multi-Purpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting and protein removal, storing soft (hydrophilic), rigid gas permeable (fluoro silicon acrylate and silicon acrylate)and PMMA contact lenses as recommended by your eye care practitioner.

    Device Description

    SOLO-Care™ Brand Multi-Purpose Solution is a sterile aqueous solution used in the care contact lenses and is indicated for cleaning, rinsing, chemical (not heat) disinfecting and protein removal. storing soft (hydrophilic), rigid gas permeable (fluoro silicon acrylate and silicon acrylate) and PMMA contact lenses as recommended by your eye care practitioner. It may also be uses as a diluent for enzymatic cleaning tablets, which are to be used in conformance to the established labeling directions of the enzymatic cleaning tablet. The solution is contains in a plastic bottle and consists of a sterile isotonic solution containing sodium chloride, polyoxyethylene polyoxypropylene block copolymer, sodium phosphate dibasic, sodium phosphate monobasic, and preserved with edetate disodium dihydrate and polyhexanide 0.0001%. Each bottle is supplied sterile and is labeled with a lot number and expiration date.

    AI/ML Overview

    The provided document is limited to the 510(k) Summary of Safety and Effectiveness for SOLO-Care™ Brand Multi-Purpose Solution. This document primarily focuses on demonstrating substantial equivalence to a predicate device for regulatory approval. As such, it does not contain the detailed acceptance criteria and study information typically associated with performance evaluations of AI/ML-driven medical devices.

    The document describes a contact lens solution, not an AI/ML device. Therefore, the questions related to AI/ML device performance, such as sample sizes for test sets, ground truth establishment by experts, MRMC studies, standalone performance, and training set details, are not applicable to this submission.

    However, I can extract the relevant information regarding the "study" mentioned and acceptance criteria within the context of a contact lens solution:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Disinfection EfficacySatisfactory microbiological testing results at time periods ranging from 10 minutes to 4 hours. (Specific reduction logs or pass/fail criteria are not detailed in this summary, but are implied to be satisfactory for contact lens disinfection.)
    Compatibility with PMMA LensesCompatibility study successfully completed.
    Substantial Equivalence to Predicate DeviceDemonstrated as having identical formulation and indications for use as the currently marketed SOLO-Care™ Brand MULTI-PURPOSE SOLUTION.
    10-minute Disinfection Cycle Time Equivalence to 4-hour Cycle TimeDemonstrated as substantially equivalent.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. The document mentions "a series of microbiological testing" and "a compatibility study" but does not provide the number of samples or lenses used in these tests.
    • Data Provenance: Not explicitly stated (e.g., country of origin). However, the studies were "pre-clinical studies" and "all studies have previously been submitted under P940042, K 982168, and under K983291," indicating they were conducted as part of previous regulatory submissions. These are retrospective in the context of this submission, as they were prior submissions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as this is a chemical solution, not an AI/ML diagnostic device requiring expert interpretation of results for ground truth.

    4. Adjudication method for the test set

    This information is not applicable as this is a chemical solution, not an AI/ML diagnostic device requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as this is a chemical solution, not an AI/ML diagnostic device involving human readers or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable as this is a chemical solution and does not involve an algorithm.

    7. The type of ground truth used

    For the microbiological testing, the "ground truth" would be established by standard microbiological methods (e.g., colony counts to determine microbial reduction). For the compatibility study, the "ground truth" would be established by observed physical or chemical integrity and performance of the PMMA lenses after exposure to the solution. These are laboratory-based scientific measurements, not expert consensus or pathology in the clinical sense.

    8. The sample size for the training set

    This information is not applicable as this is a chemical solution and does not involve a training set for an AI/ML algorithm.

    9. How the ground truth for the training set was established

    This information is not applicable as this is a chemical solution and does not involve a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K983291

    Validate with FDA (Live)

    Date Cleared
    1998-11-25

    (65 days)

    Product Code
    Regulation Number
    886.5918
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SOLO-Care ™ Brand Multi-Purpose Solution Is Indicated for use in daily cleaning, ninsing, chemical (not haat) disinfecting, protein removal and storing of soft (hydrophilic) or rigld gas permeable (fluoro sillcone acrylate and sillcone acrylate) lenses as recommended by your eye care precitioner.

    Device Description

    SOLO-Care™ Brand Multi-Purpose Solution is a sterile aqueous solution containing sodium chloride, polyoxyethylene polyoxypropylene block copolymer, sodium phosphate dibasic, sodium phosphate monobasic, and preserved with edetate disodium dihydrate 0.025% and polyhexanide 0.0001%. SOLO-Care™ Brand Multi-Purpose Solution contains multiple active ingredients in sufficient concentration to perform the function of daily cleaning, protein removal, rinsing, disinfecting, and storing soft (hydrophilic) contact lenses as recommended by your eye care practitioner. The sterile solution is contained in a plastic bottle and is labeled with a lot number and expiration date.

