Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162226

    Validate with FDA (Live)

    Date Cleared
    2017-01-25

    (170 days)

    Product Code
    Regulation Number
    876.5010
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Reliant(TM) Multistage Dilatation Balloon Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.

    Device Description

    The Reliant™ Multistage Dilatation Balloon Catheter is capable of 3 distinct and progressively larger size diameters via controlled radial expansion. Specific balloon sizes are printed on each package and hub label.

    The Reliant™ Multistage Dilatation Balloon Catheter is designed to pass through a 2.8mm or greater working channel of an endoscope. It will also accept a 0.035 in (0.89 mm) guidewire through its guidewire lumen. This catheter comes packaged with a 0.035 in (0.89mm) guidewire preloaded in the guidewire lumen. The guidewire is about 20cm longer than the balloon catheter with the excess length extending from single lumen tube.

    The guidewire locking device is attached to the guidewire hub of the catheter. The locking device will be packaged in locked.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for the performance tests in the typical sense of a target metric the device must achieve. Instead, it reports the measured values for the proposed device and compares them to the predicate device. The underlying acceptance criterion for these tests is that the new device should be substantially equivalent to the predicate device, meaning its performance characteristics are similar enough not to raise new questions of safety or effectiveness.

    Performance TestProposed Device Performance (Reliant™ Multistage Dilatation Balloon Catheter)Predicate Device Performance (K151671)
    Tensile Strength between Balloon and Double Lumen Tube (N)$67.11 \pm 2.16$$78.11 \pm 9.59$
    Tensile Strength between Double Lumen Tube and Junction (N)$109.99 \pm 2.12$$94.67 \pm 3.54$
    Biocompatibility - CytotoxicityThe test article extracts had not a cytotoxic potential. Neither SC extracts nor CSO extracts had a cytotoxic potential.The test article extracts had not a cytotoxic potential. Neither SC extracts nor CSO extracts had a cytotoxic potential.
    Biocompatibility - SensitizationNo evidence of causing sensitization in the guinea pig.No evidence of causing sensitization in the guinea pig.
    Biocompatibility - IrritationThe test article extracts have no animal intracutaneous reactivity as compared to the control.The test article extracts have no animal intracutaneous reactivity as compared to the control.

    Note: For tensile strength, the comparison indicates "similar," implying that the measured values, while not identical, are considered acceptable for demonstrating substantial equivalence given the context of the device's function and intended use. For biocompatibility, the results are identical, indicating the proposed device performs the same as the predicate in these aspects.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document does not specify the exact sample size for the tensile strength tests (e.g., how many balloons were tested). For biocompatibility, the tests are typically performed on a statistically relevant number of samples according to ISO standards, but the specific number is not disclosed.
    • Data Provenance: The document does not explicitly state the country of origin for the test data for the tensile strength and biocompatibility tests. Given that Micro-Tech (Nanjing) Co., Ltd. is based in China, it is highly probable the testing was conducted in China or by a contracted lab. The data is retrospective in the sense that these tests were performed on the finished device before this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is not applicable to this document. The "ground truth" for material property tests like tensile strength and biocompatibility is established by standardized laboratory methods and measurements, not by expert consensus or interpretation of images/clinical data. These are objective measures.

    4. Adjudication method for the test set

    This is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies or studies involving human readers/interpreters to resolve discrepancies in diagnoses or assessments. The tests described here are physical and biological material characterization tests, which do not involve such human-based adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a medical catheter (a physical instrument), not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed or is relevant for this type of device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. As mentioned above, the device is a physical medical instrument, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the tests performed (tensile strength, biocompatibility) is based on:

    • Standardized Laboratory Measurements: For tensile strength, the ground truth is the measured force required to break the material, determined by calibrated testing equipment and procedures.
    • Standardized Biological Assay Results: For biocompatibility (cytotoxicity, sensitization, irritation), the ground truth is the observable biological response (or lack thereof) in laboratory assays, compared against established biological safety criteria (e.g., ISO 10993 series standards).

    8. The sample size for the training set

    This is not applicable. The concept of a "training set" (used for machine learning or AI algorithms) does not apply to the development or testing of this physical medical device. Device development involves engineering design, material selection, and physical testing, not algorithm training.

    9. How the ground truth for the training set was established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1