510(k) Data Aggregation

    K Number
    K231621

    Validate with FDA (Live)

    Date Cleared
    2023-07-25

    (53 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed ViewFlex Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Device Description

    The Reprocessed ViewFlex Xtra ICE Catheter is a temporary intracardiac ultrasound catheter intended for use in patients to visualize cardiac structures and blood flow within the heart. The Reprocessed ViewFlex Xtra ICE Catheter is inserted into the heart via intravascular access. The Reprocessed ViewFlex Xtra ICE Catheter shaft is a 9 French catheter with a useable length of 90 cm. The shaft is compatible with a 10 French or larger introducer for insertion into the femoral or jugular veins. The catheter tip is a 64-element linear phased array transducer housed in silicone. The distal portion of the shaft is deflectable in four directions allowing for left-to-right and anteriorto-posterior deflection. The handle of the catheter has two deflection mechanisms that correspond with movement of the distal shaft in the four planes of movement. The Reprocessed ViewFlex Xtra Ice Catheter is compatible with ViewMate Z or ViewMate and Philips CX50 ultrasound consoles via the use of a compatible ViewFlex Catheter interface module.

    AI/ML Overview

    The provided text does not contain a study that proves the device meets specific acceptance criteria in terms of accuracy or clinical performance. The submission is for a reprocessed medical device (Intracardiac Echocardiography Catheter), and the studies described are primarily focused on demonstrating that the reprocessed device is as safe and effective as the predicate device and operates as originally intended.

    Specifically, the document focuses on:

    • Bench and laboratory testing: To demonstrate the performance (safety and effectiveness) of the reprocessed device.
    • Comparison to a predicate device: Asserting that the reprocessed device is substantially equivalent to legally marketed predicate devices.
    • Validation of reprocessing methods: Ensuring cleaning, sterilization, and functional integrity after reprocessing.

    Therefore, many of the requested categories for a study demonstrating clinical performance or diagnostic accuracy are not applicable to this submission.

    However, I can extract the relevant information regarding the testing performed and the general "acceptance criteria" through comparison to the predicate.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    The document does not provide a table of precise quantitative acceptance criteria (e.g., specific thresholds for image quality metrics) or their exact reported performance values. Instead, it states that the reprocessed device meets the functional and safety requirements to be considered equivalent to the predicate. The "performance" is implicitly "as safe and effective as the predicate" and "operates as originally intended."

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityDemonstrated compliance.
    Cleaning ValidationDemonstrated compliance.
    Sterilization ValidationDemonstrated compliance.
    Functional PerformanceOperates as originally intended (includes Visual Inspection, Simulated Use, Dimensional Verification, Ultrasound Transducer Testing, Image Quality Testing, Acoustic Output Testing, Mechanical Characteristics).
    EMC and Electrical SafetyDemonstrated compliance.

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for the "test set" (e.g., number of reprocessed catheters tested for each criterion). The testing would have been done on a sufficient number of reprocessed devices to validate the reprocessing methods and ensure consistency. The data provenance is from bench and laboratory testing conducted by Stryker Sustainability Solutions, which is prospective data generation for the purpose of demonstrating substantial equivalence. The country of origin of the data is not explicitly stated but is implied to be within the US (given FDA submission).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable/not provided as the submission is for a reprocessed device and describes technical and functional testing, not a clinical study involving diagnostic accuracy with expert-established ground truth.

    4. Adjudication method for the test set

    This information is not applicable/not provided as the submission is for a reprocessed device and describes technical and functional testing, not a clinical study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done

    No, a MRMC comparative effectiveness study was not conducted or described. The focus is on the substantial equivalence of the reprocessed device to its predicate through engineering and functional testing.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable/not provided. The device is an Intracardiac Echocardiography Catheter, not an AI algorithm. The functional testing described is effectively a standalone performance assessment of the physical device.

    7. The type of ground truth used

    The "ground truth" for the functional and safety testing would be:

    • Engineering specifications of the original device.
    • Industry standards and regulatory requirements for medical device safety and performance (e.g., biocompatibility standards, sterilization efficacy).
    • The performance characteristics of the new/unreprocessed predicate device.

    8. The sample size for the training set

    This information is not applicable/not provided. There is no "training set" in the context of this device submission, as it relates to a physical medical device and its reprocessing, not a machine learning algorithm.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182238

    Validate with FDA (Live)

    Date Cleared
    2019-05-08

    (261 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reprocessed ViewFlex™ Xtra ICE Catheter is indicated for use in adult and adolescent pediatric patients to visualize cardiac structures, blood flow and other devices within the heart.