    AI/ML Overview

    This document is a 510(k) Summary for the SOLO-Care™ Brand Multi-Purpose Solution, which is a contact lens solution. It describes the device's indications for use, its substantial equivalence to a predicate device, and outlines the studies conducted to demonstrate its safety and effectiveness.

    Here's the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative or tabular format for specific performance metrics (e.g., microbial kill rates, protein removal percentages with thresholds). Instead, it states that various studies were conducted to confirm safety and effectiveness, and that the product is shown to be "substantially equivalent" to a predicate device.

    However, we can infer the intent of the acceptance criteria based on the described functions and comparative claims. The performance is reported in terms of its ability to perform these functions effectively and its compatibility.

    Acceptance Criteria (Inferred from described functions and claims)Reported Device Performance
    Cleaning of soft (hydrophilic) and rigid gas permeable (RGP) contact lensesSOLO-Care™ Brand Multi-Purpose Solution successfully performs daily cleaning.
    Protein Removal for soft (hydrophilic) and RGP contact lensesSOLO-Care™ Brand Multi-Purpose Solution successfully performs protein removal. (Also noted previous approval for daily protein removal for soft lenses under K982168).
    Rinsing of soft (hydrophilic) and RGP contact lensesSOLO-Care™ Brand Multi-Purpose Solution successfully performs rinsing.
    Chemical Disinfection (not heat) of soft (hydrophilic) and RGP contact lensesSOLO-Care™ Brand Multi-Purpose Solution successfully performs chemical disinfection.
    Storing of soft (hydrophilic) and RGP contact lensesSOLO-Care™ Brand Multi-Purpose Solution successfully performs storing.
    Compatibility with rigid gas permeable (fluoro silicone acrylate and silicone acrylate) lenses under recommended care regimenLens solution compatibility tests indicated that SOLO-Care™ Brand Multi-Purpose Solution is compatible with RGP lenses, demonstrating equivalence to the predicate device (Boston Simplicity).
    Compatibility with Focus Lens DropsAn additional compatibility study was successfully conducted to determine compatibility with Focus Lens Drops.
    Safety and EffectivenessA series of preclinical and clinical studies were conducted to assess and demonstrate the safety and effectiveness of SOLO-Care™ Brand Multi-Purpose Solution.
    Substantial Equivalence to BOSTON Simplicity Multi-Action SolutionSOLO-Care™ Brand Multi-Purpose Solution is substantially equivalent for daily cleaning, rinsing, disinfecting RGP lenses.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states: "A series of preclinical and clinical studies have been conducted to assess and demonstrate the safety and effectiveness..." However, no specific sample sizes for clinical or preclinical test sets are provided within this document.

    The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given it's an FDA submission, the studies would typically be prospective for clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The evaluation of contact lens solutions typically involves laboratory testing for microbial reduction, cytotoxicity, and clinical trials for safety and comfort, which are assessed by various medical professionals and technicians, but "experts" establishing ground truth in the sense of image interpretation is not directly applicable here.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. Adjudication methods like 2+1 or 3+1 are typically used for medical image interpretation where there's subjectivity and disagreement among readers. This is not directly relevant to the testing of a contact lens solution.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done, as this device is a contact lens solution, not an AI-powered diagnostic tool. Therefore, there is no effect size related to human readers improving with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No standalone algorithm performance study was done, as this device is a chemical solution, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a contact lens solution, "ground truth" generally refers to:

    • Laboratory-based measurements: e.g., microbial kill rates using standardized organisms, protein removal efficacy, cleaning efficacy using simulated soiling, lens compatibility assessments.
    • Clinical outcomes data: e.g., slit lamp examinations for ocular health, patient reported comfort, adverse event rates, visual acuity.
    • Comparison to predicate device performance benchmarks.

    The document generically states "preclinical and clinical studies have been conducted to assess and demonstrate the safety and effectiveness." It implies that standard test methods applicable to contact lens solutions were used to establish the "ground truth" for these various performance parameters.

    8. The sample size for the training set

    This information is not provided and is not applicable in the context of a chemical contact lens solution. "Training set" refers to data used to train machine learning models, which are not involved in this device's development or evaluation.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K982168

    Validate with FDA (Live)

    Date Cleared
    1998-11-17

    (151 days)

    Product Code
    Regulation Number
    886.5928
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SOLO-Care ™ Brand Multi-Purpose Solution is indicated for use In dally cleaning, rinsing, chemical (not heat) disinfecting, protein removal and storing of soft (hydrophilic) contact lenses as recommended by your eye care practitioner.