    Device Description

    The Reprocessed ViewFlex Xtra ICE Catheter is a temporary intracardiac ultrasound catheter intended for use in patients to visualize cardiac structures and blood flow within the heart. The Reprocessed ViewFlex Xtra ICE Catheter is inserted into the heart via intravascular access. The Reprocessed ViewFlex Xtra ICE Catheter shaft is a 9 French catheter with a useable length of 90 cm. The shaft is compatible with a 10 French or larger introducer for insertion into the femoral or jugular veins. The catheter tip is a 64- element linear phased array transducer housed in silicone. The distal portion of the shaft is deflectable in four directions allowing for leftto-right and anterior-to-posterior deflection. The handle of the catheter has two deflection mechanisms that correspond with movement of the distal shaft in the four planes of movement. The Reprocessed ViewFlex Xtra Ice Catheter is compatible with ViewMate Z or ViewMate and Philips CX50 ultrasound consoles via the use of a compatible ViewFlex Catheter interface module.

    AI/ML Overview

    The provided text is a 510(k) summary for a reprocessed medical device, specifically a Reprocessed ViewFlex Xtra ICE Catheter. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of the device's performance against detailed acceptance criteria for a new AI or diagnostic device.

    Therefore, many of the requested elements (e.g., sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth for training) are not applicable to this document as it describes a reprocessed catheter and not a new diagnostic algorithm or AI system.

    However, I can extract the relevant information regarding performance testing that was conducted to demonstrate the reprocessed device's safety and effectiveness.

    Here's a breakdown of what can be extracted, and where limitations exist based on the document:

    1. A table of acceptance criteria and the reported device performance:

    The document describes the types of tests performed to demonstrate that the reprocessed device performs "as originally intended" and is "effectively identical" to the predicate. It does not provide specific quantitative acceptance criteria or detailed numerical performance results for these tests. The conclusion states the device is "at least as safe and effective as the predicate device," implying it met internal standards for each test.

    Test CategoryReported Device PerformanceAcceptance Criteria (Not explicitly stated in document)
    BiocompatibilityPassed (implied by conclusion of safety and effectiveness)(Not detailed, but typically involve ISO 10993 compliance)
    Validation of ReprocessingPassed (implied by conclusion of safety and effectiveness)(Not detailed, but involves demonstrating proper cleaning, disinfection/sterilization, and restoration of function)
    Sterilization ValidationPassed (implied by conclusion of safety and effectiveness)(Not detailed, but involves demonstrating sterility assurance level)
    Functional Performance Tests"demonstrates that Reprocessed ViewFlex device perform as originally intended."(Not detailed, would involve specific mechanical and operational parameters)
    Acoustic TestingPassed (implied by conclusion of safety and effectiveness)(Not detailed, would involve ultrasound image quality and output parameters)
    Electrical Safety TestingPassed (implied by conclusion of safety and effectiveness)(Not detailed, but typically involves IEC 60601 standards)
    Electromagnetic Compatibility TestingPassed (implied by conclusion of safety and effectiveness)(Not detailed, but typically involves IEC 60601-1-2 standards)
    Packaging ValidationPassed (implied by conclusion of safety and effectiveness)(Not detailed, but involves maintaining sterility and product integrity)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified. The document states "Each individual device is tested for appropriate function" during reprocessing, and that bench and laboratory testing was conducted. This implies testing on a statistically relevant sample of reprocessed devices, but the exact number isn't given.
    • Data Provenance: Not applicable in the context of this device and the types of tests described. These are laboratory/bench tests, not clinical data sets from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable. This document describes engineering and laboratory testing for a reprocessed catheter, not a diagnostic algorithm that requires expert ground truth for interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are used for interpreting clinical data, particularly in a reader study or when establishing ground truth from multiple experts. This is not relevant for bench testing of a catheter.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a reprocessed catheter, not an AI or diagnostic imaging device. No MRMC study was mentioned or would be relevant for this type of submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. There is no algorithm; this is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable. The "ground truth" for these engineering tests would be established by validated test methods, adherence to specifications, and comparison to the original device's performance, as measured by calibrated equipment and documented procedures. For instance, the ground truth for electrical safety is adherence to established electrical safety standards.

    8. The sample size for the training set:

    • Not applicable. This is a reprocessed physical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set for an AI model is involved.

    Summary of the Study:

    The study described is a series of bench and laboratory tests designed to confirm that the Reprocessed ViewFlex Xtra ICE Catheter is safe and effective and performs "as originally intended," thereby demonstrating substantial equivalence to the predicate device.

    The tests included:

    • Biocompatibility
    • Validation of Reprocessing
    • Sterilization Validation
    • Functional Performance Tests
    • Acoustic Testing
    • Electrical Safety Testing
    • Electromagnetic Compatibility Testing
    • Packaging Validation

    The specific detailed results or quantitative acceptance criteria for each test are not provided in this summary document, only the general statement that the testing "demonstrates that Reprocessed ViewFlex device perform as originally intended" and that the device is "at least as safe and effective as the predicate device." The study essentially aims to prove that the reprocessing process does not negatively impact the device's original validated performance and safety characteristics.