    Device Description

    SOLO-Care™ Brand Multi-Purpose Solution is a sterile aqueous solution containing sodium chloride, polyoxyethylene polyoxypropylene block copolymer, sodium phosphate dibasic, sodium phosphate monobasic, and preserved with edetate disodium dihydrate 0.025% and polyhexanide 0.0001%. SOLO-Care™ Brand Multi-Purpose Solution contains multiple active ingredients in sufficient concentration to perform the function of daily cleaning, daily protein removal, rinsing, disinfecting, and storing soft (hydrophilic) contact lenses as recommended by your eye care practitioner. The sterile solution is contained in a plastic bottle and is labeled with a lot number and expiration date.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the SOLO-CARE™ Brand MULTI-PURPOSE SOLUTION, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/GoalSOLO-CARE™ Device Performance/Outcome
    SafetyNon-toxicResults demonstrated non-toxicity
    BiocompatibleResults demonstrated biocompatibility
    No safety issues with use without rinse prior to insertionNo safety issues raised with SOLO-Care™ Multi-Purpose Solution used without a rinse prior to insertion.
    Effectiveness (General)Safe and effective as a multi-purpose soft (hydrophilic) contact lens solutionResults demonstrated the solution is safe and effective as a multi-purpose soft (hydrophilic) contact lens solution.
    Effectiveness (Specific - Protein Removal)Substantially equivalent to predicate device (ReNu® Multi-Purpose Solution / ReNu® 1 Step™ Enzymatic Cleaner) for removal of visible proteinClinical results provide valid scientific evidence that SOLO-care™ Multi-Purpose Solution is substantially equivalent to ReNu® Multi-Purpose Solution / ReNu® 1 Step™ Enzymatic Cleaner for the removal of visible protein.
    Substantial EquivalenceComparable to other currently marketed soft contact lens solutionsResults demonstrate the solution is comparable to other currently marketed soft contact lens solutions.
    Substantially equivalent to predicate device across all indicated uses (daily cleaning, protein removal, rinsing, disinfecting, storing)SOLO-Care Brand Multi-Purpose Solution is substantially equivalent to Bausch & Lomb ReNu Multi-Purpose Solution used with Bausch & Lomb ReNu 1 Step Enzymatic Cleaner in that both are indicated for use in daily cleaning, daily protein removal, rinsing, disinfecting, and storing of soft (hydrophilic) contact lenses.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: The document describes "A one month clinical evaluation" and "a series of preclinical tests." However, it does not specify the exact number of participants or samples used in either the preclinical or clinical studies.
    • Data Provenance: The document does not explicitly state the country of origin for the data. The submitter is located in Duluth, Georgia, USA, and the FDA review is from the USA, suggesting the studies likely occurred in the USA. The studies were conducted adhering to "current Good Clinical Practices and published regulations (21 CFR Parts 50, 56, 312, 812)" for clinical studies and "GLP regulation (21 CFR Part 58)" for preclinical studies, which are US regulations. The studies appear to be prospective as they were conducted to assess the product's safety and effectiveness.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document does not provide information on the number or qualifications of experts used to establish a ground truth for the test set. For a clinical evaluation of a contact lens solution, ground truth would typically be established through patient self-assessment, clinical examinations by ophthalmologists or optometrists, and potentially laboratory analysis of lens deposits. However, these details are not provided in the given text.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. This is a medical device (contact lens solution), not an AI-powered diagnostic or imaging device, so MRMC studies involving human readers and AI assistance are not applicable in this context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No standalone algorithm performance was done. This device is a solution, not an algorithm. Performance tests focused on the solution's properties and its effect on contact lenses and users.

    7. The Type of Ground Truth Used

    Based on the description:

    • For safety, ground truth would be based on clinical observation of adverse events and preclinical toxicology/biocompatibility testing standards.
    • For effectiveness (protein removal), ground truth was implicitly established by comparing the SOLO-Care™ solution's performance for "removal of visible protein" against the performance of the predicate device (ReNu® Multi-Purpose Solution / ReNu® 1 Step™ Enzymatic Cleaner). This suggests a comparative clinical outcome, likely visual assessment of protein deposits on lenses or other relevant clinical endpoints that demonstrate "removal of visible protein." The specific methodology for assessing "visible protein" is not detailed.

    8. The Sample Size for the Training Set

    The document does not refer to a training set in the context of an algorithm or AI. The term "training set" is not applicable to the development and testing of this type of medical device.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set in the context of this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1